  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Denver",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80024",
      "recalling_firm": "Genesee BioMedical, Inc.",
      "address_1": "700 W Mississippi Ave Unit D5",
      "address_2": "N/A",
      "postal_code": "80223-4509",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US nationwide, South Korea, Australia, COSTA RICA, Taiwan,  New Zealand, Italy, Canada, Netherlands, Norway, Sweden, Switzerland, Israel, Japan, Egypt, South Africa, United Kingdom",
      "recall_number": "Z-1953-2018",
      "product_description": "Sternal Valve Retractors With Atrial and Aortic Rakes; Models SRCL-1, SRCL-AR, SRLP-1, SRLP-AR, ATR-1, ATR-AR, CSR-1 and CSR-AR.  For use in cardiac surgery.",
      "product_quantity": "444 units",
      "reason_for_recall": "Failure in a retractor blade due to corrosion. In the area of the blade where the bearing/button meets the blade body, corrosion can form, weakening the welded joint and may cause the bearing/button to break.",
      "recall_initiation_date": "20180226",
      "center_classification_date": "20180529",
      "report_date": "20180606",
      "code_info": "Serial Numbers: 1144  1145  707  709  959  404  1114  1151  520  621  784  1156  542  694  452  455  940  433  560  941  718  388  960  808  1045  1007  479  484  473  390  1067  1071  469  470  1054  977  626  967  494  1140  450  783  1194  853  920  273  368  1120  1121  1126  770  773  949  950  475  1021  1022  1159  877  680  769  370  1057  751  790  665  1092  1028  1118  981  585  586  582  993  1019  995  660  1094  1000  1001  1002  1003  1004  1034  1035  440  654  684  918  957  982  996  998  997  558  841  843  1100  238  731  369  854  526  187  489  568  695  992  677  1143  815  366  1182  1077  905  1044  1134  1046  564  1026  676  1040  1066  516  726  363  389  939  971  559  852  1080  566  599  600  522  1124  1150  628  629  909  796  990  903  916  917  1125  834  836  376  377  262  456  457  1115  490  1139  1160  774  789  822  833  476  403  1161  1164  1178  766  786  538  849  1078  1088  1093  451  471  782  215  356  807  1048  1073  937  1122  624  1225  591  1090  649  817  819  174  261  745  653  581  835  367  816  929  875  891  901  1123  804  842  933  578  434  1132  1133  923  787  375  912  924  1112  1136  1179  1180  619  620  637  675  708  801  947  571  572  573  1036  1085  951  1113  1165  1168  1170  1176  1177  562  651  797  814  826  828  829  926  931  935  540  551  552  679  803  794  553  678  802  811  812  813  821  686  954  1089  1138  595  598  721  757  907  908  910  911  913  1135  567  622  625  638  652  765  800  889  922  927  928  442  445  446  447  519  545  823  449  458  459  460  461  1006  1074  1076  362  430  570  948  255  477  1142  256  472  710  738  958  976  492  974  402  409  856  1025  1027  1079  1127  1128  1119  915  574  643  644  740  741  742  743  771  839  1037  956  1038  1039  1047  1049  1050  1051  1052  1053  1055  1056  1075  1091  1104  1105  513  521  525  557  575  584  590  627  674  735  785  809  810  827  844  846  847  876  887  925  932  934  942  943  945  970  983  994  596  632  720  724  725  744  754  762  969  968  1024  1116  1115  1154  855  878  636  640  641  495  491  425  1069  1070  900  904  767  768  664  761  763  764  699  706  696  698  648  645  646  647  630  569  527  499  357  359  360  272  443  444  840  845  848  1058  1059  1064  1065  1146  1153  1167  1169  1171  1172  1173  1175"
    }
  ]
}