  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Southborough",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79762",
      "recalling_firm": "Gyrus Acmi, Incorporated",
      "address_1": "136 Turnpike Rd",
      "address_2": "N/A",
      "postal_code": "01772-2118",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) Distribution to the states of :   AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, and WY., and to the countries of :  Japan, Canada, Mexico, Australia, and EU.",
      "recall_number": "Z-1952-2018",
      "product_description": "Gyrus ACMI Reusable Active Cord for Bovie/ValleyLab Generators, Catalog No C650-129A (00821925008700)    The cord consists of a single conductor stranded wire terminated with plugs that connect to a generator on one end and a plug that connects to a Gyrus ACMI working element on the other end. They are provided non-sterile and are reusable. The cord is designed to be used with Gyrus ACMI GOCUS 640, Bovie, and other general purpose monopolar electrosurgical generators and appropriate electrodes and endoscopes.",
      "product_quantity": "6447",
      "reason_for_recall": "There is a potential for the cord to spark and cause a fire.",
      "recall_initiation_date": "20180417",
      "center_classification_date": "20180529",
      "termination_date": "20230918",
      "report_date": "20180606",
      "code_info": "All lots manufactured before December 7, 2017    UDI (00821925008700)"
    }
  ]
}