  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Branchburg",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79899",
      "recalling_firm": "Roche Molecular Systems, Inc.",
      "address_1": "1080 US Highway 202 S",
      "address_2": "N/A",
      "postal_code": "08876-3733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of: CA, GA, HI, IN, MD, ME, MI, MN, NC, NY, PA, VI, and WI.  The products were distributed to the following foreign countries: Austria, Belgium, Brazil, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, France, Germany, Guatemala, Hong Kong, Hungary, India, Israel, Italy, Japan, Mexico, Netherlands, Norway, Poland, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, UAE, United Kingdom, Vietnam.",
      "recall_number": "Z-1956-2018",
      "product_description": "MagNA Pure 24 System, GMMI: 07290519001     Product Usage:  The MagNA Pure 24 System is a fully automated clinical sample extractor. It extracts nucleic acids from various sample types, sample tubes, and volume inputs for multiple downstream applications. The main features of the system are :   Scalable purification of 1 to 24 samples   Primary/secondary tube handling   Inventory and sample management via barcodes   Efficient order management via host The MagNA Pure 24 System offers 4 run types:   Purification (rack) run   Purification (cartridge) run   Post Elution run   Sample transfer run",
      "product_quantity": "163 units",
      "reason_for_recall": "Cross-contamination of samples has been reported when running the existing protocol on the system with the rack-based purification run.",
      "recall_initiation_date": "20180502",
      "center_classification_date": "20180530",
      "termination_date": "20190430",
      "report_date": "20180606",
      "code_info": "Device Identifier:  07613336106174"
    }
  ]
}