      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74059",
      "recalling_firm": "Medtronic Neuromodulation",
      "address_1": "7000 Central Ave",
      "address_2": "N/A",
      "postal_code": "55421-1241",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-1926-2016",
      "product_description": "Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators.    Used to treat chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain.",
      "product_quantity": "60,941",
      "reason_for_recall": "Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925 volts in one to two days rather than the typical 30 days. Insufficient coupling between recharger and device during recharge session was found to be key factor.",
      "recall_initiation_date": "20160503",
      "center_classification_date": "20160607",
      "termination_date": "20170804",
      "report_date": "20160615",
      "code_info": "All lot/serial numbers are affected for both Model numbers 37714 and 97714."
    }
  ]
}