{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Ann Arbor",
      "address_1": "6200 Jackson Road",
      "reason_for_recall": "Terumo Cardiovascular Systems (Terumo CVS) received two complaints related to the Gas System for the Terumo Advanced Perfusion System 1. Both were reported instances of users experiencing difficulty adjusting the System 1 Gas System local flow control knob.",
      "address_2": "",
      "product_quantity": "1095 Units",
      "code_info": "catalog number: 801188 and Serial numbers: 0001-0006, 0008-00050, 00052-00123,00125-00152, 00154-0174,00176-00211, 00213-00241, 00245-00250, 00300-00356, 00358-00365, 00367-00449, 00451-00489, 00492-00522, 00524-00531, 00533-00649, 00651-00660, 00662-00825,0826-0830, 00831-01157, and 01160-01164.",
      "center_classification_date": "20120709",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  ARGENTINA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.",
      "state": "MI",
      "product_description": "Terumo Advanced Perfusion System 1  Electronic O2 Blender/Analyzer.    The electronic gas blender provides control and monitoring of the gas flow rate and oxygen content of the gas input to the oxygenator in the perfusion circuit. It also provides monitoring of the carbon dioxide if 95-5 (%oxygen/%carbon dioxide) is used as the blending gas.",
      "report_date": "20120718",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "recall_number": "Z-1962-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "61231",
      "termination_date": "20130219",
      "more_code_info": "",
      "recall_initiation_date": "20120622",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}