  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85553",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distributions.",
      "recall_number": "Z-1959-2020",
      "product_description": "Siemens Cios Spin VA30-mobile X-Ray system,  Material # 10308194 - Product Usage: The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.",
      "product_quantity": "46 units",
      "reason_for_recall": "Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session.  If the system displays the error message  Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and  result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector",
      "recall_initiation_date": "20200420",
      "center_classification_date": "20200511",
      "termination_date": "20210510",
      "report_date": "20200520",
      "code_info": "Serial Numbers:  50124  50135  50131  50171  50129  50032  50033  50048  50050  50051  50073  50075  50081  50088  50098  50107  50109  50111  50120  50196  50102  50180  50136  50185  50096  50155  50091  50130  50127  50077  50078  50134  50122  50101  50092  50094  50166  50049  50192  50161  50125  50128  50132  50079  50160  50030"
    }
  ]
}