  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65133",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "Worldwide distribution: USA (nationwide)  including DC and Puerto Rico and countries:  ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIDJAN, BELARUS, BELGIUM, BOLIVIA, BOSNIA-HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, FINLAND, FRANCE, GEORGIA, GERMANY, GREAT BRITAIN, GREECE, GUADELOUPE, GUATEMALA, GUYANA, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN,JORDAN, KAZAKHSTAN, KENYA, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIAN, LUXEMBOURG, MALAYSIA, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PANAMA, PARAGUAY, PERU. PHILIPPINES, POLAND, PORTUGAL, REPUBLIC OF MACEDONIA, REUNION, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAK REPUBLIC, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SUDAN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TUNISIA, TUR,EKY, UKRAINE, UNITED ARAB EMIRATES, URUGUAY, VENEZUELA, VIETNAM, YUGOSLAVIA, MALAWI, MAURITANIA, NIGERIA, UKRAINE, IVORY COST, OMAN, ZAMBIA, UZBEKISTAN, ESTONIA, IRAQ, and JAMAICA.",
      "recall_number": "Z-1943-2013",
      "product_description": "SIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY   GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM  GE SIGNA 3.0T WITH EXCITE MR SYSTEM   GE SIGNA EXCITE 1.5T MR SYSTEM  GE SIGNA EXCITE 3.0T MR SYSTEM Signa HDx   Signa HDxt   Signa HDi   Signa Vibrant    GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM (i.e. Signa Ovation 5 and Ovation HD)   GE 1.5T SIGNA HDe MR SYSTEM   GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM   GE 0.7T SIGNA OPENSPEED (4) WITH EXCITE MR SYSTEM   GE 0.7T SIGNA OPENSPEED (5) WITH EXCITE MR SYSTEM    The GE Signa Infinity line of products are whole body magnetic resonance scanners designed to support high resolution, high signal-to-noise ratio and short scan times. The Signa Infinity System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa Infinity systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.",
      "product_quantity": "7016 (2919 US; 4097 OUS)",
      "reason_for_recall": "Screen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue  due to a potential patient data mismatch.While the issue has only been observed at one customer site  on the Signa HDxT 1.5T system, internal investigation revealed that the issue also impacts multiple GE healthcare MR Systems.",
      "recall_initiation_date": "20111121",
      "center_classification_date": "20130809",
      "termination_date": "20141022",
      "report_date": "20130821",
      "code_info": "Ref. # 60808    \"Products: Signa Infinity MR System with EXCITE Technology, Signa EXCITE 1.5T and Signa EXCITE 3.0T, Signa 3T with EXCITE, Signa 3T Infinity with EXCITE, GE 1.5T AND 3.0T SIGNA HDX MR SYSTEM (Waukesha, WI)  FMI 60808\"  Mfg Lot or Serial #  00000298045MR6  00000237944MR4  00000278301MR7  00000257724MR5  00000029079MR2  00000280678MR4  00000286982MR4  00000283664MR1  00000263686MR8  00000238343MR8  00001011BD6649  00000105XF5011  00000263832MR8  00000256696MR6  00000284409MR0  00000265824MR3  00000284407MR4  00000252930MR3  00000301212MR7  00000277495MR8  00000238364MR4  00000230664MR5  00000262150MR6  00000275961MR1  00000231910MR1  00000284466MR0  00000276678MR0  00000006583YR9  00000262101MR9  00000283815MR9  00000268471MR0  00000028886MR2  00000297726MR2  00000277916MR3  00000290478MR7  00000247217MR3  00000245819MR8  00000284864MR6  00000287026MR9  00000238355MR2  00000265183MR4  00000266460MR5  00000251094MR9  00000266825MR9  00000265718MR7  00000289270MR1  00000288702MR4  00000286851MR1  00000281374MR9  000000HDMR2685  00000HDMR20256  00000236703MR5  00000297201MR6  00000234797MR9  00000265066MR1  00000278766MR1  00000255610MR8  00000280252MR8  00000281156MR0  00000011742YR4  00000011570YR9  00000279479MR0  00000009057YR1  00000284902MR4  00000287649MR7  00000239864MR2  00000235678MR0  00000241145MR2  00000255599MR3  00000028959MR7  00000235242MR5  00000257752MR6  00000256990MR3  00000275647MR6  00000246615MR9  00000300674MR9  00000011579YR0  00000286249MR8  00000279921MR1  00000011882YR8  00000262908MR7  00000271646MR2  00000290192MR4  00000254485MR6  00000259530MR4  00000288192MR8  00000245803MR2  00000246622MR5  00000278624MR2  00000261879MR1  00000937BD5ED5  00000301020MR4  00000289016MR8  00000229365MR2  00000232415MR0  00000256228MR8  00000264529MR9  00000298718MR8  00000284386MR0  00000276588MR1  00000242558MR5  00000253635MR7  00000256984MR6  00000287626MR6  00000280089MR4  00000256699MR0  00000260698MR6  00000262933MR5  00000272896MR2  00000272890MR5  00000237064MR1  00000005232YR4  00000233272MR4  00000299162MR8  00000229474MR2  00000248485MR5  00000268524MR6  00000237059MR1  00000285218MR4  00000011910YR7  00000266754MR1  00000286785MR1  00000301356MR2  00000299828MR4  00000241868MR9  00000273672MR6  00000262458MR3  00000228289MR3  00000257732MR8  00000286701MR8  00000278873MR5  00000244845MR4  00000291805MR0  00000283949MR6  00000230694MR2  00000242548MR6  00000241866MR3  00000289771MR8  00000292635MR0  00000278669MR7  00000279078MR0  00000235690MR5  00000286843MR8  00000270976MR4  00000269336MR4  00000260699MR4  00000012244YR0  00000266828MR3  00000260864MR4  00000272284MR1  00000273632MR0  00000279874MR2  00000284640MR0  00000258090MR0  00000300909MR9  00000249768MR3  00000272283MR3  00000286016MR1  00000283849MR8  00000296144MR9  00000279405MR5  00000287081MR4  00000302511MR1  00000251241MR6  00000283682MR3  00000274745MR9  00000234794MR6  00000255208MR1  00000262151MR4  00000279483MR2  00000933BD5E06  00000239042MR5  00001008BD6517  00000235236MR7  00000280524MR0  00000267180MR8  00000284599MR8  00000270386MR6  00000HDMR20043  00000287846MR0  00000268283MR9  00000269380MR2  00000235680MR6  00000251800MR9  00000294406MR4  00000245809MR9  00000255571MR2  00000264617MR2  00000265964MR7  00000234855MR5  00000HDMR20069  00000273541MR3  00000278688MR7  00000277575MR7  00000284709MR3  00000011508YR9  00000277494MR1  00000248159MR6  00000282788MR9  00000239241MR3  00000278291MR0  00000278391MR8  00000919BD5B98  00000265720MR3  00000231911MR9  00000241865MR5  00000261877MR5  00000255569MR6  00000950BD6297  00000296292MR6  00000243702MR8  00000010163YR4  00000260470MR0  00000290052MR0  00000012162YR4  00000288015MR1  00000280000MR1  00000266799MR6  00000275583MR3  00000272627MR1  00000231913MR5  00000300799MR4  00000276634MR3  00000244846MR2  00000251231MR7  00000024170MR6  00000243700MR2  00000262464MR1  00000008345YR1  00000239857MR6  00000260626MR7  00000275653MR4  00000281157MR8  00000270462MR5  00000281271MR7  00000261328MR9  00000012023YR8  00000283377MR0  00000008059YR8  00000243685MR5  00000011589YR9  00000285927MR0  00000278193MR8  00000287940MR1  00000237949MR3  00000282547MR9  00000280775MR8  0000267533MR8A  00000265827MR6  00000286853MR7  00000241861MR4  00000265596MR7  00000256704MR8  00000266073MR6  00000254665MR3  00000268049MR4  00000903BD55CD  00000242585MR8  00000261068MR1  00000297656MR1  00000243249MR0  00000262735MR4  00000269045MR1  00000286526MR9  00000271753MR6  00000281937MR3  00000223456MR5  151  000S0933BD5DFD  00000266214MR6  00000294524MR4  00000290130MR4  00000291370MR5  00000025636MR4  00000241514MR9  00000237936MR0  00000292922MR2  00000011550YR1  00000285775MR3  00000915BD5AED  00000282371MR4  00000260011MR2  00000287864MR3  00000006556YR5  00000301463MR6  00000287826MR2  00001048XF501C  00000258419MR1  00000255211MR5  00000244850MR4  00000287767MR8  00000234847MR2  00000249758MR4  00000263792MR4  00000233267MR4  00000280436MR7  00000288096MR1  00000231893MR9  00000029128MR0  00000260017MR9  00000265939MR9  00000263972MR2  00000281826MR8  00000234843MR1  00000260462MR7  00000029187MR8  00000299416MR8  00000277784MR5  00000263507MR6  00000291750MR8  00000252437MR9  00000271051MR5  00000237968MR3  00000232419MR2  00000011585YR7  00000011567YR3  00000218753MR2  00000HDMR10899  00000235237MR5  00000252422MR1  00000288852MR7  00000248132MR3  00000290370MR6  00000275967MR8  00000289763MR5  00000265062MR0  00000280249MR4  000000HDMR1086  00000241143MR7  00000237932MR9  00000028674MR1  00000241137MR9  00000253153MR1  00000255612MR4  00000283865MR4  00000278555MR8  00000263313MR9  00000011711YR9  00000919BD5BB0  00000286091MR4  00000284389MR4  00000279753MR8  00000011521YR2  000000HDMR2984  00000269047MR7  00000275648MR4  00000258403MR5  00000256367MR4  00000012355YR4  00000247727MR1  00000294694MR5  00000292964MR4  00000282393MR8  00000261069MR9  00000278678MR8  00000272891MR3  00000301858MR7  00000238867MR6  00000301804MR1  00000273195MR8  00000283321MR8  00000280462MR3  00000011015YR5  00000271561MR3  00000301565MR8  00000272101MR7  00000250019MR7  00000286886MR7  00000278769MR5  00000258089MR2  00000236831MR4  00000011793YR7  00000022936MR8  00000286767MR9  00000276591MR9  00000287398MR2  00000903BD55EB  00000233430MR8  00000252378MR5  00000262842MR8  00000247213MR2  00000280696MR6  00000234673MR2  00000249770MR9  00000241863MR0  00000238361MR0  00000254188MR6  00000264553MR9  00000255197MR6  00000246623MR3  00000266698MR0  00000259122MR0  00000260863MR6  00000235588MR1  00000242588MR2  00000237946MR9  00000285622MR7  00000245172MR2  00000265613MR0  00000238344MR6  00000257755MR9  00000233437MR3  00000251697MR9  00000233299MR7  00000248478MR0  00000284321MR7  00000245171MR4  00000268318MR3  00000267849MR8  00000298841MR8  00000291802MR7  00000254184MR5  00000247225MR6  0000000000XXXX  00000242549MR4  00000288847MR7  00000283736MR7  00000248167MR9  00000249226MR2  00000233434MR0  00000253166MR3  00000239045MR8  00000236833MR0  00000253631MR6  00000293744MR9  00000281232MR9  00000287652MR2  00000256217MR1  00000241101MR5  00000272563MR8  00000291492MR7  00000268137MR7  00000279056MR6  00000284012MR2  00000272880MR6  00000296858MR4  00000253827MR0  00000259131MR1  00000263795MR7  00000251726MR6  00000287427MR9  00000256687MR5  00000258372MR2  00000239528MR3  00000258087MR6  00000254198MR5  00000254183MR7  00000262048MR2  00000248162MR0  00000233443MR1  00000298384MR9  00000251716MR7  00000258079MR3  00000281800MR3  00000245302MR5  00000236826MR4  00000233300MR3  00000256218MR9  00000237700MR0  00000297705MR6  00000237002MR1  00000240107MR3  00000279455MR0  00000245177MR1  00000270516MR8  00000259421MR6  00000280747MR7  00000253624MR1  00000264467MR2  00000230098MR6  00000264799MR8  00000243695MR4  00000273015MR8  00000259424MR0  00000255214MR9  00000264489MR6  00000255592MR8  00000250433MR0  00000290767MR3  00000291932MR2  00000254668MR7  00000248135MR6  00000276654MR1  00000260675MR4  00000287257MR0  00000234858MR9  00000292213MR6  00000299517MR3  00000237698MR6  00000231892MR1  00000238360MR2  00000271645MR4  00000291455MR4  00000278812MR3  00000237703MR4  00000257746MR8  00000252438MR7  00000254192MR8  00000299483MR8  00000261479MR0  00000241133MR8  00000241100MR7  00000273443MR2  00001101XF5089  00000253627MR4  00000272883MR0  00000263261MR0  00000253164MR8  00000248483MR0  00000246620MR9  00000244849MR6  00000256669MR3  00000245175MR5  00000246625MR8  00000271652MR0  00000300503MR0  00000266800MR2  00000271861MR7  00000238362MR8  00000236699MR5  00000263574MR6  00000261382MR6  00000298980MR4  00000290549MR5  00000300669MR9  00000288872MR5  00000239033MR4  00000259524MR7  00000299622MR1  00000248158MR8  00000264012MR6  0000024925BMR5  00000242553MR6  00000255204MR0  00000241518MR0  00000267671MR6  00000291999MR1  00000246609MR2  00000258418MR3  00000243243MR3  00000291205MR3  00000277777MR9  00000257716MR8  00000292210MR2  00000249276MR7  00000290604MR8  00000265102MR4  00000232406MR9  00000241099MR1  00000251725MR8  00000259941MR3  00000290148MR6  00000259575MR9  00000276419MR9  00000250439MR7  00000253150MR7  00000274162MR7  00000300896MR8  00000257415MR0  00000300753MR1  00000290333MR4  00000255215MR6  00002522296MR9  00000283154MR3  00000252911MR3  00000279055MR8  00000290200MR5  00000239868MR3  00000243698MR8  00000259916MR5  00000298852MR5  00000279409MR7  00000257443MR2  00000239538MR2  00000251709MR2  00000292051MR0  00000235695MR4  00000251060MR0  00000273294MR9  00000260465MR0  00000229479MR1  00000290546MR1  00000266696MR4  00000231898MR8  00000239027MR6  00000264464MR9  00000298853MR3  00000237939MR4  00000256700MR6  00000250434MR8  00000271862MR5  00000259521MR3  00000252423MR9  00000259529MR6  00000281945MR6  00000285427MR1  00000256214MR8  00000262506MR9  00000263577MR9  00000250447MR0  00000236625MR0  00000262592MR9  00000233263MR3  00000236503MR9  41404  00000250034MR6  00000298376MR5  00000265096MR8  00000253839MR5  00000233298MR9  00000256708MR9  00000264465MR6  00000256226MR4  00000264796MR4  00000245174MR8  00000237948MR5  00000238357MR8  00000260104MR5  00000231909MR3  00000233297MR1  00000239508MR5  00000289541MR5  00000238353MR7  00000252432MR0  00000262049MR0  00000262589MR5  00000287107MR7  00000247219MR9  00000261383MR4  00000260862MR8  00000280697MR4  00000298823MR6  00000237692MR9  00000233433MR2  00000238179MR6  00000253147MR3  00000250209MR4  00000250451MR2  00000257429MR1  00000233444MR9  00000271866MR6  00000284601MR2  00000248165MR3  00000258367MR2  00000264794MR9  00000271868MR2  00000261731MR4  00000280693MR3  00000284322MR5  00000251728MR2  00000250192MR2  00000291840MR7  00000243691MR3  00000241725MR1  00000249229MR6  00000300466MR0  00000252660MR6  00000259528MR8  00000249275MR9  00000272882MR2  00000261733MR0  00000237070MR8  00000240103MR2  00000291706MR0  00000230103MR4  00000235688MR9  00000241720MR2  00000273445MR7  00000291754MR0  00000260860MR2  00000251797MR7  00000254674MR5  00000257740MR1  00000271864MR1  00000291787MR0  00000252933MR7  00000288915MR2  00000292391MR0  00000271651MR2  00000256220MR5  00000258096MR7  00000259128MR7  00000287595MR3  00000287614MR2  00000024118MR7  00000257729MR4  00000260016MR1  00000245807MR3  00000253825MR4  00000251714MR2  00000251639MR1  00000250216MR9  00000258399MR5  00000290573MR5  00000231890MR5  00000257750MR0  00000268315MR9  00000230665MR2  00000282102MR3  00000254186MR0  00000259426MR5  00000253639MR9  00000276570MR9  00000239246MR2  00000263259MR4  00000245804MR0  00000239865MR9  00000272879MR8  00000230104MR2  00000288003MR7  00000239010MR2  00000291746MR6  00000250021MR3  00000259962MR9  00000291721MR9  00000273019MR0  00000238182MR0  00000262905MR3  00000251718MR3  00000258421MR7  00000251727MR4  00000254199MR3  00000230683MR5  00000237077MR3  00000230681MR9  00000253161MR4  00000229471MR8  00000243693MR9  00000235241MR7  00000273016MR6  00000285429MR7  00000253842MR9  00000291419MR0  00000261470MR9  00000258095MR9  00000250020MR5  00000250219MR3  00000273296MR4  00000267104MR8  00000242561MR9  00000262734MR7  00000283920MR7  00000229476MR7  00000255213MR1  00000243179MR9  00000232421MR8  00000261474MR1  00000278752MR1  00000256697MR4  00000264551MR3  00000258081MR9  00000257748MR4  00000276267MR2  00000262508MR5  00000246617MR5  00000247229MR8  00000294461MR9  00000294527MR7  00000298978MR8  00000247737MR0  00000284061MR9  00000247228MR0  00000249658MR6  00000285428MR9  00000239534MR1  00000256698MR2  00000245176MR3  00000273442MR4  00000250023MR9  00000255597MR7  00000244852MR0  00000256985MR3  00000259942MR1  00000284622MR8  00000250751MR5  00000253148MR1  00000254500MR2  00000291856MR3  00000274103MR1  00000277776MR1  00000290536MR2  00000247729MR7  00000233426MR6  00000274051MR2  00000231905MR1  00000259586MR6  00000298383MR1  00000233268MR2  00000291733MR4  00000260105MR2  00000263499MR6  00000255591MR0  00000260467MR6  00000251243MR2  00000253638MR1  00000264550MR5  00000296313MR0  00000290032MR2  00000025169MR1  00000024847MR2  00000264792MR3  00000276653MR3  00000233436MR5  00000235693MR9  00000237071MR6  00000238167MR1  00000239234MR8  00000242173MR3  00000239227MR2  00000241869MR7  00000252297MR7  00000258411MR8  00000255203MR2  00000255206MR5  00000255209MR9  00000254662MR0  00000259119MR6  00000250208MR6  00000261478MR2  00000263500MR1  00000262154MR8  00000262659MR6  00000263947MR4  00000277774MR6  00000273444MR0  00000273297MR2  00000281290MR7  00000264455MR7  00000291887MR8  00000285430MR5  00000274106MR4  00000280695MR8  00000285431MR3  00000260674MR7  00000254195MR1  00000274053MR8  00000233435MR7  00000262103MR5  00000234844MR9  00000234798MR7  00000236829MR8  00000261384MR2  00000260468MR4  00000231899MR6  00000260859MR4  00000253831MR2  00000284078MR3  00000257425MR9  00000258410MR0  00000233301MR1  00000259423MR2  00000234862MR1  00000260019MR5  00000276420MR7  00000254190MR2  00000258365MR6  00000262594MR5  00000283363MR0  00000242546MR0  00000259525MR4  00000247224MR9  00000238169MR7  00000252409MR8  00000251715MR9  00000239236MR3  00000267852MR2  00000281264MR2  00000281859MR9  00000255607MR4  00000232426MR7  00000262443MR5  00000276352MR2  00000276268MR0  00000276651MR7  00000290516MR4  00000238173MR9  00000282481MR1  00000263498MR8  00000259567MR6  00000238864MR3  00000258078MR5  00000260691MR1  00000260481MR7  000A0000012MR2  00000273265MR9  00000262066MR4  00000263969MR8  00000271173MR7  00000258099MR1  00000250022MR1  00000280418MR5  00000269437MR0  00000264125MR6  00000271054MR9  00000280096MR9  00000280465MR6  00000281886MR2  00000280433MR4  00000265177MR6  00000265870MR6  00000264407MR8  00000274691MR5  00000268285MR4  00000265173MR5  00000234848MR0  00000272348MR4  00000258416MR7  00000266713MR7  00000273196MR6  00000284315MR9  00000270975MR6  00000273228MR7  00000270569MR7  00000264007MR6  00000276732MR5  00000262068MR0  00000265184MR2  00000258393MR8  00000234851MR4  00000280892MR1  00000269044MR4  00000245298MR5  00000280464MR9  00000263572MR0  00000274998MR4  00000281796MR3  00000280674MR3  00000278196MR1  00000249272MR6  00000275015MR6  00000258370MR6  00000262901MR2  00000262442MR7  00000272339MR3  00000275595MR7  00000270244MR7  00000281935MR7  00000270383MR3  00000278626MR7  00000270250MR4  00000262459MR1  00000282909MR1  00000263942MR5  00000283074MR3  00000259572MR6  00000262444MR3  00000274891MR1  00000279922MR9  00000269326MR5  00000282548MR7  00000281267MR5  00000257749MR2  00000265966MR4  00000267808MR4  00000261062MR4  00000273517MR3  00000281020MR8  00000280954MR9  00000262669MR5  00000267653MR4  00000261683MR7  00000282105MR6  00000271250MR3  00000257420MR0  00000283998MR3  00000262904MR6  00000285302MR6  00000275355MR6  00000281346MR7  00000280724MR6  00000279502MR9  00000274997MR6  00000266463MR9  00000263834MR4  00000266344MR1  00000273536MR3  00000281185MR9  00000264402MR9  00000260672MR1  00000281312MR9  00000262067MR2  00000281456MR4  00000237947MR7  00000238346MR1  00000270023MR5  0000000M3000PT  00000272629MR7  00000264744MR4  00000026259MR2  00000271052MR3  00000280234MR6  00002700253MR8  00000283663MR3  00000263314MR7  00000278247MR2  00000273487MR9  00000269038MR6  00000274072MR8  00000269438MR8  00000270459MR1  00000280056MR3  00000280009MR2  00000270218MR1  00000282253MR4  00000268288MR8  00000267111MR3  00000263655MR3  00000263684MR3  00000265801MR1  00000280276MR7  00000279925MR2  00000279887MR4  00000264783MR2  00000266194MR0  00000269343MR0  00000267657MR5  00000278472MR6  00000263977MR1  00000280085MR2  00000281445MR7  00000274835MR8  00000278722MR4  00000267468MR7  00000264093MR6  00000275097MR4  00000269377MR8  00000273459MR8  00000272621MR4  00000263880MR7  00000264406MR0  00000280735MR2  00000259380MR4  00000277185MR5  00000266753MR3  00000271865MR8  00000272430MR0  00000277489MR1  00000267804MR3  00000266461MR3  00000281774MR0  00000283316MR8  00000269384MR4  00000275060MR2  00000275469MR5  00000280385MR6  00000265722MR9  00000275020MR6  00000264554MR7  00000262462MR5  00000239530MR9  00000281889MR6  00000257745MR0  00000268286MR2  00000276728MR3  00000282859MR8  00000282032MR2  00000267978MR5  00000230695MR9  00000266068MR6  00000282695MR6  00000257421MR8  00000269324MR0  00000261677MR9  00000269040MR2  00000280046MR4  00000261867MR6  00000260650MR7  00000265612MR2  00000265965MR4  00000238366MR9  00000277963MR5  00000255651MR2  00000250752MR3  00000252932MR9  00000272271MR8  00000262075MR5  00000234792MR0  00000271663MR7  00000259917MR3  00000283452MR1  00000257742MR7  00000265061MR2  00000271127MR3  00000271083MR8  00000283468MR7  00000265592MR6  00000267805MR0  00000264403MR7  00000263266MR9  00000283345MR7  00000235586MR5  00000284360MR5  00000287579MR7  00000271234MR7  00000278727MR3  00000277939MR5  00000269456MR0  00000275959MR5  00000265914MR2  00000273365MR7  00000264458MR1  00000279628MR2  00000278245MR6  00000012158YR2  00000272341MR9  00000263948MR2  00000272343MR5  00000275846MR4  00000270385MR8  00000245401MR5  00000270388MR2  00000230687MR6  00000234577MR5  00002686687MR1  00000280435MR9  00000270607MR5  00000267208MR7  00000282830MR9  00000283077MR6  00000230696MR7  00000269576MR5  00000260099MR7  00000279500MR3  00000266677MR4  00000263935MR9  00000280057MR1  00000273200MR6  00000283653MR4  00000260649MR9  1  00000227408MRI  00000264599MR2  00000269459MR4  00000227571MR7  00000280698MR2  00000267974MR4  00000257743MR5  00000277607MR8  00000283437MR2  00000281681MR7  00000278622MR6  00000272529MR9  00000260098MR9  00000280620MR6  00000239523MR4  00000270627MR3  00000274711MR1  00000267040MR4  00000281022MR4  00000273621MR3  00000267211MR1  00000257744MR3  00000264408MR6  00000265610MR6  00000274713MR7  00000281314MR5  00000281273MR3  00000270550MR7  00000269342MR2  00000261477MR4  00000273341MR8  00000269382MR8  00000272105MR8  00000273368MR1  00000280274MR2  00000266156MR9  00000272962MR2  00000265616MR3  00000261736MR3  00000270321MR3  00000278768MR7  00000278076MR5  00000276486MR8  00000250454MR6  00000266822MR6  00000263503MR5  00000279798MR3  00000283665MR8  00000278875MR0  00000279755MR3  00000266011MR6  00000269460MR2  00000276152MR6  00000276414MR0  00000273230MR3  00000272406MR0  00000262800MR6  00000269574MR0  00000269457MR8  00000283365MR5  00000230691MR8  00000277599MR7  00000267811MR8  00000262665MR3  00000281265MR9  00000027862MR0  00000283530MR4  00000266668MR3  00000270625MR7  00000268478MR5  00000269042MR8  00000265182MR6  00000273367MR3  00000284786MR1  00000262668MR7  00000264401MR1  00000266747MR5  00000271246MR1  00000275937MR1  00000274937MR2  00000287317MR2  00000267464MR6  00000283624MR5  00000273527MR2  00000241103MR1  00000267181MR6  00000266838MR2  00000274716MR0  00000270971MR5  00000267469MR5  00000243180MR7  00000265687MR4  00000277862MR9  00000251722MR5  00000270243MR9  00000269378MR6  00000234841MR5  0  00000278726MR5  00000281227MR9  00000277550MR0  00000275086MR7  00000266857MR2  00000270968MR1  00000277965MR0  00000280094MR4  00000281347MR5  00000272077MR9  00000270461MR7  00000276330MR8  00000280702MR2  347400  00000263660MR3  00000270011MR0  00000262196MR9  00000263973MR0  00000262841MR0  00000279076MR4  00000270629MR9  00000274717MR8  00000267210MR3  00000273221MR2  00000276594MR9  00002181458MR0  891011  00000265181MR8  00000266009MR0  00000277949MR4  00000263971MR4  00000267533MR8  00000026584MR6  00000268688MR9  00000026572MR1  00000269577MR3  00000276735MR8  00000271209MR9  00000024927MR2  00000274931MR5  00000274170MR0  00000282463MR9  00000286406MR4  00000266827MR5  00000282524MR8  00000281077MR8  00000277788MR6  00000271118MR2  00000276592MR3  00000263257MR8  00000262666MR1  00000255598MR5  00000027888MR3  00000254663MR8  00000266067MR8  00000273680MR9  00000253826MR2  00000268367MR0  00002675655MR9  00000262673MR7  00000269458MR6  00000027242MR4  00000268365MR4  00000267803MR5  00000272531MR5  00000264492MR0  00000225375MR5  00000270392MR4  00000265917MR5  00000267649MR2  00000261201MR8  00000277496MR6  00000275023MR0  00000259963MR7  00000277959MR3  00000271123MR2  00000270247MR0  00000273528MR0  00000268363MR9  00000249762MR6  00000283151MR9  00000280278MR3  00000272342MR7  00000283473MR7  00000279480MR8  00000243696MR2  00000274930MR7  00000277192MR1  00000272103MR3  00000278619MR2  00000237067MR4  00000278246MR4  00000278365MR2  00000277700MR1  00000277923MR9  00000274889MR5  00000269323MR2  00000270008MR6  00000274933MR1  00000245299MR3  00000249931MR7  00000257373MR1  00000262312MR2  00000273295MR6  00000283762MR3  00000282849MR9  00000283471MR1  00000266673MR3  00000249760MR0  00000273514MR0  00000276737MR4  00000278561MR6  00000252441MR1  00000275669MR0  00000273458MR0  00000274939MR8  00000287086MR3  00000282812MR7  00000268324MR1  00000272892MR1  00000267178MR2  00000270460MR9  00000260009MR6  00000279147MR3  00000282492MR8  00000281457MR2  00000267321MR8  00000261066MR5  00000266672MR5  00000271124MR0  00000275920MR7  00000234845MR6  00000249227MR0  00000277955MR1  00000272286MR6  00000280087MR8  2333  00000245861MR0  00000267527MR0  000WNJ07102006  00000281526MR4  00000285664MR9  00000273460MR6  00000281776MR5  00000270249MR6  00000270969MR9  00000253637MR3  104  00000279607MR6  00000264797MR2  00000260696MR0  00000260015MR3  00000266163MR5  00000281703MR9  00000281188MR3  00000281488MR7  00000271269MR3  00000271647MR0  00000280374MR0  00000243186MR4  00000254667MR9  00000264087MR6  00000277188MR9  00000267322MR6  00000286309MR0  00000243182MR3  00000251056MR8  00000263874MR0  00000281442MR4  00000274839MR0  00000267467MR9  00000272620MR6  00000265007MR5  00000268691MR3  00000275597MR3  00000281829MR2  00000272895MR4  00000247222MR3  00000274709MR5  00000275474MR5  00000271268MR5  00000266837MR4  00000287063MR2  00000277193MR9  00000282792MR1  00000262504MR4  00000259584MR1  00000271526MR6  00000267472MR9  00000278623MR4  00000280069MR6  00000275062MR8  00000243251MR6  00000286391MR8  00000000BYV7UN  00000282107MR2  00000235243MR3  00000266755MR8  00000279885MR8  00000267470MR3  00000282106MR4  00000254492MR2  00000278244MR9  00000281644MR5  00000283800MR1  00000282395MR3  00000278871MR9  00000279499MR8  00000280088MR6  00000262930MR1  00000265919MR1  00000266756MR6  00000267114MR7  00000266674MR1  00000267213MR7  00000269381MR0  00000269462MR8  00000269385MR1  00000269579MR9  00000285951MR0  00000274838MR2  00000273462MR2  00000273675MR9  00000276415MR7  00000275936MR3  00000269043MR6  00000272963MR0  00000275938MR9  00000281455MR6  00000266450MR6  00000262297MR5  00000263871MR6  00000278292MR8  00000281605MR6  00000267648MR4  00000262051MR6  00000280321MR1  00000279450MR1  00000256983MR8  00000276249MR0  00000280301MR3  00000277492MR5  00000271055MR6  00000290712MR9  00000272628MR9  00000266764MR0  00000266007MR4  00000259979MR3  00000249277MR5  00000275344MR0  00000255614MR0  00000267646MR8  00000261870MR0  00000272347MR6  00000269483MR4  00000261197MR8  00000264404MR5  00000272104MR1  00000274079MR3  00000282857MR2  00000283506MR4  00000264095MR1  00000265588MR4  00000280095MR1  00000271121MR6  00000282961MR2  00000267651MR8  00000281266MR7  00000286425MR4  00000273008MR3  00000268119MR5  00000273679MR1  00000270007MR8  00000281840MR9  00000262457MR5  00000283435MR6  00000280434MR2  00000279799MR1  00000278131MR8  00000241097MR5  00000266216MR1  00000295102MR8  00000286792MR7  00000012078YR2  00000278881MR0  00000279149MR9  00000011724YR2  00000011569YR1  00000272490MR4  00000001159YR1  00000295216MR6  00000297879MR9  00000300910MR7  00000280363MR3  00000283952MR0  00000294319MR9  00000287908MR8  00000922BD5C3B  00000238178MR8  00000289202MR4  00000244851MR2  00000289023MR4  00000262906MR1  00000287383MR4  00000288100MR1  00000011862YR0  00000026188MR7  00000289340MR2  00000296303MR1  00000295233MR1  00000286305MR8  00000299327MR7  00000250210MR2  00000281340MR0  00000280947MR3  00000260868MR5  00000294554MR1  00000243184MR9  00000243760MR6  00000298504MR2  00000298223MR9  00000301340MR6  00000297185MR1  00000286528MR5  00000267859MR7  00000290056MR1  00000011927YR1  00000280253MR6  00000293343MR0  00000234576MR7  00000010705YR2  00000233303MR7  00000281841MR7  00000279608MR4  00000298930MR9  00000286307MR4  00000007498YR9  00000298945MR7  00000245327MR2  000000279162MR  00000239539MR0  00000833BD4ADE  00000267324MR2  00000006740YR5  00000284575MR8  00000278819MR8  00000815BD3A4A  00000237069MR0  00000930BD5D3B  00000009444YR1  00000283604MR7  00000273224MR6  00000011926YR3  00000028261MR4  00000283606MR2  00000295058MR2  00000006736YR3  00002891887MR7  00000011954YR5  00000246613MR4  00000251711MR8  00000011879YR4  00000295218MR2  00000012194YR7  00000294323MR1  00000298947MR3  00000285499MR0  00000293047MR7  00000262258MR7  00000011947YR9  00000296848MR5  00000289818MR7  00000012234YR1  00000286620MR0  00000HDML10300  00000297371MR7  00000247227MR2  00000286306MR6  00000011571YR7  00000268011MR4  00000012239YR0  00000295140MR8  00000295810MR6  00000295593MR8  00000298946MR5  00000244844MR7  00000294696MR0  00000011867YR9  00000294557MR4  00000011595YR6  00000029464MR7  00000285725MR8  00000283817MR5  00000287417MR0  00000284263MR1  00000280251MR0  00000295191MR1  00000011786YR1  00000922BD5C37  00000011887YR7  00000290113MR0  00000919BD5BBB  00000011203YR7  00000HDMR20258  00000283994MR2  00000276417MR3  00000296223MR1  00000012241YR6  00000288373MR4  00000278041MR9  00000289585MR2  00000011730YR9  00000270674MR5  00000300496MR7  151545  00000HDMR10743  00000280231MR2  00000011746YR5  00000011741YR6  00000247730MR5  00000287569MR8  00000012180YR6  00000011878YR6  00000256225MR4  00000009576YR0  00000282957MR0  00000012237YR4  00000295876MR7  00000011502YR2  00000302186MR2  00000012275YR4  00000252926MR1  00000258084MR3  00000297475MR6  00000262596MR0  00000292980MR0  00000012306YR7  00000281692MR4  00000012351YR3  00000276411MR6  00000290001MR7  00000277443MR8  00000273361MR6  00000011588YR1  00000028867MR3  00000238351MR1  00000289164MR6  00000288121MR7  00000287256MR2  00000288461MR7  00000285383MR6  00000012042YR8  00000285221MR8  00000007582YR0  00000266346MR6  00000287935MR1  00000286885MR9  00000238171MR3  00000008551YR4  00000263505MR0  00000287489MR9  00000263970MR6  00000268302MR7  00000239859MR2  00000274829MR1  00000233432MR4  00000011798YR6  00000286699MR4  00000288590MR3  00000257730MR2  00000281943MR1  00000918BD5B50  00000012163YR2  00000284535MR2  00000289915MR1  00000011596YR4  00000297488MR9  00000282581MR8  00000012341YR4  00000287566MR4  00000267113MR9  00000266043MR9  00000236830MR6  00000288410MR4  00000273515MR7  00000264743MR6  00000283301MR0  00000286410MR6  00000248479MR8  00000948BD61E8  00000294272MR0  00000299309MR5  00000010512YR2  00000300784MR6  00000297698MR3  00000281606MR4  00000288850MR1  00000267039MR6  1  00000289775MR9  00000262899MR8  00000264092MR8  00000282252MR6  00000267656MR7  00000267666MR7  00000011541YR0  00000280579MR4  00000243686MR3  00000286743MR0  00000009005YR0  00000288344MR5  00000281342MR6  00000278772MR9  00000287069MR9  00000902BD55AD  00000300951MR1  00000250455MR3  00000294553MR3  00000298313MR8  00000239854MR3  00000282817MR6  00000276491MR8  00000302258MR9  00000295743MR9  00000266066MR0  00000919BD5B77  00000275359MR8  00000288066MR4  00000270246MR2  00000012196YR2  00000282104MR9  00000298952MR3  00001101XF50B9  00000286985MR7  00000267112MR1  00000011501YR4  00000276495MR9  00000260451MR0  00000012229YR1  00000291169MR1  00000287078MR0  00000289956MR5  00000253632MR4  00000248169MR5  00000006757YR9  00000295806MR4  00000232411MR9  00000286507MR9  00000245672MR1  00000011219YR3  00000245676MR2  00000HDMR10762  00000229480MR9  00000277780MR3  00000011774YR7  00000293007MR1  00000262802MR2  00000007858YR4  00000279424MR6  00000239038MR3  11797  00000295025MR1  00000262935MR0  00000012387YR7  00000251721MR7  00001101XF50AE  00000247721MR4  00001101XF5099  00000010599YR9  00000278851MR1  00000295027MR7  00000288704MR0  00000012222YR6  00000288980MR6  00000298490MR4  00000293293MR7  00000231889MR7  00000288194MR4  00000012388YR5  00000011584YR0  00000289395MR6  00000270251MR2  00000288071MR4  00000293276MR2  00000286964MR2  00000282795MR4  00000283776MR3  00000286865MR1  00000266065MR2  00000266097MR5  00000252408MR0  00001011BD6684  00000284284MR7  00000011950YR3  00000301289MR5  00000292823MR2  00000286382MR7  00000298792MR3  00000011784YR6  00000299660MR1  00000302505MR3  00000247731MR3  00000286429MR6  00000011897YR6  00000284025MR4  00000286558MR2  00000012020YR4  00000294273MR8  00000011744YR0  00000286385MR0  00000281861MR5  00000298220MR5  00000283520MR5  00000287255MR4  00000292432MR2  00000282868MR9  00000281604MR9  00000287808MR0  00000288592MR9  00000294523MR6  00000011920YR6  00000286634MR1  00000286544MR2  00000287899MR9  00000257430MR9  00000011900YR8  00000289331MR1  00000298496MR1  00000282731MR9  00000285727MR4  00000297370MR9  00000285491MR7  00000253170MR5  00000262299MR1  00000236832MR2  00000284037MR9  00000282270MR8  00000288593MR7  00000288690MR1  00000294304MR1  00000287082MR2  00000253149MR9  00000282394MR6  00000273220MR4  00000298974MR7  00000299166MR9  00000239520MR0  00000297419MR4  00000296233MR0  00000275022MR2  00000279918MR7  00000301847MR0  00000279613MR4  00000287870MR0  00000238349MR5  00000285497MR4  00000296100MR1  00000297852MR6  00000916BD5B05  00000299911MR8  00000283845MR6  00000233302MR9  00000919BD5B9D  00000264090MR2  00000283864MR7  00000288123MR3  00000285965MR0  00000295512MR8  00000011958YR6  00000012236YR6  00000260471MR8  00000011892YR7  00000298413MR6  00000288825MR3  00000289792MR4  00000282828MR3  00000254672MR9  00000011885YR1  00000299762MR5  00000289913MR6  00000264091MR0  00000289533MR2  1  00000012063YR4  00000294444MR5  00000296104MR3  00000298936MR6  00000261627MR4  00000286059MR1  00000285693MR8  00000298809MR5  00000276722MR6  00000285966MR8  00000289531MR6  00000012181YR4  00000295518MR1  00000293317MR4  00000013657YR2  00000284576MR6  00000006592YR0  00000289777MR5  00000919BD5B97  00000276489MR2  00000011901YR6  00000238873MR4  00000284234MR2  00000012303YR4  00000012264YR8  00000288771MR9  00000297011MR9  00000011726YR7  00000267622MR9  00000561695MR3  00000011941YR2  00000297205MR7  00000289764MR3  00000231891MR3  00000289197MR6  00000299850MR8  00000288383MR3  00000287476MR6  00000006669YR6  00000286855MR2  00000012240YR8  00000916BD5B07  00000011953YR7  00000276245MR8  00000297627MR2  00000010722YR7  00000236924MR7  00000228285MR3  00000296367MR6  00000295176MR2  000000HDMR3130  00000006734YR8  00000243420MR7  00000298497MR9  00000011718YR4  00000288190MR2  00000292792MR9  00000284235MR9  00001101XF50A1  00000288152MR2  00000262298MR3  00000286765MR3  00000289201MR6  00000288064MR9  00000236504MR7  00000261476MR6  00000011722YR6  00001003BD6432  00000012189YR7  00000283996MR7  00000289163MR8  00000297369MR1  00000298710MR5  00000237702MR6  00000233270MR8  00000293342MR2  00000283402MR6  00000278848MR7  00000276586MR5  00000293324MR0  00000256208MR0  00000230111MR7  00000281869MR8  00000268689MR7  00000268290MR4  00000286927MR9  00000011787YR9  000000HDMR2904  00000296575MR4  00000277763MR9  00000241098MR3  00000283075MR0  00000289024MR2  00000287438MR6  00000008398YR0  00000288649MR7  00000283022MR2  00000011518YR8  00000294945MR1  00000275050MR3  00000286923MR8  00000286907MR1  00000274307MR8  00000248481MR4  00000280388MR0  00000289441MR8  00000282789MR7  00000011886YR9  00000235702MR8  00000288723MR0  00000280275MR9  00000283019MR8  00000288223MR1  00000012080YR8  00000288720MR6  00000005538YR4  00000287554MR0  00000282948MR9  000010",
      "more_code_info": "48XF500B  00000279999MR7  00000011572YR5  00000267325MR9  00000286967MR5  00000011996YR6  00000011869YR5  00000281775MR7  00000279562MR3  00000286893MR3  00000274696MR4  00000294738MR0  00000011581YR6  00000905BD566C  00000295741MR3  00000273678MR3  00000287079MR8  00000265587MR6  00000275049MR5  00000296224MR9  00000285751MR4  00000011524YR6  00000283467MR9  00000296631MR5  00000294320MR7  00000280998MR6  00000278679MR6  00000297516MR7  00000011890YR1  00000302526MR9  00000287271MR1  000S0933BD5E02  00000270564MR8  00000012402YR4  00000286448MR6  00000294994MR9  00000232424MR2  00000288703MR2  00000286948MR5  00000280007MR6  00000276630MR1  00000247720MR6  00000295173MR9  00000288722MR2  0  00000299471MR3  00000235692MR1  00000295026MR9  00000248488MR9  00000284211MR0  00000296111MR8  00000284055MR1  00000289189MR3  00000278607MR7  00000275649MR2  00000281825MR0  00000289015MR0  00000279412MR1  00000012133YR5  00000231897MR0  00000918BD5B54  00000288346MR0  00000281941MR5  00000012230YR9  00000294144MR1  00000298826MR9  00000284283MR9  00000294990MR7  00000278313MR2  00000294648MR1  00000275650MR0  00000287286MR9  00000266205MR4  00001001BD6328  00000273396MR2  00000270010MR2  00000HDMR20063  00000009525YR7  00000231904MR4  00000279463MR4  00000276593MR1  00000266783MR0  00000240102MR4  00000290141MR1  00000287849MR4  00000251724MR1  00000286947MR7  00000288757MR8  00000011086YR6  00000256237MR9  00000291979MR3  00000235674MR9  00000282082MR7  00000289203MR2  00000271251MR1  00000301818MR1  00000287551MR6  00000011934YR7  00000260480MR9  00000287880MR9  00000300498MR3  00000283514MR8  00000292820MR8  00000274888MR7  00000284715MR0  00000289789MR0  00000265716MR1  00000286018MR7  00000011563YR4  00000027552MR4  00000286560MR8  00000297700MR7  00000281258MR4  2102  00000239242MR1  00000289627MR2  00000028654MR9  00000298428MR4  00000288097MR9  00000263974MR8  00000297383MR2  00000298011MR8  00000282689MR9  00000282863MR0  00000283818MR3  00000283821MR7  00000294118MR5  00000930BD5D3C  00000278102MR9  00000012003YR0  00000284391MR0  00000289042MR4  00000282683MR2  00000263968MR0  00000284563MR4  00000283523MR9  00000278817MR2  00000283175MR8  00000012224YR2  00000919BD5BAA  00000302523MR6  00000273524MR9  00000235675MR6  00000289332MR9  00000288431MR0  00000278559MR0  00000269061MR8  00000287623MR3  00000283027MR1  00000265000MR0  00000295789MR2  00000289218MR0  00000294737MR2  00000011591YR5  00000299882MR1  00000282794MR7  00000288245MR4  00000012385YR1  00001101XF50A4  00000289335MR2  00000280670MR1  00000286492MR4  00000288721MR4  00000288224MR9  00000275670MR8  00000011224YR3  00000249771MR7  00000289347MR7  00000262667MR9  00000232413MR5  00000283434MR9  00000287553MR2  00000282996MR8  00000283126MR1  00000286094MR8  00000295807MR2  00000008380YR8  00000294605MR1  00000283699MR7  00000277491MR7  00000299640MR3  00000282829MR1  00000299422MR6  00000276266MR4  00000011946YR1  00000285754MR8  00000286850MR3  00000286009MR6  00000286433MR8  00000012232YR5  00000283847MR2  00000286776MR0  00000287461MR8  00000263875MR7  00000288004MR5  00000266203MR9  00000025604MR9  00000295898MR1  00000298502MR6  00000287616MR7  00000292290MR4  00000011884YR4  00000283639MR3  00000230108MR3  00000011578YR2  00000235686MR3  00000288110MR0  00000008550YR6  00000238183MR8  00000251229MR1  00000279618MR3  00000286744MR8  00000289027MR5  00000285055MR0  00000010964YR5  00000301615MR1  00000294519MR4  00000289761MR9  00000298976MR2  00000298187MR6  00000240106MR5  00000275463MR8  00000281228MR7  00000300984MR2  00000236702MR7  00000230099MR4  00000288822MR0  00000272897MR0  00000281485MR3  00000277919MR7  00000235681MR4  00000011547YR7  00000261385MR9  00000242180MR8  00000230107MR5  00000261941MR9  00000284903MR2  00000028606MR1  00000012406YR5  00000289762MR7  00000279415MR4  00000011883YR6  00000011504YR8  00000919BD5BC4  00000267529MR6  00000298468MR0  00000283400MR0  00000276435MR5  00000296830MR3  00000275646MR8  00000235238MR3  00000261378MR4  00000011721YR8  00000011801YR8  00000295899MR9  00000283021MR4  00000279614MR2  00000265723MR7  00000013090YR6  00000288023MR5  00000288775MR0  00000011269YR8  00000284458MR7  00000012065YR9  00000012188YR9  00000279914MR6  00000288161MR3  00000273673MR4  00000295701MR7  00000287898MR1  00000296364MR3  00000012047YR7  00000245817MR2  00000237931MR1  00000288159MR7  00000279882MR5  00000298968MR9  00000012223YR4  00000012225YR9  00000283679MR9  00000277186MR3  00000283364MR8  00000281939MR9  00000268512MR1  00000281862MR3  00000287939MR3  00000289525MR8  00000012017YR0  00000278164MR9  00000287855MR1  00000294892MR5  00000011939YR6  00000285724MR1  00000284275MR5  00000294402MR3  00000249929MR1  00000287819MR7  00000283605MR4  00000283040MR4  00000294668MR9  00000011877YR8  00000288061MR5  00000278770MR3  00000283859MR7  00000294342MR1  00000285382MR8  00000275962MR9  00000295476MR6  00000283173MR3  00000299523MR1  00000294959MR2  00000289261MR0  00000289271MR9  00000295052MR5  00000286895MR8  00000278129MR2  00000237945MR1  00000288191MR0  00000288020MR1  00000278816MR4  00000028286MR6  00000241724MR4  00000288341MR1  00000012193YR9  00000295139MR0  00000238358MR6  00000260697MR8  00000294409MR8  00000299449MR9  00000919BD5BA9  00000297447MR5  00000302624MR2  00000289318MR8  00000023524MR8  00000282340MR9  00000012167YR3  00000288377MR5  00000233622MR0  00000234840MR7  00000279070MR7  00000265552MR0  00000287878MR3  00000298810MR3  00000286587MR1  00000285490MR9  00000012198YR8  00000287617MR5  00000236835MR5  00000012185YR5  00000287656MR3  00000012157YR4  00000288021MR9  00000294394MR2  00000247460MR9  00000247725MR5  00000294442MR9  00000261869MR2  00000001206YR5  00000294327MR2  00000288342MR9  00000294158MR1  00000295742MR1  00000298716MR2  00000298657MR8  00000300863MR8  00000237701MR8  00000239028MR4  00000257374MR9  00000266449MR8  00000268419MR9  00000268282MR1  00000276589MR9  00000278167MR2  00000010140YR2  00000282523MR0  00000285796MR9  00000286998MR0  00000919BD5BB5  00000011511YR3  00000289998MR7  00000298044MR9  00000292315MR9  00000266468MR8  00000011575YR8  00000300904MR0  00000299350MR9  00000298687MR5  00000287415MR4  00000011056YR9  00000235682MR2  00000283408MR3  00000011956YR0  00000263255MR2  00000281154MR5  00000262804MR8  00000296410MR4  00000299527MR2  00000298708MR9  00000012161YR6  00000922BD5C38  00000301290MR3  00000287490MR7  00000296441MR9  00000011796YR0  00000287600MR1  00000289205MR7  00000012192YR1  00000296301MR5  00000289160MR4  00000299758MR3  00000281933MR2  00000295088MR9  00000012272YR1  00000012072YR5  00000011544YR4  00000287588MR8  00000011743YR2  00000011873YR7  00000280891MR3  00000009042YR3  00000299525MR6  00000286409MR8  00000274161MR9  00000012132YR7  00000012070YR9  00000272290MR8  00000287400MR6  00000292474MR4  00000011931YR3  00000012231YR7  00000280001MR9  00000009636YR2  00000276718MR4  00000301323MR2  00000243687MR1  00000236834MR8  00000008549YR8  00000297893MR0  00000262672MR9  00000280391MR4  00000263336MR0  00000298709MR7  00000295235MR6  00000286946MR9  000000HDMR2892  00000286966MR7  00000012346YR3  00000296519MR2  00000285626MR8  00000287062MR4  00000261061MR6  00000287391MR7  00000294160MR7  00000289014MR3  00000289262MR8  00000298429MR2  00000241721MR0  00000297382MR4  00000300923MR0  00000012006YR3  00000293299MR4  00000264221MR3  00000283528MR8  00000287363MR6  00000280767MR5  00000284693MR9  00000283622MR9  00000237481MR6  00000288255MR3  00000011932YR1  00000289336MR0  00000011555YR0  00000298949MR9  00000284785MR3  00000299662MR7  00000286906MRS  00000284690MR5  00000011955YR2  00000292901MR6  00000299424MR2  00000297486MR3  00000012015YR4  00000287117MR6  00000287479MR0  00000011717YR6  00000301299MR4  00000288098MR7  00000011506YR3  00000011551YR9  00000302625MR9  00000831BD49EF  00000286898MR2  00000283997MR5  00000011804YR2  00000011874YR5  00000289438MR4  00000011727YR5  00000011875YR2  00000271084MR6  00000276151MR8  00000300608MR7  00000254664MR6  00000299642MR9  00000280461MR5  00000011586YR5  00000011338YR1  00000011549YR3  00000301349MR7  00000284534MR5  00000286478MR3  00000007882YR4  00000007887YR3  00000300517MR0  00000300937MR0  00000301301MR8  00000300811MR7  00000300581MR6  00000283654MR2  00000282007MR4  00000011938YR8  00000266129MR6  00000239537MR7  00000277860MR3  00000012050YR1  00000012010YR5  00000012008YR9  00000241434MR3  00000273537MR1  00000277917MR1  00000280364MR1  00000274692MR3  00000284472MR8  00000272334MR4  00000283812MR6  00000026005MR8  00000286783MR6  00000283515MR5  00000266343MR3  00000300902MR4  000000HDMR2946  00000028135MR2  00000239858MR4  00000251059MR2  00000279503MR7  00000274078MR5  00000271119MR0  00000287941MR9  00000295004MR6  00000294609MR3  00000302424MR7  00000294625MR9  00000288651MR3  00000253834MR6  00000009853YR3  00000007544YR0  00000269340MR6  00000235683MR0  00000008115YR8  00000245806MR5  00000261444MR4  00000280893MR9  00000295482MR4  00000287506MR0  00000261678MR7  00000272344MR3  00000299885MR4  00000028432MR3  00000296210MR8  00000273516MR5  00000301188MR9  000000HDMR2609  00000287596MR1  00000300619MR4  00000265553MR8  00000253157MR2  00000259568MR4  00000254193MR6  00000234853MR0  00000280950MR7  00000273253MR5  00000261329MR7  00000280949MR9  00000250205MR2  00000295229MR9  00000006586YR2  00000287315MR6  00000253838MR7  00000301501MR3  00000280946MR5  00000301105MR3  00000291943MR9  00000286334MR8  00000272562MR0  00000HDMR20269  00000293315MR8  00000287645MR6  00000284262MR3  00000257754MR2  00000302472MR6  00000252410MR6  00000296145MR6  00000270382MR5  00000026717MR0  00001104XF5011  00000300884MR4  00000291681MR5  00000261330MR5  00000261381MR8  00000265805MR2  00000264619MR8  00000300950MR3  00000011872YR9  00000006726YR4  00000300577MR4  00000280328MR6  00000280229MR6  00000295190MR3  00000288459MR1  00000284411MR6  00000298533MR1  00000274744MR2  00000262670MR3  00000009837YR6  00000302510MR3  00000300457MR9  00000299949MR8  00000300970MR1  00000287877MR5  00000299962MR1  00000300769MR7  00000300458MR7  00000294515MR2  00000289222MR2  00000300702MR8  00000GE0210033  00000249266MR8  00000251090MR7  00000256198MR3  00000259105MR5  00000301098MR0  00000275094MR1  00000267807MR6  00000272532MR3  00000298755MR0  00000288075MR5  00000289884MR9  00000298098MR5  00000919BD5B9F  00000286148MR2  00000011739YR0  00000011728YR3  00000270465MR8  00000301223MR4  00000295003MR8  00000290369MR8  00000288222MR3  00000260861MR0  00000274050MR4  00000280912MR7  00000290572MR7  00000288254MR6  00000288257MR9  00001011BD6688  00000287601MR9  00000012019YR6  00000274712MR9  00000278557MR4  00000237000MR5  00000234791MR2  00000259129MR5  00000257433MR3  00000231900MR2  00000242592MR4  00000264557MR0  00000258406MR8  00000230105MR9  00000256968MR9  00000241522MR2  00000251638MR3  00000235239MR1  00000296056MR5  00000267463MR8  00000267323MR4  00000255596MR9  00000282922MR4  00000232409MR3  00000259580MR9  00000268421MR5  00000285956MR9  00000267197MR2  00000275539MR5  00000287268MR7  00000286370MR2  00000012005YR5  00000295055MR8  00000011564YR2  00000012408YR1  00000265063MR8  00000269041MR0  00000268123MR7  00000283407MR5  00000284060MR1  00000011896YR8  00000912BD59B4  00000294305MR8  00000261871MR8  00000284594MR9  00000242176MR6  00000250196MR3  00000256686MR7  00000277762MR1  00000278672MR1  00000281153MR7  00000288815MR4  00000280761MR8  00000268761MR4  00000293334MR9  00000280725MR3  00000233294MR8  00000278548MR3  00000281487MR9  00000007581YR2  00000282526MR3  00000282856MR4  00000286321MR5  00000287552MR4  00000288211MR6  00000246611MR8  00000286892MR5  00000295541MR7  00000264748MR5  000000HDMR3834  00000265594MR2  00000283774MR8  00000254376MR7  00000919BD5B8E  00000281602MR3  00000283650MR0  00000298068MR8  00000264740MR2  00000294988MR1  00000287773MR6  00000283517MR1  00000274180MR9  00000290310MR2  00000011714YR3  00000262040MR9  00000012057YR6  00000292979MR2  00000011783YR8  00000012071YR7  00000268122MR9  00000286368MR6  00000011580YR8  00000289849MR2  00000260628MR3  00000276493MR4  00001011BD6629  00000029749MR3  00000298713MR9  00001052XF5067  00000287615MR9  000000HDMR5803  00000236707MR6  00000241726MR9  00000234859MR7  00000252664MR8  00000268317MR5  00000256209MR8  00000277609MR4  00000273293MR1  00000280092MR8  00000280694MR1  00000255602MR5  00000285432MR1  00000282611MR3  00000284077MR5  00000287509MR4  00000282392MR0  00000291406MR7  00000250452MR0  00000256210MR6  00000256213MR0  00000251020MR4  00000281517MR3  00000281518MR1  00000279478MR2  00000262891MR5  00000282022MR3  00000281698MR1  00000283118MR8  00000284361MR3  00000280705MR4  00000290039MR7  00000290603MR0  00000290005MR8  00000290548MR7  00000263253MR7  00000293009MR7  00000293031MR1  00000295840MR3  00000296628MR1  00000252662MR2  00000282681MR6  00000290547MR9  00000292214MR4  00000239232MR2  00000255931MR8  00000299516MR5  00000250211MR0  00000298072MR0  00000298382MR3  00000238870MR0  00000298840MR0  00000298579MR4  00000299960MR5  00000299562MR9  00000280373MR2  00000252658MR0  00000269482MR6  00000271238MR8  00000270972MR3  00000264558MR8  00000239044MR1  00000260102MR9  00000264747MR7  00000273488MR7  00000273533MR0  00000269572MR4  00000274172MR6  00000277864MR5  00000274938MR0  00000280626MR3  00000282525MR5  00000284717MR6  00000284596MR4  00000285026MR1  00000270973MR1  00000251871MR0  00000266160MR1  00000264742MR8  00000261679MR5  00000264088MR6  00000265866MR4  00000269379MR4  00000272335MR1  00000279924MR5  00000279890MR8  00000281257MR6  00000279869MR2  00000288196MR9  00000284643MR4  00000286778MR6  00000288204MR1  00000288774MR3  00000283893MR6  00000283404MR2  00000281828MR4  00000266158MR5  00000280691MR7  00000277642MR5  00000286925MR3  00000261441MR0  00000257431MR7  00000286930MR3  00000011230YR0  00000252665MR5  00000261590MR4  00000266345MR8  00000280860MR8  00000278800MR8  00000278554MR1  00000007890YR7  00000280921MR8  00000281057MR0  00000286160MR7  00000009575YR2  00000028686MR4  00000295181MR2  00000252377MR7  00000252406MR4  00000250220MR1  00000265591MR8  00000274934MR9  00000278398MR3  00000276680MR6  00000278673MR9  00000280064MR7  00000278211MR8  00000280673MR5  00000281315MR2  00000282512MR3  00000284237MR5  00000280230MR4  00000919BD5B94  00000285591MR4  00000286736MR4  00000011790YR3  00000011792YR9  00000287961MR7  00000011562YR6  00000288934MR3  00000287559MR9  00000012009YR7  00000288460MR9  00000289975MR5  00000011909YR9  00000289912MR8  00000288932MR7  00000289791MR6  00000011898YR4  00000011507YR1  00000289793MR2  00000011598YR0  00000011599YR8  00000011895YR0  00000011948YR7  00000012383YR6  00000012268YR9  00000294910MR5  00000012354YR7  00000296203MR3  00000012266YR3  00000296112MR6  00000294610MR1  00000294604MR4  00000294627MR5  00000294908MR9  00000296359MR3  00000296645MR5  00000296355MR1  00000295230MR7  00000295514MR4  00000297625MR6  00000296633MR1  00000298262MR7  00000296201MR7  00000296832MR9  00000298824MR4  00000298219MR7  00001007BD6496  00000299950MR6  00000006742YR1  00000008327YR9  00001043XF5005  00000300812MR5  00000280580MR2  00000298977MR0  00000299335MR0  00000256206MR4  00000300124MR5  00000300699MR6  00000300554MR3  00000302253MR0  00000299495MR2  00000261576MR3  00000257727MR8  00000275596MR5  00000289527MR4  00000287598MR7  00000270630MR7  00000233623MR8  00000263292MR5  00000233440MR7  00000300553MR5  00000285119MR4  00000010954YR6  00000274168MR4  00000239043MR3  00000238361MR3  00000243412MR4  00000247728MR9  00000291846MR4  00000294918MR8  00000288849MR3  00000274932MR3  00000277779MR5  00000239230MR6  00000248174MR5  00000258085MR0  00000237708MR3  00000233446MR4  00000243409MR0  00000253630MR8  00000256216MR3  00000246610MR0  00000266072MR8  00000245816MR4  00000279482MR4  00000256969MR7  00000256221MR3  00000255212MR3  00000264462MR3  00000274074MR4  00000259527MR0  00000270563MR0  00000252931MR1  00000270631MR5  00000265960MR5  00000266690MR7  00000262593MR7  00000239527MR5  00000271666MR0  00000260013MR8  00000241678MR2  00000283451MR3  00000247732MR1  00000247734MR7  00000276164MR1  00000252381MR9  00000276274MR8  00000239228MR0  00000284316MR7  00000276733MR3  00000299202MR2  00000239046MR6  00000295763MR7  00000277918MR9  00000249761MR8  00000250213MR6  00000241875MR4  00000012340YR6  00000286154MR0  00000011889YR3  00000282908MR3  00000278747MR1  00000267431MR5  00000272107MR4  00000279069MR9  00000270572MR1  00000259915MR7  00000298581MR0  00000232429MR1  00000011944YR6  00000814BD39ED  00000297630MR6  00000278606MR9  00000281793MR0  00000246614MR2  00000007090YR4  00000279909MR6  00000284641MR8  00000242181MR6  00000253156MR4  00000283347MR3  00000301499MR0  00000011861YR2  00000286426MR2  00000284356MR3  00000268525MR3  00000259378MR8  00000268528MR7  00000299538MR9  00000299621MR3  00000300573MR3  00000261206MR7  00000272432MR6  00000278243MR1  00000287918MR7  00000299184MR2  00000295483MR2  00000255613MR2  00000232414MR3  00000271525MR8  00000256986MR1  00000261379MR2  00000230110MR9  00000249273MR4  00000249278MR3  00000239860MR0  00000264009MR2  00000264998MR6  00000254491MR4  00000264015MR9  00000293676MR3  00000236505MR4  00000239041MR7  00000257438MR2  00000261473MR3  00000249764MR2  00000265826MR8  00000242551MR0  00000268527MR9  00000256967MR1  00000260667MR1  00000262661MR2  00000246265MR3  00000250201MR1  00000235679MR8  00000249765MR9  00000257437MR4  00000254670MR3  00000261471MR7  00000267110MR5  00000270522MR6  00000293715MR9  00000243408MR2  00000252431MR2  00000252429MR6  00000255201MR6  00000026066MR9  00000267106MR3  00000270515MR0  00000270523MR4  00000259123MR8  00000273018MR2  00000256224MR7  00000294526MR9  00000238872MR6  00000238865MR0  00000241719MR4  00000243767MR1  00000249766MR7  00000261442MR8  00000260866MR9  00000267851MR4  00000252920MR4  00000241864MR8  00000245808MR1  00000247722MR2  00000251801MR7  00000262900MR4  00000261327MR1  00000259127MR9  00000268134MR4  00000267107MR1  00000268321MR7  00000270517MR6  00000266801MR0  00000257444MR0  00000254488MR0  00000256694MR1  00000255207MR3  00000241102MR3  00000243766MR3  00000250438MR9  00000257725MR2  00000260670MR5  00000261386MR7  00000258101MR5  00000261380MR0  00000264006MR8  00000271648MR8  00000252435MR3  00000235703MR6  00000252663MR0  00000255198MR4  00000255226MR3  00000265095MR0  00000267109MR7  00000239533MR3  00000268135MR1  00000242555MR1  00000241722MR8  00000254201MR7  00000264997MR8  00000262733MR9  00000261573MR0  00000265091MR9  00000259574MR2  00000249279MR1  00000282686MR5  00000281704MR7  00000270384MR1  00000272336MR9  00000276731MR7  00000278842MR0  00000280768MR3  00000283470MR3  00000286392MR6  00000287810MR6  00000265969MR6  00000261200MR0  00000261065MR7  00000262931MR9  00000264741MR0  00000265179MR2  00000267977MR7  00000271126MR5  00000266147MR8  00000272973MR9  00000274166MR8  00000267037MR0  00000279159MR8  00000289004MR4  00000287879MR1  00000280302MR1  00000288901MR2  00000262803MR0  00000277194MR7  00000279751MR2  00000283386MR1  00000282501MR6  00000288667MR9  00000285172MR3  00000282791MR3  00000283526MR2  00000282790MR5  00000288244MR7  00000286559MR0  00000270974MR9  00000272893MR9  00000278749MR7  00000280898MR8  00000280951MR5  00000285380MR2  00000283761MR5  00000286800MR8  00000289798MR1  00000288811MR3  00000273229MR5  00000274996MR8  00000282088MR4  00000284787MR9  00000285222MR6  00000260625MR9  00000265590MR0  00000276635MR0  00000280005MR0  00000283378MR8  00000286538MR4  00000273676MR7  00000272894MR7  00000277583MR1  00000279802MR3  00000280621MR4  00000285303MR4  00000282831MR7  00000283174MR1  00000286897MR4  00000264089MR4  00000274688MR1  00000277487MR5  00000284644MR2  00000265593MR4  00000273223MR8  00000275091MR7  00000275598MR1  00000277189MR7  00000279752MR0  00000280319MR5  00000276370MR4  00000288589MR5  00000279458MR4  00000280299MR9  00000281936MR5  00000284457MR9  00000267809MR2  00000282858MR0  00000261198MR6  00000285219MR2  00000283662MR5  00000267650MR0  00000281317MR8  00000273535MR5  00000287590MR4  00000278566MR5  00000280002MR7  00000008560YR5  00000283076MR8  00000279423MR8  00000281603MR1  000000HDMR5528  00000288812MR1  00000294929MR5  00000296257MR9  00000301762MR1  00000011864YR6  00000291970MR2  00000011937YR0  00000012360YR4  00000301010MR5  00000011772YR1  00000012165YR7  00000012186YR3  00000295022MR8  00000291977MR7  00000295828MR8  00000292982MR6  00000297474MR9  00001008BD652A  00001009BD654A  00000298182MR7  00000298659MR4  00000300477MR7  00000301264MR8  00000301383MR6  00000301612MR8  00001046XF501A  00000267973MR6  00000277214MR3  00000289196MR8  00000011894YR3  00000012183YR0  00000011997YR4  00000012375YR2  00000288874MR1  00000919BD5BB3  00000292591MR5  00000294936MR0  00000300949MR5  00000296439MR3  00000294944MR4  00000297384MR0  00000297466MR5  00000298759MR2  00000297806MR2  00000299273MR3  00000235694MR7  00001011BD664D  00000299269MR1  00000301928MR8  00000301168MR1  00000302700MR0  00000VALUE1162  00000287630MR8  00000298928MR3  00000298670MR1  00000012184YR8  00000011590YR7  00000292974MR3  00000295878MR3  00000299191MR7  00000293587MR2  00000289315MR4  00000293499MR0  00000296225MR6  00000298328MR6  00000300182MR3  00000300701MR0  00000300729MR1  00000295020MR2  00000296093MR8  00000297420MR2  00000290325MR0  00000290488MR6  00000296634MR9  00000297385MR7  00000298059MR7  00000298672MR7  00000011923YR0  00000299163MR6  00000299313MR7  00000011952YR9  00000289026MR7  00000012358YR8  00000919BD5BA8  00000295531MR8  00000296458MR3  00000295561MR5  00000298680MR0  00000298911MR9  00000299946MR4  00000301398MR4  00000301806MR6  00000301929MR6  00000294328MR0  00000293650MR8  00000294547MR5  00000296094MR6  00000299506MR6  00000301930MR4  00000288936MR8  00000011928YR9  00000293298MR6  00000294920MR4  00000302426MR2  00000298935MR8  00000298403MR7  00000300859MR6  00000300183MR1  00000300123MR7  00000300682MR2  00000300617MR8  00000262664MR6  00000302474MR2  00000277190MR5  00000286479MR1  00000288701MR6  00000011574YR1  00000011871YR1  00000012315YR8  00000294284MR5  00000012304YR2  00000293593MR0  00000295529MR2  00000298467MR2  00000299448MR1  00000257738MR5  00000286434MR6  00000300887MR7  00000285353MR9  00000011558YR4  00000289591MR0  00000298314MR6  00000298155MR3  00000299164MR4  00000298793MR1  00000298969MR7  00000299311MR1  00000299192MR5  00000301187MR1  00000299964MR7  00000302356MR1  00000299539MR7  00000242552MR8  00000298329MR4  00000299312MR9  00000288428MR6  00000288099MR5  00000011876YR0  00000012190YR5  00000848BD5311  00000295013MR7  00000299265MR9  00000298774MR1  00000298221MR3  00000298858MR2  00000300879MR4  00000297725MR4  00000301104MR6  00000299167MR7  00000VALUE1179  00000294685MR3  00000299677MR5  00000298330MR2  00000299275MR8  00000300922MR2  00000296468MR2  00000012125YR1  00000012317YR4  00000012270YR5  00000293586MR4  00000294928MR7  00000295059MR0  00000300747MR3  00000295236MR4  00000294051MR8  00000299165MR1  00000299507MR4  00000301392MR7  00001104XF507C  00000302643MR2  00000289523MR3  00000293164MR0  00000293163MR2  00000919BD5BA1  00000296646MR3  00001101XF50A2  00000301498MR2  00000011738YR2  00000298846MR7  00000297202MR4  00000299328MR5  00000302679MR6  00000012352YR1  00000944BD6168  00000295528MR4  00000297014MR3  00000297658MR7  00000300770MR5  00000293297MR8  00000297728MR8  00000257734MR4  00000299643MR7  00000289039MR0  00000296442MR7  00000295219MR0  00000299349MR1  00000302473MR4  00000913BD5A47  00000011314YR2  00000302188MR8  00000011866YR1  00000301397MR6  00001003BD6415  00000300952MR9  00000302568MR1  00000302532MR7  00000298934MR1  00000GE0131001  00000GE0205029  00000GE0206027  00000249264MR3  00000257756MR7  00000291468MR7  00000290648MR5  00000263263MR6  00000267326MR7  00000271122MR4  00000273006MR7  00000273540MR5  00000276418MR1  00000282959MR6  00000298553MR9  00000296302MR3  00000298214MR8  00000299271MR7  00000299883MR9  00000263264MR4  00000262660MR4  00000298412MR8  00000268306MR8  00000259585MR8  00000286696MR0  00000011899YR2  00000278850MR3  00000011936YR2  00000294626MR7  00000296267MR8  00000297835MR1  00000229478MR3  00000238871MR8  00000265916MR7  00000265589MR2  00000231894MR7  00000253625MR8  00000266782MR2  00000272288MR2  00000294628MR3  00000279884MR1  00000278677MR0  00000011865YR3  00000247724MR8  00000279461MR8  00000011054YR4  00000228373MR7  00000288385MR0  00000265611MR4  00000250218MR5  00000267670MR8  00000287850MR2  00000289474MR9  00000265820MR1  00000264011MR8  00000294306MR6  00000299450MR7  00000298686MR7  00000282691MR5  00000266744MR2  00000283822MR5  00000269339MR8  000000HDMR4304  00000276163MR3  00000280777MR4  00000282950MR5  00000283348MR1  00000287019MR4  00000010970YR2  00000275095MR8  00000286896MR6  00000919BD5B8F  00000247719MR8  00000298094MR4  00000300615MR2  00000012199YR6  00000282855MR6  00000284901MR6  00000279616MR7  00000010066YR9  00000250442MR1  00000273833MR4  00000281867MR2  00000253158MR0  00000283350MR7  00000264618MR0  00000265869MR8  00000269325MR7  00000300750MR7  00000012345YR5  00000241872MR1  00000289773MR4  00000268121MR1  00000284757MR2  00000266467MR0  00000258091MR8  00000274836MR6  00000284238MR3  00000275538MR7  00000263339MR4  00000296833MR7  00000284648MR3  00000293318MR2  00000280922MR6  00000283933MR0  00000274073MR6  00000245821MR4  00000253837MR9  00000258414MR2  00000258368MR0  00000250202MR9  00000246621MR7  00000245408MR0  00000264999MR4  00000251798MR5  00000239521MR8  00000254497MR1  00000251643MR3  00000250449MR6  00000255200MR8  00000258100MR7  00000261591MR2  00000268473MR6  00000265920MR9  00000262065MR6  00000280432MR6  00000272340MR1  00000275645MR0  00000259125MR3  00000265060MR4  00000260695MR2  00000275584MR1  00000266144MR5  00000259842MR3  00000242547MR8  00000266646MR9  00000265172MR7  00000279726MR4  00000267182MR4  00000276250MR8  00000274890MR3  00000267212MR9  00000272078MR7  00000261675MR3  00000298568MR7  00000011540YR2  00000012000YR6  00000297851MR8  00000289334MR5  00000918BD5B42  00000299763MR3  00000265868MR0  00000284465MR2  00000272964MR8  00000011543YR6  00000286626MR7  00000026639MR2  00000287631MR6  00000012007YR1  00000283932MR2  00000012064YR2  00000283073MR5  00000283772MR2  00000284872MR9  00000012018YR8  00000276730MR9  00000267327MR5  00000279165MR5  00000283521MR3  00000283508MR0  00000279630MR8  00000286338MR9  00000288043MR3  00000012056YR8  00000295193MR7  00000289820MR3  00000922BD5C3A  00000294735MR6  00000298808MR7  00000279727MR2  00000289348MR5  00000289526MR6  00000289478MR0  00000012322YR4  00001101XF50A8  00000294552MR5  00000295804MR9  00000297822MR9  00000918BD5B38  00000293588MR0  00000297809MR6  00000300125MR2  00000011024YR7  00000250445MR4  000000HDMR3570  00000264643MR8  00000287964MR1  00000296305MR6  00000011945YR3  00000286849MR5  00000011929YR7  00000011998YR2  00001105XF5020  00000299526MR4  00000280732MR9  00000294658MR0  00000294114MR4  00000295182MR0  00000282269MR0  00000011800YR0  00000277441MR2  00000011582YR4  00000283745MR8  00000294117MR7  00000294661MR4  00000286286MR0  00000919BD5B90  00000905BD5675  00000237699MR4  00000286381MR9  00000297892MR2  00000012062YR6  00000012235YR8  00000012305YR9  00000288806MR3  00000011732YR5  00000283401MR8  00000273531MR4  00000027972MR4  00000296440MR1  00000298222MR1  00000011725YR9  00000011736YR6  00000012379YR4  00000297465MR7  00000299661MR9  00000288433MR6  00000012243YR2  00000289782MR5  00000297970MR6  00000289587MR8  00000294529MR3  00000919BD5BB1  00000297834MR4  00000272535MR6  00000297446MR7  00000918BD5B43  00000297853MR4  00000297444MR2  00000288669MR5  00000243183MR1  00000919BD5BA6  00000011120YR3  00000301138MR4  00000012269YR7  00000294343MR9  00000282910MR9  00000300886MR9  00000294919MR6  00000298281MR7  00000275934MR8  00000300673MR1  00000297660MR3  00000275969MR4  00000284688MR9  00000284384MR5  00000011514YR7  00000298070MR4  00000012122YR8  00000255210MR7  00000285496MR6  00000275047MR9  00000298929MR1  00000288109MR2  00000288586MR1  00000300726MR7  000000245811MR  00000301139MR2  00001002BD6349  00000288610MR9  00000262799MR0  00000278105MR2  00000268303MR5  00000007134YR0  00000281683MR3  000000HDMR2862  00000283698MR9  00000256205MR6  00000011445YR4  00001101XF50CE  00000256211MR4  00000243246MR6  00000267466MR1  00000254189MR4  00000294636MR6  00000258409MR2  00000250446MR2  00000262316MR3  00000255603MR3  00000253828MR8  00000264014MR2  00000264549MR7  00000229475MR9  00000237710MR9  00000264222MR1  00000239238MR9  00000247733MR9  00000253168MR9  00000242590MR8  00000245407MR2  00000267101MR4  00000238365MR1  00000246616MR7  00000251641MR7  00000237073MR2  00000258093MR4  00000242591MR6  00000272561MR2  00000245668MR9  00000230689MR2  00000232403MR6  00000232404MR4  00000262932MR7  00000278304MR1  00000268763MR0  00000256703MR0  00000266465MR4  00000270633MR1  00000279080MR6  00000238181MR2  00000262052MR4  00000266448MR0  00000262805MR5  00000274690MR7  00000271139MR8  00000274171MR8  00000265180MR0  00000233295MR5  00000265922MR5  00000274837MR4  00000265006MR7  00000263340MR2  00000265092MR7  00000242545MR2  00000261063MR2  00000250191MR4  00000011888YR5  00000272971MR3  00000012066YR7  00000296050MR8  00000283954MR6  00000012251YR5  00000287632MR4  00000284355MR5  00000297628MR0  00000281702MR1  00000286894MR1  00000284051MR0  00000296403MR9  00000295522MR7  00000278195MR3  00000278394MR2  00000283951MR2  00000012044YR4  00000011520YR4  00000011911YR5  00000295508MR6  00000286922MR0  00000233422MR5  00000012164YR0  00000011546YR9  00000297372MR5  00000283591MR6  00000296146MR4  00000239029MR2  00000287650MR6  00000294419MR7  00000298071MR2  00000275635MR1  00000281794MR8  00000297659MR5  00000253169MR7  00000277579MR9  00000011594YR9  00000295507MR8  00000295699MR3  00000265105MR7  00000288668MR7  00000294624MR2  00000299676MR7  00000252298MR5  00000300939MR6  00000286265MR4  00000264005MR0  00000299664MR3  00000298489MR6  00000255192MR7  00000241144MR5  00000250199MR7  00000288591MR1  00000265724MR5  000000HDMR3707  00000278312MR4  00000280690MR9  00000280578MR6  00000243768MR9  00000236626MR8  00000258082MR7  00000259578MR3  00000246624MR1  00000262907MR9  00000261572MR2  00000268239MR1  00000229473MR4  00000253623MR3  00000255615MR7  00000245406MR4  00000250448MR8  00000260100MR3  00000237072MR4  00000233442MR3  00000252659MR8  00000232323MR6  00000263946MR6  00000241135MR3  00000239861MR8  00000261676MR1  00000271249MR5  00000272970MR5  00000267654MR2  00000264405MR2  00000271660MR3  00000266270MR8  00000270390MR8  00000282080MR1  00000267328MR3  00000265064MR6  00000011863YR8  00000289848MR4  00000272624MR8  00000285797MR7  00000298690MR9  00000006748YR8  00000277444MR6  00000242557MR7  00000272967MR1  00000286159MR9  00000286624MR2  00000237933MR7  00000237942MR8  00000256289MR0  000000HDMR3145  00000012024YR6  00000281679MR1  00000298555MR4  00000287283MR6  00000011870YR3  00000279066MR5  00000279161MR4  00000280672MR7  00000282546MR1  00000258413MR4  00001003BD6419  00000918BD5B34  00000011943YR8  000000HDMR4228  00000287441MR0  00000278393MR4  00000286156MR5  00000278565MR7  00000919BD5B92  00000280526MR5  00000007852YR7  00000008100YR0  00000260651MR5  00000270252MR0  00000284598MR0  00000008399YR8  00000280362MR5  00000278392MR6  00000281376MR4  00000280952MR3  00000278390MR0  00000279870MR0  00000286269MR6  00000287838MR7  00000011720YR0  00000269580MR7  UNK51078490003  00000265957MR1  00001101XF5026  00000277783MR7  00000272966MR3  00000011980YR2  00000297836MR9  00000296236MR3  00000281159MR4  00000268370MR4  00000296409MR6  00000293527MR8  00000280692MR5  00000262463MR3  00000261203MR4  00000300580MR8  00000290255MR9  00000284686MR3  00000283938MR9  00000274052MR0  00000282641MR0  00000283362MR2  00000278721MR6  00000262505MR1  00000252657MR2  00000258412MR6  00000264561MR2  00000276652MR5  00000271653MR8  00000274105MR6  00000280769MR1  00000242562MR7  00000281344MR2  00000012076YR6  00000274689MR9  00000282583MR4  00000276458MR7  00000011567YR5  00000286650MR7  00000950BD62AB  00000300703MR6  00000297906MR0  00000245675MR4  00000282553MR7  00000282682MR4  00000278468MR4  00000274936MR4  00000012075YR8  00000260627MR5  00000262460MR9  00000264616MR4  00000270245MR4  00000262198MR5  00000284390MR2  00000270226MR4  00000278668MR9  00000283486MR9  00000299272MR5  00000006658YR9  00000262407MR0  00000258396MR1  00000254675MR2  00000245409MR8  00000241867MR1  00000264800MR4  00000273343MR4  00000919BD5BA2  00000270970MR7  00000243683MR0  00000234793MR8  00000237004MR7  00000 233445MR6  00000240104MR0  00000292439MR7  00000268521MR2  00000283409MR1  00000254671MR1  00000243699MR6  00000299570MR2  00000251018MR8  00000276655MR8  00000253844MR5  00000237704MR2  00000297862MR5  00000262813MR9  00000262814MR7  00000262736MR2  00000265099MR2  00000243414MR0  00000296314MR8  00000241513MR1  00000245300MR9  00000253833MR8  00000256705MR5  00000242177MR4  00000237934MR5  00000272435MR9  00000284297MR9  00000286145MR8  00000272622MR2  00000265178MR4  00000274832MR5  00000270025MR0  00000286281MR1  00000264745MR1  00000265059MR6  00000261205MR9  00000267207MR9  00000273677MR5  00000287550MR8  00000298719MR6  00000298844MR2  00000296103MR5  00000012359YR6  00000011930YR5  00000012319YR0  00000248176MR0  00000284023MR9  00000272433MR4  00000007527YR5  00000287875MR9  00000286766MR1  00000012128YR5  00000012051YR9  00000284739MR0  00000011119YR5  00000280004MR3  00000298466MR4  00000292467MR8  00000011933YR9  00000288253MR8  00000012259YR8  00000300885MR1  00000296215MR7  00000294159MR9  00000011924YR8  00000011893YR5  00000294161MR5  00000298971MR3  00000295808MR0  00000298931MR7  00000287575MR5  00000012342YR2  00000010513YR0  00000296235MR5  00000297203MR2  00000297699MR1  00000286734MR9  00000250436MR3  00000300475MR1  00000295217MR4  00000296204MR1  00000012159YR0  00000012120YR2  00000297417MR8  00000012386YR9  00000294145MR8  00000298311MR2  00000282008MR2  00000274886MR1  00000276459MR5  00000009417YR7  00000284279MR7  00000278621MR8  00000278748MR9  00000280365MR8  00000286613MR5  00000301066MR7  00000296147MR2  00000922BD5C36  00000295028MR5  00000282511MR5  00000276585MR7  00000279408MR9  00000294146MR6  00000296216MR5  00000301350MR5  00000299663MR5  00000273369MR9  00000282266MR6  00000276676MR4  00000296232MR2  00000294659MR8  00000294341MR3  00000294434MR6  00000288821MR2  00000284642MR6  00000279448MR5  00000280858MR2  00000286601MR0  00000011935YR4  00000012166YR5  00000012373YR7  00000294307MR4  00000298224MR7  00000290408MR4  000000HDMR4252  00000279915MR3  00000007894YR9  00000281187MR5  00000275667MR4  00000297380MR8  00000232427MR5  00000260690MR3  00000261204MR2  00000281158MR6  00000282813MR5  00000287313MR1  00000282379MR7  00000011949YR5  00000028870MR1  00000295900MR5  00000296048MR2  00000241146MR0  00000250033MR8  00000258369MR8  00000251092MR3  00000253165MR5  00000252299MR3  00000243763MR0  00000251242MR4  00000241141MR1  00000257728MR6  00000260676MR2  00000280273MR4  00000275668MR2  00000286697MR8  00000287025MR1  00000297629MR8  00000297724MR7  00000297487MR1  00000011745YR7  00000297819MR5  00000296518MR4  00000299197MR4  00000298548MR9  00000299195MR8  00000299204MR8  00000299196MR6  00000283038MR8  00000831BD4A21  00000273538MR9  00000289266MR9  00000249930MR9  00000273530MR6  00000283156MR8  00000279612MR6  00000297657MR9  00000298495MR3  00000276197MR1  00000238347MR9  00000283023MR0  00000252405MR6  00000235244MR1  00000270567MR1  00000012060YR0  00000287360MR2  00000012302YR6  00000289821MR1  00000263872MR4  00000273622MR1  00000289774MR2  00000278799MR2  00000012197YR0  00000255594MR4  00000283574MR2  00000028626MR7  00000008364YR2  00000241136MR1  00000239031MR8  00001011BD6643  00000271659MR5  00000270389MR0  00000283398MR6  00000299425MR9  00000251245MR7  00000276660MR8  00000230106MR7  00000245400MR7  00000254499MR7  UNK2357500-201  UNK2357500-202  00000263575MR3  00000261737MR1  00000255616MR5  UNK2357500-204  00000250197MR1  00000247718MR0  UNK2357500-205  00000238359MR4  00000300710MR1  00000236502MR1  00000237075MR7  00000271650MR4  00000245404MR9  00000245818MR0  00000268323MR3  00000250441MR3  00000266675MR8  00000254202MR5  00000268117MR9  00000272075MR3  00000266204MR7  00000272287MR4  00000276138MR5  00000283792MR0  00000264096MR9  00000266743MR4  00000282730MR1  00000268115MR3  00000271780MR9  00000275738MR3  00000284337MR3  00000281858MR1  00000273226MR1  00000264094MR4  00000265065MR3  00000278613MR5  00000272079MR5  00000283039MR6  00000279452MR7  00000265614MR8  00000273534MR8  00000260452MR8  00000281866MR4  00000278558MR2  00000266070MR2  00000026397MR9  00000277952MR8  00000268114MR6  00000268510MR5  00000279454MR3  00000270628MR1  00000285690MR4  00000280902MR8  00000012068YR3  00000011925YR5  00000298583MR6  00000838BD4EC9  00000024341MR2  00000248134MR9  00000301566MR6  00000289997MR9  00000288711MR5  00000284296MR1  00000283343MR2  00000298010MR0  00000277757MR1  00000011576YR6  00000028405MR2  00000295442MR8  00000011522YR0  00000293589MR8  00000289439MR2  00000287116MR8  00000299252MR7  00000918BD5B33  00000274715MR2  00000253163MR0  00000265867MR2  00000278103MR7  00000274174MR2  00000284408MR2  00000287421MR2  00000276242MR5  00000288247MR0  00000236997MR3  00000300555MR0  00000012069YR1  00000284492MR6  00000250453MR8  00000295484MR0  00000298757MR6  00000298532MR3  00000283753MR2  00000301106MR1  00000281443MR2  00000296110MR0  00000279908MR8  00000294407MR2  00000294443MR7  00000299326MR9  00000282494MR4  00000294723MR2  00000012261YR4  00000919BD5B8D  00000287422MR0  00000267320MR0  00000273461MR4  00000011509YR7  00000283458MR8  00000299849MR0  00000277532MR8  00000297971MR4  00000298673MR5  00000238345MR3  00000247218MR1  00000250460MR3  00000230690MR0  00000258080MR1  00000232408MR5  00000256987MR9  00000238170MR5  00000237001MR3  00000250204MR5  00000247221MR5  00000257435MR8  00000256670MR1  00000239036MR7  00000235594MR9  00000237003MR9  00000235592MR3  00000252921MR2  00000233429MR0  00000252656MR4  00000250444MR7  00000234578MR3  00000245802MR4  00000261730MR6  00000255595MR1  00000253154MR9  00000251698MR7  00000245815MR6  00000247220MR7  00000255593MR6  00000235591MR5  00000241142MR9  00000235593MR1  00000233438MR1  00000231912MR7  00000232423MR4  00000232410MR1  00000234795MR3  00000231907MR7  00000237066MR6  00000235595MR6  00000232418MR4  00000234860MR5  00000237060MR9  00000237706MR7  00000235691MR3  00000235696MR2  00000237078MR1  00000234799MR5  00000241873MR9  00000236828MR0  00000237707MR5  00000237065MR8  00000250461MR1  00000238165MR5  00000250459MR5  00000237709MR1  00000232422MR6  00000234838MR1  00000237941MR0  00000254507MR7  00000238354MR5  00000239039MR1  00000259977MR7  00000238180MR4  00000239030MR0  00000238350MR3  00000249257MR7  00000242172MR5  00000236627MR6  00000239233MR0  00000241520MR6  00000241718MR6  00000243415MR7  00000242175MR8  00000242179MR0  00000240108MR1  00000242504MR9  00000243697MR0  00000243416MR5  00000241679MR0  00000243692MR1  00000248166YM1  00000248486MR3  00000245673MR9  00000248157MR0  00000245862MR8  00000243188MR0  00000246634MR0  00000243417MR3  00000251228MR3  00000249228MR8  00000248170MR3  00000248482MR2  00000248487MR1  00000248161MR2  00000247212MR4  00000251095MR6  00000251694MR6  00000247735MR4  00000243419MR9  00000250215MR1  00000251640MR9  00000252922MR0  00000252934MR5  00000253830MR4  00000254673MR7  00000247210MR8  00000251693MR8  00000247223MR1  00000250195MR5  00000253829MR6  00000251723MR3  00000253146MR5  00000255216MR4  00000253160MR6  00000250450MR4  00000256707MR1  00000257428MR3  00000258092MR6  00000259133MR7  00000259382MR0  00000257726MR0  00000258374MR8  00000258417MR5  00000256989MR5  00000252427MR0  00000259135MR2  00000261592MR0  00000260018MR7  00000261681MR1  00000255608MR2  00000259576MR7  00000262313MR0  00000258397MR9  00000257440MR8  00000262315MR5  00000265821MR9  00000258408MR4  00000259126MR1  00000262408MR8  00000263260MR2  00000270270MR2  00000242178MR2  00000242559MR3  00000234857MR1  00000239532MR5  00000255611MR6  00000294569MR9  00000258398MR7  00000247214MR0  00000296261MR1  00000298103MR3  00000237938MR6  00000273017MR4  00000260673MR9  00000292024MR7  00000296819MR6  00000264556MR2  00000259381MR2  00000263576MR1  00000292215MR1  00000253155MR6  00000237068MR2  00000270568MR9  00000259569MR2  00000260101MR1  00000272528MR1  00000260461MR9  00000269376MR0  00000270570MR5  00000272076MR1  00000266859MR8  00000270566MR3  00000273620MR5  00000273466MR3  00000272431MR8  00000257441MR6  00000262074MR8  00000269435MR4  00000272429MR2  00000237935MR2  00000265781MR5  00000278305MR8  00000261728MR0  00000268762MR2  00000272338MR5  00000270463MR3  00000280525MR7  00000270977MR2  00000277578MR1  00000264552MR1  00000263791MR6  00000272626MR3  00000278363MR7  00000279145MR7  00000269338MR0  00000270683MR6  00000262404MR7  00000238363MR6  00000273239MR4  00000262314MR8  00000278364MR5  00000302302MR5  00000302746MR3  00000271270MR1  00000267200MR4  00000267203MR8  00000268477MR7  00000268515MR4  00000269575MR7  00000270026MR8  00000272969MR7  00000272070MR4  00000273267MR5  00000271662MR9  00000263338MR6  00000263878MR1  00000263873MR2  00000261880MR9  00000265918MR3  00000262077MR1  00000265921MR7  00000266202MR1  00000266676MR6  00000268366MR2  00000263265MR1  00000268010MR6  00000267812MR6  00000268289MR6  00000268693MR9  00000268008MR0  00000268472MR8  00000268420MR7  00000272071MR2  00000269461MR0  00000271661MR1  00000271664MR5  00000272625MR5  00000269341MR4  00000268686MR3  00000266462MR1  00000272073MR8  00000272074MR6  00000267465MR3  00000272100MR9  00000272889MR7  00000272898MR8  00000272999MR4  00000270565MR5  00000273360MR8  00000271665MR2  00000275582MR5  00000273225MR3  00000275735MR9  00000275599MR9  00000272965MR5  00000275087MR5  00000275737MR5  00000276492MR6  00000266146MR0  00000266148MR6  00000276272MR2  00000266745MR9  00000274173MR4  00000269463MR6  00000279404MR8  00000279459MR2  00000270022MR7  00000272337MR7  00000273489MR5  00000272289MR0  00000273464MR8  00000266064MR5  00000275358MR0  00000270624MR0  00000279148MR1  00000278955MR0  00000277924MR7  00000278469MR2  00000279451MR9  00000283358MR0  00000283576MR7  00000276631MR9  00000233428MR2  00000236996MR5  00000237076MR5  00000242503MR1  00000237940MR2  00000236998MR1  00000239245MR4  00000245671MR3  00000251699MR5  00000251713MR4  00000276436MR3  00000260693MR7  00000262078MR9  00000261682MR9  00000262197MR7  00000261060MR8  00000262461MR7  00000260012MR0  00000261684MR5  00000263819MR5  00000012041YR0  00000287916MR1  00000243694MR7  00000287942MR7  00000298411MR0  00000284385MR2  00000294894MR1  00000237062MR5  00000289839MR3  00000011723YR4  00000255606MR6  00000296358MR5  00000283717MR7  00000297810MR4  00000011773YR9  00000294324MR9  00000286007MR0  00000301065MR9  00000282465MR4  00000012320YR8  822  00000288151MR4  00000286272MR0  00000294321MR5  00000298505MR9  00000919BD5B84  00000902BD55AE  00000263940MR9  00000289210MR7  00000011799YR4  00000302425MR4  00000289392MR3  00000249259MR3  00000299912MR6  00000284313MR4  00000296402MR1  00000285628MR4  00000294115MR1  00000267618MR7  00000295745MR4  00000294960MR0  00000298948MR1  00000012309YR1  00000295095MR4  00000296072MR2  00000930BD5D40  00000816BD3AA6  00000278127MR6  00000288073MR0  00000012156YR6  00000296213MR2  00000296273MR6  00000012124YR4  00000011891YR9  00000012002YR2  00000285728MR2  00000294652MR3  00000012318YR2  00000269584MR9  00000284759MR8  00000287629MR0  00000287881MR7  00000298076MR1  00000011737YR4  00000258402MR7  00000300614MR5  00000299844MR1  00000294647MR3  00000285048MR5  00000295748MR8  00000289790MR8  00000286543MR4  00000011922YR2  00000287460MR0  00000268837MR2  00000268047MR8  00000269332MR3  00000268417MR3  00000273266MR7  00000264749MR3  00000268517MR0  00000272098MR5  00000274169MR2  00000277580MR7  00000276244MR1  00000277960MR1  00000278040MR1  00000278801MR6  00000278556MR6  00000281375MR6  00000280675MR0  00000275527MR0  00000273007MR5  00000280366MR6  00000282337MR5  00000009032YR4  00000281316MR0  00000279906MR2  00000283773MR0  00000283485MR1  00000278221MR7  00000284392MR8  00000284464MR5  00000277533MR6  00000285051MR9  00000277643MR3  00000286617MR6  00000285627MR6  00000278608MR5  00000287393MR3  00000283734MR2  00000286691MR1  00000282003MR3  00000286968MR3  00000286596MR2  00000287269MR5  00000287070MR7  00000287399MR0  00000287361MR0  00000287592MR0  00000286469MR2  00000287900MR5  00000287851MR0  00000287866MR8  00000289657MR9  00000289194MR3  00000011597YR2  00000012048YR5  00000012058YR4  00000289320MR4  00000012053YR5  00000012061YR8  00000012055YR0  00000011777YR0  00000012079YR0  00000281934MR0  00000012182YR2  00000012074YR1  00000012130YR1  00000011912YR3  00000011505YR5  00000011807YR5  00000011261YR5  00000814BD39F1  00000012022YR0  00000293562MR5  00000011729YR1  00000012045YR1  00000919BD5BB8  00000292642MR6  00000277445MR3  00000011557YR6  00000294709MR1  00000294905MR5  00000295189MR5  00000011999YR0  00000919BD5BBE  00000012347YR1  00000285165MR7  00000922BD5C3D  00000919BD5BBA  00000292088MR2  00000012238YR2  00000275968MR6  00000278568MR1  00000277646MR6  00000278397MR5  00000278849MR5  00000276719MR2  00000283026MR3  00000281341MR8  00000283406MR7  00000283116MR2  00000276720MR0  00000276434MR8  00000283114MR7  00000281152MR9  00000011200YR3  00000012123YR6  00000011545YR1  00000011735YR8  00000292942MR0  00000284261MR5  00000280780MR8  00000296036MR7  00000294927MR9  00000295697MR7  00000295239MR8  00000298404MR5  00000294926MR1  00000283775MR5  00001101XF50A6  00000295594MR6  00000290324MR3  00000292984MR2  00000282550MR3  00000279407MR1  00000289384MR0  00000286153MR2  00000237697MR8  00000287314MR9  00000287827MR0  00000245674MR7  00000245820MR6  00000295896MR5  00000295234MR9  00000295237MR2  00000295542MR5  00000295702MR5  00000296092MR0  00000295895MR7  00000296202MR5  00000296304MR9  00000293323MR2  00000296411MR2  00000294555MR8  00000295897MR3  00000296368MR4  00000297204MR0  00000298558MR8  00000297463MR2  00000298582MR8  00000292688MR9  00000296644MR8  00000299493MR7  00000301405MR7  00000296457MR5  00000297626MR4  00001102XF506C  00000296306MR4  00001101XF50B3  00001101XF50BA  00000301243MR2  00001105XF5010  00000298186MR8  00000259647MR6  00000261680MR3  00000269436MR2  00000301351MR3  00000301859MR5  00000267728MR4  00000257427MR5  00000301833MR0  00000268690MR5  00000236827MR2  00000298857MR4  00000251700MR1  00000268470MR2  00000298954MR9  00000300495MR9  00000248175MR2  00000260014MR6  00000299473MR9  00000301338MR0  00000299524MR9  00000299829MR2  00000296361MR9  00000299186MR7  00000300697MR0  00000301339MR8  00000248484MR8  00000300185MR6  00000277691MR2  00000297904MR5  00000299827MR6  00000299268MR3  00000265104MR0  00000302652MR3  00000300610MR3  00000300730MR9  00000298538MR0  00000299244MR4  00000232405MR1  00000293974MR2  00000301910MR6  00000297710MR6  00000299553MR8  00000241134MR6  00000232401MR0  00000298541MR4  00000270528MR3  00000270097MR9  00000268301MR9  00000267860MR5  00000272968MR9  00000268692MR1  00000270467MR4  00000272536MR4  00000267652MR6  00000266824MR2  00000267471MR1  00000275964MR5  00000277961MR9  00000277920MR5  00000276679MR8  00000278163MR1  00000281364MR0  00000278674MR7  00000285602MR9  00000011995YR8  00000288160MR5  00000012004YR8  00000012016YR2  00000292288MR8  00000919BD5BB6  00000295192MR9  00000277764MR7  00000285674MR8  00000288939MR2  00000282732MR7  00000286625MR9  00000277211MR9  00000287568MR0  00000292078MR3  00000280782MR4  00000011959YR4  00000012001YR4  00000292983MR4  00000289655MR3  00000295240MR6  00000281078MR6  00000298197MR5  00000288359MR3  00000296037MR5  00000299666MR8  00000299274MR1  00000299315MR2  00000296108MR4  00000296049MR0  00000298410MR2  00000300771MR3  00000298933MR3  00000299316MR0  00000298409MR4  00000285053MR5  00000236701MR9  00000284869MR5  00000257442MR4  00000243250MR8  00000249655MR2  00000249757MR6  00000249230MR4  00000248172MR9  00000295809MR8  00000298549MR7  00000298825MR1  00000298282MR5  00000297903MR7  00000298756MR8  00000296426MR0  00000298715MR4  00000299641MR1  00000299203MR0  00000299470MR5  00000301737MR3  00000299270MR9  00000301816MR5  00000301736MR5  00000301817MR3  00000295475MR8  00000299183MR4  00000301916MR3  00000295901MR3  00000301815MR7  00000299194MR1  00000254504MR4  00000252424MR7  00000254489MR8  00000254196MR9  00000247457MR5  00000254496MR3  00000253633MR2  00000251244MR0  00000266464MR7  00000297489MR7  00000298189MR2  00000298069MR6  00000284866MR1  00000294520MR2  00000298754MR3  00000243244MR1  00000257424MR2  00000263256MR0  00000243410MR8  00000279077MR2  00000262798MR2  00000274708MR7  00000250200MR3  00000249763MR4  00000279563MR1  00000289819MR5  00000282503MR2  00000297863MR3  00000256219MR7  00000274175MR9  00000295540MR9  00000268526MR1  00000011881YR0  00000288823MR8  00000296521MR8  00000286254MR8  00000011942YR0  00000239237MR1  00000262153MR0  00000012077YR4  00000288014MR4  00000284710MR1  00000278389MR2  00000011513YR9  00000009569YR5  00000011500YR6  00000298584MR4  000000HDMR6286  00000242593MR2  00000289588MR6  00000006589YR6  1112  00000286509MR5  00000253167MR1  00000284833MR1  00000254194MR4  00000262791MR7  00000289914MR4  00000281363MR2  00000249254MR4  00000252434MR6  00000257419MR2  00000249280MR9  00000259579MR1  00000280243MR7  00000249260MR1  00000300574MR1  00000297885MR6  00000270468MR2  00000276276MR3  00000276273MR0  00000288212MR4  00000286777MR8  00000259120MR4  00000269344MR8  00000273518MR1  00000286146MR6  00000262257MR9  00000273465MR5  00000270466MR6  00000011318YR3  00000010509YR8  00000011779YR6  00000294116MR9  00000295474MR1  00000296365MR0  00000298213MR0  00000298845MR9  00000294722MR4  00000287023MR6  00000271777MR5  00000273539MR7  00000299947MR2  00000284024MR7  00000286411MR4  00000227450MR7  00000260464MR3  00000295015MR2  00000299757MR5  00000011782YR0  00000283405MR9  00000288772MR7  00000287475MR8  00000241131MR2  00000245669MR7  00000259379MR6  00000239536MR6  00000284578MR2  00000261940MR1  00000235685MR5  00000282383MR9  00000282549MR5  00000285498MR2  00000299830MR0  00000275357MR2  00000285020MR4  00000259584MR3  00000269348MR9  00000243769MR7  00000276659MR0  00000286384MR2  00000271125MR7  00000282810MR1  00000238177MR0  00000238868MR4  00000259581MR7  00000277958MR5  00000267214MR5  00000277531MR0  00000268511MR3  00000266737MR6  00000276265MR6  00000282923MR2  000000000R1997  00000252440MR3  00000278687MR9  00000255205MR7  00000269454MR5  00000280068MR8  00000012253YR1  000000HDMR2522  00000243690MR5  00000009612YR3  00000295829MR6  00000296864MR2  00000297013MR5  00000289589MR4  00000263254MR5  00000269392MR7  00000273344MR2  00000293239MR0  00000284236MR7  00000284867MR9  00000284019MR7  00000011733YR3  00000297808MR8  00000294525MR1  00000280778MR2  00000282865MR5  00000287275MR2  00000291833MR2  00000250440MR5  00000298373MR2  00000291331MR7  00000298851MR7  00000258373MR0  00000291674MR0  00000292364MR7  00000277602MR9  00000262671MR1  00000258400MR1  00000247216MR5  00000918BD5B3A  00000300938MR8  00000919BD5BA4  00000299168MR5  00000011921YR4  00000286158MR1  00000288705MR7  00000289361MR8  00000299185MR9  00000010985YR0  00000283948MR8  00000283127MR9  00000286275MR3  00000285049MR3  00000290000MR9  00000278560MR8  00000011868YR7  00000012052YR7  00000012349YR7  00000919BD5BBD  00000293023MR8  00000296362MR7  00000298276MR7  00000298679MR2  00000298955MR6  00000300798MR6  00000298970MR5  00000301385MR1  00000251057MR6  00000298758MR4  00000278463MR5  00000261574MR8  00000293279MR6  00000235701MR0  00000290947MR1  00000267093MR3  00000268136MR9  00000298377MR3  00000283351MR5  00000290033MR0  00000291875MR3  00000268009MR8  00000287382MR6  00000296237MR1  00000012191YR3  00000296464MR1  00000287230MR7  00000288462MR5  00000233271MR6  00000276591MR5  00000273366MR5  00000279460MR0  00000272533MR1  00000239869MR1  00000268012MR2  00000300982MR6  00000300810MR9  00000281446MR5  00000011187YR2  00000298531MR5  00000301848MR8  00000294660MR6  00000012348YR9  00000255199MR2  00000282897MR8  00000291505MR6  00000290570MR1  00000257735MR1  00000282029MR8  00000251799MR3  00000286263MR9  00000288041MR7  00000011542YR8  00000294692MR9  00000011785YR3  00000229366MR0  00000265607MR2  00000293051MR9  00000265101MR6  00000252924MR6  00000265098MR4  00000260671MR3  00000248171MR1  00000258420MR9  00000267103MR0  00000271658MR7  00000273227MR9  00000274071MR0  00000276494MR2  00000273457MR2  00000268284MR7  00000273009MR1  00000275845MR6  00000289845MR0  00000011781YR2  00000286798MR4  00000286273MR8  00000011771YR3  00000254185MR2  00000278549MR1  00000011795YR2  00000287316MR4  00000288122MR5  00000286471MR8  00000284287MR0  00000918BD5B53  00000011802YR6  00000011554YR3  00000283843MR1  00000011559YR2  00000297012MR7  00000288387MR4  00000011794YR5  00000275017MR2  00000285054MR3  00000012271YR3  00000294693MR7  00000298042MR3  00000248163MR8  00000302358MR7  00000294946MR9  00000293651MR6  00000280923MR4  00000284714MR3  00000289765MR0  00000299253MR5  00000294895MR8  00000295090MR5  00000298684MR2  00000299910MR0  00000011517YR0  00000282382MR1  00000263944MR1  00000233424MR1  00000268522MR0  00000243185MR6  00000256207MR2  00000266670MR9  00000285301MR8  00000283755MR7  00000282464MR7  00000282958MR8  00000282083MR5  00000286687MR9  00000295700MR9  00000295877MR5  00000286965MR9  00000286786MR9  00000295762MR9  00000255202MR4  00000278627MR5  00000011515YR4  00000294130MR0  00000285673MR0  00000011593YR1  00000905BD567D  00000296091MR2  00000289976MR3  00000011307YR6  00000300745MR7  00000298707MR1  00000240111MR5  00000252426MR2  00000011063YR5  00000278844MR6  00000283620MR3  00000288888MR1  00000269046MR9  00000274070MR2  00000285352MR1  00000012049YR3  00000294392MR6  000000HDMR2875  00000011713YR5  00000011560YR0  00000012323YR2  00000272427MR6  00000238341MR2  00000298647MR9  00000284387MR8  00000285018MR8  00001043XF5019  00000250207MR8  00000268529MR5  00000262792MR5  00000270626MR5  00000268474MR4  00000287379MR2  00000288019MR3  00000298972MR1  00000011503YR0  00000275465MR3  00000246619MR1  00000250206MR0  00000257423MR4  00000246635MR7  00000259134MR5  00000233425MR8  00000237005MR4  00000240109MR9  00000246618MR3  00000249657MR8  215804  00000250420MR7  00000250193MR0  00000250457MR9  00000251719MR1  00000251717MR5  00000252925MR3  00000253841MR1  00000256204MR9  00000254661MR2  00000256693MR3  00000261628MR2  00000261732MR2  00000262100MR1  00000264459MR9  00000265608MR0  00000256223MR9  00000263241MR7  00000267853MR0  00000270524MR2  00000271863MR3  00000257741MR9  00000283361MR4  00000287765MR2  00000243407MR4  00000233441MR5  00000288144MR9  00000234805MR0  00000233266MR6  00000237950MR1  00000251712MR6  00000230663MR7  00000239243MR9  00000294475MR9  00000270272MR8  00000293052MR7  00000237691MR1  00000254197MR7  00000298726MR1  00000264008MR4  00000275356MR4  00000276590MR7  00000271322MR0  00000260692MR9  00000261202MR6  00000263341MR0  00000267806MR8  00000270464MR1  00000270391MR6  00000273362MR4  00000274150MR2  00000273522MR3  00000275847MR2  00000275935MR5  00000276332MR4  00000282084MR3  00000283300MR2  00000285914MR8  00000285300MR0  00000286060MR9  00000286058MR3  00000286061MR7  00000291852MR2  00000264746MR9  00000300776MR2  00000284863MR8  00000290008MR2  00000289220MR6  00000274740MR0  00000256706MR3  00000262807MR1  00000273363MR2  00000272530MR7  00000273674MR2  00000276734MR1  00000277215MR0  00000279920MR3  00000280861MR6  00000282685MR7  00000283791MR2  00000286551MR7  00000285435MR4  00000285917MR1  00000287120MR0  00000287576MR3  00000283152MR7  00000288207MR4  00000289360MR0  00000289198MR4  00000011712YR7  00000012311YR7  00000845BD51D2  00000294322MR3  00000293027MR9  00000294420MR5  00000294506MR1  00000294606MR9  00000294961MR8  00000295089MR7  00000295051MR7  00000292643MR4  00000918BD5B55  00000296642MR2  00000296363MR5  00000281255MR0  00000294395MR9  00000297821MR1  00000297008MR5  00000297697MR5  00000297878MR1  00000298487MR0  00000299651MR0  00000300476MR9  00000300504MR8  00000301764MR7  00000302644MR0  000000HDMR5719  00000GE0208031  00000GE0249066  00000288193MR6  00000287425MR3  00000289339MR4  00000265005MR9  00000297421MR0  00000281056MR2  00000243187MR2  00000289319MR6  00000261729MR8  00000270271MR0  00000270269MR4  00000242556MR9  00000268530MR3  00000258394MR6  00000245670MR5  00000284362MR1  00000242584MR1  UNK22943020002  UNK22943020003  00000246627MR4  00000290550MR3  00000239235MR5  00000254498MR9  00000290571MR9  00000287812MR2  00000272881MR4  00000250198MR9  00000281746MR8  00000250194MR8  00000262149MR8  00000282046MR2  00000263796MR5  00000295560MR7  00000218379MR6  00000283359MR8  00000288898MR0  00000258404MR3  00000290355MR7  00000253162MR2  00000248168MR7  00000234831MR6  00000255588MR6  00000262050MR8  00000265818MR5  00000261939MR3  00000245942MR8  00000241517MR2  00000254486MR4  00000241521MR4  00000238176MR2  00000265822MR7  00000250443MR9  00000243762MR2  00000243418MR1  00000235587MR3  00000241516MR4  00000252918MR8  00000290647MR7  00000290038MR9  00000264559MR6  00000268316MR7  00000274164MR3  00000261734MR8  00000272099MR3  00000268476MR9  00000280619MR8  00000279800MR7  00000280713MR9  00000280759MR2  00000280734MR5  00000281313MR7  00000276199MR7  00000267036MR2  00000275473MR7  00000281865MR6  00000275016MR4  00000275965MR2  00000281857MR3  00000278210MR0  00000278307MR4  00000268418MR1  00000272436MR7  00000275642MR7  00000282079MR3  00000268320MR9  00000260026MR0  00000277921MR3  00000268291MR2  00000277582MR3  00000283150MR1  00000011788YR7  00000289357MR6  00000285692MR0  00000286621MR8  00000297905MR2  00000919BD5B93  00000289887MR2  00000298658MR6  00000293416MR4  00000280700MR6  00000300700MR2  00000275466MR1  00000289658MR7  00000262455MR9  00000274288MR0  00000284577MR4  00000280781MR6  00000294285MR2  00000288435MR1  00000286157MR3  00000918BD5B57  00000294396MR7  00000295591MR2  00000283433MR1  00000288012MR8  00000288614MR1  00000012134YR3  00000295530MR0  00000283516MR3  00000012011YR3  00000280582MR8  00000289847MR6  00000299314MR5  00000299254MR3  00000299310MR3  00001048XF5021  00000299648MR6  00000289394MR9  00000294329MR8  00000280327MR8  00000287847MR8  00000012046YR9  00000286287MR8  00000299650MR2  00000299260MR0  00000254375MR9  00000286250MR6  00000237063MR3  00000283320MR0  00000266069MR4  00000027810MR1  00000297404MR6  00000296047MR4  00000292641MR8  00000012187YR1  00000295902MR1  00000012127YR7  00000282999MR2  00000245814MR9  00000258077MR7  00000262064MR9  00000283302MR8  00000290007MR4  00000241874MR7  00000284233MR4  UNK51078490008  00000259130MR3  00000231908MR5  00000240110MR7  00000264547MR1  00000266821MR8  00000265719MR5  00000267810MR0  00000011734YR1  00000242174MR1  00000266820MR0  00000272285MR8  00000241871MR3  00000283866MR2  00000274714MR5  00000263943MR3  00000277200MR2  00000247211MR6  00000259124MR6  00000257439MR0  00000274935MR6  00000298804MR6  00000298265MR0  00000298790MR7  00000298806MR1  00000298791MR5  00000301038MR6  00000233423MR3  00000252430MR4  00000245940MR2  00000250458MR7  00000256236MR1  00000252915MR4  00000257737MR7  00000259425MR7  00000272565MR3  00000257751MR8  00000261475MR8  00000267105MR5  00000261199MR4  00000275014MR9  00000266669MR1  00000266826MR7  00000269581MR5  00000264620MR6  00000274697MR2  00000275699MR7  00000277600MR3  00000271239MR6  00000301750MR6  00000274694MR9  00000912BD59A1  00000268369MR6  00000268685MR5  00000278038MR5  00000275533MR8  00000279464MR2  00000279615MR9  00000284533MR7  00000285677MR1  00000286649MR9  00000289195MR0  00000288613MR3  00000289162MR0  00000288062MR3  00000011553YR5  00000011512YR1  00000289473MR1  00000011561YR8  00000010254YR3  00000012059YR2  00000296238MR9  00000296865MR9  00000298215MR5  00000299871MR4  00000301186MR3  00000281186MR7  00000012121YR0  00000294344MR7  00000298557MR0  00000277535MR1  00001004BD6441  00000259106MR3  00000GE0151026  00000GE0209032  00000267672MR4  00000265058MR8  00000275736MR7  00000283697MR1  00000287306MR5  00000279447MR7  00000293217MR6  00000298534MR9  00000300778MR8  00000254503MR6  00000249927MR5  00000251093MR1  00000296846MR9  00000297381MR6  00000296847MR7  00000296558MR0  00000296863MR4  00000295174MR7  00000241723MR6  00000282690MR7  00000289785MR8  00000298975MR4  00000287024MR4  00000246608MR4  00000256235MR3  00000270514MR3  00000260103MR7  00000264793MR1  00000259582MR5  00000264798MR0  00000270518MR4  00000260097MR1  00000270521MR8  00000260466MR8  00000272564MR6  00000270525MR9  00000256215MR5  00000258395MR3  00000261387MR5  00000264490MR4  00000266697MR2  00000274107MR2  00000268523MR8  00000253152MR3  00000263941MR7  00000264548MR9  00000283317MR6  00000266757MR4  00000279892MR4  00000278667MR1  00000280903MR6  00000273681MR7  00000279083MR0  00000283357MR2  00000280676MR8  00000276248MR2  00000271120MR8  00000274830MR9  00000262456MR7  00000266746MR7  00000283385MR3  00000285384MR4  00000274892MR9  00000283696MR3  00000285023MR8  00000286267MR0  00000275093MR3  00000282911MR7  00000284214MR4  00000261868MR4  00000277195MR4  00000275475MR2  00000275580MR9  00000286149MR0  00000294435MR3  00000011719YR2  00000296360MR1  00000294271MR2  00000286585MR5  00000277641MR7  00000296404MR7  00000294441MR1  00000012014YR7  00000010796YR1  00000299348MR3  00000299451MR5  00000011026YR2  00000299417MR6  00000295056MR6  00000295477MR4  00000294607MR7  00000287122MR6  00000298093MR6  00000011084YR1  00000295014MR5  00000298843MR4  00000297179MR4  00000284904MR0  00000286012MR0  00000287308MR1  00000287396MR6  000000HDMR5258  00000012013YR9  00000294510MR3  00000298097MR7  00000838BD4ECB  00000299494MR5  00000300186MR4  00000294907MR1  00000297973MR0  00000919BD5B86  00000294514MR5  00000011552YR7  000001106XF022  00000295481MR6  00000286290MR2  00000298488MR8  00000296366MR8  00000287389MR1  00000288979MR8  00000288891MR5  00000301039MR4  00000299675MR9  00000273525MR6  00000286282MR9  00000275844MR9  00000262300MR7  00000008340YR2  00000280857MR4  00000286262MR1  00000286686MR1  00000297445MR9  00000299674MR2  00000295513MR6  00001009BD65FE  00000230686MR8  00000257733MR6  00000263945MR8  00000296463MR3  00000266145MR2  00000274693MR1  00000277191MR3  00000277692MR0  00000276153MR4  00000280627MR1  00000276632MR7  00000290191MR6  00000914BD5A7E  00000294436MR1  00001002BD639A  00000286692MR9  00000287113MR5  00000931BD5D7B  00000294274MR6  00000300616MR0  00000300618MR6  00000011731YR7  00000286248MR0  00000288650MR5  00000011775YR4            REF # 67626    \"Products: GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM (Signa Ovation 5 and Ovation HD)  (Beijing, China)  FMI 67626\"  Mfg Lot or Serial #  000SPI275  138161HM5  154416HM2  167478HM7  169099HM9  171687HM7  176546HM0  188506HM0  192710HM2  193122HM9  202350HM5  213385HM8  226223HM6  238928HM6  241057HM9  266117HM1  139841HM1  210567HM4  150713HM6  160804HM1  162558HM1  164061HM4  165767HM5  172417HM8  172812HM0  172864HM1  173903HM6  177138HM5  177789HM5  178516HM1  179358HM7  180124HM0  181095HM1  181525HM7  182066HM1  182244HM4  182592HM6  183128HM8  187360HM3  188807HM2  189100HM1  191022HM3  191158HM5  196644HM9  200573HM4  202345HM5  204158HM0  205908HM7  206819HM5  210513HM8  213193HM6  214041HM6  214712HM2  215676HM8  216125HM5  217358HM1  218474HM5  219021HM3  219406HM6  222105HM9  223599HM2  224698HM1  224707HM0  228098HM0  229042HM7  230858HM3  231784HM0  234987HM6  242129HM5  243112HM0  244329HM9  245315HM7  245639HM0  250988HM3  252871HM9  256318HM7  257108HM1  262327HM0  265044HM8  267712HM8  270517HM6  270524HM2  271390HM7  271741HM1  157409HM4  148416HM1  174343HM4  175020HM7  199950HM7  205562HM2  218302HM8  155505HM1  212655HM5  155506HM9  195286HM0  223710HM5  231594HM3  233074HM4  157408HM6  147520HM1  189940HM0  198004HM4  198790HM8  224716HM1  245556HM6  224689HM0  229582HM2  199942HM4  181827HM7  227474HM4  226814HM2  214907HM8  239726HM3  241050HM4  223171HM0  247892HM3  271397HM2  271404HM6  271750HM2  276214HM4  146883HM4  158636HM1  147900HM5  169828HM1  170935HM1  190433HM3  191901HM8  193217HM7  193933HM9  195769HM5  197642HM2  224432HM5  259738HM3  270004HM5  144793HM7  150146HM9  221033HM4  170104HM4  181826HM9  195287HM8  208285HM7  219747HM3  220369HM3  247222HM3  257835HM9  221645HM5  168318HM4  181520HM8  220560HM7        Ref # 67815    \"Product: GE 0.7T SIGNA OPENSPEED WITH EXCITE MR SYSTEM, (HFO5), (Signa Openspeed 5)   (Hino, Japan)  FMI 67815D\"  Mfg Lot or Serial #  00000000102YR4  00000000127YR1  00000000133YR9  00000000142YR0  00000000144YR6  00000000146YR1  00000000149YR5  00000000151YR1  00000000155YR2  00000003011YR4  00000003022YR1  00000003023YR9  00000003024YR7  00000003025YR4  00000003026YR2  00000003027YR0  00000003028YR8  00000003029YR6  00000003030YR4  00000003031YR2  00000003033YR8  00000003034YR6  00000003035YR3  00000003036YR1  00000003039YR5  00000003041YR1  00000003042YR9  00000003043YR7  00000003044YR5  00000003050YR2  00000003051YR0   00000003052YR8  00000003053YR6  00000003049YR4  00000000122YR2  00000003047YR8    \"Product: GE 0.7T SIGNA OPENSPEED WITH EXCITE MR SYSTEM (HFO4), (Signa Openspeed 4)  CR No. 9612283-XX/XX/XX-00X-C (Hino, Japan)  FMI 67815C\"  Mfg Lot or Serial #  00000000100YR8  00000000109YR9  00000000113YR1  00000000118YR0  00000000119YR8  00000000120YR6  00000000128YR9  00000000129YR7  00000000132YR1  00000000134YR7  00000000136YR2  00000000150YR3  00000000152YR9  00000943610YM6  00000951021YM5  00000962592YM2  00000000124YR8  00000000137YR0    \"Product: GE 0.35T SIGNA OVATION WITH EXCITE (MFO4) MR SYSTEM (Signa Ovation 4)  CR No. 9612283-XX/XX/XX-00X-C (Hino, Japan)  FMI 67815B\"  Mfg Lot or Serial #  00000000123YR0  00000000125YR5  00000000130YR5  00000000138YR8  00000000139YR6  00000000561YR1  00000000563YR7  00000000564YR5  00000000586YR6  00000000756YR7  00000000757YR5  00000000758YR3  00000000759YR1  00000000760YR9  00000000768YR2  00000000770YR8  00000000771YR6  00000000772YR4  00000000773YR2  00000000775YR7  00000000784YR9  00000000785YR6  00000000786YR4  00000000788YR0  00000000796YR3  00000000798YR9  00000000809YR4  00000000812YR8  00000000816YR9  00000000818YR5  00000000825YR0  00000000826YR8  00000000827YR6  00000000835YR9  00000963007YM0  00000000804YR5  00000000778YR1  00000000779YR9  00000000780YR7  00000000789YR8  00000000790YR6  00000000793YR0  00000000795YR5  00000000799YR7  00000000800YR3  00000000801YR1  00000000803YR7  00000000805YR2  00000000807YR8  00000000808YR6  00000000810YR2  00000000819YR3  00000000828YR4  00000000830YR0  00000000831YR8  00000000832YR6  00000000834YR2  00000000836YR7  00000000837YR5  00000000838YR3  00000000839YR1  00000000840YR9  00000000841YR7  00000000811YR0  00000000140YR4  00000000141YR2  00000000565YR2  00000000763YR3  00000000766YR6  00000000767YR4  00000000769YR0  00000000787YR2  00000000761YR7  00000000116YR4  00000000117YR2  00000000777YR3  00000000143YR8  00000000781YR5  00000000776YR5  00000000765YR8  00000000782YR3  00000000791YR4  00000000823YR5  00000000824YR3  00000000797YR1  00000000794YR8    \"Product: GE 1.5T SIGNA HDe MR SYSTEM   CR No. 9612283-XX/XX/XX-00X-C (Hino, Japan)  FMI 67815A\"  Mfg Lot or Serial #  00000004984YR1  00000005068YR2  00000005079YR9  00000005080YR7  00000005081YR5  00000005648YR1  00000005656YR4  00000005659YR8  00000005664YR8  00000006155YR6  00000006157YR2  00000006172YR1  00000006176YR2  00000006177YR0  00000006188YR7  00000006194YR5  00000006198YR6  00000006200YR0  00000006204YR2  00000006514YR4  00000006515YR1  00000006516YR9  00000006519YR3  00000006522YR7  00000006525YR0  00000006528YR4  00000006532YR6  00000006535YR9  00000006538YR3  00000006544YR1  00000007947YR5  00000007954YR1  00000007969YR9  00000007970YR7  00000007971YR5  00000008692YR6  00000008698YR3  00000008699YR1  00000008701YR5  00000008703YR1  00000008709YR8  00000008715YR5  00000008716YR3  00000008720YR5  00000008726YR2  00000008727YR0  00000008728YR8  00000008729YR6  00000008732YR0  00000008734YR6  00000009833YR5  00000010024YR8  00000010025YR5  00000010114YR7  00000010578YR3  00000010583YR3  00000010686YR4  00000010794YR6  00000010795YR3  00000010830YR8  00000010831YR6  00000010834YR0  00000010856YR3  00000010863YR9  00000011210YR2  00000011428YR0  00000011429YR8  00000011495YR9  00000011499YR1  00000011637YR6  00000011648YR3  00000011650YR9  00000011669YR9  00000011686YR3  00000011826YR5  00000011829YR9  00000011830YR7  00000011842YR2  00000012085YR7  00000012098YR0  00000012099YR8  00000012100YR4  00000012102YR0  00000012142YR6  00000012172YR3  00000012206YR9  00000012207YR7  00000012221YR8  00000012227YR5  00000012286YR1  00000012287YR9  00000012289YR5  00000012330YR7  00000012393YR5  00000012434YR7  00000012435YR4  00000012503YR9  00000012507YR0  00000012576YR5  00000012651YR6  00000012652YR4  00000012727YR4  00000012728YR2  00000012755YR5  00000012756YR3  00000012818YR1  00000012820YR7  00000012940YR3  00000012943YR7  00000012988YR2  00000012992YR4  00000012994YR0  00000012995YR7  00000013056YR7  00000013061YR7  00000013177YR1  00000013180YR5  00000013185YR4  00000013239YR9  00000013241YR5  00000013246YR4  00000013301YR7  00000013306YR6  00000013310YR8  00000013375YR1  00000013379YR3  00000013384YR3  00000013385YR0  00000013399YR1  00000013453YR6  00000013799YR2  00000013800YR8  00000013802YR4  00000013804YR0  00000013878YR4  00000013880YR0  00000014113YR5  00000014114YR3  00000014214YR1  00000014276YR0  00000014277YR8  00000014520YR1  00000014524YR3  00000014801YR5  00000014806YR4  00000014858YR5  00000014859YR3  00000014909YR6  00000014910YR4  00000014956YR7  00000011535YR2  00000012209YR3  00000014414YR7  00000006174YR7  00000006507YR8  00000010827YR4  00000011960YR2  00000012144YR2  00000012497YR4  00000012575YR7  00000012753YR0  00000012991YR6  00000013693YR7  00000013875YR0  00000014415YR4  00000014419YR6  00000014620YR9  00000006547YR4  00000014907YR0  00000011692YR1  00000012367YR9  00000012508YR8  00000012550YR0  00000012722YR5  00000012942YR9  00000013305YR8  00000013451YR0  00000013626YR7  00000013805YR7  00000013876YR8  00000013877YR6  00000012796YR9  00000007964YR0  00000008704YR9  00000009948YR1  00000010023YR0  00000010685YR6  00000010833YR2  00000011827YR3  00000011824YR0  00000013057YR5  00000006181YR2  00000006183YR8  00000006184YR6  00000006185YR3  00000006186YR1  00000007951YR7  00000007952YR5  00000008688YR4  00000008689YR2  00000008705YR6  00000008706YR4  00000008707YR2  00000008708YR0  00000008710YR6  00000008711YR4  00000008712YR2  00000008730YR4  00000008731YR2  00000008733YR8  00000009832YR7  00000010227YR7  00000010231YR9  00000010250YR9  00000010253YR3  00000010579YR1  00000010754YR0  00000010755YR7  00000010857YR1  00000010860YR5  00000010862YR1  00000011186YR4  00000011211YR0  00000011212YR8  00000011238YR3  00000011290YR4  00000011343YR1  00000011349YR8  00000011537YR8  00000011539YR4  00000011638YR4  00000011639YR2  00000011649YR1  00000011652YR5  00000011653YR3  00000011685YR5  00000011689YR7  00000011690YR5  00000011828YR1  00000011840YR6  00000011961YR0  00000011963YR6  00000011964YR4  00000011987YR5  00000011991YR7  00000011992YR5  00000012103YR8  00000012117YR8  00000012118YR6  00000012119YR4  00000012141YR8  00000012147YR5  00000012171YR5  00000012366YR1  00000012368YR7  00000012389YR3  00000012390YR1  00000012391YR9  00000012449YR5  00000012551YR8  00000012579YR9  00000012580YR7  00000012581YR5  00000012582YR3  00000012723YR3  00000012729YR0  00000012875YR1  00000012876YR9  00000012879YR3  00000012880YR1  00000012936YR1  00000012937YR9  00000012993YR2  00000013055YR9  00000013092YR2  00000013135YR9  00000013238YR1  00000013240YR7  00000013244YR9  00000013245YR6  00000013249YR8  00000013303YR3  00000013309YR0  00000013347YR0  00000013377YR7  00000013397YR5  00000013445YR2  00000013446YR0  00000013448YR6  00000013452YR8  00000013576YR4  00000013577YR2  00000013578YR0  00000013579YR8  00000013580YR6  00000013582YR2  00000013585YR5  00000013617YR6  00000013618YR4  00000013619YR2  00000013620YR0  00000013622YR6  00000013695YR2  00000013696YR0  00000013697YR8  00000013699YR4  00000013700YR0  00000013701YR8  00000013702YR6  00000013810YR7  00000013914YR7  00000013915YR4  00000013916YR2  00000013935YR2  00000013936YR0  00000013937YR8  00000014058YR2  00000014059YR0  00000014069YR9  00000014104YR4  00000014105YR1  00000014109YR3  00000014110YR1  00000014112YR7  00000014212YR5  00000014213YR3  00000014215YR8  00000014216YR6  00000014217YR4  00000014278YR6  00000014327YR1  00000014329YR7  00000014330YR5  00000014331YR3  00000014332YR1  00000014349YR5  00000014350YR3  00000014351YR1  00000014402YR2  00000014413YR9  00000014416YR2  00000014457YR6  00000014515YR1  00000014519YR3  00000014525YR0  00000014526YR8  00000014554YR0  00000014555YR7  00000014623YR3  00000014654YR8  00000014655YR5  00000014656YR3  00000014657YR1  00000014658YR9  00000014698YR5  00000014699YR3  00000014700YR9  00000014701YR7  00000014702YR5  00000014703YR3  00000014704YR1  00000014749YR6  00000014750YR4  00000014751YR2  00000014752YR0  00000014754YR6  00000014800YR7  00000014802YR3  00000014803YR1  00000014804YR9  00000014860YR1  00000014861YR9  00000014862YR7  00000014863YR5  00000014911YR2  00000014957YR5  00000014981YR5  00000014982YR3  00000014983YR1  00000014984YR9  00000014999YR7  00000015000YR3  00000015002YR9  00000015041YR7  00000015131YR6  00000012578YR1  00000013806YR5  00000014107YR7  00000009947YR3  00000013183YR9  00000006164YR8  00000006509YR4  00000012759YR7  00000013879YR2  00000006202YR6  00000006170YR5  00000008717YR1  00000013625YR9  00000005078YR1  00000005655YR6  00000008690YR0  00000009954YR9  00000006527YR6  00000006536YR7  00000008693YR4  00000009955YR6  00000010756YR5  00000010828YR2  00000011288YR8  00000011347YR2  00000011497YR5  00000011636YR8  00000011768YR9  00000011825YR7  00000012821YR5  00000013134YR2  00000013181YR3  00000006161YR4  00000006167YR1  00000012291YR1  00000012501YR3  00000012549YR2  00000013311YR6  00000010582YR5  00000011042YR9  00000010832YR4  00000011285YR4  00000011651YR7  00000011966YR9  00000012086YR5  00000012169YR9  00000012288YR7  00000012327YR3  00000012329YR9  00000012506YR2  00000012655YR7  00000012989YR0  00000013060YR9  00000013242YR3  00000013247YR2  00000013248YR0  00000014908YR8  00000011038YR7  00000012654YR0  00000014516YR9  00000013382YR7  00000013623YR4  00000013627YR5  00000013628YR3  00000013703YR4  00000014061YR6  00000014064YR0  00000011766YR3  00000013130YR0  00000005073YR2  00000006190YR3  00000006512YR8  00000006520YR1  00000006549YR0  00000010864YR7  00000011839YR8  00000011965YR1  00000012168YR1  00000012293YR7  00000012436YR2  00000012496YR6  00000013395YR9  00000013398YR3  00000006552YR4  00000004672YR2  00000005065YR8  00000005067YR4  00000005069YR0  00000005076YR5  00000005084YR9  00000005645YR7  00000005650YR7  00000005652YR3  00000005653YR1  00000005654YR9  00000005658YR0  00000005660YR6  00000005662YR2  00000006158YR0  00000006159YR8  00000006160YR6  00000006163YR0  00000006165YR5  00000006166YR3  00000006168YR9  00000006178YR8  00000006179YR6  00000006180YR4  00000006182YR0  00000006187YR9  00000006192YR9  00000006193YR7  00000006201YR8  00000006505YR2  00000006506YR0  00000006508YR6  00000006513YR6  00000006517YR7  00000006518YR5  00000006524YR3  00000006526YR8  00000006529YR2  00000006531YR8  00000006537YR5  00000006540YR9  00000006546YR6  00000006548YR2  00000007945YR9  00000007946YR7  00000007948YR3  00000007949YR1  00000007950YR9  00000007953YR3  00000007955YR8  00000007956YR6  00000007957YR4  00000007958YR2  00000007959YR0  00000007960YR8  00000007961YR6  00000007962YR4  00000007963YR2  00000007965YR7  00000007972YR3  00000007973YR1  00000007974YR9  00000008685YR0  00000008686YR8  00000008687YR6  00000008691YR8  00000008696YR7  00000008718YR9  00000008723YR9  00000008725YR4  00000009949YR9  00000009956YR4  00000010021YR4  00000010251YR7  00000010580YR9  00000010684YR9  00000010687YR2  00000010858YR9  00000010859YR7  00000010861YR3  00000011040YR3  00000011041YR1  00000011289YR6  00000011344YR9  00000011345YR6  00000011346YR4  00000011348YR0  00000011430YR6  00000011672YR3  00000011694YR7  00000011695YR4  00000011764YR8  00000011841YR4  00000011843YR0  00000011988YR3  00000011990YR9  00000011993YR3  00000012082YR4  00000012084YR0  00000012143YR4  00000012148YR3  00000012170YR7  00000012324YR0  00000012325YR7  00000012326YR5  00000012331YR5  00000012653YR2  00000012657YR3  00000012726YR6  00000012730YR8  00000012804YR1  00000012822YR3  00000012939YR5  00000012941YR1  00000013027YR8  00000013028YR6  00000013093YR0  00000013178YR9  00000013182YR1  00000013186YR2  00000013205YR0  00000013250YR6  00000013251YR4  00000013287YR8  00000013344YR7  00000013345YR4  00000013387YR6  00000013644YR0  00000013583YR0  00000014805YR6  00000012292YR9  00000005657YR2  00000006171YR3  00000006173YR9  00000006197YR8  00000007966YR5  00000007967YR3  00000010829YR0  00000010978YR5  00000011039YR5  00000011534YR5  00000011646YR7  00000011691YR3  00000011994YR1  00000012208YR5  00000012392YR7  00000012498YR2  00000012502YR1  00000012505YR4  00000012724YR1  00000013302YR5  00000013447YR8  00000013694YR5  00000013704YR2  00000013803YR2  00000014108YR5  00000014418YR8  00000014454YR3  00000011765YR5  00000006189YR5  00000011286YR2  00000011986YR7  00000012795YR1  00000012817YR3  00000013184YR7  00000013386YR8  00000013396YR7  00000013449YR4  00000013624YR2  00000014062YR4  00000014111YR9  00000014401YR4  00000014403YR0  00000014455YR0  00000014517YR7  00000014912YR0  00000013095YR5  00000015091YR2  00000011767YR1  00000010977YR7  00000011538YR6  00000012226YR7  00000012577YR3  00000012658YR1  00000013094YR8  00000013203YR5  00000012573YR2  00000013381YR9  00000013645YR7  00000012205YR1  00000012824YR9  00000012938YR7  00000013801YR6  00000014063YR2  00000011035YR3  00000011498YR3  00000011962YR8  00000011989YR1  00000013243YR1  00000013343YR9  00000006539YR1  00000006551YR6  00000008719YR7  00000008721YR3  00000010254YR1  00000011037YR9  00000012754YR8  00000014060YR8  00000014071YR5  00000015092YR0  00000006156YR4  00000010225YR1  00000012547YR6  00000012819YR9  00000013807YR3  00000005077YR3  00000006169YR7  00000008694YR2  00000008713YR0  00000008714YR8  00000010226YR9  00000011432YR2  00000012210YR1  00000004773YR8  00000005082YR3  00000006196YR0  00000006521YR9  00000006545YR8  00000006550YR8  00000010230YR1  00000012823YR1  00000013304YR1  00000013307YR4  00000013376YR9  00000010229YR3  00000014211YR7  00000012083YR2  00000013179YR7  00000012328YR1  00000012990YR8  00000013450YR2  00000014753YR8  00000015001YR1  00000010252YR5  00000012797YR7  00000013380YR1  00000005072YR4  00000005074YR0  00000006203YR4  00000006510YR2  00000006511YR0  00000006523YR5  00000006530YR0  00000006533YR4  00000006534YR2  00000006541YR7  00000006542YR5  00000006543YR3  00000006553YR2  00000006554YR0  00000010939YR7  00000010976YR9  00000012552YR6  00000012553YR4  00000012574YR0  00000012793YR6  00000012794YR4  00000012878YR5  00000012996YR5  00000013058YR3  00000013059YR1  00000013204YR3  00000013378YR5  00000013629YR1  00000013757YR0  00000014328YR9  00000014417YR0  00000014456YR8  00000011838YR0  00000006199YR4  00000010115YR4  00000011536YR0  00000012290YR3  00000014518YR5  00000011532YR9  00000006175YR4  00000006195YR2  00000008700YR7  00000008722YR1  00000008724YR7  00000010581YR7  00000011496YR7  00000011668YR1  00000012101YR2  00000012145YR9  00000012146YR7  00000012149YR1  00000012792YR8  00000013308YR2  00000014065YR7  00000014458YR4  00000014622YR5  00000014621YR7  00000013698YR6"
    }
  ]
}