  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Irvine",
      "address_1": "15353 Barranca Pkwy",
      "reason_for_recall": "Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.",
      "address_2": "",
      "product_quantity": "430",
      "code_info": "Model Number: CNS-6201    Software Version 02-26 only.  Serial Numbers:  00485,   00488,   00489,   00490,   00491,   00361,   00429,   00430,  00432,   00436,   00443,   00447,   00478,   00486,   00546,   00564,   00573,   00574,   00575,   00576,   00592,   00626,  00638,  00643,   00651,   00674,   00675,   00677,   00678,  00679,  00687.",
      "center_classification_date": "20140702",
      "distribution_pattern": "Worldwide Distribution - US in the state of  Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and  South Africa.",
      "state": "CA",
      "product_description": "CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201     For cardiac and vital signs monitoring for multiple patients.",
      "report_date": "20140709",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Nihon Kohden America Inc",
      "recall_number": "Z-1979-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68459",
      "termination_date": "20150428",
      "more_code_info": "",
      "recall_initiation_date": "20140603",
      "postal_code": "92618-2216",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}