  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "San Jose",
      "address_1": "5900 Optical Ct",
      "reason_for_recall": "The gravity steam, Ethylene Oxide (EO) and STERRAD 100S sterilization methods are being removed from the \"Instructions for Use\"; due to the firm's evaluation that identified original validation was not done at worst case conditions.",
      "address_2": "",
      "product_quantity": "6388 units",
      "code_info": "All serial/lot numbers",
      "center_classification_date": "20120709",
      "distribution_pattern": "Worldwide Distribution--USA (nationwide)",
      "state": "CA",
      "product_description": "Stryker brand Arthroscope and Hardware Set Tray (sterilization tray); Model/Part Number: 233-032-116; Product is manufactured and distributed by Stryker Endoscopy, San Jose, CA      The Arthroscope and Hardware Set Tray is used to hold and protect surgical devices during the sterilization process. Specifically this tray is used for arthroscopes and arthroscopic hardware devices. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held.",
      "report_date": "20120718",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Endoscopy",
      "recall_number": "Z-1961-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62346",
      "termination_date": "20121023",
      "more_code_info": "",
      "recall_initiation_date": "20120619",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}