  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Maple Grove",
      "address_1": "2 Scimed Pl",
      "reason_for_recall": "Boston Scientific is initiating a voluntary removal of two lots of JETSTREAM SC 1.85mm and one lot of JETSTREAM XC 2.1mm Atherectomy Catheters due to these units being mislabeled. Some 1.85mm JETSTREAM SC catheters may have been labeled as 2.1mm catheters, and some 2.1mm JETSTREAM XC catheters may have been labeled as 1.85mm catheters.",
      "address_2": "",
      "product_quantity": "137",
      "code_info": "18775553, 18775552, 18822086",
      "center_classification_date": "20170502",
      "distribution_pattern": "US: AL, AZ, CT, FL, HI, IL IN, IA, KY, LA, ME, MI, MO, NJ,NY, OH, OR, PA, SC, TN, TX, VA, WA, WI.    OUS; GERMANY, GREAT BRITAIN, POLAND, NETHERLANDS",
      "state": "MN",
      "product_description": "The JETSTREAM is a rotational atherectomy catheter system designed for use in debulking and treating vascular disease in the peripheral vasculature. The  System includes multiple distal ports located at the catheter tip, which are designed to provide independent infusion and aspiration functions for the active removal of fluid, excised tissue and thrombus from the peripheral treatment site.",
      "report_date": "20170510",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Boston Scientific Corporation",
      "recall_number": "Z-1922-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76955",
      "termination_date": "20190510",
      "more_code_info": "",
      "recall_initiation_date": "20170405",
      "postal_code": "55311-1565",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}