  "results": [
    {
      "status": "Terminated",
      "city": "Addison",
      "state": "TX",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85106",
      "recalling_firm": "Inlightened Partners LLC",
      "address_1": "16200 Addison Rd Ste 105",
      "address_2": "N/A",
      "postal_code": "75001-5302",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of CA, TX, NC, ID, CO, VA, FL, VT, AK, DC, MA, IL, AZ, MN, IA, TN, WI, WA, OK, MI, OH, GA, NY, IN, PA, NV, NJ, AR, CO , LA, MT, OR, SD, ND, MD, UT, MO, KY, NM, KS, AS, NE, WY, PR, MS and the country of Canada.",
      "recall_number": "Z-1943-2021",
      "product_description": "inLight Medical LED Pads used with Polychromatic light technology system",
      "product_quantity": "926",
      "reason_for_recall": "LED pads with cord connectors that have v-shaped notches may not have been assembled properly. This issue could lead to a power interruption, causing the controller and power supply to shut down.",
      "recall_initiation_date": "20190604",
      "center_classification_date": "20210624",
      "termination_date": "20210625",
      "report_date": "20210630",
      "code_info": "All serial numbers with cord connectors that have v-shaped notches"
    }
  ]
}