      "country": "United States",
      "city": "Murrysville",
      "address_1": "1001 Murry Ridge Ln",
      "reason_for_recall": "The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the device s the air pathway and be ingested or inhaled by the user, and the PE-PUR foam  may off-gas certain chemicals.",
      "address_2": "",
      "product_quantity": "15,357,775 (10,307,186 in US, 5,039,748 OUS) in total",
      "code_info": "All Serial Numbers    Material Numbers: AC250S  AC550S  AH250HS  AH250S  AH260HS  AH260S  AH260TS  AH450HS  AH450S  AH550HS  AH550S  AH650HS  AH650S  AH750HS  AH750S  AU151HS  AU151S  AU251HS  AU251S  AU451HS  AU451S  AU551HS  AU551S  AU651HS  AU651S  AU751HS  AU751S  CA151HS  CA151S  CA251HS  CA251S  CA451HS  CA451S  CA551HS  CA551S  CA651HS  CA651S  CA751HS  CA751S  CN157HS  CN157S  CN257HS  CN257S  CN457HS  CN457S  CN557HS  CN557S  CN657HS  CN657S  CN757HS  CN757S  DE452HS  DE452S  DE552HS  DE552S  DE752HS  DE752S  DS150HS  DS150S  DS250HS  DS250S  DS450HS  DS450S  DS550HS  DS550S  DS650HS  DS650S  DS750HS  DS750S  FR355HS  FR355S  FR555HS  FR555S  HL454S  HL554S  IN151HS  IN151S  IN251HS  IN251S  IN451HS  IN451S  IN551HS  IN551S  IN651HS  IN651S  IN751HS  IN751S  JP256HS  JP456HS  JP456S  JP556HS  JP556S  JP656HS  JP756HS  JP756S  KP250HS  KP250S  KP450HS  KP450S  KP550HS  KP550S  KP750HS  KP750S  LA151HS  LA151S  LA251HS  LA251S  LA451HS  LA451S  LA551HS  LA551S  LA651HS  LA651S  LA751HS  LA751S  RIN551S  RJP556S  SH558HS  SH558S  UODS150S  UODS250S  UODS450S  UODS550S  UODS560S  UODS650S  UODS750S  RTODS650HS  151BR  LA151SBR  151P  157P  250P  251BR  LA251SBR  RBRLA251SBR  251P  RBRLA251HS  RBRLA251S  255P  FR255HS  FR255S  257P  355P  450P  451BR  LA451SBR  RBRLA451SBR  451P  RIN451S  RBRLA451HS  RBRLA451S  452P  UDE452S  454P  456P  RJP456S  457P  550P  551BR  LA551SBR  RBRLA551SBR  551P  RBRLA551HS  RBRLA551S  552P  UDE552S  554P  555P  556P  557P  651BR  LA651SBR  RBRLA651S  RBRLA651SBR  651P  657P  750P  751BR  LA751SBR  RBRLA751SBR  751P  RBRLA751HS  RBRLA751S  752P  756P  RJP756S  757P  AU151H  CA151  CA151H  CA451HC  CA451W  CA551C  CA551HC  CN157  DS150  DS450C  DS450D  DS550C  DS550HC  KP450C  KP450HC  KP550C  KP550HC  RDS150S  RDS250S  RDS450S  RDS520S  RDS550HS  RDS550S  RDS650S  RDS750HS  RDS750S  RIN751S  UDS150S  UDS250S  UDS450S  UDS520S  UDS550S  UDS650S  UDS750S  AHX200H11  AHX200H11C  AHX200H11W  AHX200S11  AHX200T11  AHX200T11C  AHX200T11W  AHX400H11  AHX400H11C  AHX400H11W  AHX400S11  AHX400T11  AHX400T11C  AHX400T11W  AHX500H11  AHX500H11C  AHX500H11W  AHX500S11  AHX500T11  AHX500T11C  AHX500T11W  AHX600H11  AHX600H11C  AHX600H11W  AHX600S11  AHX600T11  AHX600T11C  AHX600T11W  AHX700H11  AHX700H11C  AHX700H11W  AHX700S11  AHX700T11  AHX700T11C  AHX700T11W  ARX200H15  ARX200S15  ARX401H15  ARX401S15  ARX500H15  ARX500S15  ARX600H15  ARX600S15  ARX700H15  ARX700S15  AUX400H15  AUX400S15  AUX400T15  AUX400T15C  AUX500H15  AUX500S15  AUX500T15  AUX500T15C  AUX700H15  AUX700S15  AUX700T15  AUX700T15C  BLX200H15  BLX200S15  BLX201S15  BLX201S15C  BLX400H15  BLX400H15C  BLX400S15  BLX400S15C  BLX500H15  BLX500H15C  BLX500H15MQ  BLX500S15  BLX500S15C  BLX600H15  BLX600S15  BLX700H15  BLX700S15  BRX200S18  BRX400S18  BRX404S18  BRX500S18  BRX504S18  BRX600S18  BRX700S18  CAX200H12  CAX200S12  CAX200T12  CAX400H12  CAX400H12C  CAX400H12W  CAX400S12  CAX400T12  CAX400T12C  CAX400T12W  CAX500H12  CAX500H12C  CAX500H12W  CAX500S12  CAX500S12C  CAX500S12W  CAX500T12  CAX500T12C  CAX500T12W  CAX501H12  CAX501H12C  CAX501H12W  CAX501T12  CAX501T12C  CAX501T12W  CAX600H12  CAX600H12C  CAX600H12W  CAX600S12  CAX600T12  CAX600T12C  CAX600T12W  CAX700H12  CAX700H12C  CAX700H12W  CAX700S12  CAX700S12C  CAX700S12W  CAX700T12  CAX700T12C  CAX700T12W  CNX200H17  CNX200T17  CNX400H17  CNX400T17  CNX500H17  CNX500T17  CNX600H17  CNX600T17  CNX700H17  CNX700T17  DEX400S13  DEX500S13  DEX700S13  DNX500H11C  DNX500T11C  DSX200H11  DSX200H11C  DSX200H11W  DSX200S11  DSX200S11C  DSX200S11W  DSX200T11  DSX200T11C  DSX200T11W  DSX400H11  DSX400H11C  DSX400H11W  DSX400S11  DSX400T11  DSX400T11C  DSX400T11W  DSX500H11  DSX500H11C  DSX500H11CEP  DSX500H11EP  DSX500H11W  DSX500H11WEP  DSX500S11  DSX500S11C  DSX500S11W  DSX500T11  DSX500T11C  DSX500T11CEP  DSX500T11CMK  DSX500T11EP  DSX500T11W  DSX500T11WEP  DSX500T11WMK  DSX600H11  DSX600H11C  DSX600H11W  DSX600S11  DSX600T11  DSX600T11C  DSX600T11W  DSX700H11  DSX700H11C  DSX700H11CEP  DSX700H11EP  DSX700H11W  DSX700H11WEP  DSX700S11  DSX700S11C  DSX700S11W  DSX700T11  DSX700T11C  DSX700T11CEP  DSX700T11CMK  DSX700T11EP  DSX700T11W  DSX700T11WEP  DSX700T11WMK  EEX200H15  EEX200S15  EEX400H15  EEX400S15  EEX401H15  EEX401S15  EEX500H15  EEX500S15  EEX600H15  EEX600S15  EEX700H15  EEX700S15  ESX200H15  ESX200S15  ESX201S15  ESX201S15C  ESX400H15  ESX400S15  ESX401H15  ESX401H15C  ESX401S15  ESX401S15C  ESX401S15PR  ESX500H15  ESX500H15C  ESX500S15  ESX500S15C  ESX600H15  ESX600S15  ESX700H15  ESX700S15  EUX200H15  EUX200S15  EUX400H15  EUX400S15  EUX401H15  EUX401S15  EUX500H15  EUX500S15  EUX600H15  EUX600S15  EUX700H15  EUX700S15  FRX400S14  FRX400S14W  FRX401S14  FRX500S14  FRX500S14W  FRX501H14W  FRX501S14  FRX501S14W  FRX700S14  FRX700S14W  GBX200H15  GBX200S15  GBX400H15  GBX400S15  GBX401H15  GBX401S15  GBX500H15  GBX500S15  GBX600H15  GBX600S15  GBX700H15  GBX700S15  INX200H15  INX200S15  INX200T15  INX400H15  INX400S15  INX400T15  INX500H15  INX500S15  INX500T15  INX600H15  INX600S15  INX600T15  INX700H15  INX700S15  INX700T15  ITX200H15  ITX200S15  ITX400H15  ITX400H15W  ITX400S15  ITX500H15  ITX500H15W  ITX500S15  ITX600H15  ITX600S15  ITX700H15  ITX700S15  JPX400S16  JPX400S16C  JPX500S16  JPX500S16C  JPX501S16  JPX700S16  JPX700S16C  KRX200H15  KRX200S15  KRX200T15  KRX400H15  KRX400S15  KRX400T15  KRX500H15  KRX500S15  KRX500T15  KRX600H15  KRX600S15  KRX600T15  KRX700H15  KRX700S15  KRX700T15  LAX200H15  LAX200S15  LAX200T15  LAX400H15  LAX400S15  LAX400T15  LAX500H15  LAX500S15  LAX500T15  LAX600H15  LAX600S15  LAX600T15  LAX700H15  LAX700S15  LAX700T15  NDX200H15  NDX200S15  NDX400H15  NDX400S15  NDX500H15  NDX500H15C  NDX500S15  NDX500S15C  NDX600H15  NDX600S15  NDX700H15  NDX700S15  PPX100S10  PTX500H15PR  PTX500H15PRC  PTX500H15PRW  PWX401H15  PWX401S15  RDEX500S13  RDSX200S11  RDSX400S11  RDSX500S11  RDSX600S11  RDSX700S11  RGBX500H15  RGBX500S15  RGBX700S15  RJPX500S16  SHX500S11  TRX200H15  TRX200S15  TRX400H15  TRX400S15  TRX500H15  TRX500S15  TRX600H15  TRX600S15  TRX700H15  TRX700S15  UAX400S15  UAX500S15  UAX700S15  UCDSX400S11  UCDSX500S11  UCDSX700S11  UFRX500S14  VAX700H11  VAX700H11C  VAX700T11  VAX700T11C  ZAX200H15  ZAX200S15  ZAX400H15  ZAX400S15  ZAX500H15  ZAX500S15  ZAX600H15  ZAX600S15  ZAX700H15  ZAX700S15  DSX219S11  FRX500H14W  INX219S15  RBRX700S18  RDEX400S13  RDEX700S13  RFRG503S14  RFRX1130S14  RFRX501S14  RFRX700S14  RJPX400S16  RJPX700S16  VAX500H11  VAX500H11C  VAX500T11  VAX500T11C  ARX200H15C  ARX401H15C  ARX500H15C  ARX600S15C  ARX700S15C  LAX200H15C  LAX400H15C  LAX500H15C  LAX600H15C  LAX700H15C  BRX200H18C  BRX404H18C  BRX504H18C  BRX600H18C  BRX700H18C  UDSX200S11  UDSX400S11  UDSX500S11  UDSX600S11  UDSX700S11  ARG502S15  AUG400H15  AUG400S15  AUG500H15  AUG500S15  BLG500S15  BLG502S15  BRG502S18  CAG400H12  CAG400S12  CAG500H12  CAG500S12  CHG502S15  DEG500S13  DEG502S13  DSG400B11  DSG400H11  DSG400S11  DSG500H11  DSG500S11  ESG500S15  ESG502S15  EUG400S15  EUG500S15  EUG502S15  FRG501S14  FRG503S14  GBG500S15  GBG502S15  HKG502S15  IAG502S15  IDG502S15  ITG502S15  JPG500S16  LAG502S15  MYG502S15  NDG500S15  NDG502S15  PLG502S15  RDSG400S11  RDSG500S11  SAG502S15  SHG500S11  TRG502S15  TWG502S15  UDSG400S11  UDSG500S11  DEG400S13  RJPG500S16  DEV402  DEV402H  AUV501  AUV501H  BRV509  CAV501  CAV501H  CNV507H  DEV502  DEV502H  INDV501  INDV501H  INV501  INV501H  LAV501  LAV501H  TRV501  TRV501H  CA521HS  CA521NTS  CA521S  CA521TS  DS520HS  DS520S  DS520TS  CNV507HN1  RDEV402  RDEV502",
      "center_classification_date": "20210713",
      "distribution_pattern": "Global distribution",
      "state": "PA",
      "product_description": "SystemOne (Q-Series), DreamStation, \tDreamStation Go, Dorma 400, Dorma 500, REMstar SE Auto Noncontinuous Ventilators    The Philips Respironics E30 ventilator is intended to provide invasive and non-invasive ventilatory support for individuals with Respiratory Insufficiency. It is specifically for the care of adult and pediatric patients >7 years of age and >18kgs.",
      "report_date": "20210721",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Philips Respironics, Inc.",
      "recall_number": "Z-1974-2021",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "88058",
      "more_code_info": "",
      "recall_initiation_date": "20210614",
      "postal_code": "15668-8517",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}