  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80003",
      "recalling_firm": "Maquet Datascope Corp - Cardiac Assist Division",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Other",
      "distribution_pattern": "The devices were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Guam.    The devices were distributed to the following foreign countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Brunei, Bulgaria, Canada, Chile, China, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrica, Taiwan, Thailand, Turkey, UAE, UK, Vietnam, Yemen.",
      "recall_number": "Z-1928-2018",
      "product_description": "CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP), Part Number 0998-00-0800-XX & 0998-UC-0800-XX (excluding 0998-00-0800-83, 0998-UC-0800-83 & 0998-00-0800-75) and cart 0997-00-1179",
      "product_quantity": "4300",
      "reason_for_recall": "Ingress of fluids into the device can affect various electronic circuit boards which would prevent initiation or continuation of therapy.",
      "recall_initiation_date": "20180503",
      "center_classification_date": "20180531",
      "report_date": "20180606",
      "code_info": "All Serial Numbers distributed from 03/06/2012 to  04/26/2018"
    }
  ]
}