{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82897",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution. US Nationwide, AFGHANISTAN , HUNGARY , PANAMA , ALGERIA , ICELAND , PARAGUAY , ARGENTINA , INDIA , PERU , AUSTRALIA , INDONESIA , PHILIPPINES , AUSTRIA , IRAN , POLAND , AZERBAIJAN , IRAQ , PORTUGAL , BAHRAIN , IRELAND , QATAR , BANGLADESH , ISRAEL , ROMANIA , BELARUS , ITALY , RUSSIA , BELGIUM , JAPAN , SAUDI ARABIA , BOLIVIA , JORDAN , SERBIA , BOSNIA AND HERZEGOVINA , KAZAKHSTAN , SINGAPORE , BRAZIL , KENYA , SLOVAKIA , BRUNEI , KOSOVO , SLOVENIA , BULGARIA , KUWAIT , SOUTH AFRICA , CAMBODIA , LATVIA , SOUTH KOREA , CANADA , LEBANON , SPAIN , CHILE , LIBYA , SRI LANKA , CHINA , LITHUANIA , SWEDEN , COLOMBIA , LUXEMBOURG , SWITZERLAND , COSTA RICA , MACAU , SYRIA , CROATIA , MACEDONIA , TAIWAN , CYPRUS , MALAYSIA , TANZANIA , CZECH REPUBLIC , MALDIVES , THAILAND , DENMARK , MALTA , TRINIDAD , DOMINICAN REP , MARTINIQUE , TUNISIA , ECUADOR , MAURITIUS , TURKEY , EGYPT , MEXICO , TURKMENISTAN , EL SALVADOR , MOROCCO , UGANDA , ESTONIA , MYANMAR , UKRAINE , FINLAND , NAMIBIA , UNITED ARAB EMIRATES , FRANCE , NEPAL , UNITED KINGDOM , GEORGIA , NETHERLANDS , URUGUAY , GERMANY , NEW ZEALAND , UZBEKISTAN , GHANA , NIGERIA , VENEZUELA , GREECE , NORTH KOREA , VIETNAM , GUATEMALA , NORWAY , YEMEN , GUYANA , OMAN , ZAMBIA , HONDURAS , PAKISTAN , HONG KONG , and PALESTINE.",
      "recall_number": "Z-1939-2019",
      "product_description": "Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), 0998-00-0800-53  For cardiovascular use.",
      "product_quantity": "6,197",
      "reason_for_recall": "Battery short-run times and unanticipated stoppage may be due to user not performing battery maintenance and handling as needed.  Inadequate labeling is being targeted for this recall, and design updates will be handled in a subsequent recall.",
      "recall_initiation_date": "20190617",
      "center_classification_date": "20190715",
      "report_date": "20190724",
      "code_info": "All serial numbers affected"
    }
  ]
}