  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Best",
      "state": "N/A",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87978",
      "recalling_firm": "Philips Medical Systems Nederlands",
      "address_1": "Veenpluis 4-6",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution. US states of: AL, AZ, FL, GA, IL, MA, MO, NJ, OR, and VA;  Netherlands, United Kingdom, Poland, Spain, France, Czech Republic, India, Brazil, Singapore, Japan, Colombia, United Arab Emirates, China, Russian Federation, Oman, and Mexico.",
      "recall_number": "Z-1904-2021",
      "product_description": "Philips Allura Xper System: Interventional Fluoroscopic X-Ray System  Product No.: 722026, 722027, 722028, 722029, 722035, 722038, and 722058.",
      "product_quantity": "42 devices",
      "reason_for_recall": "Certain low-voltage DC power supplies (DCPS) used in these products may be subject to an increased probability of failure. Failure of a DCPS may result in the sudden loss of imaging functionality or mechanical movement, depending on what subsystems the DCPS is powering.",
      "recall_initiation_date": "20160623",
      "center_classification_date": "20210621",
      "termination_date": "20230208",
      "report_date": "20210630",
      "code_info": "Product No.: 722026, 722027, 722028, 722029, 722035, 722038, and 722058.  Date(s) of Distribution:   16-SEP-2014  -  04-MAY-2016.    System equipment numbers:  64937259  65862117  63304032  63151474  63250793  63303889  63304149  64140063  63855729  63516113  64077887  66407452  66350303  65342973  63210187  63511276;    66923399  66422060  64085323  61099178  64103799  67936786  66730784  66736708  66628236  63883590  64916638  64927462  65345691  66517161  66403768  63218881  67360962  69856976  65258983  66167983  64778594  61655917  65206981  67874859"
    }
  ]
}