{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Chelmsford",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83109",
      "recalling_firm": "ZOLL Medical Corporation",
      "address_1": "269 & 271 Mill Road",
      "address_2": "N/A",
      "postal_code": "01824",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US nationwide.    The products were distributed to the following foreign countries:  Austria, Australia, Bangladesh, Brazil, Canada, China, Denmark, France, Germany, Great Britain, Indonesia, Italy, Japan, Luxemburg, New Zealand, Panama, Russia, Saudi Arabia, Singapore, Spain, Thailand, Taiwan.",
      "recall_number": "Z-1931-2019",
      "product_description": "ZOLL AED PRO Automated External Defibrillator    Product Usage:  The AED PRO unit is intended to defibrillate victims of ventricular fibrillation or pulseless ventricular tachycardia, for ECG monitoring, and for CPR monitoring of patients. The CPR monitoring function provides a metronome designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of at least 2 inches (5 cm) for adult patients. The AED PRO system is indicated for use on victims of cardiac arrest with apparent lack of circulation as indicated by:   Unconsciousness   Absence of breathing, and   Absence of pulse and other signs of circulation. The device is also indicated for use when ECG monitoring is indicated to evaluate the patient s heart rate or ECG morphology. The AED PRO system is indicated for adult and pediatric patients.",
      "product_quantity": "1,969",
      "reason_for_recall": "There is a potential for the device to fail to deliver energy to the patient.",
      "recall_initiation_date": "20190612",
      "center_classification_date": "20190703",
      "report_date": "20190710",
      "code_info": "Part Numbers:  90010200499991010,   90010400499991010,   70010400499991010,   90010202499991010,   90010600499991010;        GTIN:  00847946016128,   00847946016142,   00847946015138,   00847946016135,   00847946016166    Serial Numbers:  AA17C041915  AA17C041916  AA17J043664  AA17J043666  AA17J043670  AA17J043671  AA17J043672  AA17J043673  AA17J043674  AA17J043675  AA17J043682  AA17J043683  AA17J043725  AA17J043813  AA17J043814  AA17J043817  AA17J043820  AA17J043825  AA17L044265  AA18B044867  AA18B044868  AA18B044869  AA18B044870  AA18B044871  AA18B044872  AA18B044873  AA18B044874  AA18B044875  AA18B044877  AA18B044878  AA18B044879  AA18B044880  AA18B044881  AA18B044882  AA18B044883  AA18B044884  AA18B044885  AA18B044886  AA18C044887  AA18C044888  AA18C044889  AA18C044890  AA18C044891  AA18C044892  AA18C044893  AA18C044894  AA18C044895  AA18C044896  AA18C044897  AA18C044898  AA18C044899  AA18C044900  AA18C044901  AA18C044902  AA18C044903  AA18C044904  AA18C044905  AA18C044906  AA18C044907  AA18C044908  AA18C044909  AA18C044910  AA18C044911  AA18C045011  AA18C045012  AA18C045013  AA18C045014  AA18C045015  AA18C045016  AA18C045017  AA18C045018  AA18C045019  AA18C045020  AA18C045021  AA18C045022  AA18C045023  AA18C045024  AA18C045025  AA18C045026  AA18C045027  AA18C045028  AA18C045029  AA18C045030  AA18C045031  AA18C045032  AA18C045033  AA18C045034  AA18C045035  AA18C045036  AA18C045037  AA18C045038  AA18C045039  AA18C045040  AA18C045041  AA18C045042  AA18C045043  AA18C045044  AA18C045045  AA18C045046  AA18C045047  AA18C045048  AA18C045049  AA18C045050  AA18C045051  AA18C045052  AA18C045053  AA18C045054  AA18C045055  AA18C045056  AA18C045057  AA18C045058  AA18C045059  AA18C045060  AA18C045061  AA18C045062  AA18C045063  AA18C045064  AA18C045065  AA18C045070  AA18C045071  AA18C045072  AA18C045073  AA18C045074  AA18C045075  AA18C045076  AA18C045077  AA18C045078  AA18C045079  AA18C045080  AA18C045081  AA18C045082  AA18C045083  AA18C045084  AA18C045085  AA18C045086  AA18C045087  AA18C045088  AA18C045089  AA18C045090  AA18C045091  AA18C045092  AA18C045093  AA18C045094  AA18C045095  AA18C045096  AA18C045097  AA18C045098  AA18C045099  AA18C045100  AA18C045101  AA18C045102  AA18C045103  AA18C045104  AA18C045105  AA18C045106  AA18C045107  AA18C045108  AA18C045109  AA18C045110  AA18C045111  AA18C045112  AA18C045113  AA18C045114  AA18C045115  AA18C045116  AA18C045117  AA18C045118  AA18C045119  AA18C045120  AA18C045121  AA18C045122  AA18C045123  AA18C045124  AA18C045125  AA18C045126  AA18C045127  AA18C045128  AA18C045129  AA18C045130  AA18C045131  AA18C045144  AA18C045145  AA18C045146  AA18C045147  AA18C045148  AA18C045149  AA18C045150  AA18C045151  AA18C045152  AA18C045153  AA18C045251  AA18C045252  AA18C045253  AA18C045254  AA18C045255  AA18C045256  AA18C045257  AA18C045258  AA18C045259  AA18C045260  AA18C045261  AA18C045262  AA18C045263  AA18C045264  AA18D045265  AA18D045266  AA18D045267  AA18D045268  AA18D045269  AA18D045270  AA18D045271  AA18D045272  AA18D045273  AA18D045274  AA18D045275  AA18D045276  AA18D045277  AA18D045278  AA18D045279  AA18D045280  AA18D045281  AA18D045282  AA18D045283  AA18D045284  AA18D045285  AA18D045286  AA18D045287  AA18D045288  AA18D045289  AA18D045290  AA18D045291  AA18D045292  AA18D045293  AA18D045294  AA18D045295  AA18D045296  AA18D045297  AA18D045298  AA18D045299  AA18D045300  AA18D045301  AA18D045302  AA18D045303  AA18D045304  AA18D045305  AA18D045306  AA18D045307  AA18D045308  AA18D045309  AA18D045310  AA18D045311  AA18D045312  AA18D045313  AA18D045314  AA18D045315  AA18D045332  AA18D045333  AA18D045334  AA18D045335  AA18D045336  AA18D045337  AA18D045338  AA18D045339  AA18D045340  AA18D045341  AA18D045342  AA18D045343  AA18D045344  AA18D045345  AA18D045349  AA18D045350  AA18D045351  AA18D045352  AA18D045353  AA18D045354  AA18D045355  AA18D045356  AA18D045357  AA18D045358  AA18D045359  AA18D045360  AA18D045361  AA18D045362  AA18D045363  AA18D045364  AA18D045365  AA18D045366  AA18D045367  AA18D045368  AA18D045369  AA18D045370  AA18D045371  AA18D045372  AA18D045373  AA18D045374  AA18D045375  AA18D045376  AA18D045377  AA18D045378  AA18D045379  AA18D045380  AA18D045381  AA18D045382  AA18D045383  AA18D045384  AA18D045385  AA18D045386  AA18D045387  AA18D045388  AA18D045389  AA18D045390  AA18D045391  AA18D045392  AA18D045393  AA18D045394  AA18D045395  AA18D045396  AA18D045397  AA18D045398  AA18D045399  AA18D045400  AA18D045401  AA18D045402  AA18D045403  AA18D045404  AA18D045405  AA18D045406  AA18D045407  AA18D045408  AA18D045409  AA18D045410  AA18D045411  AA18D045412  AA18D045413  AA18D045414  AA18D045415  AA18D045416  AA18D045417  AA18D045418  AA18D045419  AA18D045420  AA18D045421  AA18D045422  AA18D045423  AA18D045424  AA18D045425  AA18D045426  AA18D045427  AA18D045428  AA18D045429  AA18D045430  AA18D045431  AA18D045432  AA18D045433  AA18D045434  AA18D045435  AA18D045436  AA18D045437  AA18D045438  AA18D045439  AA18D045440  AA18D045441  AA18D045442  AA18D045443  AA18D045444  AA18D045445  AA18D045446  AA18D045447  AA18D045448  AA18D045449  AA18D045450  AA18D045451  AA18D045452  AA18D045453  AA18D045454  AA18D045455  AA18D045456  AA18D045457  AA18D045458  AA18D045459  AA18D045460  AA18D045461  AA18D045462  AA18D045463  AA18D045464  AA18D045465  AA18D045505  AA18D045506  AA18D045507  AA18D045509  AA18D045510  AA18D045511  AA18D045512  AA18D045513  AA18D045514  AA18D045515  AA18D045516  AA18D045517  AA18D045518  AA18D045519  AA18D045520  AA18D045521  AA18D045522  AA18D045523  AA18D045524  AA18D045525  AA18D045526  AA18D045527  AA18D045528  AA18D045529  AA18D045530  AA18D045534  AA18D045535  AA18D045536  AA18D045537  AA18D045538  AA18D045539  AA18D045540  AA18D045541  AA18D045542  AA18D045543  AA18D045544  AA18D045545  AA18D045546  AA18D045547  AA18D045548  AA18D045549  AA18D045550  AA18D045551  AA18D045552  AA18D045553  AA18D045554  AA18D045555  AA18D045556  AA18D045557  AA18D045558  AA18D045559  AA18D045560  AA18D045561  AA18D045562  AA18D045563  AA18D045564  AA18D045565  AA18D045566  AA18D045567  AA18D045568  AA18D045617  AA18D045634  AA18D045635  AA18D045636  AA18D045637  AA18D045638  AA18E045663  AA18E045664  AA18E045665  AA18E045666  AA18E045677  AA18E045678  AA18E045679  AA18E045680  AA18E045681  AA18E045682  AA18E045683  AA18E045685  AA18E045686  AA18E045687  AA18E045688  AA18E045689  AA18E045690  AA18E045767  AA18E045769  AA18E045770  AA18E045771  AA18E045772  AA18E045773  AA18E045774  AA18E045775  AA18E045776  AA18E045777  AA18E045778  AA18E045779  AA18E045780  AA18E045781  AA18E045782  AA18E045783  AA18E045784  AA18E045785  AA18E045786  AA18E045787  AA18E045788  AA18E045789  AA18E045790  AA18E045791  AA18E045792  AA18E045793  AA18E045794  AA18E045795  AA18E045796  AA18E045797  AA18E045798  AA18E045799  AA18E045800  AA18E045801  AA18E045802  AA18E045803  AA18E045804  AA18E045805  AA18E045806  AA18E045807  AA18E045808  AA18E045809  AA18E045810  AA18E045811  AA18E045823  AA18E045824  AA18E045825  AA18E045826  AA18E045827  AA18E045828  AA18E045829  AA18E045830  AA18E045831  AA18E045832  AA18E045834  AA18E045835  AA18E045836  AA18E045837  AA18E045838  AA18E045839  AA18E045840  AA18E045841  AA18E045842  AA18E045974  AA18E045975  AA18E045976  AA18E045977  AA18E045978  AA18E045979  AA18E045980  AA18E045981  AA18E045982  AA18E045983  AA18E045984  AA18E045985  AA18E045986  AA18E045987  AA18E045988  AA18E045989  AA18E045990  AA18E045991  AA18E045992  AA18E045993  AA18E045994  AA18E045995  AA18E045996  AA18E045997  AA18E045998  AA18E045999  AA18E046000  AA18E046001  AA18E046002  AA18E046003  AA18E046004  AA18E046007  AA18E046008  AA18E046009  AA18F046058  AA18F046059  AA18F046060  AA18F046061  AA18F046062  AA18F046063  AA18F046064  AA18F046065  AA18F046066  AA18F046067  AA18F046068  AA18F046069  AA18F046070  AA18F046071  AA18F046072  AA18F046073  AA18F046074  AA18F046075  AA18F046076  AA18F046077  AA18F046078  AA18F046079  AA18F046080  AA18F046081  AA18F046082  AA18F046083  AA18F046084  AA18F046085  AA18F046086  AA18F046087  AA18F046088  AA18F046089  AA18F046090  AA18F046091  AA18F046092  AA18F046093  AA18F046094  AA18F046095  AA18F046096  AA18F046097  AA18F046098  AA18F046099  AA18F046100  AA18F046101  AA18F046102  AA18F046103  AA18F046104  AA18F046105  AA18F046106  AA18F046107  AA18F046108  AA18F046109  AA18F046110  AA18F046111  AA18F046112  AA18F046113  AA18F046114  AA18F046115  AA18F046116  AA18F046117  AA18F046118  AA18F046119  AA18F046120  AA18F046121  AA18F046122  AA18F046123  AA18F046124  AA18F046125  AA18F046126  AA18F046127  AA18F046128  AA18F046129  AA18F046130  AA18F046131  AA18F046132  AA18F046133  AA18F046134  AA18F046135  AA18F046136  AA18F046137  AA18F046138  AA18F046139  AA18F046140  AA18F046141  AA18F046142  AA18F046143  AA18F046144  AA18F046145  AA18F046146  AA18F046147  AA18F046148  AA18F046150  AA18F046151  AA18F046152  AA18F046153  AA18F046154  AA18F046155  AA18F046156  AA18F046157  AA18F046158  AA18F046159  AA18F046160  AA18F046161  AA18F046162  AA18F046163  AA18F046164  AA18F046165  AA18F046166  AA18F046167  AA18F046168  AA18F046169  AA18F046170  AA18F046171  AA18F046172  AA18F046173  AA18F046174  AA18F046193  AA18F046194  AA18F046195  AA18F046196  AA18F046197  AA18F046198  AA18F046199  AA18F046200  AA18F046201  AA18F046202  AA18F046203  AA18F046204  AA18F046205  AA18F046206  AA18F046207  AA18F046208  AA18F046209  AA18F046210  AA18F046211  AA18F046212  AA18F046213  AA18F046214  AA18F046215  AA18F046216  AA18F046217  AA18F046218  AA18F046219  AA18F046220  AA18F046221  AA18F046222  AA18F046223  AA18F046224  AA18F046225  AA18F046226  AA18F046227  AA18F046228  AA18F046229  AA18F046230  AA18F046257  AA18F046258  AA18F046259  AA18F046262  AA18F046263  AA18F046264  AA18F046265  AA18G046505  AA18G046506  AA18G046507  AA18G046508  AA18G046509  AA18G046510  AA18G046511  AA18G046512  AA18G046513  AA18G046514  AA18G046515  AA18G046516  AA18G046517  AA18G046518  AA18G046519  AA18G046554  AA18G046555  AA18G046556  AA18G046557  AA18G046558  AA18G046559  AA18G046560  AA18G046561  AA18G046562  AA18G046563  AA18G046564  AA18G046565  AA18G046566  AA18G046567  AA18G046568  AA18H046706  AA18H046707  AA18H046708  AA18H046709  AA18H046710  AA18H046711  AA18H046712  AA18H046713  AA18H046714  AA18H046715  AA18H046716  AA18H046717  AA18H046718  AA18H046719  AA18H046720  AA18H046757  AA18H046758  AA18H046759  AA18H046760  AA18H046761  AA18H046762  AA18H046763  AA18H046764  AA18H046765  AA18H046766  AA18H046767  AA18H046768  AA18H046769  AA18H046770  AA18H046771  AA18H046772  AA18H046773  AA18H046774  AA18H046775  AA18H046776  AA18H046777  AA18H046778  AA18H046779  AA18H046780  AA18H046781  AA18H046782  AA18H046783  AA18H046784  AA18H046785  AA18H046786  AA18H046787  AA18H046788  AA18H046789  AA18H046790  AA18H046791  AA18H046792  AA18H046793  AA18H046794  AA18H046795  AA18H046799  AA18H046800  AA18H046801  AA18H046802  AA18H046803  AA18H046804  AA18H046842  AA18H046843  AA18H046844  AA18H046845  AA18H046846  AA18H046847  AA18H046848  AA18H046850  AA18H046851  AA18H046852  AA18H046853  AA18H046854  AA18H046855  AA18H046856  AA18H046857  AA18H046858  AA18H046859  AA18H046860  AA18H046861  AA18H046862  AA18H046863  AA18H046864  AA18H046865  AA18H046866  AA18H046867  AA18H046868  AA18H046869  AA18H046870  AA18H046871  AA18H046872  AA18H046873  AA18H046874  AA18H046875  AA18H046876  AA18H046877  AA18H046878  AA18H046879  AA18H046880  AA18H046881  AA18H046882  AA18H046883  AA18H046884  AA18H046885  AA18H046886  AA18H046887  AA18H046888  AA18H046889  AA18H046890  AA18H046891  AA18H046892  AA18H046893  AA18H046894  AA18H046895  AA18H046896  AA18H046897  AA18H046898  AA18H046899  AA18H046900  AA18H046901  AA18H046902  AA18H046903  AA18H046904  AA18H046905  AA18H046906  AA18H046907  AA18H046908  AA18H046909  AA18H046910  AA18H046911  AA18H046912  AA18H046913  AA18H046914  AA18H046915  AA18H046916  AA18H046917  AA18H046918  AA18H046919  AA18H046920  AA18H046921  AA18H046922  AA18H046923  AA18H046924  AA18H046925  AA18H046926  AA18H046927  AA18H046928  AA18H046929  AA18H046930  AA18H046931  AA18H046932  AA18H046933  AA18H046934  AA18H046935  AA18H046936  AA18H046937  AA18H046938  AA18H046939  AA18H046940  AA18I046941  AA18I046942  AA18I046943  AA18I046944  AA18I046945  AA18I046946  AA18I046947  AA18I046948  AA18I046949  AA18I046954  AA18I046955  AA18I046956  AA18I046957  AA18I046996  AA18I046997  AA18I046998  AA18I046999  AA18I047000  AA18I047001  AA18I047002  AA18I047003  AA18I047004  AA18I047005  AA18I047007  AA18I047008  AA18I047009  AA18I047010  AA18I047011  AA18I047056  AA18I047057  AA18J047110  AA18J047111  AA18J047112  AA18J047147  AA18J047148  AA18J047149  AA18J047150  AA18J047151  AA18J047152  AA18J047153  AA18J047154  AA18J047155  AA18J047157  AA18J047158  AA18J047159  AA18J047160  AA18J047161  AA18J047162  AA18J047163  AA18J047164  AA18J047165  AA18J047166  AA18J047167  AA18J047168  AA18J047169  AA18J047170  AA18J047171  AA18J047172  AA18J047173  AA18J047174  AA18J047175  AA18J047176  AA18J047177  AA18J047178  AA18J047179  AA18J047180  AA18J047181  AA18J047182  AA18J047183  AA18J047184  AA18J047185  AA18J047186  AA18J047187  AA18J047188  AA18J047189  AA18J047190  AA18J047191  AA18J047192  AA18J047194  AA18J047195  AA18J047196  AA18J047197  AA18J047198  AA18J047199  AA18J047200  AA18J047201  AA18J047202  AA18J047203  AA18J047206  AA18J047207  AA18J047208  AA18J047209  AA18J047210  AA18J047264  AA18J047265  AA18J047299  AA18J047300  AA18J047301  AA18J047302  AA18J047303  AA18J047304  AA18J047305  AA18J047306  AA18J047307  AA18J047308  AA18J047309  AA18J047310  AA18J047311  AA18J047312  AA18J047313  AA18J047314  AA18J047315  AA18J047316  AA18J047317  AA18J047318  AA18J047319  AA18J047320  AA18J047322  AA18J047323  AA18J047324  AA18J047325  AA18J047326  AA18J047327  AA18J047328  AA18J047329  AA18J047330  AA18J047331  AA18J047332  AA18J047333  AA18J047334  AA18K047389  AA18K047390  AA18K047391  AA18K047392  AA18K047393  AA18K047394  AA18K047395  AA18K047396  AA18K047397  AA18K047398  AA18K047399  AA18K047400  AA18K047401  AA18K047402  AA18K047403  AA18K047404  AA18K047405  AA18K047406  AA18K047407  AA18K047408  AA18K047409  AA18K047410  AA18K047411  AA18K047412  AA18K047413  AA18K047421  AA18K047422  AA18K047423  AA18K047424  AA18K047425  AA18K047426  AA18K047427  AA18K047428  AA18K047429  AA18K047433  AA18K047445  AA18K047446  AA18K047447  AA18K047448  AA18K047449  AA18K047450  AA18K047451  AA18K047452  AA18K047453  AA18K047454  AA18K047455  AA18K047456  AA18K047457  AA18K047458  AA18K047460  AA18K047603  AA18K047634  AA18K047635  AA18K047636  AA18K047637  AA18K047638  AA18K047639  AA18K047641  AA18K047642  AA18K047643  AA18K047644  AA18K047645  AA18K047646  AA18K047647  AA18K047648  AA18K047649  AA18K047650  AA18K047651  AA18K047652  AA18K047653  AA18K047654  AA18K047655  AA18K047656  AA18K047657  AA18K047658  AA18K047659  AA18K047660  AA18K047661  AA18K047662  AA18K047663  AA18K047664  AA18L047709  AA18L047710  AA18L047711  AA18L047712  AA18L047713  AA18L047714  AA18L047715  AA18L047716  AA18L047717  AA18L047718  AA18L047719  AA18L047720  AA18L047721  AA18L047722  AA18L047733  AA18L047734  AA18L047742  AA18L047743  AA18L047744  AA18L047745  AA18L047746  AA18L047747  AA18L047748  AA18L047749  AA18L047750  AA18L047751  AA18L047752  AA18L047753  AA18L047754  AA18L047755  AA18L047756  AA18L047757  AA18L047758  AA18L047759  AA18L047760  AA18L047761  AA18L047762  AA18L047763  AA18L047764  AA18L047765  AA18L047766  AA18L047767  AA18L047768  AA18L047769  AA18L047770  AA18L047771  AA18L047772  AA18L047773  AA18L047774  AA18L047775  AA18L047776  AA18L047777  AA18L047778  AA18L047779  AA18L047780  AA18L047784  AA18L047785  AA18L047786  AA18L047787  AA18L047788  AA18L047789  AA18L047790  AA18L047791  AA18L047792  AA18L047793  AA18L047794  AA18L047795  AA18L047796  AA18L047797  AA18L047798  AA18L047799  AA18L047800  AA18L047801  AA18L047802  AA18L047803  AA18L047804  AA18L047805  AA18L047806  AA18L047807  AA18L047808  AA18L047809  AA18L047810  AA18L047811  AA18L047812  AA18L047813  AA18L047814  AA18L047815  AA18L047816  AA18L047817  AA18L047818  AA18L047819  AA18L047820  AA18L047821  AA18L047822  AA18L047823  AA18L047824  AA18L047825  AA18L047826  AA18L047827  AA18L047828  AA18L047829  AA18L047830  AA18L047831  AA18L047832  AA18L047833  AA18L047834  AA18L047835  AA18L047836  AA18L047837  AA18L047838  AA18L047839  AA18L047851  AA18L047852  AA18L047853  AA18L047854  AA18L047855  AA18L047856  AA18L047857  AA18L047858  AA18L047859  AA18L047860  AA19A047888  AA19A047889  AA19A047890  AA19A047891  AA19A047892  AA19A047893  AA19A047894  AA19A047895  AA19A047896  AA19A047897  AA19A047898  AA19A047899  AA19A047902  AA19A047903  AA19A047904  AA19A047905  AA19A047906  AA19A047907  AA19A047908  AA19A047910  AA19A047911  AA19A047913  AA19A047914  AA19A047915  AA19A047916  AA19A047917  AA19A047918  AA19A047919  AA19A047920  AA19A047921  AA19A047922  AA19A047923  AA19A047924  AA19A047925  AA19A047926  AA19A047927  AA19A047928  AA19A047929  AA19A047930  AA19A047931  AA19A047932  AA19A047933  AA19A047934  AA19A047935  AA19A047936  AA19A047937  AA19A047938  AA19A047939  AA19A047940  AA19A047941  AA19A047942  AA19A047943  AA19A047944  AA19A047945  AA19A047946  AA19A047947  AA19A047948  AA19A047949  AA19A047950  AA19A047951  AA19A047952  AA19A047953  AA19A047954  AA19A047955  AA19A047956  AA19A047957  AA19A047958  AA19A047959  AA19A047960  AA19A047961  AA19A047962  AA19A047963  AA19A047964  AA19A047965  AA19A047966  AA19A047967  AA19A047968  AA19A047969  AA19A047970  AA19A047971  AA19A047972  AA19A047973  AA19A047974  AA19A047975  AA19A047976  AA19A047977  AA19A047978  AA19A047979  AA19A047980  AA19A047981  AA19A047982  AA19A047983  AA19A047984  AA19A047985  AA19A047986  AA19A047987  AA19A047988  AA19A047989  AA19A047990  AA19A047991  AA19A047992  AA19A047993  AA19A047994  AA19A047995  AA19A047996  AA19A047997  AA19A047998  AA19A047999  AA19A048000  AA19A048001  AA19A048002  AA19A048003  AA19A048004  AA19A048005  AA19A048006  AA19A048007  AA19A048008  AA19A048009  AA19A048010  AA19A048011  AA19A048012  AA19A048013  AA19A048014  AA19A048015  AA19A048016  AA19A048017  AA19A048018  AA19A048019  AA19A048020  AA19A048021  AA19A048022  AA19A048023  AA19A048024  AA19A048025  AA19A048026  AA19A048027  AA19A048028  AA19A048029  AA19A048030  AA19A048031  AA19A048032  AA19A048033  AA19A048034  AA19A048035  AA19A048036  AA19A048037  AA19A048038  AA19A048039  AA19A048040  AA19A048041  AA19A048042  AA19A048043  AA19A048044  AA19A048045  AA19A048046  AA19A048047  AA19A048048  AA19A048049  AA19A048050  AA19A048051  AA19A048052  AA19A048053  AA19A048054  AA19A048055  AA19A048056  AA19A048057  AA19A048058  AA19A048059  AA19A048060  AA19A048061  AA19A048062  AA19A048063  AA19A048064  AA19A048065  AA19A048066  AA19A048067  AA19A048068  AA19A048069  AA19A048070  AA19A048071  AA19A048072  AA19A048074  AA19A048075  AA19A048076  AA19A048077  AA19A048078  AA19A048079  AA19A048080  AA19A048081  AA19A048082  AA19A048083  AA19A048084  AA19A048085  AA19A048086  AA19A048087  AA19A048088  AA19A048089  AA19A048090  AA19A048091  AA19A048092  AA19A048093  AA19A048094  AA19A048095  AA19A048096  AA19A048097  AA19A048098  AA19A048099  AA19A048100  AA19A048101  AA19A048102  AA19A048103  AA19A048104  AA19A048105  AA19A048106  AA19A048107  AA19A048108  AA19A048109  AA19A048110  AA19A048111  AA19A048112  AA19A048113  AA19A048114  AA19A048115  AA19A048116  AA19A048117  AA19A048118  AA19A048119  AA19A048120  AA19A048121  AA19A048122  AA19A048123  AA19A048124  AA19A048125  AA19A048126  AA19A048127  AA19A048128  AA19B048179  AA19B048180  AA19B048181  AA19B048182  AA19B048183  AA19B048184  AA19B048185  AA19B048186  AA19B048187  AA19B048188  AA19B048189  AA19B048190  AA19B048191  AA19B048192  AA19B048194  AA19B048195  AA19B048196  AA19B048197  AA19B048198  AA19B048199  AA19B048200  AA19B048201  AA19B048202  AA19B048203  AA19B048204  AA19B048205  AA19B048206  AA19B048207  AA19B048208  AA19B048209  AA19B048210  AA19B048211  AA19B048212  AA19B048213  AA19B048214  AA19B048215  AA19B048216  AA19B048217  AA19B048218  AA19B048219  AA19B048220  AA19B048221  AA19B048222  AA19B048223  AA19B048224  AA19B048225  AA19B048226  AA19B048227  AA19B048228  AA19B048229  AA19B048230  AA19B048231  AA19B048232  AA19B048233  AA19B048234  AA19B048235  AA19B048236  AA19B048237  AA19B048238  AA19B048239  AA19B048240  AA19B048241  AA19B048242  AA19B048243  AA19B048244  AA19B048245  AA19B048246  AA19B048247  AA19B048248  AA19B048249  AA19B048250  AA19B048251  AA19B048253  AA19B048254  AA19B048255  AA19B048256  AA19B048257  AA19B048258  AA19B048259  AA19B048260  AA19B048261  AA19B048262  AA19B048263  AA19B048264  AA19B048265  AA19B048266  AA19B048267  AA19B048268  AA19B048269  AA19B048270  AA19B048271  AA19B048272  AA19B048273  AA19B048274  AA19B048275  AA19B048276  AA19B048277  AA19B048278  AA19B048279  AA19B048280  AA19B048281  AA19B048282  AA19B048283  AA19B048284  AA19B048285  AA19B048286  AA19B048287  AA19B048288  AA19B048289  AA19B048290  AA19B048291  AA19B048292  AA19B048293  AA19B048298  AA19B048299  AA19B048300  AA19B048301  AA19B048302  AA19B048303  AA19B048304  AA19B048305  AA19B048306  AA19B048307  AA19B048308  AA19B048309  AA19B048310  AA19B048311  AA19B048312  AA19B048313  AA19B048314  AA19B048315  AA19B048316  AA19B048317  AA19B048318  AA19B048319  AA19B048320  AA19B048321  AA19B048322  AA19B048323  AA19B048324  AA19B048325  AA19B048326  AA19B048327  AA19B048328  AA19B048329  AA19B048330  AA19B048331  AA19B048332  AA19B048333  AA19B048334  AA19B048335  AA19B048336  AA19B048337  AA19B048338  AA19B048339  AA19B048340  AA19B048341  AA19B048342  AA19B048343  AA19B048344  AA19B048345  AA19B048346  AA19B048347  AA19B048348  AA19B048349  AA19B048350  AA19B048351  AA19B048352  AA19B048353  AA19B048354  AA19B048355  AA19B048356  AA19B048357  AA19B048358  AA19B048359  AA19B048360  AA19B048361  AA19B048362  AA19B048363  AA19B048364  AA19B048365  AA19B048366  AA19B048367  AA19B048368  AA19B048369  AA19B048370  AA19B048371  AA19B048372  AA19B048373  AA19B048374  AA19B048375  AA19B048376  AA19B048377  AA19B048378  AA19B048379  AA19B048380  AA19B048381  AA19B048382  AA19B048383  AA19B048384  AA19B048385  AA19B048386  AA19B048387  AA19B048388  AA19B048389  AA19B048390  AA19B048391  AA19B048392  AA19B048393  AA19B048394  AA19B048395  AA19B048396  AA19B048397  AA19B048398  AA19B048399  AA19B048400  AA19B048401  AA19B048402  AA19B048403  AA19B048404  AA19B048405  AA19B048406  AA19B048407  AA19B048408  AA19B048409  AA19B048410  AA19B048411  AA19B048412  AA19B048413  AA19B048414  AA19B048415  AA19B048416  AA19B048417  AA19B048418  AA19B048419  AA19B048420  AA19B048421  AA19B048422  AA19B048423  AA19B048424  AA19B048425  AA19B048432  AA19B048433  AA19B048434  AA19B048435  AA19B048436  AA19B048437  AA19B048438  AA19B048439  AA19B048441  AA19B048442  AA19B048443  AA19B048444  AA19B048445  AA19B048446  AA19B048447  AA19B048448  AA19B048449  AA19B048450  AA19B048451  AA19B048452  AA19B048453  AA19B048454  AA19B048455  AA19B048456  AA19B048457  AA19B048458  AA19B048459  AA19B048460  AA19B048461  AA19B048462  AA19B048463  AA19B048464  AA19B048465  AA19B048466  AA19B048467  AA19B048468  AA19B048469  AA19B048470  AA19B048471  AA19B048472  AA19B048473  AA19B048474  AA19B048475  AA19B048476  AA19B048477  AA19B048478  AA19B048479  AA19B048480  AA19B048481  AA19B048482  AA19B048483  AA19B048484  AA19B048485  AA19B048486  AA19B048487  AA19B048488  AA19B048489  AA19B048490  AA19B048491  AA19B048492  AA19B048494  AA19B048495  AA19B048496  AA19B048497  AA19B048498  AA19B048499  AA19B048500  AA19B048501  AA19B048502  AA19B048503  AA19B048504  AA19B048505  AA19B048506  AA19B048507  AA19B048508  AA19B048509  AA19B048510  AA19B048511  AA19B048512  AA19B048513  AA19B048514  AA19B048515  AA19B048516  AA19B048517  AA19B048518  AA19B048519  AA19B048520  AA19B048521  AA19B048522  AA19B048523  AA19B048524  AA19B048525  AA19B048526  AA19B048527  AA19B048528  AA19B048529  AA19B048530  AA19B048531  AA19B048532  AA19B048533  AA19B048534  AA19B048535  AA19B048536  AA19B048537  AA19B048538  AA19B048539  AA19B048540  AA19B048541  AA19B048542  AA19B048543  AA19B048544  AA19B048545  AA19B048546  AA19B048547  AA19B048548  AA19B048549  AA19B048550  AA19B048551  AA19B048553  AA19C048627  AA19C048630  AA19C048632  AA19C048633  AA19C048634  AA19C048635  AA19C048637  AA19C048639  AA19C048642  AA19C048643  AA19C048645  AA19C048646  AA19C048647  AA19C048648  AA19C048649  AA19C048650  AA19C048651  AA19C048652  AA19C048654  AA19C048655  AA19C048656  AA19C048657  AA19C048658  AA19C048659  AA19C048660  AA19C048661  AA19C048662  AA19C048663  AA19C048664  AA19C048665  AA19C048666  AA19C048667  AA19C048668  AA19C048669  AA19C048670  AA19C048671  AA19C048672  AA19C048673  AA19C048674  AA19C048675  AA19C048677  AA19C048678  AA19C048680  AA19C048681  AA19C048682  AA19C048683  AA19C048684  AA19C048685  AA19C048686  AA19C048687  AA19C048688  AA19C048689  AA19C048690  AA19C048691  AA19C048692  AA19C048693  AA19C048702  AA19C048723  AA19C048724  AA19C048725  AA19C048726  AA19C048727  AA19C048728  AA19C048729  AA19C048730  AA19C048731  AA19C048732  AA19C048733  AA19C048734  AA19C048735  AA19C048736  AA19C048737  AA19C048765  AA19C048766  AA19C048767  AA19C048768  AA19C048769  AA19C048770  AA19C048771  AA19C048772  AA19C048773  AA19C048774  AA19C048775  AA19C048776  AA19C048777  AA19C048778  AA19C048779  AA19C048780  AA19C048781  AA19C048782  AA19C048783  AA19C048784  AA19C048785  AA19C048787  AA19C048788  AA19C048789  AA19C048790  AA19C048791  AA19C048792  AA19D048811  AA19D048812  AA19D048813  AA19D048814  AA19D048816  AA19D048817  AA19D048818  AA19D048819  AA19D048820  AA19D048821  AA19D048823  AA19D048825  AA19D048826  AA19D048827  AA19D048829  AA19D048830  AA19D048831  AA19D048832  AA19D048833  AA19D048834  AA19D048836  AA19D048837  AA19D048838  AA19D048839  AA19D048840  AA19D048844  AA19D048845"
    }
  ]
}