  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Brea",
      "address_1": "250 S Kraemer Blvd",
      "reason_for_recall": "Due to the lack of updated thermal components, these instruments may be more susceptible to the internal case temperature exceeding the 18 to 36 degrees Celsius operating range, causing SYS flags. The instrument will not report results until the internal case temperature returns to the operating range. This situation may cause a delay in the reporting of patient results.",
      "address_2": "",
      "product_quantity": "4 units total (1 unit in US)",
      "code_info": "Serial No. 505143, 504805, 501368, 504231",
      "center_classification_date": "20160613",
      "distribution_pattern": "US, Morocco, Venezuela",
      "state": "CA",
      "product_description": "Access 2 Immunoassay System, Remanufactured, Catalog No. 386220 is an invitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analytes concentrations found in human body fluids.",
      "report_date": "20160622",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Beckman Coulter Inc.",
      "recall_number": "Z-1956-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74212",
      "termination_date": "20160826",
      "more_code_info": "",
      "recall_initiation_date": "20160517",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}