  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98682",
      "recalling_firm": "Cook Incorporated",
      "address_1": "750 N Daniels Way",
      "address_2": "",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay",
      "recall_number": "Z-1936-2026",
      "product_description": "Beacon Tip Centimeter Sizing Catheter  RPN/GPN  NR5.0-35-70-P-10S-0-CSC-20  G31213  NR5.0-35-100-P-10S-0-CSC-20  G31214  NR5.0-35-100-P-10S-PIG-CSC-20  G31216  NR5.0-35-70-P-10S-VCF-CSC-20  G31220  NR5.0-35-100-P-10S-VCF-CSC-20  G31221    Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).",
      "product_quantity": "4,245",
      "reason_for_recall": "Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.",
      "recall_initiation_date": "20260402",
      "center_classification_date": "20260501",
      "report_date": "20260513",
      "code_info": "G31213 UDI-DI 00827002312135 Lot 16805456  G31214 UDI-DI 00827002312142 Lots 16886459 16874181 16855513 16867505 16815763 16796665 16802282 16766437 16734169 16695772  G31216 UDI-DI 00827002312166 Lots 16886460 16891299 16912494 16831676 16836358 16836359 16836360 16864534 16864652 16881655 16882508 16912493 16872559 16878260 16824312 16831675 16874182 16855514 16868889 16859732 16858079 16878261 16806270 16804619 16805454 16804399 16815762 16803920 16803921 16802096 16802097 16814369 16814370 16806333 16804620 16831528 16822289 16822290 16782945 16782946 16795733 16796300 16779437 16795734 16800424 16802632 16824313 16764526 16763024 16800730 16802633 16802133 16800731 16821105 16824314 16711110 16717839 16730326 16720153 16736740 16739291 16693653 16690051 16673569 16680240 16704378  G31220 UDI-DI 00827002312203 Lots 16881657 16872241 16827163 16852999 16805455 16804400 16802132 16779438 16730325 16739292 16700199 16690050 16661312 16674400  G31221 UDI-DI 00827002312210 Lots 16839296 16879977 16814371 16821104",
      "more_code_info": ""
    }
  ]
}