{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Indianapolis",
      "address_1": "9115 Hague Rd",
      "reason_for_recall": "Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers.  Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine  plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.",
      "address_2": "",
      "product_quantity": "6255",
      "code_info": "Catalog Numbers: 05401488190; 0439980319; 05168805190; 04399803190, 04878205160.",
      "center_classification_date": "20150627",
      "distribution_pattern": "US (nationwide) including PR.",
      "state": "IN",
      "product_description": "HDL-Cholesterol plus gen.3.; and Small Supersized    Intended for use on COBAS Integra 800 and 400 plus systems, cobas c111, and cobas c systems for the quantitative determination of HDL-cholesterol concentration in serum and plasma.",
      "report_date": "20150708",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "recall_number": "Z-1925-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71182",
      "termination_date": "20170324",
      "more_code_info": "",
      "recall_initiation_date": "20150507",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}