  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Parsippany",
      "address_1": "10 Pomeroy Rd",
      "reason_for_recall": "Zimmer Biomet is initiating a recall of specific lots of foil pouched, sterile Trabecular Metal implants that were manufactured between April 2011 and October 2012. The foil pouches are part of a dual barrier system to provide sterile integrity.",
      "address_2": "",
      "product_quantity": "19,884 total",
      "code_info": "00588606510 TM MONO TIB STR GRN SZ 5 10MM 61769554 61780786 61803934  00588606512 TM MONO TIB STR GRN SZ 5 12MM 61769555 61785358  00588606514 TM MONO TIB STR GRN SZ 5 14MM 61746710 61769556 61785359 61803935  00588606517 TM MONO TIB STR GRN SZ 5 17MM 61739874 61780787 61785360 61792655 61803936  61809005  00588606610 TM MONO TIB STR GRN SZ 6 10MM 61770839 61780793 61785374 61809020 61820718  61839382  00588606612 TM MONO TIB STR GRN SZ 6 12MM 61739882 61733397 61759778 61770840 61785375  61809018 61844376  00588606614 TM MONO TIB STR GRN SZ 6 14MM 61739885 61759779 61780795 61785376 61809015  00588606617 TM MONO TIB STR GRN SZ 6 17MM 61739883 61785377 61820719  00588606710 TM MONO TIB STR BLUE SZ 7 10MM 61769582 61792678 61797213 61851763  00588606712 TM MONO TIB STR BLUE SZ 7 12MM 61785386 61792681 61797214 61857235  00588606714 TM MONO TIB STR BLUE SZ 7 14MM 61780797 61785387 61809058  00588606717 TM MONO TIB STR BLUE SZ 7 17MM 61785388 61814249  00588607510 TM MONO TIB STR GRN SZ 5 10MM 61785366 61803928 61820713 61844346  00588607512 TM MONO TIB STR GRN SZ 5 12MM 61770823 61792661 61803929 61851753  00588607514 TM MONO TIB STR GRN SZ 5 14MM 61739865 61785369 61803931  00588607517 TM MONO TIB STR GRN SZ 5 17MM 61785370",
      "center_classification_date": "20160610",
      "distribution_pattern": "Nationwide and Foreign.",
      "state": "NJ",
      "product_description": "TM MONO TIB STR GRN SZ 5 - 00588606510  TM MONO TIB STR GRN SZ 5 12MM - 00588606512  TM MONO TIB STR GRN SZ 5 14MM - 00588606514   TM MONO TIB STR GRN SZ 5 17MM  - 00588606517  TM MONO TIB STR GRN SZ 6 10MM - 00588606610     TM MONO TIB STR GRN SZ 6 12MM - 00588606612     TM MONO TIB STR GRN SZ 6 14MM - 00588606614  TM MONO TIB STR GRN SZ 6 17MM - 00588606617  TM MONO TIB STR BLUE SZ 7 10MM - 00588606710  TM MONO TIB STR BLUE SZ 7 12MM - 00588606712  TM MONO TIB STR BLUE SZ 7 14MM - 00588606714  TM MONO TIB STR BLUE SZ 7 17MM - 00588606717   TM MONO TIB STR GRN SZ 5 10MM - 00588607510  TM MONO TIB STR GRN SZ 5 12MM - 00588607512  TM MONO TIB STR GRN SZ 5 14MM - 00588607514   TM MONO TIB STR GRN SZ 5 17MM - 00588607517",
      "report_date": "20160622",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Trabecular Metal Technology, Inc.",
      "recall_number": "Z-1953-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "73536",
      "termination_date": "20171129",
      "more_code_info": "",
      "recall_initiation_date": "20160415",
      "postal_code": "07054-3722",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}