  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Vernon Hills",
      "address_1": "353 Corporate Woods Pkwy",
      "reason_for_recall": "E-Line Cutting Electrode package 8416.0305 contains E-Line Hook Electrode 8416.09 instead of E-Line Cutting Electrode 8416.03.",
      "address_2": "",
      "product_quantity": "5 boxes (5 electrodes each)",
      "code_info": "Product Number: 8416.0305; Batch Number: 1390827; Expiration Dates 10/15/2023 and 03/11/2024",
      "center_classification_date": "20190710",
      "distribution_pattern": "US nationwide distribution in the states of South Carolina, Rhode Island, and Missouri.",
      "state": "IL",
      "product_description": "E-Line Cutting Electrode 11.5Fr Mono 0 degrees, Sterile.  One electrode is placed in a plastic tray and then inserted into a Tyvek pouch.  Five packaged electrodes are placed in a white box. Product Usage: Resectoscopes are used for endoscopically controlled ablation of tissue. They are used, in combination with endoscopic accessories, for examination, diagnosis, and/or therapy in various medical disciplines such as urology and gynecology.",
      "report_date": "20190717",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Richard Wolf Medical Instruments Corp.",
      "recall_number": "Z-1957-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "82946",
      "termination_date": "20200804",
      "more_code_info": "",
      "recall_initiation_date": "20190523",
      "postal_code": "60061-3110",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}