  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saint Paul",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65798",
      "recalling_firm": "Boston Scientific CRM Corp",
      "address_1": "4100 Hamline Ave N",
      "address_2": "N/A",
      "postal_code": "55112-5700",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution in the state of Tennessee.",
      "recall_number": "Z-1979-2013",
      "product_description": "Boston Scientific, ENERGEN DR ICD, Model E143. The device is an Implantable Cardioverter Defibrillator.",
      "product_quantity": "1   (4 devices were implanted and are not part of this action)",
      "reason_for_recall": "Boston Scientific CRM manufacturing quality system recently discovered test artifacts (related to shock charge time) in a small number of implantable defibrillators (ENERGEN DR ICD model E143) which require further investigation.  While these devices successfully passed all manufacturing tests and met all requirements, they showed a slightly longer charge time than other devices during a manufactu",
      "recall_initiation_date": "20130715",
      "center_classification_date": "20130816",
      "termination_date": "20131112",
      "report_date": "20130828",
      "code_info": "Serial # 108063"
    }
  ]
}