  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Boulder",
      "address_1": "6325 Gunpark Dr",
      "reason_for_recall": "JustRight Surgical is recalling various lots of JustRight 5mm Stapler product because they may deliver incomplete staple formation and incomplete tissue closure.",
      "address_2": "Suite G",
      "product_quantity": "1452",
      "code_info": "Model Number JR-REL25-2.0  Lot numbers 75ED0114, 75JD1403, 75KD0304, 75KD0726, 75LD1224, 75AE0707, 75AE1406, 75BE1310, 75BE1313, 75CE2605, 75DE0802, and 75DE1524.",
      "center_classification_date": "20150702",
      "distribution_pattern": "Distribution - US Nationwide and Ireland.",
      "state": "CO",
      "product_description": "JustRight 5mm Reload, JR-REL25-2.0 Reload is packaged in cases of 12., placed in a tray and sealed with a Tyvek lid.    Product  Usage:  The JustRight 5mm Stapler is intended for use in abdominal, gynecologic, pediatric, and thoracic surgery for the resection, transection, and creation of anastomosis.",
      "report_date": "20150708",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Justright Surgical, LLC",
      "recall_number": "Z-1956-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71521",
      "termination_date": "20150826",
      "more_code_info": "",
      "recall_initiation_date": "20150611",
      "postal_code": "80301-3592",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}