  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74080",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "N/A",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US nationwide distribution.",
      "recall_number": "Z-1912-2016",
      "product_description": "DePuy Synthes 8.0 mm Flexible Shaft Reamer, 360 mm, Orthopedic manual surgical instrument.  Part number 351.43",
      "product_quantity": "244 units",
      "reason_for_recall": "Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions.  The high growth inhibition levels observed during testing could be attributed to corrosion of the device at solder points.  This could potentially be reproduced during use and reprocessing.",
      "recall_initiation_date": "20160502",
      "center_classification_date": "20160606",
      "termination_date": "20161213",
      "report_date": "20160615",
      "code_info": "Catalog ID # 351.43  Lot numbers: XXX5071, 3945479, 4738380, XXX5072, 3997000, 4816738, XXX5073,   3996999, 4816739, A3CN449, 3997338, 4441056, 3007721, 4326050, 5407442, 3148892, 4307973, 5582819, 3175916, 4387338, 5722664, 3175202, 4441055, 5813837, 3175201, 4441079, 5786778, 3921562, 4559672, 5800844, 3928332, 4561383, 5903413, 3935789, 4619549, 5910995."
    }
  ]
}