      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80027",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Distributed to the states of CT, IL, MN, NV, SD, and WA.",
      "recall_number": "Z-1954-2018",
      "product_description": "Carina Ventilator, ventilator. Catalog no. 5704110.    Long-term ventilator for treatment in hospitals by a qualified medical personnel.",
      "product_quantity": "8 units",
      "reason_for_recall": "The valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in the affected devices.",
      "recall_initiation_date": "20180426",
      "center_classification_date": "20180529",
      "termination_date": "20180907",
      "report_date": "20180606",
      "code_info": "Serial nos. ASKM-0065  ASKM-0066  ASKM-0106  ASKM-0134  ASKM-0137  ASKM-0150  ASKM-0151  ASKN-0009"
    }
  ]
}