  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74196",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including states of: AL, CA, CT, FL, GA, HI, IL, IN, KS, NE, NY,NC, OH, OR, PA, TN, TX, UT, A, WA, WI; and countries of: Australia, Belgium, Brazil, Canada, China, Denmark, Estonia, France, Germany, Hong Kong, Indonesia, Italy, Japan, Korea (Republic of), Kuwait, Mexico, New Zealand, Norway, Peru, Poland, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1915-2016",
      "product_description": "GE Revolution CT    The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications. The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different angles.",
      "product_quantity": "197 (US = 68; OUS = 129)",
      "reason_for_recall": "GEHC has become aware of a potential safety issue in the Revolution CT Patient Table that can result in unintended motion in cases involving large patients. The Revolution CT cradle can descend on its own after activation by the user. This can put the patient at risk for harm. The harm can involve the impinging or crushing of a hand if caught between the cradle and CT inner gantry bore. No injurie",
      "recall_initiation_date": "20160509",
      "center_classification_date": "20160606",
      "termination_date": "20180904",
      "report_date": "20160615",
      "code_info": "Mfg Lot or Serial # System ID  00000437624CN0 N/A  00000439694CN1 N/A  00000439777CN4 N/A  00000439930CN9 N/A  00000441224CN3 203688ERREV1  00000441575CN8 N/A  00000441624CN4 N/A  00000441648CN3 N/A  00000441916CN4 469841REVO  00000442032CN9 N/A  00000442124CN4 N/A  00000442219CN2 N/A  00000442443CN8 N/A  00000443993CN1 N/A  00000444530CN0 N/A  00000445085CN4 N/A  00000438756CN9 309655REVCT  00000439113CN2 212241REVCT  00000439205CN6 415476MCT3  00000439287CN4 415476MCT2  00000439836CN8 215662CTREV  00000440713CN6 815226RVCT  00000441459CN5 206320REV  00000441716CN8 808522RCT  00000442228CN3 281274REV  00000442994CN0 919684REVCT  00000443379CN3 608263REVCT  00000443664CN8 409772CT7  00000443900CN6 402481BEREVCT  00000444165CN5 AM14CFI65  00000444175CN4 801507REV  00000444253CN9 704384REV  00000444745CN4 808KMCCT  00000444872CN6 316268REVCT  00000445059CN9 310222REVO  438369CN1 206598REV1  438902CN9 801662REV1  439823CN6 408356RV1  439835CN0 503216SREV  441040cn3 215952CTREV  441064CN3 256HHRCT1  441066CN8 414805REVCT  441262CN3 RMHCTREV  441990CN9 206598ERREV  442592CN2 205699CTREV  442706CN8 727398REVO  443069CN0 404616CTR  443227CN4 940898REVO  443327CN2 858824REVCT  444398CN2 731541REVCT  444409CN7 386REVCT3  444840CN3 760940REVCT  444913CN8 206288SCREV  00000437624CN0 786596WKRIVO  00000439777CN4 703698REVOCT  00000439930CN9 715847RCT1  00000442124CN4 408498RV2  443327CN2 858824REVCT  00000441624CN4 217876REVO1  00000445085CN4 713792ACT35  00000441575CN8 847437ACT7  442032CN9 512DELLCT  00000442994CN0 N/A  3993CN1 972566OPREVO  445084CN7 708216CTREV  442592CN2 AM13KPM21  444165CN5 317338STVREVO  444970CN8 630527REVO  00000436942CN7 83119511730  00000439799CN8 83119074723  00000441381CN1 83130644883  00000441694CN7 83132979115  00000442968CN4 83141966438  00000439505CN9 190059CT06  00000440315CN0 220078CT04  00000441578CN2 030355CT07  00000441683CN0 CT441683CN0  00000443038CN5 CT443038CN5  00000445060CN7 204787RCT1  00000437410CN4 519646REVO1  440051CN1 E416249CT1  441839CN8 E416249CTE  442044CN4 403944RCT1  442094CN9 604988RCT1  442163CN2 416864RCT1  442378CN6 E416249CT2  443342CN1 905521RCT1  441839CN8 N/A  442219CN2 416284CT1  442443CN8 604682RCT1  00000440672CN4 82421040350  00000440940CN5 82421040345  00000440986CN8 82421050106  00000441519CN6 82421040346  00000441805CN9 82421040359  00000441899CN2 82421120479  00000442052CN7 82421030199  00000442479CN2 82421190324  00000442749CN8 82421040369  00000443071CN6 82421300072  00000443167CN2 82421190317  00000443253CN0 82421120484  00000443308CN2 82421040370  00000443741CN4 82421200266  00000444092CN1 82421120524  00000444206CN7 82421160425  00000444243CN0 82421150167  00000444317CN2 N/A  00000444470CN9 82421040372  00000444648CN0 82421100836  00000444706CN6 N/A  00000444712CN4 82421100835  00000444888CN2 82421100846  00000444926CN0 N/A  00000444969CN0 N/A  00000445019CN3 82421130179  00000445199CN3 N/A  00000442412CN3 DK1051CT04  00000444278CN6 EE4019CT05  00000436925CN2 M2322254  00000440136CN0 P4192906  00000441105CN4 X41944012  00000441768CN9 M41980112  00000442256CN4 E4191734  00000442507CN0 M44827197  00000442792CN8 M5480791  00000443967CN5 M2800929  DUMFMI25465B02 CC0618CT08  00000439368CN2 PC3383CT01  00000441210CN2 HC5284CT04  00000443675CN4 410014CT04  00000438247CN9 828210096  00000440778CN9 835210259  00000442887CN6 5973CT0019  00000440478CN6 A5624142  00000443869CN3 A51773126  00000437849CN3 REV001  00000439234CN6 REV006  00000439407CN8 REV002  00000439537CN2 REV003  00000439754CN3 REV004  00000439931CN7 REV008  00000440024CN8 REV005  00000440622CN9 REV007  00000441177CN3 REV009  00000441322CN5 REV011  00000441959CN4 REV012  00000442334CN9 REV020  00000442785CN2 REV015  00000443471CN8 REV014  00000443501CN2 REV013  00000445156CN3 REV022  00000444523CN5 N/A  00000442622CN7 850210885  00000442940CN3 850210884  00000443590CN5 850210890  00000445179CN5 N/A  00000441549CN3 KW1098CT01  00000442802CN5 KW1011CT03  00000444791CN8 KW1011CT04  00000444110CN1 CT444110CN01  00000439632CN1 85109909185  00000442289CN5 NO1031CT01  00000442868CN6 NO1146CT02  00000445041CN7 N/A  00000442748CN0 CT442748CN0  00000444156CN4 CT444156CN4  00000439834CN3 PL1008CT04  00000440221CN0 PL1150CT03  00000440441CN4 SA1012CT10  00000443703CN4 N/A  00000444345CN3 N/A  00000444375CN0 SA1012CT12  00000443956CN8 6073CT0013  00000443448CN6 DPN88638  00000444082CN2 MPX30114  00000439660CN2 D007CT07  00000440739CN1 A010CT16  00000441292CN0 A010CT17  00000443380CN1 A201CT02  00000443400CN7 A201CT01  00000443540CN0 A201CT03  00000443952CN7 R002CT09  00000444016CN0 A007CT04  00000444672CN0 N/A  00000438438CN4 900010CT06  00000439169CN4 260010CT19  00000443556CN6 260010CT20  00000442350CN5 853210292  00000438994CN6 AE1477CT01  00000444038CN4 AE1015CT02  DUMFMI25465B01 00216CTS13  00000440706CN0 00160CTS03  00000440906CN6 00487CTS03  00000440928CN0 00243CTS11  00000443409CN8 00145CTS01  00000443994CN9 00095CTS04"
    }
  ]
}