      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "Failed to meet established requirements for sterile barrier integrity due to a breach of the inner cavity.  If the outer tray is compromised in addition to the inner tray a periprosthetic infection may occur. The risk associated with a periprosthetic infection is a revision or potential multi-stage revision to treat the infection.",
      "address_2": "",
      "product_quantity": "352",
      "code_info": "item 6215-12-185  Lots:  7884165  7884736  7884737  7884738  7890330  11011448  11012660  60394169  60415420  60515793  60564530  60601129  60614014  60797044  60803056  60841848  60922655  60922656  60927853  60989112  61017152  61104192  61135872  61200531  61284571  61366919  61464308  61505310  61533025  61566733  61742151  61780361  61784264  61834192  61834778  61840307  61930376  61947884  61947885  62097910  62146925  62195165  62249431  62281270  62305675  62317987  62358643  62448678  62536901  62536902  62536903  62546527  62560372  62587577  62610989  62624715  62624716  62624717  62624718  62709664  62709665  62709667  62709670",
      "center_classification_date": "20150701",
      "distribution_pattern": "US Nationwide Distribution, including DC and the states of AL, AZ, CA, CO, FL, GA, IL, IN, KS, LA, MD, MI, MN, MO, MT, NC, ND, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, and WY.",
      "state": "IN",
      "product_description": "Natural Knee II Revision Femoral Stem, 125mm, item 6215-12-185.  packaged in configuration 803/791.  Knee Prosthesis component.",
      "report_date": "20150708",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer, Inc.",
      "recall_number": "Z-1948-2015",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "71531",
      "termination_date": "20150701",
      "more_code_info": "",
      "recall_initiation_date": "20141002",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}