  "results": [
    {
      "status": "Ongoing",
      "city": "Valencia",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98678",
      "recalling_firm": "Advanced Bionics, LLC",
      "address_1": "28515 Westinghouse Pl",
      "address_2": "",
      "postal_code": "91355-1398",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "",
      "distribution_pattern": "Worldwide - U.S Nationwide distribution including in the state of AK, AL, AR, AZ, CA, CO, CT, CD, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Canada, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, China, Chile, Croatia, Cuba, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guam, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kazakhstan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Trinidad and Tobago, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1950-2026",
      "product_description": "Brand Name: HiResolution\" Bionic Ear System  Product Name: M Zn-Air Battery Pak  Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; CI-5501-240; CI-5501-250; CI-5501-260; CI-5501-270; CI-5501-280  Software Version: Not Applicable  Product Description: The M Zn-Air Battery Pak is a disposable power source for the Naida\" CI M and Sky CI\" M Sound Processors. The type of disposable battery to be used with the M Zn-Air Battery Pak is a 675 power implant battery (P675). Regular 675 hearing instrument batteries will not be powerful enough for a cochlear implant. The suggested manufacturer and battery is PowerOne P675 Implant Plus battery.    The M Zn-Air Battery Pak is available with non-tamperproof and tamperproof cartridges.  Component: Not Applicable",
      "product_quantity": "45173",
      "reason_for_recall": "The non-tamperproof battery door supplied with the sound processor fails to conform to IEC 60601-1-11:2015 by not requiring the use of a tool to replace the button/coin cell battery.",
      "recall_initiation_date": "20260323",
      "center_classification_date": "20260427",
      "report_date": "20260506",
      "code_info": "Lot Code: All lots manufactured up to February 3, 2026.  Refer to Exhibit_0006",
      "more_code_info": ""
    }
  ]
}