  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Deerfield",
      "address_1": "1 Baxter Pkwy",
      "reason_for_recall": "If the universal ingredient (UI) in an active configuration is changed using the Configuration Editor, a flush of the outlet pump tube will not be initiated by the software.  It could result in the original UI remaining in the tube which could be delivered into the next bag.",
      "address_2": "",
      "product_quantity": "1758",
      "code_info": "model numbers 1200-DY, 1200-DX, 1200-DYR  and 1200-DXR; EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX",
      "center_classification_date": "20140701",
      "distribution_pattern": "US Nationwide",
      "state": "IL",
      "product_description": "EXACTAMIX 1200 Compounder, model numbers 1200-DY, 1200-DX, 1200-DYR and 1200-DXR; EM1200 DY Display, Replacement EM 1200 DY Display.  EXACTAMIX 2400 Compounder model numbers 2400-DY, 2400-DX, 2400-DYR and 2400-DYX.  Baxa Corporation.  An automated pumping system that compounds multiple sterile ingredients into a finished solution in a single patient IV administration bag.",
      "report_date": "20140709",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Baxter Healthcare Corp.",
      "recall_number": "Z-1963-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68506",
      "termination_date": "20170222",
      "more_code_info": "",
      "recall_initiation_date": "20140610",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}