{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87960",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US: Alabama,  Alaska,  Arizona,  Arkansas,  California,  Florida,  Georgia,  Illinois,  Iowa,  Kentucky,  Louisiana,  Maine,  Maryland,  Massachusetts,  Michigan,  Mississippi,  Missouri,  New Hampshire,  New Jersey,  New York,  North Dakota,  Ohio,  Oklahoma,  Pennsylvania,  Puerto Rico,  Tennessee,  Texas,  Virginia,  West Virginia,  Wisconsin,  Wyoming,    OUS: Andorra, Austria, Croatia, Czech Republic, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, Israel, Italy, Japan, Jordan, Poland, Portugal, San Marino, Slovakia, Slovenia, Spain, United Arab Emirates",
      "recall_number": "Z-1905-2021",
      "product_description": "Access SARS CoV-2 lgG II Reagent, REF: C69057, IVD CE UDI: (01)15099590742744",
      "product_quantity": "13,337 units",
      "reason_for_recall": "SARS-CoV-2 lgG II numerical results  may potentially  be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.",
      "recall_initiation_date": "20210507",
      "center_classification_date": "20210621",
      "termination_date": "20230327",
      "report_date": "20210630",
      "code_info": "All Lots"
    }
  ]
}