      "country": "United States",
      "city": "Hartland",
      "address_1": "900 Walnut Ridge Dr",
      "reason_for_recall": "The units may display the alarm \"sampling line clogged\" after the flow zeroing procedure that is conducted automatically 1 hour after start up.  This could result in the potential of a small bias added to the CO2 measurement and affect the alarm notification.",
      "address_2": "",
      "product_quantity": "23 sensors (PhaseIn Units)",
      "code_info": "ISA CO2 Sensor serial numbers:    107349, 107357, 107358, 107363, 107370, 107377, 107390, 107407, 107435,   107476, 107509, 107510, 107556, 107659, 107700, 107707, 107768, 107776,   107831, 107832, 107836, 107839, and 107841.",
      "center_classification_date": "20170502",
      "distribution_pattern": "US Distribution was made to medical facilities located in AL, FL, ID, LA, MO, NC, OK, SC, TX, and UT.  There was government distribution and no foreign/military distribution.",
      "state": "WI",
      "product_description": "Schiller AG ISA CO2 Sensor (AKA: PhaseIn Unit), Merge Part Number HW-HEMO-00001.  The PhaseIn Unit is an optional accessory (sold separately) to the Merge Hemo Patient Data Module (PDM). The PDM is a major component of the Merge Hemo system.",
      "report_date": "20170510",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Merge Healthcare, Inc.",
      "recall_number": "Z-1923-2017",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "76946",
      "termination_date": "20170505",
      "more_code_info": "",
      "recall_initiation_date": "20150330",
      "postal_code": "53029-8347",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}