  "results": [
    {
      "status": "Terminated",
      "city": "Trumbull",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83105",
      "recalling_firm": "CooperSurgical, Inc.",
      "address_1": "95 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "06611-1350",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution to state of:  AZ, CA, GA, IN, LA, MA, NC, NH, NJ, NY, OH, PA, VA, and WA..",
      "recall_number": "Z-1978-2019",
      "product_description": "Advincula Delineator with Ultem Plastic Soft Cup size 3.0 cm; Part No. AD750SC-KE30    The Advincula Delineator Uterine Manipulator is a single-use disposable device designed specifically for Total Laparoscopic Hysterectomy (TLH), Laparoscopic Assisted Vaginal Hysterectomy (LAVH), and/or Laparoscopic Supra-Cervical Hysterectomy (LSCH).",
      "product_quantity": "213 units in total",
      "reason_for_recall": "The soft, flexible cup of the Advincula Delineator may be unvalidated and could potentiall crack when subjected to excessive compressions during use, leading to serious health consequences, such as lacerations.",
      "recall_initiation_date": "20190614",
      "center_classification_date": "20190715",
      "termination_date": "20200519",
      "report_date": "20190724",
      "code_info": "Lot 021-19"
    }
  ]
}