      "status": "Ongoing",
      "city": "Therwil",
      "state": "",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92355",
      "recalling_firm": "SenTec AG",
      "address_1": "Ringstrasse 39",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CA, FL, IL, MA, MS, MT, NV, and TX.",
      "recall_number": "Z-1980-2023",
      "product_description": "V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS).",
      "product_quantity": "11 in the U.S.",
      "reason_for_recall": "The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode \"production\" and were not reset to \"released\" mode.",
      "recall_initiation_date": "20230502",
      "center_classification_date": "20230620",
      "report_date": "20230628",
      "code_info": "Serial numbers 338947, 338952, 338953, 338963, 338967, 338989, 338991, 338993, 338994, 338996, 338997, UDI-DI 07640121880513.",
      "more_code_info": ""
    }
  ]
}