  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "East Walpole",
      "address_1": "333 Coney St",
      "reason_for_recall": "ADVIA Centaur Homocysteine dilution recovery on the diluent identified the percent recovery for the 1:10 patient sample dilutions lower than reported in the Instructions for Use (IFU).",
      "address_2": "",
      "product_quantity": "51 units",
      "code_info": "Lot #/Exp. Date:  88D1524 01 Jun 2014  88D3303 26 Nov 2013",
      "center_classification_date": "20140702",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and the countries of : Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Greece, Hungary,  Ireland, Italy Kazakhstan, Latvia, Liechtenstein, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland ,Turkey, Unit.Arab Emir and Vatican",
      "state": "MA",
      "product_description": "ADVIA Centaur HCY Diluent - 10mL Bottle Catalog number 09877493 (124533), SMN 10310370, and Ancillary Pack Catalog number 03302138 (124581), SMN 10378677.   Used on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur CP platforms.   For in vitro diagnostic use in the quantitative determination of total homocysteine (HCY) in serum or plasma.",
      "report_date": "20140709",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc",
      "recall_number": "Z-1960-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65562",
      "termination_date": "20160531",
      "more_code_info": "",
      "recall_initiation_date": "20130625",
      "postal_code": "02032-1516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}