  "results": [
    {
      "country": "United States",
      "city": "Tempe",
      "address_1": "1625 W 3rd St Bldg 1",
      "reason_for_recall": "Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.",
      "address_2": "",
      "product_quantity": "Total Devices=23,029 (U.S.=16,582; O.U.S. =6,447)",
      "code_info": "Product Code/GTIN:  VENEL14040 /00801741102295;  VENEL14060 /00801741102301;  VENEL14080 /00801741102318;  VENEL14100 /00801741102325;  VENEL14120 /00801741102332;  VENEL14140 /00801741102349;  VENEL14160 /00801741102356;  VENEM14040 /00801741102226;  VENEM14060 /00801741102233;  VENEM14080 /00801741102240;  VENEM14100 /00801741102257;  VENEM14120 /00801741102264;  VENEM14140 /00801741102271;  VENEM14160 /00801741102288;  VENUL14040 /00801741103483;  VENUL14060 /00801741103490;  VENUL14080 /00801741103506;  VENUL14100 /00801741103513;  VENUL14120 /00801741103520;  VENUL14140 /00801741103537;  VENUL14160 /00801741103544;  VENUM14040 /00801741103414;  VENUM14060 /00801741103421;  VENUM14080 /00801741103438;  VENUM14100 /00801741103445;  VENUM14120 /00801741103452;  VENUM14140 /00801741103469;  VENUM14160 /00801741103476;    U.S. Device Serial Numbers:  ANEP2484;  ANEP2485;  ANEP2487;  ANEP3962;  ANEP3964;  ANEP3966;  ANEP3968;  ANEP3972;  ANEP3973;  ANEP3979;  ANEP3982;  ANEP3986;  ANEP5357;  ANEP5362;  ANEP5363;  ANEP5364;  ANEP5367;  ANEP5368;  ANEP5370;  ANEP5373;  ANEP5375;  ANEP5378;  ANEQ0213;  ANEQ0216;  ANEQ0217;  ANEQ0219;  ANEQ0220;  ANEQ0221;  ANEQ1381;  ANEQ1382;  ANEQ1383;  ANEQ1384;  ANEQ1402;  ANEQ1403;  ANEQ1404;  ANEQ1405;  ANEQ1406;  ANEQ1407;  ANEQ1408;  ANEQ1409;  ANEQ1410;  ANEQ1411;  ANEQ2758;  ANEQ2759;  ANEQ2777;  ANEQ2778;  ANEQ3075;  ANEQ3076;  ANEQ4080;  ANEQ4081;  ANEQ4082;  ANEQ4083;  ANEQ4092;  ANEQ4093;  ANEQ4094;  ANEQ4106;  ANEQ4387;  ANEQ4747;  ANEQ4748;  ANEQ4749;  ANEQ4752;  ANEQ4981;  ANEQ4982;  ANEQ4983;  ANEQ4984;  ANEQ4985;  ANEQ4998;  ANEQ4999;  ANEQ5000;  ANEQ5001;  ANEQ5002;  ANER0479;  ANER0480;  ANER0481;  ANER0492;  ANER0493;  ANER0533;  ANER0534;  ANER1497;  ANER1503;  ANER2390;  ANER2391;  ANER2392;  ANER2393;  ANER2394;  ANER2395;  ANER2798;  ANER2799;  ANER2800;  ANER2817;  ANER2818;  ANER2819;  ANER2837;  ANER2845;  ANER2851;  ANER2852;  ANER2853;  ANER2854;  ANER4202;  ANER4205;  ANER4210;  ANER4211;  ANER4213;  ANER4216;  ANER4219;  ANER4221;  ANER4225;  ANER4232;  ANER5121;  ANER5122;  ANER5128;  ANER5137;  ANER5138;  ANER5143;  ANER5154;  ANER5161;  ANER5170;  ANER5176;  ANER5177;  ANER5178;  ANES1034;  ANES1050;  ANES2225;  ANES2226;  ANES2237;  ANES3023;  ANES3039;  ANES3040;  ANES3051;  ANES3052;  ANES3057;  ANES3058;  ANES3061;  ANES3062;  ANES3063;  ANES3064;  ANES3130;  ANES4104;  ANES4119;  ANET0569;  ANET0570;  ANET0571;  ANET0572;  ANET0573;  ANET0574;  ANET1485;  ANET1486;  ANET1487;  ANET1488;  ANET1498;  ANET1502;  ANET1503;  ANET1504;  ANET1505;  ANET2538;  ANET2542;  ANET2543;  ANET2544;  ANET2560;  ANET2568;  ANET2569;  ANET2570;  ANET2571;  ANET2572;  ANET3801;  ANET3802;  ANET3803;  ANET3819;  ANET3820;  ANET3821;  ANET3822;  ANET3823;  ANET3842;  ANET3879;  ANET3880;  ANEU0744;  ANEU0749;  ANEU0750;  ANEU0751;  ANEU0752;  ANEU0753;  ANEU0754;  ANEU0772;  ANEU1595;  ANEU1607;  ANEU1608;  ANEU1609;  ANEU1610;  ANEU1627;  ANEU2434;  ANEU2435;  ANEU2436;  ANEU2443;  ANEU2444;  ANEU2445;  ANEU2446;  ANEU2447;  ANEU2448;  ANEU3279;  ANEU3291;  ANEU3292;  ANEU3293;  ANEU3294;  ANEU4228;  ANEU4233;  ANEU4234;  ANEU4241;  ANEU4242;  ANEU4247;  ANEU4248;  ANEU4249;  ANEU4257;  ANEV0589;  ANEV0597;  ANEV0600;  ANEV0604;  ANEV0615;  ANEV1719;  ANEV1720;  ANEV1721;  ANEV1722;  ANEV1741;  ANEV1742;  ANEV1743;  ANEV1744;  ANEV1745;  ANEV2618;  ANEV3071;  ANEV3072;  ANEV3083;  ANEV3084;  ANEV3085;  ANEV3497;  ANEV4026;  ANEV4027;  ANEV4028;  ANEV4029;  ANEV4030;  ANEV4047;  ANEV4048;  ANEV4049;  ANEV4050;  ANEV4051;  ANEV4074;  ANEW0555;  ANEW0567;  ANEW0568;  ANEW0569;  ANEW0570;  ANEW0571;  ANEW0572;  ANEW0573;  ANEW0574;  ANEW0596;  ANEW0597;  ANEW1518;  ANEW1586;  ANEW2008;  ANEW2009;  ANEW2010;  ANEW2011;  ANEW2035;  ANEW2036;  ANEW2236;  ANEW2237;  ANEW2530;  ANEW2774;  ANEW3382;  ANEW3383;  ANEW3384;  ANEW3385;  ANEW3398;  ANEW3399;  ANEW3400;  ANEW3423;  ANEW3425;  ANEW3426;  ANEX0698;  ANEX0714;  ANEX0718;  ANEX0725;  ANEX0756;  ANEX1728;  ANEX1729;  ANEX1731;  ANEX1732;  ANEX1733;  ANEX1735;  ANEX1738;  ANEX1741;  ANEX1745;  ANEX1749;  ANEX3102;  ANEX3103;  ANEX3106;  ANEX3107;  ANEX3108;  ANEX3109;  ANEX3110;  ANEX3111;  ANEX3114;  ANEX3128;  ANEX3133;  ANEX3134;  ANEX3140;  ANEX4723;  ANEX4724;  ANEX4727;  ANEX4728;  ANEX4729;  ANEX4739;  ANEX4741;  ANEX4749;  ANEX4754;  ANEX4757;  ANEX4771;  ANEX4774;  ANEX4776;  ANEX6029;  ANEX6030;  ANEX6031;  ANEX6040;  ANEX6041;  ANEX6043;  ANEX6044;  ANEX6047;  ANEX6050;  ANEX6051;  ANEX6058;  ANEX6069;  ANEY1238;  ANEY1239;  ANEY1245;  ANEY1252;  ANEY1253;  ANEY1260;  ANEY1261;  ANEY1270;  ANEY1271;  ANEY1281;  ANEY2057;  ANEY2058;  ANEY2059;  ANEY2060;  ANEY2061;  ANEY2062;  ANEY2063;  ANEY3215;  ANEY3216;  ANEY3219;  ANEY3220;  ANEY3221;  ANEY3222;  ANEY3223;  ANEY3224;  ANEY3227;  ANEY3228;  ANEY3229;  ANEY3230;  ANEY3233;  ANEY3234;  ANEY3259;  ANEY4238;  ANEY4330;  ANEY4331;  ANEY4332;  ANEY4333;  ANEY4334;  ANEY4335;  ANEY4336;  ANEY4337;  ANEY4338;  ANEY4339;  ANEY4340;  ANEY4341;  ANEY4342;  ANEY4343;  ANEY4348;  ANEY4559;  ANEY4561;  ANEZ0934;  ANEZ0935;  ANEZ0936;  ANEZ0938;  ANEZ0939;  ANEZ0940;  ANEZ0941;  ANEZ0942;  ANEZ0943;  ANEZ0946;  ANEZ0947;  ANEZ0948;  ANEZ0949;  ANEZ0950;  ANEZ0951;  ANEZ0952;  ANEZ0953;  ANEZ1475;  ANEZ2108;  ANEZ2114;  ANEZ2115;  ANEZ2116;  ANEZ2117;  ANEZ2125;  ANEZ2126;  ANEZ2127;  ANEZ2128;  ANEZ2129;  ANEZ2130;  ANEZ2141;  ANEZ2142;  ANEZ2143;  ANEZ2144;  ANEZ2145;  ANEZ2146;  ANEZ3440;  ANEZ3441;  ANEZ3456;  ANEZ3457;  ANEZ3458;  ANEZ3459;  ANEZ3460;  ANEZ3461;  ANEZ3462;  ANEZ3463;  ANEZ3499;  ANEZ3501;  ANEZ3507;  ANEZ3508;  ANEZ3509;  ANEZ3510;  ANEZ3511;  ANEZ3512;  ANEZ3842;  ANEZ3843;  ANEZ3844;  ANEZ3845;  ANEZ3846;  ANFN1461;  ANFN1462;  ANFN1463;  ANFN1464;  ANFN1465;  ANFN1466;  ANFN1467;  ANFN1468;  ANFN1469;  ANFN1470;  ANFN1471;  ANFN1472;  ANFN1473;  ANFN1474;  ANFN1476;  ANFN1477;  ANFN1479;  ANFN1711;  ANFN2401;  ANFN2402;  ANFN2403;  ANFN2404;  ANFN2794;  ANFN2795;  ANFN2796;  ANFN2797;  ANFN2798;  ANFN2799;  ANFN2807;  ANFN2811;  ANFN2813;  ANFN2815;  ANFN2826;  ANFN2827;  ANFN2828;  ANFN2830;  ANFN2833;  ANFN2834;  ANFN2856;  ANFN3995;  ANFN4003;  ANFN4004;  ANFN4005;  ANFN4006;  ANFP0898;  ANFP0899;  ANFP0900;  ANFP0924;  ANFP1521;  ANFP1522;  ANDS0732;  ANDS0731;  ANDS1276;  ANDS1290;  ANDS1303;  ANDS3953;  ANDS1316;  ANDS1315;  ANDS1321;  ANDS1329;  ANDS2463;  ANDT1238;  ANDT1239;  ANDT1237;  ANDS4789;  ANDS4855;  ANDS3954;  ANDS4791;  ANDS4790;  ANDT1074;  ANDT1075;  ANDT1076;  ANDT1690;  ANDT2909;  ANDT2913;  ANDT2915;  ANDT2910;  ANDT2914;  ANDT2912;  ANDU0099;  ANDU0125;  ANDT2911;  ANDT2916;  ANDU0124;  ANDU0100;  ANDT2917;  ANDU0126;  ANDU1127;  ANDU1128;  ANDU0127;  ANDU1115;  ANDU1129;  ANDU1863;  ANDU1857;  ANDU1864;  ANDU1885;  ANDU1868;  ANDU1886;  ANDU1895;  ANDU1901;  ANDU1908;  ANDV0722;  ANDU1907;  ANDU3116;  ANDU3146;  ANDU3117;  ANDU4030;  ANDU4037;  ANDU4034;  ANDU4033;  ANDU4044;  ANDV0725;  ANDU4077;  ANDV3780;  ANDU4061;  ANDV0724;  ANDV3177;  ANDV3176;  ANDU4056;  ANDV3954;  ANDV3956;  ANDV3955;  ANDV3783;  ANDU4081;  ANDV0255;  ANDV3953;  ANDV0266;  ANDV0267;  ANDV0277;  ANDV0254;  ANDV0276;  ANDV0288;  ANDV0301;  ANDV0302;  ANDV1213;  ANDV1212;  ANDV1248;  ANDV1238;  ANDV1242;  ANDV1250;  ANDV1251;  ANDV1255;  ANDV1260;  ANDV1262;  ANDV2543;  ANDV2523;  ANDV2544;  ANDV2524;  ANDV2539;  ANDV2525;  ANDV2540;  ANDV2538;  ANDV2521;  ANDV2541;  ANDV1259;  ANDV3578;  ANDV3611;  ANDV2542;  ANDV3610;  ANDV2522;  ANDV3607;  ANDV3612;  ANDV3608;  ANDV3577;  ANDV3609;  ANDV4575;  ANDV4571;  ANDV4572;  ANDV4556;  ANDV4574;  ANDV4570;  ANDV4557;  ANDV4573;  ANDW1491;  ANDW1492;  ANDW1494;  ANDW1495;  ANDW1493;  ANDW1496;  ANDW2049;  ANDW2052;  ANDW2048;  ANDW2051;  ANDW2050;  ANDW1497;  ANDW2054;  ANDX1478;  ANDW2053;  ANDX1477;  ANDX1389;  ANDX1439;  ANDX1431;  ANDX1437;  ANDX2329;  ANDW3627;  ANDX1432;  ANDX2428;  ANDX2330;  ANDX1433;  ANDX1436;  ANDX1438;  ANDX1434;  ANDX1435;  ANDW4380;  ANDW4370;  ANDW4363;  ANDW4357;  ANDW4354;  ANDW4337;  ANDW4343;  ANDW4394;  ANDW4397;  ANDW4377;  ANDW4351;  ANDW4352;  ANDW4404;  ANDX6143;  ANDX6145;  ANDX0705;  ANDW4418;  ANDX0753;  ANDW4411;  ANDW4387;  ANDX0732;  ANDX0731;  ANDX0744;  ANDX2176;  ANDX2175;  ANDX2196;  ANDX2179;  ANDX2195;  ANDX2197;  ANDX2177;  ANDX2178;  ANDX2194;  ANDX2174;  ANDX3090;  ANDX3087;  ANDX3108;  ANDX3110;  ANDX3091;  ANDX3109;  ANDX3089;  ANDX3088;  ANDX5333;  ANDX5334;  ANDX5332;  ANDX5391;  ANDX5393;  ANDX5404;  ANDX5403;  ANDX5401;  ANDX5392;  ANDX5402;  ANDY1239;  ANDY1248;  ANDY2472;  ANDY1251;  ANDY1250;  ANDY2489;  ANDY2474;  ANDZ2798;  ANDY1249;  ANDY2492;  ANDY3312;  ANDY2473;  ANDY3314;  ANDY3327;  ANDY2490;  ANDY2491;  ANDY2471;  ANDY3313;  ANDY3329;  ANDY3328;  ANDY3332;  ANDY3330;  ANDZ1259;  ANDZ1264;  ANDZ1261;  ANDY3331;  ANDZ1378;  ANDZ1382;  ANDZ1263;  ANDZ1260;  ANDZ1383;  ANDZ1381;  ANDZ1380;  ANDZ1379;  ANDZ1262;  ANDZ1359;  ANDZ2310;  ANDZ2311;  ANDZ1384;  ANDZ2330;  ANDZ2332;  ANDZ2327;  ANDZ2328;  ANDZ1385;  ANDZ2331;  ANDZ2329;  ANDZ2312;  ANDZ2313;  ANDZ3842;  ANDZ3865;  ANDZ3860;  ANDZ3871;  ANDZ3846;  ANDZ3866;  ANEN1097;  ANEN1087;  ANEN1117;  ANEN1116;  ANEN3443;  ANEN1115;  ANEN1113;  ANEN1096;  ANEN2172;  ANEN2170;  ANEN2162;  ANEN1114;  ANEN2161;  ANEN1095;  ANEN2175;  ANEN2168;  ANEN2169;  ANEN2178;  ANEN2171;  ANEN3415;  ANEN3425;  ANEN3416;  ANEP4060;  ANEN3414;  ANEP3568;  ANEN3437;  ANEP3567;  ANEP4062;  ANEN3413;  ANEN3412;  ANEN2167;  ANEN3418;  ANEN4913;  ANEN3417;  ANEP4061;  ANEN4904;  ANEN4926;  ANEN4903;  ANEN4905;  ANEN4914;  ANEP0879;  ANEN4910;  ANEN4923;  ANEN4902;  ANEN4912;  ANEN4925;  ANEP0899;  ANEP0884;  ANEN4930;  ANEP0909;  ANEP0904;  ANEP0880;  ANEP0881;  ANEP0896;  ANEP0883;  ANEP0893;  ANEP0905;  ANEP0907;  ANEP0902;  ANEP0891;  ANEP2470;  ANEP2466;  ANEP2465;  ANEP2454;  ANEP2474;  ANEP2468;  ANEP2467;  ANEP2462;  ANEP2472;  ANEP2458;  ANEP2459;  ANEP0898;  ANEP2482;    O.U.S. Device Serial Numbers:  ANCY1572;  ANCY1572;  ANDS1298;  ANDS4185;  ANDS4185;  ANDX3065;  ANDX5353;  ANDY1226;  ANDY1226;  ANEQ1357;  ANEQ2737;  ANEQ2737;  ANEQ4070;  ANFN1494;  ANDQ3038;  ANDQ3878;  ANDQ5252;  ANDS1298;  ANDS1298;  ANDS4185;  ANDX0765;  ANDZ2279;  ANDZ2279;  ANDZ3828;  ANEN1065;  ANEN2142;  ANEN3391;  ANES1029;  ANES4089;  ANEX1705;  ANEZ3425;  ANDV4538;  ANFN1494;  ANEQ4965;  ANEQ2737;  ANDX0765;  ANDX5353;  ANEQ4965;  ANDS4185;  ANEQ2737;  ANEQ4965;  ANCY1569;  ANDS4185;  ANEQ4965;  ANBZ2682;  ANDS1298;  ANDS1298;  ANEN1065;  ANEQ4965;  ANEQ2737;  ANEQ4965;  ANDQ3038;  ANFN1494;  ANDQ3038;  ANDS4185;  ANDS4185;  ANES4089;  ANDX3065;  ANDX5353;  ANEQ2737;  ANCY1572;  ANDS1298;  ANDX2427;  ANDX5353;  ANEQ2737;  ANEQ4965;  ANEZ0985;  ANCY1572;  ANEQ1357;  ANDX5353;  ANEQ4070;  ANEQ4965;  ANDX5353;  ANDZ3828;  ANEQ4965;  ANEQ1357;  ANEQ2737;  ANEQ4965;  ANEZ0985;  ANFN1494;  ANEZ3425;  ANFN2845;  ANEQ2737;  ANDS4185;  ANEQ2737;  ANCY1572;  ANDX2427;  ANFP0913;  ANEZ0985;  ANDS1298;  ANFN2845;  ANCV0499;  ANDX5353;  ANEQ2737;  ANCV0499;  ANDS4185;  ANDQ3878;  ANEQ4965;  ANEQ4965;  ANEQ2737;  ANEZ3425;  ANEZ3425;  ANES1029;  ANDS4185;  ANEQ4070;  ANEQ4070;  ANES4089;  ANEQ4965;  ANES1029;  ANFN2845;  ANDQ3038;  ANES1029;  ANDS1298;  ANCY1572;  ANEQ2737;  ANDS4185;  ANDX5353;  ANEQ4965;  ANDZ2279;  ANDZ3828;  ANEQ2737;  ANES1029;  ANDZ3828;  ANCY1572;  ANCV0499;  ANDS1298;  ANDW2003;  ANEN1065;  ANEQ1357;  ANEQ4070;  ANDW2003;  ANDZ2279;  ANEQ2737;  ANES4089;  ANEZ0985;  ANDX5353;  ANES1029;  ANEZ0985;  ANFP2378;  ANDS1298;  ANDX2427;  ANDX2427;  ANES1029;  ANEQ1357;  ANDX2427;  ANDZ2279;  ANEN1065;  ANDZ3828;  ANDS4185;  ANDX2427;  ANDQ3038;  ANDQ3038;  ANDQ3038;  ANDQ3878;  ANDQ5252;  ANDS1298;  ANDS4185;  ANDV4538;  ANDW2003;  ANDW3567;  ANDX2427;  ANDX5353;  ANDX5353;  ANDX5353;  ANDY1226;  ANDZ2279;  ANEN2142;  ANEN2142;  ANEX1705;  ANEY2032;  ANEY2032;  ANEY2032;  ANEZ3425;  ANFN1494;  ANFN2845;  ANFN2845;  ANEN3391;  ANEQ2737;  ANES1029;  ANES1029;  ANEY3213;  ANEZ3425;  ANFN1494;  ANFP0913;  ANFP0913;  ANDQ5252;  ANDX3065;  ANEQ4070;  ANES4089;  ANEY3213;  ANEQ4965;  ANDS4185;  ANEQ4965;  ANES1029;  ANEQ2737;  ANEQ4965;  ANDS4185;  ANDS4185;  ANCV0499;  ANCY1569;  ANDS4185;  ANEQ2737;  ANEQ2737;  ANCY0520;  ANDS1298;  ANDZ2279;  ANEQ1357;  ANEQ4070;  ANEQ2737;  ANES1029;  ANES4089;  ANFN2845;  ANDY1226;  ANES1029;  ANDQ3038;  ANDX2427;  ANDX3065;  ANDX5353;  ANDY1226;  ANEN1065;  ANEN2142;  ANEN1065;  ANDX2427;  ANDZ3828;  ANDS4185;  ANDS4185;  ANDW3567;  ANFP2378;  ANCY1569;  ANDQ3878;  ANEQ4070;  ANEY2032;  ANCS3349;  ANDY1226;  ANEZ3425;  ANDQ5252;  ANDS4185;  ANCT4325;  ANEQ2737;  ANCV0499;  ANEQ2737;  ANEQ4965;  ANES1029;  ANEN2142;  ANDX5353;  ANEQ4070;  ANEZ0985;  ANEZ0985;  ANEZ0985;  ANEY3213;  ANEQ2737;  ANEY2032;  ANEY2032;  ANEY2032;  ANEY2032;  ANEY3213;  ANFP0913;  ANFP2378;  ANEQ4965;  ANDV4538;  ANDX3065;  ANDQ0319;  ANDS1311;  ANDS4773;  ANDS4773;  ANDS4840;  ANDS4840;  ANDV0709;  ANDW3559;  ANDY4958;  ANDZ1324;  ANDZ2275;  ANDZ2275;  ANDZ3838;  ANEQ4753;  ANEQ4966;  ANEZ3932;  ANFN2788;  ANFN3978;  ANDQ3870;  ANDR2269;  ANDR4198;  ANDX0711;  ANDX2029;  ANDZ2275;  ANDZ3838;  ANEN3392;  ANEN4876;  ANES1044;  ANEZ3413;  ANEZ3932;  ANEZ3932;  ANFN2788;  ANEQ4753;  ANER0528;;  ANFP2354  ANDQ5253;  ANES2213;  ANDX5372;  ANEQ0218;  ANEQ4753;  ANFN1446;  ANEN3392;  ANEP3925;  ANEQ0218;  ANDQ0319;  ANDQ5253;  ANDS4773;  ANDX5372;  ANEQ2738;  ANDQ3870;  ANDY4958;  ANDV0709;  ANDX3059;  ANDY4958;  ANEN4876;  ANEP3925;  ANEQ4753;  ANEQ4753;  ANEQ4753;  ANEQ4753;  ANEQ4966;  ANFN1446;  ANEP2432;  ANEQ0218;  ANDP3984;  ANDQ0318;  ANEQ4966;  ANDQ0318;  ANDR4198;  ANEQ1358;  ANEQ1358;  ANDW3559;  ANEQ1358;  ANEQ4753;  ANEQ4753;  ANFP0881;  ANEZ3932;  ANDR2269;  ANDV0709;  ANEZ3932;  ANEQ4753;  ANDQ3870;  ANEQ4753;  ANFP2354;  ANDQ3870;  ANDX3059;  ANDQ0318;  ANDQ0319;  ANDZ2275;  ANEZ2101;  ANDQ0319;  ANDQ5253;  ANDR4198;  ANEP2432;  ANDQ0319;  ANER0528;  ANDS1311;  ANES2213;  ANFP2354;  ANDQ5253;  ANEZ2101;  ANDX2029;  ANEQ2738;  ANDS1311;  ANEP3925;  ANEZ0925;  ANEZ0925;  ANDQ0318;  ANCQ1442;  ANDY1218;  ANEN4876;  ANDX3059;  ANDZ2275;  ANEQ4753;  ANDQ5253;  ANDV0709;  ANDX3059;  ANEN3392;  ANEP3925;  ANEQ4753;  ANEQ4966;  ANEZ2101;  ANFN3978;  ANEP2432;  ANDR4198;  ANEP3925;  ANFP0881;  ANDQ5253;  ANDV0709;  ANEZ3413;  ANER0528;  ANEP3925;  ANEZ2101;  ANDS1311;  ANDS4840;  ANDX5372;  ANEZ3413;  ANDQ5253;  ANDS4840;  ANDV0709;  ANEN4876;  ANDR4198;  ANFP2354;  ANCN1544;  ANDX3059;  ANEQ4753;  ANEQ4753;  ANDX5372;  ANEQ4966;  ANDQ5253;  ANEZ3413;  ANEZ3413;  ANDX2029;  ANEN4876;  ANEP0871;  ANEP2432;  ANEZ0925;  ANDN0812;  ANEQ0218;  ANFP2354;  ANDQ5253;  ANDS1311;  ANDQ0319;  ANDV0709;  ANDN0598;  ANEQ2738;  ANDQ3870;  ANDX2029;  ANEP0871;  ANDS1311;  ANDQ3870;  ANEZ0925;  ANDS1311;  ANES2213;  ANEQ4966;  ANEZ3932;  ANEP2432;  ANEP3925;  ANEQ4753;  ANDR4198;  ANEQ0218;  ANEP2432;  ANEQ4966;  ANDP3984;  ANEQ4753;  ANDQ5253;  ANDR4198;  ANEP2432;  ANEP3925;  ANEQ1358;  ANEQ1358;  ANEQ2738;  ANEQ2738;  ANFP2354;  ANDX5372;  ANEP2432;  ANEQ4753;  ANEQ4966;  ANDQ5253;  ANEP3925;  ANEQ1358;  ANEQ4753;  ANFN2788;  ANFP2354;  ANDV0709;  ANEP3925;  ANDS4840;  ANDZ1324;  ANEQ0218;  ANEQ4753;  ANDZ3838;  ANEN4876;  ANDV0709;  ANEP2432;  ANDR4198;  ANDV0709;  ANDX5372;  ANEP2432;  ANEQ0218;  ANFN2788;  ANEP0871;  ANEQ1358;  ANDQ0319;  ANDQ5253;  ANDQ5253;  ANDR2269;  ANDS1311;  ANDX0711;  ANDX3059;  ANDY1218;  ANEN3392;  ANEP0871;  ANEQ2738;  ANEQ4966;  ANEQ4966;  ANES2213;  ANEZ3932;  ANDZ3838;  ANER0528;  ANDX3059;  ANFN1446;  ANFN1446;  ANDQ0318;  ANDQ0318;  ANDQ0319;  ANDQ3870;  ANDS4773;  ANDS4840;  ANDW3559;  ANDW3559;  ANDX2029;  ANDX2029;  ANDX3059;  ANDX3059;  ANDX3059;  ANDY4958;  ANEN3392;  ANEN3392;  ANEN4876;  ANEP3925;  ANEQ4966;  ANEQ4966;  ANES1044;  ANES1044;  ANES1044;  ANES2213;  ANEZ2101;  ANEZ3413;  ANEZ3413;  ANEZ3413;  ANEZ3932;  ANEZ3932;  ANFN2788;  ANFN3978;  ANFN3978;  ANEN4876;  ANEP0871;  ANEP2432;  ANEP3925;  ANEQ0218;  ANEZ0925;  ANEZ3932;  ANEZ3932;  ANFN3978;  ANDN0812;  ANDP3984;  ANDQ0319;  ANDQ0319;  ANDQ3870;  ANDQ3870;  ANDR4198;  ANDS4773;  ANDV0709;  ANDX2029;  ANEP2432;  ANEP3925;  ANEP3925;  ANEP3925;  ANEQ1358;  ANEQ2738;  ANEQ2738;  ANEQ4753;  ANEQ4753;  ANEQ4966;  ANEQ4966;  ANER0528;  ANES2213;  ANEZ3932;  ANFN1446;  ANDS4840;  ANDS4840;  ANDS4840;  ANEN4876;  ANES2213;  ANES2213;  ANDS1311;  ANEQ4753;  ANDQ5253;  ANEP0871;  ANDX2029;  ANDR4198;  ANDS1311;  ANDY4958;  ANEZ3413;  ANCQ1442;  ANDQ0318;  ANDP3951;  ANDP3984;  ANDX5372;  ANEP2432;  ANEP3925;  ANEQ1358;  ANEQ4753;  ANFN1446;  ANDQ0319;  ANDQ0319;  ANDV0709;  ANEP2432;  ANEQ0218;  ANFP0881;  ANDW3559;  ANEZ3413;  ANDQ5253;  ANDR4198;  ANDX3059;  ANEQ4753;  ANER0528;  ANEQ4753;  ANDS4840;  ANDY1218;  ANDX2029;  ANEQ0218;  ANEQ4753;  ANEP2432;  ANER0528;  ANDW3559;  ANDW3559;  ANCQ1442;  ANEQ4966;  ANDN0598;  ANDX2029;  ANES2213;  ANEQ4753;  ANEZ3413;  ANDY1218;  ANDQ0318;  ANDX2029;  ANDX5372;  ANDX5372;  ANDY1218;  ANDY4958;  ANDZ1324;  ANDZ2275;  ANDZ3838;  ANDS4840;  ANES2213;  ANDQ0318;  ANDQ5253;  ANEN3392;  ANES2213;  ANDS4840;  ANDV0709;  ANDV0709;  ANEQ4753;  ANES1044;  ANES2213;  ANES2213;  ANFP0881;  ANFP2354;  ANDR4198;  ANDX5372;  ANDW3559;  ANDZ1324;  ANDQ0319;  ANEQ0218;  ANER0528;  ANCT1579;  ANDS4840;  ANEZ2101;  ANDQ0318;  ANDQ0319;  ANCS3350;  ANEP3925;  ANEQ4753;  ANES2213;  ANES1044;  ANDS4840;  ANDP0510;  ANDS1311;  ANFN2788;  ANDX3059;  ANDQ0319;  ANCX3903;  ANDP0510;  ANDX5372;  ANEN3392;  ANEQ4966;  ANES1044;  ANES2213;  ANEQ4753;  ANDQ0318;  ANDQ0319;  ANDQ5253;  ANEZ2101;  ANEZ2101;  ANEZ2101;  ANEZ2101;  ANFN3978;  ANFP0881;  ANEQ4966;  ANDP3984;  ANDW3559;  ANDW3559;  ANDZ1324;  ANEP2432;  ANDQ3026;  ANDS4775;  ANDS4775;  ANDT1680;  ANDU3123;  ANDU3123;  ANDV3568;  ANDW1478;  ANDY3282;  ANDZ3821;  ANEN4871;  ANDQ1719;  ANDQ3026;  ANDQ5219;  ANDS2470;  ANDW1478;  ANDW3563;  ANDY3282;  ANDZ2282;  ANDZ3821;  ANEN1062;  ANER1509;  ANES4090;  ANEZ3414;  ANEZ3933;  ANFN2789;  ANFN3979;  ANDQ3873;  ANDQ3026;  ANER0521;  ANDX3061;  ANDZ3821;  ANEQ0215;  ANEQ4967;  ANDV3568;  ANDS3974;  ANDQ5219;  ANDY1237;  ANER0521;  ANEN4871;  ANDN0815;  ANDZ3821;  ANEQ1359;  ANFN3979;  ANEQ0215;  ANDZ3821;  ANER1509;  ANER1509;  ANEN4871;  ANFN3979;  ANDZ3821;  ANEQ4967;  ANFN3979;  ANDQ3873;  ANEQ0215;  ANER0521;  ANDY1237;  ANEN4871;  ANEQ4967;  ANEQ4967;  ANER1509;  ANES2214;  ANDP1358;  ANEQ0215;  ANDS3974;  ANDX1479;  ANDZ1325;  ANEZ3933;  ANEQ0215;  ANER1509;  ANDP0509;  ANDQ3873;  ANDZ3821;  ANFN3980;  ANEP2434;  ANCT3621;  ANDX3061;  ANDP1358;  ANEQ1359;  ANDZ3821;  ANFP2366;  ANDZ2282;  ANDQ3026;  ANDW1478;  ANER0521;  ANCX1884;  ANDQ3026;  ANDX1479;  ANES1045;  ANEZ3414;  ANDZ3821;  ANEQ4967;  ANER0521;  ANDY2449;  ANEN1062;  ANEN3390;  ANER1509;  ANEZ3933;  ANFN3979;  ANDP0509;  ANER1509;  ANDZ3821;  ANEZ3414;  ANCQ0143;  ANES1045;  ANDW1478;  ANEQ4967;  ANER1509;  ANEQ4967;  ANEZ3414;  ANEZ3414;  ANEN4871;  ANER0521;  ANDU3123;  ANDS4775;  ANFP2366;  ANDY1237;  ANEQ4967;  ANES1045;  ANDQ3026;  ANDQ3026;  ANER1509;  ANEQ1359;  ANCU3713;  ANCQ1441;  ANEQ0215;  ANER0521;  ANEQ4967;  ANEZ3933;  ANES1045;  ANDS2470;  ANEN4871;  ANEQ4967;  ANEN3390;  ANEP2434;  ANEQ0215;  ANEQ0215;  ANEQ1359;  ANEQ4967;  ANER0521;  ANEZ3933;  ANDW1478;  ANES1045;  ANES1045;  ANDY1237;  ANDV3568;  ANDZ3821;  ANDQ3873;  ANDT1680;  ANDZ2282;  ANEN4871;  ANEN3390;  ANEN4871;  ANDS3974;  ANDT1680;  ANDT1680;  ANDZ3821;  ANDZ3821;  ANEN4871;  ANEZ3414;  ANEN4871;  ANEQ1359;  ANDQ1719;  ANDQ1719;  ANDS3974;  ANDZ2282;  ANDZ3821;  ANEN1062;  ANEN4871;  ANER0521;  ANES1045;  ANES4090;  ANES4090;  ANEZ3933;  ANER0521;  ANEZ3414;  ANDN0815;  ANDP1358;  ANDQ1719;  ANDQ3026;  ANDQ3026;  ANDQ3873;  ANDQ5219;  ANDS2470;  ANDS3974;  ANDS4775;  ANDT1680;  ANDV4535;  ANDW1478;  ANDW3563;  ANDW3563;  ANDX1479;  ANDX3061;  ANDX3061;  ANDY1237;  ANDY1237;  ANDY2449;  ANDY2449;  ANDY3282;  ANDY3282;  ANDZ3821;  ANEN1062;  ANEN3390;  ANES1045;  ANES1045;  ANES1045;  ANEZ2102;  ANEZ3414;  ANEZ3414;  ANEZ3933;  ANEZ3933;  ANFN3979;  ANFN3979;  ANEN3390;  ANEN4871;  ANEN4871;  ANEP2434;  ANEQ0215;  ANEZ0926;  ANEZ0926;  ANEZ3414;  ANEZ3933;  ANFN3979;  ANCX1884;  ANDP1358;  ANDQ3026;  ANDS2470;  ANDS3974;  ANDV3568;  ANDW1478;  ANEN4871;  ANEN4871;  ANEP2434;  ANEQ1359;  ANER0521;  ANES1045;  ANES1045;  ANES1045;  ANES2214;  ANES2214;  ANEZ3414;  ANEZ3933;  ANDS3974;  ANDT1680;  ANEN4871;  ANER0521;  ANDS3974;  ANDT1680;  ANDU3123;  ANDZ3821;  ANES2214;  ANEQ1359;  ANEN4871;  ANEQ4967;  ANDQ3026;  ANDW1478;  ANEZ3933;  ANCX1884;  ANDZ2282;  ANEQ0215;  ANER1509;  ANBY3260;  ANDP0509;  ANDQ3026;  ANDV3568;  ANDQ3026;  ANEN4871;  ANDX3061;  ANEQ4967;  ANEQ4967;  ANEQ4967;  ANEQ4967;  ANER0521;  ANDP1358;  ANDS3974;  ANEN4871;  ANEQ0215;  ANFN3980;  ANDX3061;  ANEQ1359;  ANEQ4967;  ANEW3418;  ANFN3980;  ANER0521;  ANDP0509;  ANDP0509;  ANEQ0215;  ANEQ4967;  ANEN4871;  ANEN4871;  ANDS3974;  ANEQ0215;  ANDX1479;  ANCT2425;  ANCT3621;  ANDY1237;  ANDZ3821;  ANEN4871;  ANES4090;  ANCV3482;  ANDS4775;  ANDQ1719;  ANEZ3414;  ANDS2470;  ANDX1479;  ANDY2449;  ANDY3282;  ANDZ1325;  ANDZ3821;  ANDZ3821;  ANDQ3026;  ANES2214;  ANEQ0215;  ANDQ1719;  ANDT1680;  ANEN1062;  ANER1509;  ANES4090;  ANEZ3414;  ANFN3979;  ANES2214;  ANFN3980;  ANFN3980;  ANFP2366;  ANDQ3026;  ANDX1479;  ANEQ1359;  ANES2214;  ANES2214;  ANEZ3933;  ANDT1680;  ANEP2434;  ANDT1680;  ANEN1062;  ANEZ3414;  ANDQ3026;  ANDQ5219;  ANDW1478;  ANDX1479;  ANEZ3933;  ANEQ0215;  ANDP1358;  ANDQ1719;  ANDX1479;  ANDS3974;  ANEN4871;  ANDS3974;  ANDS3974;  ANDS3974;  ANES4090;  ANES1045;  ANEQ4967;  ANDZ3821;  ANDQ1719;  ANDY2449;  ANDQ3873;  ANDQ3026;  ANDQ3026;  ANDS3974;  ANEN1062;  ANDX3061;  ANDQ5219;  ANDS4775;  ANEQ0215;  ANES2214;  ANFP2366;  ANDP1358;  ANCX1884;  ANDZ2282;  ANEN1062;  ANEQ1359;  ANES1045;  ANEZ3414;  ANER0521;  ANEN1062;  ANEN3390;  ANEQ1359;  ANEQ4967;  ANEZ2102;  ANEZ2102;  ANEZ3933;  ANEZ3933;  ANFN3979;  ANFN3980;  ANER0521;  ANES1045;  ANEZ3933;  ANDS4775;  ANDW1478;  ANDX1479;  ANDZ3821;  ANEQ1359;  ANES2214;  ANEZ3933;  ANEZ3414;  ANDS3977;  ANDS3977;  ANDV4537;  ANDV4537;  ANDX2033;  ANDX2033;  ANDX2033;  ANDZ1329;  ANDZ1329;  ANDZ3837;  ANEN2136;  ANEN2136;  ANES1030;  ANFP0882;  ANDQ0323;  ANDQ0324;  ANDR3256;  ANDV4537;  ANDX0723;  ANDX2033;  ANDZ1329;  ANDZ3837;  ANEN3400;  ANER1536;  ANEY2033;  ANEZ3934;  ANDT1681;  ANFP2355;  ANER1536;  ANEY2033;  ANDQ0323;  ANDN0818;  ANEP2426;  ANCX1888;  ANDR3256;  ANDS0754;  ANDX0723;  ANES1030;  ANDX2033;  ANEQ4968;  ANEP2426;  ANEQ4968;  ANDZ3837;  ANDZ1329;  ANEP0870;  ANES1030;  ANEP0870;  ANEQ4968;  ANDX0723;  ANEZ3934;  ANEQ4968;  ANDX0723;  ANEQ2739;  ANEQ4968;  ANES1030;  ANET0557;  ANFN1447;  ANFN1447;  ANDW4347;  ANDX0723;  ANES1030;  ANES1030;  ANFP2355;  ANDQ0323;  ANDZ3837;  ANDX5382;  ANFP2355;  ANCQ0145;  ANCW1433;  ANDW4347;  ANEP0870;  ANEP0870;  ANEQ2739;  ANDS0754;  ANDR3256;  ANDT1681;  ANEQ4968;  ANEQ4968;  ANDV4537;  ANDX5382;  ANDZ3837;  ANDX2033;  ANEP0870;  ANEQ4968;  ANDQ0324;  ANDS3977;  ANDV4537;  ANEP0870;  ANDQ0324;  ANEQ4968;  ANES1030;  ANEP0870;  ANDS0754;  ANDQ0324;  ANEQ2739;  ANEZ3934;  ANDP3986;  ANDR3256;  ANEZ3415;  ANFN2790;  ANEQ4968;  ANDS3977;  ANEZ0927;  ANDQ0324;  ANEY2033;  ANDX5300;  ANEP2426;  ANER1536;  ANEZ3934;  ANEZ2103;  ANDR3256;  ANEN4879;  ANDV4537;  ANEZ3934;  ANFN3981;  ANEP0870;  ANCX1888;  ANDS0754;  ANDX2033;  ANEP0870;  ANEP2426;  ANEQ2739;  ANEQ4968;  ANEQ4968;  ANER1536;  ANES1030;  ANET0557;  ANET0557;  ANET0557;  ANEZ3934;  ANEZ3934;  ANFN1447;  ANDQ0323;  ANDQ0324;  ANDQ5220;  ANDR3256;  ANDS3977;  ANDS4778;  ANDS4778;  ANDT1681;  ANDW2000;  ANDW2000;  ANDW3565;  ANDX0723;  ANDX0723;  ANDX2033;  ANDX2033;  ANDX3064;  ANDX5300;  ANDX5300;  ANDX5382;  ANDX5382;  ANDZ1329;  ANDZ2284;  ANDZ2284;  ANDZ3837;  ANEN2136;  ANEN2136;  ANEN3400;  ANEN4879;  ANEP2426;  ANEQ1360;  ANER0529;  ANES1030;  ANES1030;  ANES1030;  ANEZ0927;  ANEZ3415;  ANEZ3415;  ANEZ3934;  ANEZ3934;  ANFN2790;  ANFN3981;  ANFN3981;  ANFP0882;  ANEN3400;  ANEN4879;  ANEP0870;  ANEP0870;  ANEQ1360;  ANEQ2739;  ANEZ3934;  ANEZ3934;  ANFN2790;  ANDQ0324;  ANDQ5220;  ANDN0597;  ANDN3149;  ANDS0754;  ANDS3977;  ANDT1681;  ANDV4537;  ANDX2033;  ANDX2033;  ANDX5300;  ANEN3400;  ANEP2426;  ANEQ1360;  ANEQ4968;  ANEQ4968;  ANER0529;  ANER1536;  ANES1030;  ANES1030;  ANEZ3934;  ANFP2355;  ANDS3977;  ANDT1681;  ANCX1888;  ANDS0754;  ANDR3256;  ANEZ3934;  ANDR3256;  ANDV4537;  ANDX0723;  ANDY4957;  ANDQ0324;  ANDX2033;  ANEQ4968;  ANDS3977;  ANFP2355;  ANDS0754;  ANDW4347;  ANDZ1329;  ANET0557;  ANET0557;  ANFP0882;  ANER0529;  ANDQ0324;  ANDZ1329;  ANEQ4968;  ANEQ4968;  ANDT1681;  ANEQ1360;  ANEQ2739;  ANDX2033;  ANDR3256;  ANDS3977;  ANDX0723;  ANEZ0927;  ANEZ2103;  ANEZ3934;  ANEP2426;  ANES1030;  ANES1030;  ANDQ0323;  ANDR3256;  ANDV4537;  ANDZ2284;  ANDR3256;  ANDT1681;  ANDV0706;  ANDW3565;  ANDW4347;  ANDX3064;  ANDX5300;  ANDX5300;  ANDX5382;  ANDY1212;  ANDY4957;  ANDY4957;  ANDZ1329;  ANEN3400;  ANER1536;  ANDT1681;  ANDY1212;  ANEP2426;  ANEQ1360;  ANEZ0927;  ANEY2033;  ANFP2355;  ANFP2355;  ANFP2355;  ANDS0754;  ANDT1681;  ANDW4347;  ANDZ2284;  ANEN2136;  ANEN4879;  ANEP0870;  ANEQ2739;  ANER0529;  ANER0529;  ANER1536;  ANEZ3934;  ANFN3981;  ANDX2033;  ANDX2033;  ANEP0870;  ANES1030;  ANDZ1329;  ANEQ4968;  ANDQ0323;  ANDS0754;  ANEQ4968;  ANER1536;  ANDZ2284;  ANEY2033;  ANDZ1329;  ANDS4778;  ANDS3977;  ANFN1447;  ANDX0723;  ANCQ0145;  ANCT2432;  ANDR3256;  ANDQ0323;  ANDQ0324;  ANDQ5220;  ANDS0754;  ANDS3977;  ANEN2136;  ANDX5300;  ANFN1447;  ANDT1681;  ANDZ2284;  ANEQ4968;  ANEP0870;  ANEP2426;  ANDQ0323;  ANEN2136;  ANEQ2739;  ANEN3400;  ANEN4879;  ANEP0870;  ANEY2033;  ANEY2033;  ANEZ2103;  ANEZ2103;  ANEZ2103;  ANEZ3934;  ANFP0882;  ANFP2355;  ANEZ3934;  ANDW4347;  ANEN2136;  ANES1030;  ANCU3712;  ANDP1355;  ANDP1355;  ANDQ3869;  ANDQ3869;  ANDS2482;  ANDS2482;  ANDW4362;  ANDX0716;  ANDX0716;  ANDX3058;  ANDX5314;  ANDY1213;  ANDY1213;  ANDY1213;  ANDZ2281;  ANDZ3816;  ANEP2415;  ANEQ4071;  ANER1534;  ANER1534;  ANER1534;  ANDP2689;  ANDP2689;  ANDQ3869;  ANDS2482;  ANDS2482;  ANDT4128;  ANDX3058;  ANDX5314;  ANDY1213;  ANDZ2281;  ANEN2149;  ANEN2149;  ANEN4884;  ANER1534;  ANEZ0928;  ANEZ3416;  ANFN2791;  ANFN3982;  ANER1534;  ANEQ2740;  ANER1534;  ANDX3058;  ANDY1213;  ANDY1213;  ANDZ1323;  ANDZ2281;  ANDZ3816;  ANEP2415;  ANDY1213;  ANDZ3816;  ANEQ1361;  ANER1534;  ANDX3058;  ANDP2689;  ANDZ3816;  ANES2215;  ANDX5314;  ANEP2415;  ANEQ1361;  ANEQ1361;  ANEQ2740;  ANEQ2740;  ANEQ2740;  ANEQ4071;  ANDX5314;  ANDX3058;  ANDX5314;  ANDZ1323;  ANEQ4071;  ANDP2689;  ANEP2415;  ANEQ2740;  ANEQ4071;  ANDX5314;  ANDS2482;  ANDT4128;  ANDP2689;  ANCW2271;  ANEQ2740;  ANDU1113;  ANER1534;  ANDU1113;  ANDY1213;  ANEN2149;  ANEP2415;  ANEP2415;  ANEQ2740;  ANDP1355;  ANDZ1323;  ANDZ3816;  ANEQ1361;  ANEZ3416;  ANEQ4071;  ANES2215;  ANFN3982;  ANDP2689;  ANDX3058;  ANDX5314;  ANEQ2740;  ANEQ2740;  ANEQ4071;  ANER1534;  ANDS2482;  ANDT4128;  ANEQ2740;  ANDP2689;  ANDP2689;  ANDT4128;  ANDP2689;  ANDX3058;  ANDX5314;  ANEQ2740;  ANDU1113;  ANDU1113;  ANDZ3816;  ANEP2415;  ANEP2415;  ANEZ3416;  ANEQ4071;  ANDP1355;  ANDX5314;  ANDS2482;  ANDP1355;  ANDX5314;  ANDX3058;  ANEN4884;  ANER1534;  ANDT4128;  ANDT4128;  ANDU1113;  ANDW4362;  ANDX0716;  ANDX3058;  ANDX5314;  ANDX5314;  ANDY1213;  ANDY1213;  ANDY2460;  ANDZ1323;  ANEN2149;  ANDP1355;  ANDT4128;  ANDX0716;  ANDX5314;  ANEN2149;  ANEP2415;  ANEQ2740;  ANDP2689;  ANDT4128;  ANDT4128;  ANDX5314;  ANDZ3816;  ANEN2149;  ANEN4884;  ANEP2415;  ANER1534;  ANES2215;  ANFN3982;  ANES2215;  ANDQ3869;  ANDS2482;  ANDT4128;  ANDU1113;  ANDW4362;  ANEQ4071;  ANES2215;  ANES2215;  ANES2215;  ANES2215;  ANES2215;  ANES2215;  ANES2215;  ANES2215;  ANEZ0928;  ANFN3982;  ANCT1149;  ANCV4708;  ANDZ1323;  ANDZ2281;  ANDP1355;  ANDP2689;  ANDS2482;  ANDW4362;  ANEP2415;  ANEP2415;  ANFN3982;  ANFP2379;  ANDY1213;  ANDQ3869;  ANDZ3816;  ANDS2482;  ANDX5314;  ANCU3712;  ANCV4708;  ANDS2482;  ANES2215;  ANEZ3416;  ANEQ2740;  ANES2215;  ANDY2460;  ANDQ3869;  ANDT4128;  ANEN2149;  ANEZ3416;  ANDP2689;  ANEP2415;  ANFP2379;  ANDY1213;  ANDP1355;  ANEQ4071;  ANDY2460;  ANEN2149;  ANEQ1361;  ANEZ0928;  ANEZ0928;  ANEZ3416;  ANFN3982;  ANEZ3416;  ANDW4362;  ANDU1113;  ANER1534;  ANES2215;  ANDQ3032;  ANDQ3875;  ANDQ3875;  ANDS2452;  ANDS2452;  ANDS2452;  ANDS2452;  ANDT4129;  ANDT4129;  ANDX5322;  ANDX5322;  ANDX5322;  ANDZ1354;  ANEQ2741;  ANEQ4072;  ANEZ3428;  ANDQ3032;  ANDZ3822;  ANER0522;  ANEP0872;  ANDZ3822;  ANEN2155;  ANDW4379;  ANDQ3875;  ANEQ2741;  ANDQ3032;  ANDZ3822;  ANEN2155;  ANDS2452;  ANDT4129;  ANDX5322;  ANDX5322;  ANDQ3032;  ANDZ3822;  ANEN2155;  ANEZ3428;  ANEZ3428;  ANEZ3428;  ANEQ2741;  ANEQ2741;  ANDQ3032;  ANDW4379;  ANDX5322;  ANEZ3428;  ANDQ1714;  ANDX5322;  ANDX5383;  ANEZ3428;  ANEP0872;  ANES2216;  ANEZ3428;  ANDQ3032;  ANDQ3032;  ANDX5322;  ANDX5322;  ANDX5322;  ANDZ1354;  ANDZ1354;  ANDQ3032;  ANDQ3032;  ANEP0872;  ANEN2155;  ANEN2155;  ANEQ2741;  ANDX5322;  ANDQ3032;  ANDS2452;  ANEN2155;  ANDQ1714;  ANDQ3032;  ANDQ3875;  ANDS2452;  ANDS3975;  ANDS4777;  ANDT4129;  ANDW4379;  ANDX5383;  ANDX5322;  ANDX5383;  ANEQ4072;  ANET0558;  ANCQ3791;  ANDQ3032;  ANDS4777;  ANDT4129;  ANDT4129;  ANDW4379;  ANEP0872;  ANER0522;  ANER0522;  ANER0522;  ANER0522;  ANES2216;  ANES2216;  ANES2216;  ANES2216;  ANET0558;  ANET0558;  ANEZ3428;  ANDS2452;  ANDX5322;  ANDX5322;  ANEN2155;  ANDW4379;  ANCQ3791;  ANDQ3032;  ANDS2452;  ANDX5322;  ANDX5383;  ANDX5383;  ANDS4777;  ANES2216;  ANDQ3032;  ANDZ3822;  ANDQ3032;  ANDQ3032;  ANDX5322;  ANEQ2741;  ANEZ3428;  ANFP2332;  ANER0522;  ANDQ0325;  ANDQ0325;  ANDQ0325;  ANDQ1720;  ANDQ1720;  ANDU1900;  ANDU1900;  ANDU1900;  ANDU1900;  ANDU3137;  ANDU3137;  ANDU3137;  ANDW4396;  ANDY2461;  ANDY3277;  ANDY3277;  ANEP0853;  ANEP0853;  ANEP5311;  ANEP5311;  ANEQ1362;  ANEQ2742;  ANEQ2742;  ANER2794;  ANEZ2104;  ANEZ3423;  ANEZ3423;  ANCR1294;  ANDQ1720;  ANDU1900;  ANDY3277;  ANEP0853;  ANEQ1362;  ANDQ0325;  ANDQ1720;  ANDQ1720;  ANDY4969;  ANDY4969;  ANDQ1720;  ANDY4969;  ANEQ1362;  ANEQ1362;  ANEQ1362;  ANEP0853;  ANEP5311;  ANEZ2104;  ANEZ3423;  ANEZ3423;  ANEZ3423;  ANDU3137;  ANDQ0325;  ANDQ1720;  ANEZ3423;  ANDU1900;  ANEQ2742;  ANEQ2742;  ANER0523;  ANER0523;  ANER2794;  ANEZ2104;  ANDQ1720;  ANDZ1321;  ANER0523;  ANDY4969;  ANER0523;  ANET3796;  ANET3796;  ANEZ3423;  ANEZ2104;  ANEZ2104;  ANDQ0325;  ANDY3277;  ANDQ1720;  ANDU1900;  ANDU1900;  ANER0523;  ANER2794;  ANDQ0325;  ANDQ1720;  ANDQ0325;  ANDQ1720;  ANEP5311;  ANEQ1362;  ANDQ1720;  ANES4091;  ANDU1900;  ANDQ0325;  ANDQ1720;  ANDU1107;  ANDU1900;  ANDU1900;  ANDW4396;  ANDW4396;  ANDY2461;  ANDY2461;  ANDY3277;  ANDY4969;  ANDY4969;  ANDZ1321;  ANEP0853;  ANDU1107;  ANDU3137;  ANDW4396;  ANDY2461;  ANER2794;  ANER2794;  ANES4091;  ANET3796;  ANEZ3423;  ANDQ1720;  ANDY3277;  ANER0523;  ANES4091;  ANES4091;  ANES4091;  ANCR1294;  ANCR1294;  ANDU1900;  ANEQ2742;  ANER2794;  ANET3796;  ANDQ0325;  ANDQ1720;  ANDY3277;  ANDY3277;  ANEP5311;  ANEP0853;  ANEP5311;  ANDU3137;  ANDY4969;  ANEQ1362;  ANDQ3036;  ANDQ3036;  ANDS0902;  ANDS4184;  ANDT2943;  ANDT2943;  ANDV0250;  ANDV1245;  ANDV2515;  ANDY4962;  ANEP3937;  ANEQ4949;  ANER0505;  ANER0505;  ANES1035;  ANET1856;  ANET1857;  ANEX6012;  ANEZ3424;  ANFN2785;  ANFP0862;  ANDQ3036;  ANDQ3036;  ANDY4962;  ANDS4184;  ANDY4962;  ANEP3937;  ANDP1359;  ANDP2700;  ANDT2943;  ANDT2943;  ANDV0250;  ANEP3937;  ANEP3937;  ANEQ4949;  ANDT2943;  ANDS4184;  ANEZ2083;  ANFN1435;  ANDP1359;  ANDS4184;  ANFN1435;  ANDU1106;  ANFP2362;  ANDV0250;  ANES1035;  ANDQ3036;  ANFN2785;  ANEQ4044;  ANDQ3036;  ANEZ0909;  ANFN2785;  ANFP2362;  ANEX6012;  ANDS0902;  ANES4066;  ANEQ4044;  ANEW3368;  ANDT2943;  ANEZ2083;  ANDQ3036;  ANDV0250;  ANDQ3036;  ANDQ3036;  ANDT2943;  ANEQ4044;  ANES4066;  ANFP0862;  ANDV0250;  ANDY4962;  ANDT2943;  ANDU1106;  ANDV2515;  ANEQ4044;  ANET1857;  ANEX6012;  ANEZ0909;  ANET1856;  ANEW3368;  ANEX6012;  ANEZ0909;  ANEZ2083;  ANEZ2083;  ANEZ3424;  ANDW2001;  ANDW4407;  ANDW4407;  ANDX1404;  ANDQ3036;  ANDQ3036;  ANDT2943;  ANDT2943;  ANDX1404;  ANDY4962;  ANEQ4949;  ANES1035;  ANET1856;  ANEW3368;  ANEW3368;  ANEW4221;  ANFN1435;  ANFN1435;  ANFN2785;  ANET1857;  ANES4066;  ANDT2943;  ANDS0902;  ANDS4184;  ANDU1106;  ANEQ4949;  ANEZ0909;  ANDQ3036;  ANDS4184;  ANDT2943;  ANDY4962;  ANEQ4044;  ANFN1435;  ANET1856;  ANET1856;  ANEZ2083;  ANFP2362;  ANDW2001;  ANDW2001;  ANDY4962;  ANES4066;  ANET1856;  ANEZ0909;  ANFN1435;  ANES4066;  ANES4066;  ANCU1842;  ANDS0902;  ANDS0902;  ANDT2943;  ANEQ4044;  ANDQ3036;  ANDT2943;  ANFN1435;  ANDY4962;  ANEQ4949;  ANDV0250;  ANDX1404;  ANET1856;  ANFN1435;  ANDX1404;  ANCY3304;  ANCY3304;  ANDT2943;  ANDT2943;  ANDU1106;  ANDU1106;  ANDV0250;  ANEQ4044;  ANDS0902;  ANDV0250;  ANEZ2083;  ANEP3937;  ANDQ3036;  ANDS0902;  AN",
      "center_classification_date": "20210617",
      "distribution_pattern": "Worldwide distribution. US Nationwide, ALGERIA,   ANDORRA,  ARGENTINA,  AUSTRALIA,  AUSTRIA,  AZERBAIJAN,  BAHRAIN,  BELGIUM,  BRAZIL,  BRUNEI,  BULGARIA,  CHILE,  COSTA RICA,  CYPRUS,  CZECH REPUBLIC,  DENMARK,  DEUTSCHLAND,  EGYPT,  FINLAND,  FRANCE,  GEORGIA,  GREECE,  HONG KONG,  HUNGARY,  INDIA,  IRAN,  IRAQ,  IRELAND,  ISRAEL,  ITALY,  JORDAN,  KAZAKHSTAN,  MALAYSIA,  MEXICO,  NETHERLANDS,  NORWAY,  OMAN,  PANAMA,  POLAND,  PORTUGAL,  ROMANIA,  RUSSIA,  SAUDIA ARABIA,  SINGAPORE,  SLOVAKIA (SLOVAK REPUBLIC),  SLOVENIA,  SOUTH AFRICA,  SOUTH KOREA,  SPAIN,  SWEDEN,  SWITZERLAND,  TAIWAN,  TURKEY,  UNITED ARAB EMIRATES (UAE),  UNITED KINGDOM,  VIETNAM",
      "state": "AZ",
      "product_description": "VENOVO Venous Stent System 9F    Product Description(REF)/dimension:  VENEL14040\t/Venovo 9F 14/40/1200mm OUS;  VENEL14060\t/Venovo 9F 14/60/1200mm OUS;  VENEL14080\t/Venovo 9F 14/80/1200mm OUS;  VENEL14100\t/Venovo 9F 14/100/1200mm OUS;  VENEL14120\t/Venovo 9F 14/120/1200mm OUS;  VENEL14140\t/Venovo 9F 14/140/1200mm OUS;  VENEL14160\t/Venovo 9F 14/160/1200mm OUS;  VENEM14040\t/Venovo 9F 14/40/800mm OUS;  VENEM14060\t/Venovo 9F 14/60/800mm OUS;  VENEM14080\t/Venovo 9F 14/80/800mm OUS;  VENEM14100\t/Venovo 9F 14/100/800mm OUS;  VENEM14120\t/Venovo 9F 14/120/800mm OUS;  VENEM14140\t/Venovo 9F14/140/800mm OUS;  VENEM14160\t/Venovo 9F 14/160/800mm OUS;  VENUL14040\t/Venovo 9F 14/40/1200mm US;  VENUL14060\t/Venovo 9F 14/60/1200mm US;  VENUL14080\t/Venovo 9F 14/80/1200mm US;  VENUL14100\t/Venovo 9F 14/100/1200mm US;  VENUL14120\t/Venovo 9F 14/120/1200mm US;  VENUL14140\t/Venovo 9F 14/140/1200mm US;  VENUL14160\t/Venovo 9F 14/160/1200mm US;  VENUM14040\t/Venovo 9F 14/40/800mm US;  VENUM14060\t/Venovo 9F 14/60/800mm US;  VENUM14080\t/Venovo 9F 14/80/800mm US;  VENUM14100\t/Venovo 9F 14/100/800mm US;  VENUM14120\t/Venovo 9F 14/120/800mm US;  VENUM14140\t/Venovo 9F 14/140/800mm US;  VENUM14160\t/Venovo 9F 14/160/800mm US;",
      "report_date": "20210623",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "recall_number": "Z-1899-2021",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "87208",
      "more_code_info": "EQ4044;  ANBX1573;  ANDY4962;  ANFN2785;  ANDV0250;  ANES4066;  ANEW3368;  ANDU1106;  ANES1035;  ANDT2943;  ANDU1106;  ANDT2943;  ANCP0212;  ANDS4184;  ANEX6012;  ANEZ0909;  ANDS4184;  ANDW2001;  ANEX6012;  ANDT2943;  ANEX6012;  ANEZ2083;  ANDU1106;  ANDP1363;  ANDP2695;  ANDQ3888;  ANDS2479;  ANDV0690;  ANDV1256;  ANDX2039;  ANDX5305;  ANEN2152;  ANEP0854;  ANEW3369;  ANEX3082;  ANEY1224;  ANEZ2084;  ANDP1363;  ANDQ3888;  ANDS0757;  ANEP0854;  ANEP3569;  ANEP3569;  ANEP3931;  ANBZ1274;  ANCP1853;  ANDT2945;  ANCQ2601;  ANEP3569;  ANEX1702;  ANEX6023;  ANEP3569;  ANDS1269;  ANEP0854;  ANDS1269;  ANEW2002;  ANDT2945;  ANEY1224;  ANDS1269;  ANCX3880;  ANDV0251;  ANFP0863;  ANFP0863;  ANCP1853;  ANDN0832;  ANDX0759;  ANEP0854;  ANEP3569;  ANEZ0910;  ANDQ1724;  ANFN3955;  ANEX1702;  ANEP0854;  ANDQ1724;  ANDU2246;  ANFN3955;  ANDX2039;  ANEZ0910;  ANDQ1724;  ANEP0854;  ANEP3569;  ANEP3931;  ANCW1425;  ANEX4716;  ANES4067;  ANDP2695;  ANEP3569;  ANDS1269;  ANDT2945;  ANDX0759;  ANDP1363;  ANEY1224;  ANEW4222;  ANEY1224;  ANCW1425;  ANDS1269;  ANEP0854;  ANEP3569;  ANEP3569;  ANFN3955;  ANDX2039;  ANDX0759;  ANDX2039;  ANDX2039;  ANDV0690;  ANDX2039;  ANEZ3400;  ANEZ2084;  ANEP0854;  ANEP3569;  ANDP2695;  ANDX2039;  ANDX2039;  ANDU2246;  ANDT2945;  ANDS2479;  ANEX6023;  ANDX2039;  ANEP0854;  ANEP3569;  ANEY1224;  ANEY1224;  ANEZ2084;  ANDY1220;  ANDT2945;  ANDU2246;  ANEP0854;  ANDQ3888;  ANDQ3888;  ANDS1269;  ANDS2479;  ANDS4824;  ANDT2945;  ANDT2945;  ANDV1256;  ANDX0759;  ANDX2039;  ANDY1220;  ANEP0854;  ANET0544;  ANEW2002;  ANES4067;  ANET0544;  ANEU4252;  ANEW2002;  ANEW3369;  ANEX3082;  ANEX3082;  ANEX4716;  ANEX6023;  ANEZ0910;  ANEZ3400;  ANEZ3400;  ANEZ3400;  ANFN3955;  ANFP0863;  ANDQ1724;  ANDQ3052;  ANDS1269;  ANDX2039;  ANDX3072;  ANDX5305;  ANDN0832;  ANDP2695;  ANDQ1724;  ANDQ3052;  ANDQ3888;  ANDS1269;  ANDS4824;  ANDU2246;  ANDV0251;  ANDV0690;  ANDX0759;  ANDX2039;  ANDX2039;  ANDX2039;  ANDX5305;  ANDX5305;  ANEN1073;  ANEN2152;  ANEP0854;  ANEP3569;  ANEP3569;  ANEP3569;  ANEP3569;  ANEP3569;  ANEP3569;  ANEP3569;  ANEW4222;  ANEX1702;  ANEX6023;  ANEZ0910;  ANEZ2084;  ANET0544;  ANEP0854;  ANEP3931;  ANES4067;  ANEW4222;  ANEP3569;  ANEY1224;  ANEZ2084;  ANDS4824;  ANDV1256;  ANDV0251;  ANDX2039;  ANDX2039;  ANDY1220;  ANDU2246;  ANDX0759;  ANEP0854;  ANES4067;  ANDT2945;  ANET0544;  ANEY1224;  ANDS1269;  ANDS1269;  ANDQ1724;  ANEZ2084;  ANEX4716;  ANEZ3400;  ANDV1256;  ANDX0759;  ANEP0854;  ANES4067;  ANEX6023;  ANDT2945;  ANEN2152;  ANEP3931;  ANDS0757;  ANES2192;  ANDY1220;  ANEW4222;  ANCX3880;  ANDQ1724;  ANDS1269;  ANDY1220;  ANDY1220;  ANDX0759;  ANEP3569;  ANEW3369;  ANDS2479;  ANDV0251;  ANEN2152;  ANEP3569;  ANEZ0910;  ANDS2479;  ANDU2246;  ANDV0251;  ANEP3569;  ANFN3955;  ANDV0251;  ANDX0759;  ANDT2945;  ANDU2246;  ANEN1073;  ANEP3931;  ANEZ2084;  ANDX0759;  ANCX3880;  ANDP2695;  ANDQ3052;  ANDS1269;  ANDT2945;  ANDX0759;  ANEP3569;  ANEZ2084;  ANDU2246;  ANDV0690;  ANEP3569;  ANEY1224;  ANEZ3400;  ANDT2945;  ANDX0759;  ANEP0854;  ANES2192;  ANEW4222;  ANES4067;  ANDX0759;  ANDX2039;  ANDT2945;  ANET0544;  ANEP3569;  ANDQ3052;  ANDS1269;  ANDS2479;  ANDU2246;  ANDX0759;  ANDX0759;  ANDX2039;  ANEP3569;  ANES2192;  ANEU4252;  ANEX3082;  ANEZ3400;  ANDQ1724;  ANEP0854;  ANCS0649;  ANDX2039;  ANCQ2601;  ANCX3880;  ANDU2246;  ANDQ3052;  ANDQ3052;  ANDS1269;  ANDS2479;  ANEY1224;  ANFP0863;  ANCX3880;  ANDS1269;  ANDT2945;  ANDV0251;  ANEY1224;  ANFN3955;  ANCW1425;  ANDV0690;  ANDS1269;  ANDS2479;  ANDU2246;  ANES4067;  ANET0544;  ANDY1220;  ANDS0757;  ANDV0251;  ANCW4218;  ANDV0690;  ANEP3569;  ANFP0863;  ANFP0863;  ANCX3880;  ANDV0251;  ANDX0759;  ANDX2039;  ANEY1224;  ANEY1224;  ANDS2479;  ANDX2039;  ANEP3931;  ANES2192;  ANDV0251;  ANDX0759;  ANET0544;  ANEY1224;  ANDY1220;  ANFN3955;  ANEW4222;  ANDP1362;  ANDP1362;  ANDP1362;  ANDP1362;  ANDP2694;  ANDP2694;  ANDX5311;  ANDY2462;  ANDY4973;  ANDZ1352;  ANDZ3839;  ANEN2144;  ANEP5326;  ANCP2409;  ANCP3371;  ANDX5311;  ANCR0695;  ANCS1159;  ANDR0757;  ANDP1362;  ANEP5326;  ANDW4341;  ANEN2144;  ANER1540;  ANEZ3957;  ANET3788;  ANEW2046;  ANDS0756;  ANDS1283;  ANFN3956;  ANFN3956;  ANDX5311;  ANDP2694;  ANDR0757;  ANDX3071;  ANEX1700;  ANEX1700;  ANDS0756;  ANET0545;  ANCW3265;  ANCW4219;  ANDX1380;  ANEN2144;  ANEP5326;  ANCP2409;  ANDX5311;  ANES2193;  ANEY1225;  ANEZ3401;  ANEZ3957;  ANCY2189;  ANDX2038;  ANDX3071;  ANEN2144;  ANET3788;  ANEY1225;  ANEZ3957;  ANEZ3957;  ANFN3956;  ANDS1283;  ANEX3083;  ANDS1283;  ANDR0757;  ANER1540;  ANDS1283;  ANDX5311;  ANEW2046;  ANEZ3957;  ANDR0757;  ANFP2315;  ANDX3071;  ANDX5311;  ANDX1380;  ANDX3071;  ANEQ4045;  ANEQ4950;  ANER1540;  ANEZ3957;  ANDX5311;  ANDP2694;  ANDR0757;  ANDS1283;  ANDS4825;  ANDX1380;  ANEP5326;  ANET3788;  ANEZ3401;  ANEZ3957;  ANDR0757;  ANDX5311;  ANEZ2085;  ANDW4341;  ANDP2694;  ANCW3265;  ANCX3893;  ANEZ3401;  ANER1540;  ANDX5311;  ANCW3265;  ANDR0757;  ANEN2144;  ANDX5311;  ANDR0757;  ANEZ2085;  ANEZ3957;  ANDY2462;  ANDS4825;  ANDP2694;  ANDR0757;  ANDR2276;  ANDR2276;  ANDS0756;  ANDS1283;  ANDS1283;  ANDX1380;  ANDX5311;  ANEN2144;  ANEN3396;  ANET0545;  ANET3788;  ANET3788;  ANES2193;  ANES4068;  ANET0545;  ANET3788;  ANEX1700;  ANEX3083;  ANEX3083;  ANEX6022;  ANEY4328;  ANEZ0911;  ANEZ2085;  ANEZ3401;  ANEZ3401;  ANEZ3957;  ANEZ3957;  ANFN2848;  ANDP2694;  ANDQ3887;  ANDR2276;  ANDR2276;  ANDY1222;  ANCY3305;  ANDR0757;  ANDR0757;  ANDR2276;  ANDR2276;  ANDS4825;  ANDW4341;  ANDW4341;  ANDX1380;  ANDX1412;  ANDX2038;  ANDX3071;  ANDX5311;  ANDX5311;  ANDX5311;  ANDY1222;  ANDY4973;  ANDY4973;  ANDZ1352;  ANEN3396;  ANEP3935;  ANEP5326;  ANEP5326;  ANEP5326;  ANEQ4045;  ANEQ4045;  ANEQ4045;  ANEQ4045;  ANEQ4950;  ANEQ4950;  ANET3788;  ANEX3083;  ANEX3083;  ANEX6022;  ANEY1225;  ANEY1225;  ANEZ0911;  ANEZ0911;  ANEZ3957;  ANFN2848;  ANDS4825;  ANES4068;  ANDP1362;  ANER1540;  ANEW3416;  ANDX5311;  ANDX5311;  ANEZ3957;  ANEN2144;  ANEP5326;  ANDR2276;  ANDS4825;  ANDS4825;  ANDX1412;  ANDX3071;  ANDY4973;  ANDZ3839;  ANEP3935;  ANEP5326;  ANEP5326;  ANES2193;  ANES4068;  ANET3788;  ANET3788;  ANET3788;  ANEX6022;  ANEX6022;  ANEX6022;  ANEZ3957;  ANDS0756;  ANDX2038;  ANEY1225;  ANDX2038;  ANDX5311;  ANEP5326;  ANDS0756;  ANDX5311;  ANDY2462;  ANEN3396;  ANEY1225;  ANEY1225;  ANDS1283;  ANDR0757;  ANDS4825;  ANDX3071;  ANEN2144;  ANEP5326;  ANEQ4950;  ANER1540;  ANEW2046;  ANFN2848;  ANDS0756;  ANDS1283;  ANDS1283;  ANDX1380;  ANEZ2085;  ANDR2276;  ANER1540;  ANER1540;  ANER1540;  ANEW2046;  ANER1540;  ANDS0756;  ANDX3071;  ANDX5311;  ANEN2144;  ANEW3416;  ANEQ4950;  ANEZ2085;  ANDY4973;  ANEZ0911;  ANEZ3957;  ANCV2915;  ANCX3893;  ANCY2189;  ANDX5311;  ANDX5311;  ANDS4825;  ANDX1380;  ANDX5311;  ANDY1222;  ANEN2144;  ANEP5326;  ANCW4219;  ANDR2276;  ANDX3071;  ANEZ0911;  ANEZ2085;  ANEZ3957;  ANDS1283;  ANEN3396;  ANEZ3957;  ANEN2144;  ANEN3396;  ANER1540;  ANCW4219;  ANEX3083;  ANDX3071;  ANDP2694;  ANDS1283;  ANDS4825;  ANDW4341;  ANDX2038;  ANEN2144;  ANEZ3957;  ANDX2038;  ANEN3396;  ANDP2694;  ANEX3083;  ANDR0757;  ANDR0757;  ANDS4825;  ANER1540;  ANEY1225;  ANDX1412;  ANEN3396;  ANDR2276;  ANDP1362;  ANDX5311;  ANEP5326;  ANEQ4045;  ANEX3083;  ANDX2038;  ANDS0756;  ANEQ4950;  ANEY1225;  ANEZ3957;  ANDZ3839;  ANEY1225;  ANFP1519;  ANER1540;  ANEQ4950;  ANES4068;  ANEN2144;  ANDS0756;  ANDS1283;  ANEW2046;  ANDS4825;  ANEN3396;  ANEQ4045;  ANEZ3401;  ANDP1362;  ANDS0756;  ANDS1283;  ANDP1362;  ANDS4825;  ANDY4973;  ANER1540;  ANES2193;  ANEY1225;  ANFN3956;  ANDR2276;  ANEW3416;  ANEZ3401;  ANEP5326;  ANEQ4950;  ANEY1225;  ANEX3083;  ANDS1283;  ANEP5326;  ANEP5326;  ANDS4825;  ANDY2462;  ANEN3396;  ANEQ4950;  ANEY1225;  ANEZ3957;  ANEQ4950;  ANEZ3957;  ANDR0755;  ANDR3258;  ANDS1296;  ANDV3569;  ANDW1480;  ANDX5302;  ANEN1074;  ANEN4899;  ANET0546;  ANET3789;  ANEX3094;  ANEZ0912;  ANEZ3402;  ANEZ3956;  ANDP3962;  ANDP3962;  ANDP3962;  ANDP3962;  ANDP3988;  ANDP3988;  ANDV0711;  ANDX5302;  ANDY3309;  ANDZ1342;  ANDZ2276;  ANEP3927;  ANEP5325;  ANCS1564;  ANEZ2086;  ANDW1480;  ANDX1408;  ANDS1296;  ANDZ2276;  ANEN4899;  ANEX0747;  ANEZ3956;  ANCV1585;  ANDV0711;  ANDX1408;  ANDX1408;  ANER1537;  ANEW2003;  ANDR0755;  ANER1537;  ANDX1408;  ANEZ3956;  ANDR3258;  ANDT1683;  ANDW1480;  ANEP5325;  ANEZ2086;  ANCV1585;  ANDR2273;  ANDP3988;  ANEP5325;  ANCS3443;  ANCX3894;  ANDX1408;  ANDV0711;  ANDZ2276;  ANCX3894;  ANDR3258;  ANDV1209;  ANEP5325;  ANDP1361;  ANDP3988;  ANDR0755;  ANDW1480;  ANDX5302;  ANEP0866;  ANEP0866;  ANEP3927;  ANDV1209;  ANDX1408;  ANDV0711;  ANDW1480;  ANEP0866;  ANEW3370;  ANEZ0912;  ANDX1408;  ANDV0711;  ANCQ4288;  ANDX1408;  ANDX1408;  ANCX0892;  ANEP5325;  ANEQ4951;  ANDP3962;  ANDV3569;  ANDX2036;  ANDV0711;  ANCS3443;  ANDR3258;  ANCS3443;  ANDV1209;  ANEP5325;  ANDR3258;  ANDT1683;  ANDT1683;  ANDV1209;  ANDW1480;  ANDX2036;  ANDX2036;  ANDZ3829;  ANEN1074;  ANER1537;  ANES2194;  ANES4069;  ANEW2003;  ANEW3370;  ANEX0747;  ANEX6019;  ANEZ0912;  ANEZ2086;  ANEZ2086;  ANEZ3402;  ANEZ3956;  ANDP3962;  ANDQ3883;  ANDR0755;  ANDR0755;  ANDX1408;  ANDX2036;  ANDP3988;  ANDR0755;  ANDR0755;  ANDS4826;  ANDT1683;  ANDV0711;  ANDV1209;  ANDV1209;  ANDV3569;  ANDX1408;  ANDX1408;  ANDX2036;  ANDX2036;  ANDX5302;  ANDX5302;  ANDZ1342;  ANEP0866;  ANEP3927;  ANEP5325;  ANEP5325;  ANEP5325;  ANEQ4046;  ANEQ4951;  ANEX0747;  ANEX0747;  ANEX0747;  ANEX0747;  ANEX1709;  ANEX3094;  ANEZ0912;  ANDT1683;  ANEW3370;  ANEZ3956;  ANEW3370;  ANDS4826;  ANDW1480;  ANDW1480;  ANDX5302;  ANEP0866;  ANEQ4046;  ANEQ4951;  ANEW3370;  ANDS4826;  ANEQ4951;  ANES2194;  ANEX6019;  ANDQ3883;  ANDR2273  ANDS1296;  ANDT1683;  ANDT1683;  ANDV0711;  ANDX5302;  ANDZ2276;  ANDZ3829;  ANEN1074;  ANEP3927;  ANEP5325;  ANEQ4951;  ANEQ4951;  ANER1537;  ANER1537;  ANES2194;  ANEV1712;  ANEX0747;  ANEZ0912;  ANDT1683;  ANDT1683;  ANEP0866;  ANER1537;  ANEX6019;  ANCP1854;  ANDR0755;  ANEV1712;  ANEW3370;  ANER1537;  ANEW3370;  ANER1537;  ANEZ3956;  ANDR2273;  ANDV0711;  ANDR0755;  ANDS1296;  ANCV1585;  ANDX1408;  ANEX6019;  ANEZ2086;  ANDW1480;  ANER1537;  ANCX3894;  ANET3789;  ANEP5325;  ANEX0747;  ANER1537;  ANCY3306;  ANDX1408;  ANEV1712;  ANEZ2086;  ANET0546;  ANCQ4288;  ANDX5302;  ANDZ2276;  ANDZ3829;  ANEZ3956;  ANDR2273;  ANDT1683;  ANDV3569;  ANDP3962;  ANDS1296;  ANEP5325;  ANDP3962;  ANES4069;  ANEZ0912;  ANEZ2086;  ANDZ1342;  ANER1537;  ANET3789;  ANEX1709;  ANEP5325;  ANEN1074;  ANER1537;  ANEX6019;  ANDR0755;  ANDX1408;  ANCS1564;  ANDV1209;  ANDX1408;  ANEZ0912;  ANEZ3956;  ANDR3258;  ANDR3258;  ANDV1209;  ANDX2036;  ANEQ4951;  ANES4069;  ANET0546;  ANEZ2086;  ANEZ2086;  ANEX1709  ANDV3569;  ANET3789;  ANDP3962;  ANEQ4951;  ANES4069;  ANET3789;  ANDX1408;  ANEP3927;  ANEV1712;  ANDV1209;  ANDW1480;  ANDX1408;  ANDX2036;  ANDX5302;  ANDX5302;  ANDR3258;  ANDZ2276;  ANDX1408;  ANEN4899;  ANDT1683;  ANDX5302;  ANEQ4046;  ANEX6019;  ANEZ3956;  ANEQ4951;  ANET0546;  ANDP1361;  ANDS1296;  ANDS4826;  ANEW3370;  ANDT1683;  ANES4069;  ANDR2273;  ANEP5325;  ANDX2036;  ANDS4826;  ANDX1408;  ANDZ3829;  ANER1537;  ANES2194;  ANEX6019;  ANEN4899;  ANEQ4046;  ANEX1709;  ANDW1480;  ANDX1408;  ANEQ4046;  ANEQ4951;  ANEX0747;  ANDS1296;  ANDX1408;  ANET3789;  ANEZ3956;  ANEV1712;  ANEZ3956;  ANCY4323;  ANCY4323;  ANDN1926;  ANDN1926;  ANDV4541;  ANDW1486;  ANDX3075;  ANDX5319;  ANDY1225;  ANDY3298;  ANDZ2278;  ANEN1075;  ANCV3481;  ANCV3481;  ANDR4203;  ANDV4541;  ANCS2424;  ANCS3347;  ANCV0501;  ANCX3895;  ANDX1382;  ANBZ2678;  ANEX0754;  ANDV0698;  ANDV4541;  ANEQ4952;  ANEQ4047;  ANEQ4952;  ANFN2850;  ANDV4541;  ANDV4541;  ANDZ2278;  ANEQ4952;  ANER1543;  ANDV4541;  ANDZ2278;  ANDX3075;  ANEP0878;  ANDZ3834;  ANEP0878;  ANEP2428;  ANEQ2721;  ANER1543;  ANER1543;  ANEZ0913;  ANEZ0913;  ANEZ2087;  ANEZ3403;  ANDX3075;  ANDR4203;  ANCV2923;  ANDZ2278;  ANEN1075;  ANEP0878;  ANER1543;  ANEZ2087;  ANFN2850;  ANDV3572;  ANDV4541;  ANEQ4047;  ANDV0698;  ANDV3572;  ANDV4541;  ANCS2424;  ANDW1486;  ANDR3263;  ANEP3932;  ANCX3895;  ANDW1486;  ANEX0754;  ANCS3347;  ANDV4541;  ANDV4541;  ANEQ4047;  ANES2195;  ANES2195;  ANES4070;  ANDV0698;  ANEP3932;  ANEQ2721;  ANEQ4047;  ANDS1310;  ANDS4827;  ANDV4541;  ANDV4541;  ANDW1486;  ANDX1416;  ANDX5319;  ANDN1926;  ANDQ3055;  ANDR4203;  ANDS1310;  ANDX1382;  ANDX1416;  ANDX3075;  ANCS2424;  ANCV0501;  ANCX0894;  ANCY3307;  ANCY4323;  ANDQ3055;  ANDQ3055;  ANDR3263;  ANDR4203;  ANDS4827;  ANDV3572;  ANDV4541;  ANDV4541;  ANDW1486;  ANDX3075;  ANDX3075;  ANDX3075;  ANDY1225;  ANDZ2278;  ANDZ3834;  ANEN1075;  ANEP0878;  ANEP2428;  ANEP3932;  ANEQ4952;  ANER1543;  ANEX0754;  ANEZ0913;  ANEZ0913;  ANDS4827;  ANDS4827;  ANEQ4952;  ANER1543;  ANES4070;  ANEX0754;  ANCX0894;  ANDR3263;  ANDS4827;  ANDS4827;  ANDW1486;  ANES2195;  ANDR4203;  ANDS4827;  ANDV0698;  ANDV3572;  ANDV3572;  ANDX1416;  ANDZ2278;  ANEN1075;  ANEP3932;  ANEQ4047;  ANEQ4952;  ANEQ4952;  ANEQ4952;  ANEQ4952;  ANES2195;  ANES2195;  ANES4070;  ANES4070;  ANER1543;  ANDR4203;  ANDS1310;  ANEQ4952;  ANEZ0913;  ANDS4827;  ANEP0878;  ANDW1486;  ANCV2923;  ANER1543;  ANET0547;  ANET0547;  ANEX0754;  ANER1543;  ANEX0754;  ANEZ3403;  ANDV0698;  ANDV3572;  ANDW1486;  ANDZ3834;  ANEQ4952;  ANDV3572;  ANEX0754;  ANEZ0913;  ANDS4827;  ANDV3572;  ANDX1416;  ANDY1225;  ANDZ3834;  ANEP3932;  ANER1543;  ANEX0754;  ANEZ3403;  ANEP2428;  ANDX5319;  ANDX5319;  ANDZ2278;  ANEQ2721;  ANEQ4047;  ANES4070;  ANDV4541;  ANDR3263;  ANDR4203;  ANEQ2721;  ANEQ4952;  ANDV4541;  ANEZ3403;  ANDR3263;  ANDW1486;  ANDX1416;  ANDX3075;  ANDX5319;  ANEP0878;  ANEQ4952;  ANEZ0913;  ANEZ3403;  ANDS1310;  ANDS4827;  ANDX1416;  ANEN1075;  ANEN1075;  ANEQ2721;  ANES2195;  ANDR4203;  ANDS4827;  ANDZ2278;  ANES4070;  ANCV3481;  ANDQ3055;  ANDV0698;  ANEQ4952;  ANDX1416;  ANEP2428;  ANEP2428;  ANCY3307;  ANDR3263;  ANEP2428;  ANFN2850;  ANEQ4952;  ANES4070;  ANDW1486;  ANDX1382;  ANDX5319;  ANDR4203;  ANDS4827;  ANEZ0913;  ANCX0894;  ANEN1075;  ANEP0878;  ANDV0698;  ANDR4203;  ANDX1416;  ANDS4827;  ANDW1486;  ANDZ3834;  ANFN3959;  ANDX1416;  ANEP3932;  ANEX0754;  ANEZ0913;  ANDX1416;  ANFP3405;  ANEZ3403;  ANDV2510;  ANDX3079;  ANDY1219;  ANDY4971;  ANDZ3835;  ANEN3397;  ANDU2247;  ANDU3135;  ANDR4206;  ANDR4206;  ANEQ4048;  ANDV3575;  ANDU3135;  ANDY4971;  ANEQ4048;  ANEQ4048;  ANEQ4953;  ANES2196;  ANES2196;  ANEY2035;  ANEY2035;  ANDR4206;  ANDR4206;  ANDU3135;  ANDR4206;  ANDU3135;  ANEZ2088;  ANDR4206;  ANEQ2722;  ANEQ2722;  ANDR4206;  ANDV3575;  ANDW4356;  ANDU2247;  ANDV0710;  ANDV2510;  ANDV0710;  ANDV0710;  ANDV2510;  ANDY2463;  ANDR4206;  ANDR4206;  ANDW4356;  ANDX1372;  ANDX1386;  ANCY3308;  ANDR4206;  ANDR4206;  ANDR4206;  ANDU2247;  ANDU3135;  ANDV0710;  ANDV0710;  ANDW4356;  ANDX1372;  ANDY1219;  ANDY4971;  ANEP3919;  ANEP3919;  ANEP3919;  ANEY2035;  ANCY3308;  ANDU2247;  ANEQ4048;  ANES2196;  ANET0548;  ANET0548;  ANEQ4953;  ANEQ4953;  ANDR4206;  ANES2196;  ANES2196;  ANEX1704;  ANEQ2722;  ANEQ4953;  ANEQ4048;  ANEQ4953;  ANEY2035;  ANEQ4048;  ANDR4206;  ANDU3135;  ANDY4971;  ANEQ4048;  ANDR4206;  ANDU2247;  ANDV3575;  ANDY1219;  ANDW4356;  ANDY2463;  ANDY2463;  ANDY2463;  ANDY4971;  ANEQ4953;  ANEQ4953;  ANDV3575;  ANES2196;  ANDY2463;  ANET0548;  ANCR0697;  ANDR4206;  ANDV3575;  ANDX3079;  ANDY4971;  ANEN3397;  ANEP3919;  ANEQ2722;  ANEQ2722;  ANEQ2722;  ANEY2035;  ANEZ2088;  ANDU2247;  ANDV2510;  ANET0548;  ANDY4971;  ANDR4206;;  ANDV2510;  ANDV3575;  ANDY2463;  ANEN3397;  ANDU2247;  ANDW4356;  ANCX0896;  ANDW4356;  ANDX3079;  ANDY4971;  ANEY2035;  ANFN3960;  ANDZ3835;  ANEP3919;  ANDV2510;  ANEZ2088;  ANEX1704;  ANDV1221;  ANDV2511;  ANDP3975;  ANDP3975;  ANDQ3058;  ANDQ3058;  ANDV2511;  ANDX3077;  ANDX5320;  ANDV0298;  ANDR4205;  ANDV2511;  ANDR2278;  ANDQ3058;  ANCV3480;  ANDW1487;  ANEQ0214;  ANEQ0214;  ANEZ3437;  ANDV0298;  ANDS3982;  ANDV0298;  ANDV0298;  ANDV2511;  ANEQ0214;  ANDV1221;  ANCV3480;  ANDQ3058;  ANEN2139;  ANCY3309;  ANDP3975;  ANDV1221;  ANDS3982;  ANDV2511;  ANDX5320;  ANEQ0214;  ANDQ3058;  ANDQ3058;  ANDR0759;  ANDR2278;  ANDR4205;  ANDS0761;  ANDX1418;  ANCT4330;  ANCT4330;  ANDP3975;  ANDR0759;  ANDR2278;  ANDS3982;  ANDU3096;  ANDX1418;  ANDX2042;  ANEN2139;  ANEN2139;  ANEN2139;  ANEN2139;  ANEP0868;  ANEQ0214;  ANEQ0214;  ANEQ0214;  ANEQ0214;  ANDS4785;  ANDP3975;  ANDQ3058;  ANDR0759;  ANDS4828;  ANDU2248;  ANDU2248;  ANDV2511;  ANDW1487;  ANDW1487;  ANDX3077;  ANDX3077;  ANDX5320;  ANDY1216;  ANDY1216;  ANDY1216;  ANEN2139;  ANEQ0214;  ANEQ1349;  ANEQ1349;  ANER1545;  ANER1545;  ANER1545;  ANER1545;  ANER1545;  ANER1545;  ANER1545;  ANER1545;  ANES4071;  ANES4071;  ANEZ0914;  ANEZ3437;  ANER1545;  ANDS3982;  ANDS3982;  ANDU2248;  ANEN2139;  ANEP0868;  ANEP0868;  ANEQ0214;  ANEQ0214;  ANER1545;  ANER1545;  ANEQ0214;  ANER1545;  ANEQ0214;  ANDP3975;  ANDQ3058;  ANDR4205;  ANDS4785;  ANDS4828;  ANDU2248;  ANDU2248;  ANDU3096;  ANDW1487;  ANDX2042;  ANDX5320;  ANDY1216;  ANEN2139;  ANER1545;  ANER1545;  ANDR2278;  ANDS3982;  ANDV2511;  ANDW1487;  ANEZ2089;  ANDS3982;  ANDS3982;  ANDU2248;  ANDX2042;  ANDR0759;  ANDX2042;  ANEP0868;  ANDV0298;  ANDW1487;  ANEP0868;  ANEP0868;  ANEZ0914;  ANDR0759;  ANDR0759;  ANDS0761;  ANDU2248;  ANDV2511;  ANDX1418;  ANDX3077;  ANDX5320;  ANEN2139;  ANEQ0214;  ANER1545;  ANES4071;  ANEQ0214;  ANDP3975;  ANCT4330;  ANCV3480;  ANDU2248;  ANDV1221;  ANEP0868;  ANEP0868;  ANDU3096;  ANEN2139;  ANEQ0214;  ANEZ2089;  ANDP3975;  ANDS0761;  ANDS4785;  ANDX2042;  ANEN2139;  ANEN2139;  ANEP0868;  ANES4071;  ANEP0868;  ANDX2042;  ANEQ0214;  ANES4071;  ANDS3982;  ANDS4785;  ANDV0298;  ANDW1487;  ANEN2139;  ANES4071;  ANEZ3437;  ANEN2139;  ANDR4205;  ANDW1487;  ANDW1487;  ANDX3077;  ANER1545;  ANEZ0914;  ANDR4205;  ANEZ0914;  ANER1545;  ANEZ0914;",
      "recall_initiation_date": "20201214",
      "postal_code": "85281-2438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}