  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83082",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1900-2019",
      "product_description": "Siemens Artis Q biplane---Interventional Fluoroscopic X-Ray System  Model Number: 10848282",
      "product_quantity": "73",
      "reason_for_recall": "Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off,  can result in severe injuries to patients and operating personnel",
      "recall_initiation_date": "20190606",
      "center_classification_date": "20190628",
      "termination_date": "20210209",
      "report_date": "20190710",
      "code_info": "Serial Numbers:  121013  121057  121056  121058  121006  121185  121016  121027  121020  121321  121448  121040  121049  121015  121094  121082  121089  121383  121144  121319  121137  121085  121155  121193  121136  121165  121140  121141  121151  121168  121169  121154  121211  121188  121415  121183  121209  121187  121309  121200  121313  121206  121361  121366  121359  121393  121301  121407  121406  121352  121379  121364  121342  121471  121520  121411  121412  121402  121430  121502  121421  121472  121460  121457  121515  121444  121516  121510  121491  121525  121530  121527"
    }
  ]
}