  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "West Chester",
      "address_1": "1302 Wrights Ln E",
      "reason_for_recall": "certain lots of the Titanium Recon Screws (6.5mm ) were found to contain the incorrect insert GP0760-A (Synthes Titanium Buttress Locking Plate System), instead of the correct insert GP0706-E (Synthes Titanium Intramedullary Nails and Components).",
      "address_2": "",
      "product_quantity": "276",
      "code_info": "Part Numbers :   04.003.024  04.003.025  04.003.026  04.003.027  04.003.028  04.003.029  04.003.031    with Lot Numbers:   8638958  8638957  8534954  8638955  8556020  8556018  8653032  8653036",
      "center_classification_date": "20150626",
      "distribution_pattern": "Nationwide Distribution.",
      "state": "PA",
      "product_description": "DePuy Synthes 6.5 mm Titanium Recon Screws w/T25 Stardrive (70mm,  75mm, 80mm, 85mm, 90mm, 95mm, 105mm); part of the Expert Lateral Femur Nail System and indicated to stabilize Subtrochanteric fractures, Ipsilateral neck/shaft fractures, Femoral shaft fractures, Impending pathologic fractures, and malunions/Non-unions.",
      "report_date": "20150708",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Synthes, Inc.",
      "recall_number": "Z-1915-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71361",
      "termination_date": "20151218",
      "more_code_info": "",
      "recall_initiation_date": "20150527",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}