      "country": "United States",
      "city": "Indianapolis",
      "address_1": "9115 Hague Rd",
      "reason_for_recall": "Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers.  Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine  plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.",
      "address_2": "",
      "product_quantity": "6255",
      "code_info": "Catalog Numbers:04657608190; 03183807190; 05171857190; 03183807190; 11875426216; 11929429216; 11929437216.",
      "center_classification_date": "20150627",
      "distribution_pattern": "US (nationwide) including PR.",
      "state": "IN",
      "product_description": "Uric Acid plus ver.2    In vitro test for the quantitative determination of uric acid in human serum, plasma and urine or Roche/Hitachi MODULAR systems, cobas c111, COBAS Integra 800 and 400 plus systems, and cobas c systems.",
      "report_date": "20150708",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "recall_number": "Z-1931-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71182",
      "termination_date": "20170324",
      "more_code_info": "",
      "recall_initiation_date": "20150507",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}