  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87812",
      "recalling_firm": "Adler MicroMed, Inc.",
      "address_1": "6842 Elaine Way",
      "address_2": "N/A",
      "postal_code": "92120-3933",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US distribution: FL, IL, MA, and PA.",
      "recall_number": "Z-1901-2021",
      "product_description": "Sapi Med sterile disposable colorectal anoscopes and proctoscopes.    REF/Product Description:  A.4083/THE BEAK - Surgical proctoscope. Disposable & sterile. Transparent. Self light;   A.4519/MPR Short - Multi Purpose Rectoscope. 190mm long. Sterile & Disposable;  A.4086/EPOFLIER - Surgical kit made of a 34,3mm wide anoscope and an opened retractor named GULL. Disposable & sterile;  A.4084/THE BEAK Diagnostic- Disposable proctoscope. Transparent. Sterile & Self lighting;   A.4522/ MPR - Multi Purpose Rectoscope. 220mm long. Sterile & Disposable",
      "product_quantity": "1,858 devices",
      "reason_for_recall": "Due to issue with product sterility.  Sterilization service provider had falsified records.",
      "recall_initiation_date": "20210423",
      "center_classification_date": "20210621",
      "termination_date": "20221004",
      "report_date": "20210630",
      "code_info": "REF Numbers/Lot Numbers:  REF A.4083 /Lot Numbers: 18D004l, 19C010l, 19l018l, 20l027l;  REF A.4084/Lot Number: 19l015l;  REF A.4519/Lot Numbers: 18l013l, 18l012l, 19D002l, 19F012l;  REF A.4522/Lot Numbers: 18G003l, 19K023l;  REF A.4086/Lot Number: 18F027l;    Sterilization Batch Numbers:  18042311;  18092611;  18071211_1;  19040512;  18073112;  19092713;  19120612;  19063012;  20102212;  19063012;"
    }
  ]
}