  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92214",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "International distribution in the country of Germany.",
      "recall_number": "Z-1907-2023",
      "product_description": "Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Model Number 0998-00-0800-55, Part Number 0998-00-0800-XX",
      "product_quantity": "4 units",
      "reason_for_recall": "The digital IFUs provided with the Cardiosave IABP during the software update to version D.00 incorrectly annotated compliance to standard 60601-1-12:2014 within the updated IFU.",
      "recall_initiation_date": "20230504",
      "center_classification_date": "20230614",
      "report_date": "20230621",
      "code_info": "UDI-DI: 10607567108414; Serial Numbers: CB286971I8 CB283804G8 CB283867G8 CB283522G8"
    }
  ]
}