{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85553",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distributions.",
      "recall_number": "Z-1958-2020",
      "product_description": "Siemens Cios Alpha VA30-mobile X-Ray system,   Material # 11105200 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.",
      "product_quantity": "16 units",
      "reason_for_recall": "Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session.  If the system displays the error message  Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and  result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector",
      "recall_initiation_date": "20200420",
      "center_classification_date": "20200511",
      "termination_date": "20210510",
      "report_date": "20200520",
      "code_info": "Serial Numbers:  40057  40047  40048  40107  40108  40111  40120  40126  40134  40135  40032  40104  40075  40102  40124  40070"
    }
  ]
}