  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Morrisville",
      "address_1": "9401 Globe Center Drive",
      "reason_for_recall": "A firmware bug causes incorrect values to be displayed for specific measured luminescence values.",
      "address_2": "Suite 140",
      "product_quantity": "868 in total",
      "code_info": "Material Number 30034301 and 30061442.    Serial numbers 904004848  807001164  1003005106  808056058  1006007069  1108006370  906001927  811005550  1002002705  1007007321  1110004989  1009005297  910008892  1005008056  1103008195  1009003991  912003102  1007005488  912003960  1008001833  808052054  1005002244  1103001072  909005818  1005008057  1010005921  1004003572  1103008191  1001006206  809004189  907004270  1006007071  1010001688  1010001689  1110004987  1112000054  1107006152  1103008189  902000004  810003913  908006454  1002002702  912003958  1009002602  1007003513  1001007642  1110006961  911008985  908001687  1011003332  909007802  901005834  901005835  1009007093  906000659  812003360  1110004988  1009002603  1001006205  1003005108  1005008055  1112000053  1001001581  1005008053  1003000532  909000550  911000226  910004186  1009003980  1110006962  907000033  907000720  908004917  1001007641  910008890  808052053  808054758  901005836  911006512  1002002703  1003005107  1106008462  912000502  1010005918  909000551  807003935  807003936  907004267  1007007320",
      "center_classification_date": "20150702",
      "distribution_pattern": "US Nationwide Distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NE NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, including DC",
      "state": "NC",
      "product_description": "Infinite M1000 in vitro diagnostic.      Product Usage:  The Tecan Infinite M1000 is a multifunctional monochromator based microplate reader that provides high performance for the vast majority of today\u0019s microplate applications and research. Besides offering absorbance and fluorescence intensity measurements, the Infinite M1000 allows the measurement of fluorescence polarization and luminescence. The Infinite M1000 is also robotic compatible and offers a built-in stacker option as well as a external injector module. The Infinite M1000 is intended as a general purpose laboratory instrument for professional use, supporting common microplates conforming to the ANS/SBS standards.",
      "report_date": "20150708",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Tecan US, Inc.",
      "recall_number": "Z-1969-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71352",
      "termination_date": "20170818",
      "more_code_info": "",
      "recall_initiation_date": "20150529",
      "postal_code": "27560",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}