      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83052",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed to the following US states:  AL, CA, CT, FL, GA, IL, KY, MD, MN, MO, OH, PA, TN, TX, VA, and Puerto Rico.",
      "recall_number": "Z-1967-2019",
      "product_description": "BD Microtainer Tube w/ BD Microgard Closure, K2EDTA additive, Model Numbers 365974 and 365975    Product Usage:  The Microtainer¿ Brand Tube with EDTA(K2) Reorder No. 365974, is used to collect, anticoagulate, transport and store skin puncture blood specimens for testing the following hematological parameters: WBC, RBC ,HGB, HCT, MCV, MCH, MCHC, Platelets, 5 part differential including : Neutrophils, Lymphocytes, Monocytes, Eosinophils, Basophils, Manual WBC Differentials, Reticulocytes, and Spun Hematocrit.",
      "product_quantity": "1707804",
      "reason_for_recall": "The products have been confirmed to have reduced or no additive within the tube reservoir.  The tubes may develop visible clots within the tube samples or micro clots that are not easily detected during visual inspection of the tubes. A clotted sample may lead to recollection of samples or, retesting of patients, resulting in delayed reporting of test results and patient treatment.",
      "recall_initiation_date": "20190603",
      "center_classification_date": "20190712",
      "termination_date": "20201208",
      "report_date": "20190724",
      "code_info": "Lot Numbers: 9017588  9017590  9017591  9017592  9017593  9017596  9017600  9028803  9045873  9046822  9050660  9052823  9056896"
    }
  ]
}