  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Arvada",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80072",
      "recalling_firm": "LivaNova USA",
      "address_1": "14401 W 65th Way",
      "address_2": "N/A",
      "postal_code": "80004-3503",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "MN, TX, WA, OR, IL, OH, SC, AK, PA, CA, WI, GA, NJ, FL, AZ, KY, MS, MO, LA, AR, MA, MI, MT, CO, TN, NE, NC, VA, NV, HI, PR, WY, AL, OK, IN, CT, ME, IA, SC, KS, NM, SD, DC, MD, ND,",
      "recall_number": "Z-1901-2018",
      "product_description": "Sorin Stockert Heater-Cooler System 3T,     a) Item #: 16-02-81, 240 v/60 Hz;    b) Item #: 16-02-82, 208 v/60 Hz;    c) Item #: 16-02-85, 120 v/60 Hz",
      "product_quantity": "2408",
      "reason_for_recall": "Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.",
      "recall_initiation_date": "20180420",
      "center_classification_date": "20180518",
      "termination_date": "20250205",
      "report_date": "20180530",
      "code_info": "Serial #:  a) 16S10743 - 6S11708    b) 16S10772 - 16S16840    C) 16S10958 - 16S16847"
    }
  ]
}