  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65797",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution in the state of CT.",
      "recall_number": "Z-1967-2013",
      "product_description": "The Discovery NM/CT 570c  CT Standalone Scan    The Discovery NM/CT 570c hybrid system is a combination of a cardiac dedicated nuclear imaging subsystem and LightSpeed GE CT subsystem. This system is intended for cardiac Nuclear Medicine (NM) diagnostic imaging applications, head and body CT diagnostic imaging applications, as well as attenuation correction and anatomical localization. The Discovery product is a SPECT-CT system. It consists of two back-to-back gantries, the NM Gantry, carrying solid state Cadmium Zinc Telluride (CZT) detectors, the CT Gantry, a single patient table, power distribution unit (PDU), NMV integrated power supply (IPS) , two operator consoles with two acquisition systems (Nuclear and CT), interconnecting cables and associated accessories (ECG gating, etc.).",
      "product_quantity": "10 Systems",
      "reason_for_recall": "GE Healthcare has recently become aware of an incident where the wrong part of the patient body was  scanned during a CT standalone exam with the Discovery NM/CT 570c",
      "recall_initiation_date": "20130410",
      "center_classification_date": "20130820",
      "termination_date": "20140121",
      "report_date": "20130828",
      "code_info": "Part numbers H3000SA UFC Cardiac SPECT System, H3001JA UFC Gantry, H3001JE UFC VCT Connecting Kit, H3001JF DISCOVERY NMV/CT 570C Connecting Kit."
    }
  ]
}