  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Tempe",
      "address_1": "1625 W 3rd St Bldg 1",
      "reason_for_recall": "Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.",
      "address_2": "",
      "product_quantity": "Total Devices=12,258 (U.S.= 7,234; O.U.S.=5,024)",
      "code_info": "Product Code/GTIN:  VENEL10040 /00801741102004;  VENEL10060 /00801741104664;  VENEL10080 /00801741104671;  VENEL10100 /00801741102035;  VENEL10120 /00801741102042;  VENEL10140 /00801741102059;  VENEL10160 /00801741102066;  VENEL12040 /00801741102141;  VENEL12060 /00801741102158;  VENEL12080 /00801741102172;  VENEL12100 /00801741102189;  VENEL12120 /00801741102196;  VENEL12140 /00801741102202;  VENEL12160 /00801741102219;  VENEM10040 /00801741101939;  VENEM10060 /00801741101946;  VENEM10080 /00801741101953;  VENEM10100 /00801741101960;  VENEM10120 /00801741101977;  VENEM10140 /00801741101984;  VENEM10160 /00801741101991;  VENEM12040 /00801741102073;  VENEM12060 /00801741102080;  VENEM12080 /00801741102097;  VENEM12100 /00801741102103;  VENEM12120 /00801741102110;  VENEM12140 /00801741102127;  VENEM12160 /00801741102134;  VENUL10040 /00801741103209;  VENUL10060 /00801741103216;  VENUL10080 /00801741103223;  VENUL10100 /00801741103230;  VENUL10120 /00801741103247;  VENUL10140 /00801741103254;  VENUL10160 /00801741103261;  VENUL12040 /00801741103346;  VENUL12060 /00801741103353;  VENUL12080 /00801741103360;  VENUL12100 /00801741103377;  VENUL12120 /00801741103384;  VENUL12140 /00801741103391;  VENUL12160 /00801741103407;  VENUM10040 /00801741103131;  VENUM10060 /00801741103148;  VENUM10080 /00801741103155;  VENUM10100 /00801741103162;  VENUM10120 /00801741103179;  VENUM10140 /00801741103186;  VENUM10160 /00801741103193;  VENUM12040 /00801741103278;  VENUM12060 /00801741103285;  VENUM12080 /00801741103292;  VENUM12100 /00801741103308;  VENUM12120 /00801741103315;  VENUM12140 /00801741103322;  VENUM12160 /00801741103339;    U.S. Device; Serial Numbers:  ANEQ4079;  ANEQ4090;  ANEQ4091;  ANEQ4101;  ANEQ4104;  ANEQ4105;  ANEQ4109;  ANEQ4978;  ANEQ4979;  ANEQ4980;  ANEQ4996;  ANEQ4997;  ANEQ5023;  ANER0477;  ANER0478;  ANER0490;  ANER0491;  ANER0510;  ANER0515;  ANER0532;  ANER1502;  ANER1528;  ANER1529;  ANER1530;  ANER1533;  ANER2795;  ANER2796;  ANER2797;  ANER2812;  ANER2813;  ANER2814;  ANER2815;  ANER2816;  ANER2838;  ANER2841;  ANER2855;  ANER4217;  ANER4228;  ANER4236;  ANER4238;  ANER4239;  ANER5126;  ANER5127;  ANER5136;  ANER5141;  ANER5142;  ANER5153;  ANER5159;  ANER5160;  ANER5169;  ANER5175;  ANES2223;  ANES2224;  ANES2233;  ANES2234;  ANES2235;  ANES2236;  ANES3045;  ANES3050;  ANES3067;  ANES4102;  ANES4103;  ANES4111;  ANES4112;  ANES4113;  ANES4114;  ANES4115;  ANES4116;  ANES4117;  ANES4118;  ANET0561;  ANET0566;  ANET0567;  ANET0568;  ANET1490;  ANET1508;  ANET1509;  ANET2533;  ANET2534;  ANET2535;  ANET2536;  ANET2537;  ANET2540;  ANET2541;  ANET3797;  ANET3798;  ANET3799;  ANET3800;  ANET3814;  ANET3815;  ANET3816;  ANET3817;  ANET3818;  ANET3877;  ANET3878;  ANEU0748;  ANEU1594;  ANEU1601;  ANEU1602;  ANEU1603;  ANEU1604;  ANEU1605;  ANEU1606;  ANEU2442;  ANEU3275;  ANEU3276;  ANEU3277;  ANEU3278;  ANEU3284;  ANEU3285;  ANEU3286;  ANEU3287;  ANEU3288;  ANEU3289;  ANEU3290;  ANEU3306;  ANEU4231;  ANEU4232;  ANEU4238;  ANEU4239;  ANEU4240;  ANEU4246;  ANEU4253;  ANEU4256;  ANEU4260;  ANEU4261;  ANEU4264;  ANEU4265;  ANEV0586;  ANEV0587;  ANEV0591;  ANEV0594;  ANEV0598;  ANEV0599;  ANEV0603;  ANEV0606;  ANEV0611;  ANEV0614;  ANEV1714;  ANEV1715;  ANEV1716;  ANEV1717;  ANEV1718;  ANEV1737;  ANEV1738;  ANEV1739;  ANEV1740;  ANEV3078;  ANEV3079;  ANEV3080;  ANEV3081;  ANEV3082;  ANEV3494;  ANEV4019;  ANEV4020;  ANEV4021;  ANEV4022;  ANEV4023;  ANEV4024;  ANEV4025;  ANEV4043;  ANEV4044;  ANEV4045;  ANEV4046;  ANEV4071;  ANEV4076;  ANEW0553;  ANEW0554;  ANEW0560;  ANEW0561;  ANEW0562;  ANEW0563;  ANEW0564;  ANEW0565;  ANEW0566;  ANEW0599;  ANEW0600;  ANEW2005;  ANEW2006;  ANEW2007;  ANEW2028;  ANEW2029;  ANEW2030;  ANEW2031;  ANEW2032;  ANEW2033;  ANEW2034;  ANEW2239;  ANEW2772;  ANEW2773;  ANEW3378;  ANEW3379;  ANEW3380;  ANEW3381;  ANEW3394;  ANEW3395;  ANEW3396;  ANEW3397;  ANEW3422;  ANEW4229;  ANEW4230;  ANEW4231;  ANEW4232;  ANEW4233;  ANEW4242;  ANEW4243;  ANEW4244;  ANEW4245;  ANEW4246;  ANEW4247;  ANEW4276;  ANEW4280;  ANEW4282;  ANEW4291;  ANEW4297;  ANEW4298;  ANEW4300;  ANEX0704;  ANEX0711;  ANEX0713;  ANEX0720;  ANEX0729;  ANEX0732;  ANEX0735;  ANEX0736;  ANEX0739;  ANEX0740;  ANEX0741;  ANEX0742;  ANEX0743;  ANEX0744;  ANEX0745;  ANEX1737;  ANEX1743;  ANEX1744;  ANEX1746;  ANEX1747;  ANEX1750;  ANEX1752;  ANEX1753;  ANEX1754;  ANEX1755;  ANEX3120;  ANEX3121;  ANEX3123;  ANEX3131;  ANEX3135;  ANEX3136;  ANEX3137;  ANEX3144;  ANEX3145;  ANEX3147;  ANEX3148;  ANEX3149;  ANEX3151;  ANEX3153;  ANEX3154;  ANEX3156;  ANEX4737;  ANEX4740;  ANEX4744;  ANEX4745;  ANEX4751;  ANEX4753;  ANEX4761;  ANEX4762;  ANEX4764;  ANEX4773;  ANEX4779;  ANEX4780;  ANEX4782;  ANEX4783;  ANEX4784;  ANEX4785;  ANEX4786;  ANEX4787;  ANEX4788;  ANEX4789;  ANEX6042;  ANEX6046;  ANEX6049;  ANEX6056;  ANEX6057;  ANEX6065;  ANEX6066;  ANEX6068;  ANEY1246;  ANEY1247;  ANEY1251;  ANEY1257;  ANEY1258;  ANEY1259;  ANEY1262;  ANEY1263;  ANEY1264;  ANEY1265;  ANEY1266;  ANEY1267;  ANEY1268;  ANEY1269;  ANEY1278;  ANEY1279;  ANEY1280;  ANEY2049;  ANEY2050;  ANEY2051;  ANEY2052;  ANEY2053;  ANEY2054;  ANEY2055;  ANEY2056;  ANEY3241;  ANEY3243;  ANEY3245;  ANEY3255;  ANEY3256;  ANEY3257;  ANEY3258;  ANEY3261;  ANEY3537;  ANEY3538;  ANEY4346;  ANEY4354;  ANEY4355;  ANEY4356;  ANEY4357;  ANEY4358;  ANEY4562;  ANEZ0957;  ANEZ0958;  ANEZ0959;  ANEZ0967;  ANEZ0972;  ANEZ0973;  ANEZ0974;  ANEZ0980;  ANEZ0996;  ANEZ0997;  ANEZ1000;  ANEZ1001;  ANEZ1002;  ANEZ1003;  ANEZ1004;  ANEZ2107;  ANEZ2112;  ANEZ2113;  ANEZ2121;  ANEZ2122;  ANEZ2123;  ANEZ2124;  ANEZ2137;  ANEZ2138;  ANEZ2139;  ANEZ2140;  ANEZ3453;  ANEZ3454;  ANEZ3455;  ANEZ3502;  ANEZ3939;  ANEZ3949;  ANFN1482;  ANFN1483;  ANFN1484;  ANFN1485;  ANFN1486;  ANFN1487;  ANFN1488;  ANFN1490;  ANFN1492;  ANFN1493;  ANFN2809;  ANFN2817;  ANFN2821;  ANFN2822;  ANFN2824;  ANFN2825;  ANFN2835;  ANFN2836;  ANFN2837;  ANFN2838;  ANFN2839;  ANFN2840;  ANFN2841;  ANFN2842;  ANFN2843;  ANFN3989;  ANFN3990;  ANFN3991;  ANFN3992;  ANFN3993;  ANFN3994;  ANFN3999;  ANFN4000;  ANFN4001;  ANFN4002;  ANFP0891;  ANFP0896;  ANFP0897;  ANFP0915;  ANFP0923;  ANDS0730;  ANDS0722;  ANDS0729;  ANDS1275;  ANDS1312;  ANDS1302;  ANDS2454;  ANDS4842;  ANDS4851;  ANDS4854;  ANDS4852;  ANDT1073;  ANDT1713;  ANDU2241;  ANDT1721;  ANDS4853;  ANDT2908;  ANDT2907;  ANDT2906;  ANDT2905;  ANDU0121;  ANDU1126;  ANDU1125;  ANDU0123;  ANDU0122;  ANDV0714;  ANDU1894;  ANDU1906;  ANDU1861;  ANDU1862;  ANDU1890;  ANDU1856;  ANDU1884;  ANDU1873;  ANDU1124;  ANDU1114;  ANDU1905;  ANDU1878;  ANDU1867;  ANDU3084;  ANDU3144;  ANDU3145;  ANDU3139;  ANDU3140;  ANDU3104;  ANDU3098;  ANDU3128;  ANDU3103;  ANDU3127;  ANDU3089;  ANDV3781;  ANDU3090;  ANDV1215;  ANDV1227;  ANDU4041;  ANDV1261;  ANDU3111;  ANDV3603;  ANDV3602;  ANDV3606;  ANDV3576;  ANDV3604;  ANDV3601;  ANDV4566;  ANDV4554;  ANDV4551;  ANDV4567;  ANDV4546;  ANDV4552;  ANDV4549;  ANDV4550;  ANDV4553;  ANDV4568;  ANDV4545;  ANDV3600;  ANDV4555;  ANDV1234;  ANDV3605;  ANDV4547;  ANDW2044;  ANDW2042;  ANDW2028;  ANDW2043;  ANDW2046;  ANDW2047;  ANDW2045;  ANDX1371;  ANDX1388;  ANDX1429;  ANDX1428;  ANDW3626;  ANDX1427;  ANDX1423;  ANDX1426;  ANDV4569;  ANDX1425;  ANDW3625;  ANDW4360;  ANDW4367;  ANDW4383;  ANDX1430;  ANDW4369;  ANDW4400;  ANDX1424;  ANDW4340;  ANDX0708;  ANDW4419;  ANDX0725;  ANDX0720;  ANDW4402;  ANDW4353;  ANDX2191;  ANDX2172;  ANDX2171;  ANDX2192;  ANDX2190;  ANDX3082;  ANDX3103;  ANDX3105;  ANDX3102;  ANDX3101;  ANDX3086;  ANDX3081;  ANDX2173;  ANDX3100;  ANDX3085;  ANDX3083;  ANDX2193;  ANDX3104;  ANDX3099;  ANDX3084;  ANDX3106;  ANDX3107;  ANDX5400;  ANDX5399;  ANDY1247;  ANDZ0784;  ANDY2488;  ANDY2486;  ANDY2487;  ANDY2470;  ANDZ2797;  ANDZ2796;  ANDZ1376;  ANDZ2803;  ANDY3323;  ANDY3310;  ANDZ1255;  ANDY3311;  ANDY3324;  ANDZ1243;  ANDZ1254;  ANDZ1372;  ANDZ1357;  ANDZ1373;  ANDY3325;  ANDZ1358;  ANDZ1375;  ANDY3326;  ANDZ1256;  ANDZ1244;  ANDZ1377;  ANDZ1258;  ANDZ1374;  ANDZ2326;  ANDZ1257;  ANDZ2325;  ANDZ3877;  ANDZ3844;  ANDZ3870;  ANDZ3861;  ANEN1090;  ANEN1107;  ANEN1112;  ANEN1088;  ANEN1111;  ANEN1093;  ANEN1108;  ANEN1109;  ANEN1094;  ANEN1089;  ANEN1110;  ANEN2194;  ANEN2193;  ANEN2189;  ANEN2190;  ANEN2185;  ANEN2182;  ANEN2186;  ANEN2187;  ANEN3444;  ANEN3421;  ANEN3438;  ANEN2188;  ANEN3430;  ANEN3436;  ANEN3422;  ANEN3441;  ANEN3433;  ANEN2180;  ANEN3423;  ANEN3431;  ANEN4928;  ANEN4916;  ANEP0914;  ANEP0919;  ANEP0915;  ANEP0912;  ANEP0906;  ANEP0918;  ANEP2441;  ANEP2455;  ANEP2450;  ANEP2442;  ANEP2461;  ANEQ3077;  ANEP2443;  ANEP3957;  ANEP3963;  ANEP3944;  ANEP3974;  ANEP3967;  ANEP3984;  ANEP3969;  ANEP3942;  ANEP3949;  ANEQ3078;  ANEP3985;  ANEP5356;  ANEP5334;  ANEP5340;  ANEP5386;  ANEP5359;  ANEP5358;  ANEP5338;  ANEP5365;  ANEP5380;  ANEP5385;  ANEP5361;  ANEP5374;  ANEP5341;  ANEP5335;  ANEP5352;  ANEQ1378;  ANEQ1397;  ANEQ1379;  ANEQ1399;  ANEQ1398;  ANEQ1376;  ANEQ1377;  ANEQ1401;  ANEQ1396;  ANEQ1380;  ANEQ1375;  ANEQ2773;  ANEQ2774;  ANEQ1400;  ANEQ2776;  ANEQ2775;  ANEQ2756;  ANEQ2757;  ANEQ2755;  ANEQ4078;  ANFQ2793;    O.U.S. Device; Serial Numbers:  ANDU2240;  ANES2208;  ANDU2240;  ANDY3304;  ANEP2429;  ANEP5328;  ANEP2429;  ANDZ3827;  ANDN0820;  ANDU2240;  ANEP2429;  ANDU2240;  ANEP2429;  ANEP2429;  ANDU2240;  ANEP5328;  ANEP2429;  ANES1042;  ANDZ2288;  ANEP5328;  ANDU2240;  ANDY3304;  ANDZ3827;  ANEP2429;  ANDN0820;  ANES1042;  ANDZ2288;  ANEP2429;  ANEP5328;  ANEP2429;  ANEP5328;  ANES1042;  ANDU2240;  ANDY3304;  ANDZ2288;  ANDZ3827;  ANDZ3827;  ANCV0493;  ANDU2240;  ANDU2240;  ANES1042;  ANES1042;  ANEP5328;  ANEP2429;  ANEP2429;  ANDN0820;  ANDU2240;  ANEP5328;  ANDY3304;  ANDN0820;  ANDZ2288;  ANDN0820;  ANEP5328;  ANCV0493;  ANDZ3827;  ANEP2429;  ANEP5328;  ANES2208;  ANEP5328;  ANES1042;  ANES2208;  ANEY3214;  ANEY4323;  ANFP2666;  ANFP2666;  ANDU2240;  ANEY3214;  ANEY3214;  ANEY3214;  ANEY3214;  ANEP5328;  ANEP2429;  ANDR1257;  ANDR2270;  ANDR2270;  ANDU1102;  ANDV1222;  ANDY2458;  ANDY2458;  ANDY3301;  ANDY3301;  ANDZ1338;  ANDZ3819;  ANEN1079;  ANEN3405;  ANER5167;  ANEX0679;  ANEZ2095;  ANEZ3410;  ANFN3972;  ANCR2676;  ANDR2270;  ANDV1222;  ANDV2517;  ANDY2458;  ANEZ3410;  ANEZ3410;  ANEZ2095;  ANDX5352;  ANEQ4065;  ANEQ4961;  ANEN3405;  ANDV1222;  ANDX5352;  ANDZ1338;  ANDV2517;  ANDV1222;  ANFN3972;  ANEN3405;  ANFN3972;  ANFN3972;  ANFN3972;  ANEN3405;  ANDV2517;  ANDY2458;  ANDR2270;  ANEQ4065;  ANDV2517;  ANEQ4065;  ANEN3405;  ANEQ4961;  ANFN3972;  ANDV2517;  ANEZ2095;  ANCV1579;  ANEN3405;  ANDR2270;  ANEN3405;  ANDX5352;  ANDZ3819;  ANEN1079;  ANEQ4961;  ANER5167;  ANEX0679;  ANEX4720;  ANEY4324;  ANEX4720;  ANEY4324;  ANEZ0921;  ANEZ0921;  ANEZ2095;  ANEZ2095;  ANEZ3410;  ANDR2270;  ANEQ4065;  ANEQ4961;  ANDR2270;  ANEQ4961;  ANEQ4961;  ANEQ4961;  ANEQ4065;  ANEQ4065;  ANDR1257;  ANCV1579;  ANDZ1338;  ANEQ4065;  ANER5167;  ANDR1257;  ANDV1222;  ANDV2517;  ANDX5352;  ANDX5352;  ANDY2458;  ANDR1257;  ANDR1257;  ANDR1257;  ANDZ1338;  ANDZ1338;  ANCV1579;  ANDY2458;  ANEQ4065;  ANFN3972;  ANEX4720;  ANEZ0921;  ANFN3972;  ANEQ4961;  ANDV2517;  ANDY2458;  ANDQ3035;  ANDQ3035;  ANDQ3035;  ANDQ3877;  ANDU1105;  ANDW4412;  ANDX0715;  ANDX5308;  ANDZ2292;  ANDZ3826;  ANEN1080;  ANEQ4066;  ANES2209;  ANET2531;  ANET2531;  ANEX1699;  ANEX1699;  ANEY4325;  ANEZ0982;  ANBY2401;  ANCT0397;  ANCT0397;  ANDZ1330;  ANES2209;  ANDX5308;  ANEY4325;  ANDW4412;  ANCW4225;  ANDV0287;  ANEQ2730;  ANEQ2730;  ANEQ2730;  ANEQ4066;  ANEX1699;  ANEZ0982;  ANEX1699;  ANFP0878;  ANFP0878;  ANET2531;  ANDQ3035;  ANDQ3035;  ANDQ3877;  ANEQ4066;  ANET2531;  ANEN1080;  ANEY4325;  ANDX5308;  ANDX5308;  ANDQ3035;  ANEQ2730;  ANCW3270;  ANES2209;  ANDX5308;  ANDZ1330;  ANDZ2292;  ANDZ3826;  ANEN1080;  ANEX1699;  ANEY3212;  ANEY4325;  ANEZ0982;  ANFP0878;  ANET2531;  ANEX1699;  ANEX1699;  ANEY3212;  ANEY4325;  ANEZ0982;  ANDQ3035;  ANDQ3035;  ANEZ0982;  ANEY4325;  ANCW3270;  ANCW4225;  ANEY4325;  ANET2531;  ANES2209;  ANET2531;  ANDQ3035;  ANDQ3877;  ANDZ3826;  ANDQ3877;  ANES2209;  ANET2531;  ANDV0287;  ANDX0715;  ANDX5308;  ANDZ1330;  ANDZ2292;  ANDQ3877;  ANDX5308;  ANEN1080;  ANEY4325;  ANDQ3877;  ANEQ4066;  ANEZ0982;  ANCW3270;  ANEQ4066;  ANES2209;  ANEQ2730;  ANEY4325;  ANDQ3041;  ANCU2726;  ANCX3902;  ANDQ3880;  ANCX3902;  ANDQ3041;  ANDQ3041;  ANEP5323;  ANEP5323;  ANEQ2731;  ANEP5323;  ANEP5323;  ANEQ4067;  ANDQ3880;  ANDY3305;  ANDQ3880;  ANCX3902;  ANDY3305;  ANEP0856;  ANEP0856;  ANEP0856;  ANEP5323;  ANEQ2731;  ANES2210;  ANES2210;  ANEZ0922;  ANFN3973;  ANFN3973;  ANEZ0922;  ANFN1442;  ANFN3973;  ANFN3973;  ANDQ3041;  ANCX3902;  ANDY3305;  ANDY3305;  ANEQ2731;  ANDY3305;  ANEP5323;  ANDQ3880;  ANDY3305;  ANCX3902;  ANDQ3041;  ANDQ3041;  ANDY3305;  ANDQ3880;  ANDQ3880;  ANDQ3880;  ANEQ2731;  ANER0516;  ANCX3902;  ANDQ3880;  ANEP5323;  ANEP5323;  ANEP5323;  ANDQ3880;  ANEP0856;  ANEP5323;  ANEZ0922;  ANFN3973;  ANDQ3023;  ANDT1058;  ANDT1058;  ANDT1058;  ANDU4075;  ANDU4075;  ANDZ1348;  ANEN1081;  ANEN1081;  ANEP0857;  ANEP0857;  ANEP3924;  ANET3794;  ANEX0680;  ANEX0680;  ANEX0680;  ANEZ2096;  ANEP3924;  ANEZ2096;  ANEP3924;  ANEP3924;  ANDQ3023;  ANDQ3023;  ANEP3924;  ANEP3924;  ANDU4075;  ANDQ3023;  ANDU4075;  ANDZ1348;  ANCU3706;  ANDZ2290;  ANCU3706;  ANDZ2290;  ANDZ2290;  ANDZ2290;  ANDZ2296;  ANEP5332;  ANDZ2296;  ANCU3716;  ANEP5332;  ANDZ2296;  ANDZ2296;  ANDR1253;  ANDU2239;  ANEQ2716;  ANDU2239;  ANES2187;  ANCU0847;  ANDX5374;  ANDW4406;  ANDW4406;  ANES2187;  ANDU2239;  ANCU0847;  ANDR1253;  ANEQ2716;  ANDR1253;  ANDW4406;  ANES2187;  ANCY3323;  ANDU2239;  ANDU2239;  ANDW4406;  ANDU2239;  ANES2187;  ANDW4406;  ANEP3938;  ANDQ3029;  ANDQ3029;  ANDW4406;  ANDX5374;  ANDQ3029;  ANDR1253;  ANDR1253;  ANDX5374;  ANES2187;  ANDR1253;  ANDR1253;  ANEQ2716;  ANDW4406;  ANCU0847;  ANCU0847;  ANDR1253;  ANDU2239;  ANDU2239;  ANDW4406;  ANDW4406;  ANEQ2716;  ANCU0847;  ANCY3323;  ANDR1253;  ANDR1253;  ANDX5374;  ANDX5374;  ANDU2239;  ANEQ2716;  ANES2187;  ANDR1253;  ANDU2239;  ANDW4406;  ANEP3938;  ANDR1253;  ANDU2239;  ANDQ3037;  ANDQ3037;  ANDT1057;  ANDT1057;  ANDW4338;  ANDW4338;  ANDX5309;  ANDX5309;  ANDZ2308;  ANEP0852;  ANEP0852;  ANER5149;  ANET3785;  ANEZ0906;  ANDT1057;  ANDQ3037;  ANDX5309;  ANEQ4945;  ANDT1057;  ANDT1057;  ANDX5309;  ANCW3260;  ANDT1057;  ANEP3936;  ANEP3936;  ANDX5309;  ANEQ4945;  ANDT1057;  ANEX3092;  ANEQ4036;  ANCY3332;  ANDU4074;  ANDX5309;  ANDZ2308;  ANEQ4036;  ANEZ0906;  ANDZ2308;  ANEP3936;  ANDX5309;  ANEP0852;  ANEQ4945;  ANEQ4945;  ANET3785;  ANEZ0906;  ANFP2368;  ANEQ4945;  ANET3785;  ANEX0671;  ANEX0671;  ANEX3092;  ANEX3092;  ANEZ0906;  ANDQ3037;  ANDX5309;  ANDQ3037;  ANDT1057;  ANDX5309;  ANDX5309;  ANDZ2308;  ANEP3936;  ANEP3936;  ANEP3936;  ANEQ4036;  ANEQ4036;  ANEQ4036;  ANDT1057;  ANDT1057;  ANEZ0906;  ANEQ4945;  ANDT1057;  ANEQ4036;  ANDW4338;  ANCR1284;  ANDZ2308;  ANER5149;  ANET3785;  ANEX0671;  ANER5149;  ANDX5309;  ANEX3092;  ANDX5309;  ANER5149;  ANEP3936;  ANDQ3037;  ANEQ4945;  ANEQ4945;  ANDX5309;  ANDU4074;  ANEZ3430;  ANEZ0906;  ANCY3303;  ANDP3952;  ANDP3952;  ANDP3952;  ANDU4027;  ANDU4027;  ANDV4526;  ANDW4339;  ANDW4339;  ANDX5324;  ANDX5385;  ANDZ3818;  ANDZ3818;  ANEN1086;  ANEQ2717;  ANEQ4946;  ANEQ4946;  ANER5165;  ANET2530;  ANEU0741;  ANEU3303;  ANEX0749;  ANEZ0907;  ANEZ0907;  ANCQ1427;  ANCQ1427;  ANET2530;  ANFN3944;  ANDZ3818;  ANEQ2717;  ANEQ4037;  ANCQ1427;  ANEQ2717;  ANEQ2717;  ANER5165;  ANCQ1427;  ANDW4339;  ANDX5385;  ANCQ1427;  ANCU1844;  ANEX0749;  ANDZ3818;  ANEQ2717;  ANEQ2717;  ANEQ2717;  ANET2530;  ANFN3944;  ANEU0741;  ANEU3303;  ANEX0749;  ANEX0749;  ANEZ0907;  ANDP3952;  ANDQ3871;  ANDQ3871;  ANDX5324;  ANDX5324;  ANDX5324;  ANDX5385;  ANDZ3818;  ANEQ4037;  ANEQ4037;  ANER5165;  ANER5165;  ANEZ0907;  ANFN3944;  ANCY3303;  ANEZ0907;  ANEU0741;  ANEU3303;  ANDU4027;  ANDX5324;  ANDX5324;  ANEN1086;  ANDX5385;  ANDZ3818;  ANDX5385;  ANEQ2717;  ANET2530;  ANEQ4037;  ANDX5324;  ANEQ4037;  ANEQ2717;  ANER5165;  ANFN3944;  ANER5165;  ANEX0749;  ANDX5324;  ANDZ3818;  ANEQ2717;  ANEQ4037;  ANDT0505;  ANDT0505;  ANDT1054;  ANDT1054;  ANDY2469;  ANDY2469;  ANDY4963;  ANDY4963;  ANDY4963;  ANDY4963;  ANDZ1345;  ANEW2047;  ANEW4217;  ANEY4322;  ANEY4322;  ANFN1431;  ANDT1054;  ANEP2417;  ANDZ1345;  ANDT1054;  ANDT1054;  ANDT0505;  ANDT1054;  ANDT1054;  ANDT1054;  ANDY2469;  ANEP2417;  ANDY4963;  ANEP0862;  ANDY2469;  ANEP2417;  ANEW2047;  ANEW4217;  ANEY3209;  ANEY3209;  ANEY4322;  ANDN0807;  ANDQ0320;  ANDQ1722;  ANDQ3030;  ANDQ3030;  ANDT0505;  ANDT0505;  ANDT1054;  ANDT1054;  ANDT1054;  ANDY2469;  ANDY4963;  ANEP0862;  ANEP2417;  ANDT1054;  ANDT1054;  ANEY4322;  ANDT1054;  ANEP0862;  ANEP0862;  ANEW4217;  ANDT0505;  ANDT1054;  ANCU3703;  ANDT1054;  ANDY4963;  ANDY4963;  ANEP0862;  ANFN1431;  ANDT0505;  ANCR2675;  ANDQ0320;  ANDQ1722;  ANDT0505;  ANEP0862;  ANEP0862;  ANEP0862;  ANEP2417;  ANFN1431;  ANDN0808;  ANDQ1726;  ANEQ1347;  ANEQ1347;  ANDQ3031;  ANEQ1347;  ANES2188;  ANDQ1726;  ANDQ3031;  ANDN0808;  ANDQ1726;  ANDQ3031;  ANEQ1347;  ANEQ1347;  ANEQ1347;  ANDQ3031;  ANDQ3031;  ANDQ3031;  ANEQ1347;  ANEQ1347;  ANDQ3031;  ANDQ3031;  ANDN0808;  ANES2188;  ANDQ1726;  ANEQ1347;  ANDU1108;  ANDY3280;  ANCY1571;  ANDU1108;  ANEN3385;  ANEN3385;  ANEQ1348;  ANEN3385;  ANCY1571;  ANEN3385;  ANCY1571;  ANCY1571;  ANCY1571;  ANEN3385;  ANDU1108;  ANEN3385;  ANDW1996;  ANCW3261;  ANET1855;  ANDN0606;  ANDR1266;  ANDR1266;  ANDW1996;  ANDW1996;  ANDW1996;  ANDY2448;  ANDW1996;  ANET1855;  ANDW1996;  ANDY2448;  ANCW3261;  ANCW3261;  ANDY2448;  ANET1855;  ANDY2448;  ANDS4772;  ANDS4772;  ANDS4772;  ANDS4772;  ANDT0508;  ANDW4345;  ANDY4968;  ANDY4968;  ANDY4968;  ANEQ2732;  ANEZ3411;  ANDQ3024;  ANDR1268;  ANDS4772;  ANDT0508;  ANDX5315;  ANDX5384;  ANDY4968;  ANDZ3817;  ANDZ3817;  ANEN4877;  ANEN4877;  ANEP0861;  ANER5152;  ANET0555;  ANET3795;  ANEW3410;  ANEZ3411;  ANFN3974;  ANDT0508;  ANCT4323;  ANCY2200;  ANER5152;  ANDZ3817;  ANDX5384;  ANEQ2732;  ANET0555;  ANFN1443;  ANFN3974;  ANDT0508;  ANEQ2732;  ANCS1161;  ANDS4772;  ANEW3410;  ANEY1223;  ANDS4772;  ANDX5315;  ANEQ2732;  ANDT0508;  ANDZ3817;  ANDX5315;  ANEQ2732;  ANEQ2732;  ANDX5384;  ANDX5384;  ANDX5315;  ANFN1443;  ANEQ2732;  ANDZ3817;  ANEP0861;  ANEQ2732;  ANEW3410;  ANEY1223;  ANFP0911;  ANDZ3817;  ANFP2377;  ANFP0911;  ANEQ2732;  ANER5152;  ANEQ2732;  ANDZ3817;  ANDY4968;  ANEQ2732;  ANCY3322;  ANDS4772;  ANDQ3024;  ANEQ2732;  ANET0555;  ANER5152;  ANER5152;  ANEY1223;  ANFP0911;  ANDQ3024;  ANER5152;  ANEW3410;  ANEQ2732;  ANDX5384;  ANER5152;  ANEY1223;  ANER5152;  ANCS1161;  ANEY1223;  ANDQ3024;  ANDS4772;  ANDX5315;  ANEQ2732;  ANEQ2732;  ANEQ2732;  ANER5152;  ANEW3410;  ANEQ2732;  ANFN3974;  ANFP2377;  ANDX5315;  ANFP0911;  ANDX5384;  ANEQ2732;  ANEP0861;  ANDZ3817;  ANFN1443;  ANDQ3024;  ANDQ3024;  ANDR1268;  ANDT0508;  ANDT0508;  ANDX5315;  ANDX5315;  ANDX5384;  ANEN4877;  ANER5152;  ANET0555;  ANEY1223;  ANEZ3411;  ANEZ3411;  ANFN1443;  ANFP0911;  ANDQ1716;  ANDX5384;  ANDX5384;  ANET0555;  ANET0555;  ANET3795;  ANEY1223;  ANEZ3411;  ANDT0508;  ANDT0508;  ANER5152;  ANEN4877;  ANEQ2732;  ANEQ2732;  ANEW3410;  ANEZ3411;  ANCS1161;  ANDS4772;  ANDX5315;  ANDX5384;  ANER5152;  ANFN1443;  ANCY3322;  ANEQ2732;  ANEQ2732;  ANER5152;  ANER5152;  ANDX5315;  ANEW3410;  ANDY4968;  ANDR1268;  ANDW4345;  ANDX5315;  ANDX5384;  ANDX5384;  ANDY4968;  ANDZ2301;  ANDZ3817;  ANDS4772;  ANFP2377;  ANDX5315;  ANCY2200;  ANER5152;  ANEY1223;  ANFN1443;  ANDT0508;  ANDZ3817;  ANEW3410;  ANEZ3411;  ANEP0861;  ANDX5384;  ANDX5384;  ANDQ3024;  ANEP0861;  ANDQ1716;  ANEQ2732;  ANFP2377;  ANEQ2732;  ANEW3410;  ANDS3973;  ANDS3973;  ANDS4774;  ANDS4774;  ANDS4774;  ANDS4774;  ANDT2946;  ANDW1997;  ANDW4361;  ANDX0760;  ANDX0760;  ANDX0760;  ANDY1234;  ANEP2420;  ANDQ5251;  ANDS2451;  ANDS3973;  ANDS4774;  ANDW4361;  ANDX2782;  ANDZ3820;  ANEN1082;  ANEN2140;  ANER1507;  ANEZ0923;  ANFN1444;  ANFN1444;  ANFN2844;  ANDQ5251;  ANDS3973;  ANFN3975;  ANDZ3820;  ANEN4896;  ANER1507;  ANER1507;  ANDQ5251;  ANDQ3872;  ANDX5373;  ANEQ4962;  ANER1507;  ANDQ3025;  ANDZ2291;  ANDX5373;  ANEQ2733;  ANER1507;  ANCN1543;  ANEQ1353;  ANDT2946;  ANDT2946;  ANDX2782;  ANDX5373;  ANEQ2733;  ANDT2946;  ANDX3060;  ANDX5373;  ANEQ1353;  ANEQ1353;  ANEQ4962;  ANER1507;  ANDS3973;  ANDW1997;  ANEQ2733;  ANFN1444;  ANFN2844;  ANDT2946;  ANEQ2733;  ANER0517;  ANDS3973;  ANDS4774;  ANER1507;  ANER1507;  ANDZ2291;  ANFP2318;  ANES4086;  ANDS3973;  ANDX0760;  ANDX5373;  ANER1507;  ANFN1444;  ANES2211;  ANER1507;  ANDS3973;  ANFP2318;  ANDT2946;  ANDZ3820;  ANEQ2733;  ANFN2844;  ANEN4896;  ANER1507;  ANDT2946;  ANDW4361;  ANDP1356;  ANEN4896;  ANFN1444;  ANEQ2733;  ANER1507;  ANDZ3820;  ANER1507;  ANDT2946;  ANFN3975;  ANDX2782;  ANDX5373;  ANDW4361;  ANEZ2098;  ANFN1444;  ANDZ2291;  ANER1507;  ANEQ1353;  ANDQ5251;  ANFN1444;  ANDX5373;  ANER1507;  ANER1507;  ANER1507;  ANER1507;  ANFN3975;  ANDS3973;  ANEN4896;  ANDQ3872;  ANDW4361;  ANDX2782;  ANEQ4962;  ANDT2946;  ANER1507;  ANDQ5251;  ANDX5373;  ANDY1234;  ANDX5373;  ANER1507;  ANDS4774;  ANER1507;  ANDS4774;  ANDX0760;  ANDY1234;  ANER1507;  ANER1507;  ANDW1997;  ANDX0760;  ANDX5373;  ANDZ3820;  ANER1507;  ANER1507;  ANDV0687;  ANDY1234;  ANER1507;  ANFN2844;  ANFP2318;  ANEZ2098;  ANDU1877;  ANDW1997;  ANDX5373;  ANDW4361;  ANDS3973;  ANEN4896;  ANDW4361;  ANDS2451;  ANDP1356;  ANDW4361;  ANER1507;  ANEQ4962;  ANDQ3025;  ANDS2451;  ANES2211;  ANES4086;  ANES4086;  ANEZ0923;  ANEZ2098;  ANER0517;  ANFN1444;  ANDQ3025;  ANDQ3872;  ANDS2451;  ANDS2451;  ANDS4774;  ANDT2946;  ANDT2946;  ANDV0687;  ANDW1997;  ANDW3561;  ANDW4361;  ANDX0760;  ANDX0760;  ANDX2782;  ANDX2782;  ANDX2782;  ANDX3060;  ANDX5373;  ANDY1234;  ANDZ2291;  ANDZ3820;  ANDZ3820;  ANEN2140;  ANEQ1353;  ANER1507;  ANES1043;  ANES1043;  ANES2211;  ANEZ0923;  ANFN1444;  ANFN1444;  ANFN2844;  ANFN3975;  ANFN3975;  ANFP0879;  ANEN2140;  ANEN4896;  ANEP2420;  ANEQ1353;  ANEQ2733;  ANES2211;  ANES2211;  ANEZ2098;  ANEZ2098;  ANEZ3412;  ANFN1444;  ANFP0879;  ANFP2318;  ANDQ3025;  ANDQ5251;  ANDS3973;  ANDS4774;  ANDT2946;  ANDX0760;  ANDX2782;  ANDX3060;  ANER1507;  ANER1507;  ANER1507;  ANER1507;  ANES2211;  ANDW4361;  ANDS3973;  ANDV0687;  ANES4086;  ANDX5373;  ANEP2420;  ANES4086;  ANDQ3872;  ANER1507;  ANDX5373;  ANDS2451;  ANDT2946;  ANEN4896;  ANER1507;  ANDQ3025;  ANDS4774;  ANER1507;  ANDQ3025;  ANEP2420;  ANER1507;  ANDQ3025;  ANDS2451;  ANEN4896;  ANEQ1353;  ANEQ4962;  ANER1507;  ANFN1444;  ANDS2451;  ANER0517;  ANER1507;  ANDT2946;  ANER1507;  ANER1507;  ANFP0879;  ANDT2946;  ANDY1234;  ANDQ3025;  ANDU1877;  ANDV0687;  ANDX5373;  ANDY1234;  ANES4086;  ANEN2140;  ANFP2317;  ANFP2318;  ANDS4774;  ANDT2946;  ANDW1997;  ANEQ1353;  ANDT2946;  ANDX5373;  ANCW1737;  ANDQ3025;  ANER1507;  ANDT2946;  ANDS4774;  ANFN1444;  ANDX3060;  ANDY1234;  ANDX5373;  ANDT2946;  ANDY1234;  ANER0517;  ANDP0508;  ANDP0508;  ANDP1356;  ANDP1356;  ANEP2420;  ANEQ1353;  ANEQ2733;  ANEQ4962;  ANER0517;  ANER1507;  ANEZ2098;  ANFP2318;  ANER0517;  ANDT2946;  ANDT2946;  ANES4086;  ANCX3883;  ANCX3883;  ANDV0702;  ANDV2519;  ANDV2519;  ANDX0727;  ANDX0727;  ANDX2426;  ANDX2426;  ANDX3063;  ANDX3063;  ANDY3291;  ANDY3291;  ANDZ1331;  ANDQ3033;  ANDS0753;  ANDS0753;  ANDT4130;  ANDV0702;  ANDX2032;  ANDY3291;  ANDZ1331;  ANDZ2297;  ANDZ3824;  ANDZ3824;  ANEN4886;  ANEP0860;  ANER5168;  ANEZ2099;  ANEZ3952;  ANFN3976;  ANDW4378;  ANEP0860;  ANDV0702;  ANDZ3824;  ANEQ1354;  ANFN3976;  ANDX5307;  ANDY1221;  ANEQ1354;  ANEP3934;  ANDX2426;  ANDY3291;  ANEN2145;  ANEP3934;  ANEQ1354;  ANDQ0322;  ANDS0753;  ANDX2426;  ANEQ4069;  ANEZ3952;  ANEQ4069;  ANFP2336;  ANDZ2297;  ANEQ4069;  ANDV1233;  ANDY1221;  ANEP0860;  ANEQ1354;  ANFN3976;  ANEQ1354;  ANEP0860;  ANFN3976;  ANFN3976;  ANEQ4069;  ANFN3976;  ANDT4130;  ANFP0912;  ANEQ4069;  ANEP3934;  ANEP3934;  ANEQ4069;  ANEP0860;  ANEZ3952;  ANDQ3033;  ANEQ1354;  ANCS3348;  ANDT4130;  ANEP0860;  ANEP3934;  ANEQ4069;  ANDV0702;  ANDY2464;  ANEP0860;  ANDV1233;  ANEQ4069;  ANCS3348;  ANDV0702;  ANDV1233;  ANDX2426;  ANDY3291;  ANEN2145;  ANEN4886;  ANEP0860;  ANDW4378;  ANDS0753;  ANEP0860;  ANEQ1354;  ANEQ1354;  ANDS0753;  ANDV0702;  ANDX2426;  ANDX2426;  ANEN2145;  ANEP0860;  ANEP0860;  ANEQ1354;  ANER5168;  ANDQ0322;  ANDQ3033;  ANDS0753;  ANDS0753;  ANDU3110;  ANDV0300;  ANDW4378;  ANDW4378;  ANDX0727;  ANDX2426;  ANDX5307;  ANDX5307;  ANDY1221;  ANDY2464;  ANDZ1331;  ANDZ2297;  ANEN2145;  ANEN3389;  ANEN4886;  ANEP0860;  ANEP3934;  ANER5168;  ANES2212;  ANES2212;  ANEZ0924;  ANEZ2099;  ANEZ2099;  ANEZ2099;  ANEZ3952;  ANEZ3952;  ANFN2787;  ANFN3976;  ANFN3976;  ANFP0912;  ANEZ0924;  ANEZ3952;  ANFP0912;  ANDQ3033;  ANDS0753;  ANDT4130;  ANDV1233;  ANDY1221;  ANEP3934;  ANEP3934;  ANEQ1354;  ANEQ1354;  ANEQ4069;  ANEQ4069;  ANEZ3952;  ANDZ2297;  ANEQ4069;  ANEQ4069;  ANDS0753;  ANDT4130;  ANDV0702;  ANDX2032;  ANDX5307;  ANEN2145;  ANEN3389;  ANEQ4069;  ANDQ0322;  ANDS0753;  ANDT4130;  ANDU3110;  ANEQ1354;  ANEQ1354;  ANER5168;  ANDX5307;  ANEQ1354;  ANES2212;  ANEN2145;  ANEQ4069;  ANDV2519;  ANDX2426;  ANDX5307;  ANDY2464;  ANDZ1331;  ANDZ1331;  ANDZ3824;  ANEN1064;  ANEN2145;  ANEN2145;  ANEQ1354;  ANDZ1331;  ANER5168;  ANFP0912;  ANFP2336;  ANDV2519;  ANEQ1354;  ANES2212;  ANDT4130;  ANCX3883;  ANDV0702;  ANDY2464;  ANER5168;  ANDZ2297;  ANEQ1354;  ANEN3389;  ANDT4130;  ANDT4130;  ANES2212;  ANDQ0322;  ANDT4130;  ANDZ3824;  ANEN3389;  ANDV0702;  ANDV1233;  ANEZ0924;  ANEZ3952;  ANFN3976;  ANCT4331;  ANDQ0322;  ANDU3110;  ANEQ4069;  ANEP0860;  ANEQ4069;  ANER5168;  ANDV2519;  ANEQ1354;  ANDY1221;  ANEQ1354;  ANCX3884;  ANDR0754;  ANDS2461;  ANDS3976;  ANDS3976;  ANDS3976;  ANDS4839;  ANDT4131;  ANDY1231;  ANDY3284;  ANDY3284;  ANDY3284;  ANDR0754;  ANDS2461;  ANDS3976;  ANDX2031;  ANDX3062;  ANDX5363;  ANDX5363;  ANDY1231;  ANDZ3823;  ANEN2147;  ANEN3398;  ANER1508;  ANET0556;  ANEZ3429;  ANCS0655;  ANEQ1355;  ANER0518;  ANEQ1355;  ANDS3976;  ANDS3976;  ANDS4839;  ANDX5323;  ANDX5363;  ANDX5363;  ANER0518;  ANDX5363;  ANER0518;  ANDX5363;  ANER0518;  ANER0518;  ANET0556;  ANER0518;  ANER1508;  ANFP2365;  ANDS4839;  ANDX5323;  ANDX5323;  ANDX5363;  ANEQ1355;  ANDU1889;  ANDX2031;  ANER1508;  ANER1508;  ANET0556;  ANER0518;  ANEQ1355;  ANDX5363;  ANDX5363;  ANDT4131;  ANER0518;  ANDX5323;  ANER1508;  ANDS3976;  ANDQ0321;  ANDW1999;  ANET0556;  ANDS2461;  ANDR0754;  ANEQ1355;  ANER1508;  ANES4087;  ANEZ2100;  ANDQ0321;  ANDQ3876;  ANDS2461;  ANDS2461;  ANDS3976;  ANDT4131;  ANDT4131;  ANDU1889;  ANDW1999;  ANDW1999;  ANDX0739;  ANDX0739;  ANDX3062;  ANDX5323;  ANDX5323;  ANDX5323;  ANDX5363;  ANDY3284;  ANDY4974;  ANDZ3823;  ANEN2147;  ANEN2147;  ANEQ1355;  ANEQ2734;  ANEQ4963;  ANER1508;  ANES4087;  ANES4087;  ANET0556;  ANET0556;  ANEY2031;  ANEY2031;  ANEZ3429;  ANEZ3429;  ANFN1445;  ANFN1445;  ANFN3977;  ANFN3977;  ANFN3977;  ANFP2365;  ANEN3398;  ANEQ1355;  ANEQ4963;  ANER1508;  ANET0556;  ANEY2031;  ANEZ2100;  ANFP2365;  ANDQ3876;  ANDX2031;  ANDX5323;  ANER0518;  ANER0518;  ANER0518;  ANER1508;  ANER1508;  ANDU1889;  ANDT4131;  ANER0518;  ANER0518;  ANDQ0321;  ANDS3976;  ANDS4839;  ANER1508;  ANDX5363;  ANDX5363;  ANDS3976;  ANDX5323;  ANDQ3876;  ANDT4131;  ANDU1889;  ANDW1999;  ANDW4395;  ANDW4395;  ANDX5363;  ANDX5363;  ANDY3284;  ANDY4974;  ANDZ3823;  ANET0556;  ANDY3284;  ANDR0754;  ANDU1889;  ANEN3398;  ANER1508;  ANET0556;  ANFP2365;  ANDS4839;  ANDX2031;  ANET0556;  ANER0518;  ANDX5363;  ANDR0754;  ANDX2031;  ANDS4839;  ANDS3976;  ANDR0754;  ANDX2031;  ANDX5323;  ANFN1445;  ANDZ3823;  ANDR0754;  ANDS4839;  ANDU1889;  ANEN3398;  ANER1508;  ANDU1889;  ANDS4839;  ANDX5363;  ANEN2147;  ANER0518;  ANDR0754;  ANDR0754;  ANDS4839;  ANDZ3823;  ANEN3398;  ANEQ4963;  ANER0475;  ANER0518;  ANER1508;  ANER0518;  ANDU4028;  ANDV0275;  ANDV0275;  ANDW4413;  ANDW4413;  ANDX5301;  ANDX5301;  ANDY4972;  ANDY4972;  ANDU4028;  ANDV0275;  ANEN1083;  ANEN1083;  ANEQ2735;  ANDW4413;  ANEQ2735;  ANER5120;  ANEQ2735;  ANDQ3028;  ANES4088;  ANEQ2735;  ANES4088;  ANER5120;  ANES4088;  ANES4088;  ANDV0275;  ANDQ3028;  ANDU4028;  ANEP5329;  ANES4088;  ANES4088;  ANES4088;  ANFP0880;  ANFP0880;  ANDQ3028;  ANDW4413;  ANEP5329;  ANEP5329;  ANES4088;  ANES4088;  ANDX5301;  ANEN1083;  ANEQ2735;  ANEQ2735;  ANES4088;  ANDX5301;  ANDV0275;  ANDQ3028;  ANDW4413;  ANDW4413;  ANDX0751;  ANDX5301;  ANDX5301;  ANDY4972;  ANER5120;  ANDU4028;  ANDV0275;  ANDX5301;  ANDX5301;  ANER5120;  ANER5120;  ANES4088;  ANER5120;  ANDU4028;  ANDX5301;  ANDV0275;  ANDV0275;  ANDV0275;  ANEQ2735;  ANEQ2735;  ANEN1083;  ANEN1083;  ANEN4883;  ANEN4883;  ANEP5329;  ANEQ2735;  ANEQ2735;  ANFP0880;  ANFP0880;  ANFP0880;  ANFP2331;  ANER5120;  ANDQ3027;  ANDQ3027;  ANDU4040;  ANDU4040;  ANDX5327;  ANDX5327;  ANDX5389;  ANDX5389;  ANEP5309;  ANEP5309;  ANEP5309;  ANEQ1356;  ANEQ2736;  ANEP5309;  ANEP5309;  ANER0519;  ANDQ3874;  ANDX5327;  ANER1535;  ANEP5309;  ANDQ3027;  ANDX5327;  ANEP5309;  ANEQ1356;  ANDX5327;  ANDQ3874;  ANDU1103;  ANDU4040;  ANDX5389;  ANDZ1326;  ANEP5309;  ANER1535;  ANER1535;  ANER1535;  ANDQ3874;  ANDQ3874;  ANDU1103;  ANER1535;  ANDU1103;  ANCX3885;  ANEP5309;  ANEP5309;  ANEQ1356;  ANDU1103;  ANEP5309;  ANEP5309;  ANEQ2736;  ANDU1104;  ANDY2452;  ANDY2452;  ANDY3299;  ANEN4894;  ANDQ3034;  ANDQ3034;  ANEN1084;  ANEN1084;  ANEQ4964;  ANEN1084;  ANEN1084;  ANDY2452;  ANDZ3825;  ANEN1084;  ANCQ4294;  ANDQ3034;  ANDU1104;  ANDY2452;  ANDZ3825;  ANEN1084;  ANER0520;  ANER0520;  ANER0520;  ANER0520;  ANCV1583;  ANER0520;  ANDU1104;  ANEN1084;  ANDT1055;  ANDR1267;  ANDT1055;  ANDW1998;  ANDW1998;  ANDW1998;  ANDX1402;  ANDY4964;  ANDY4964;  ANES2189;  ANET0542;  ANEW4218;  ANEX1697;  ANEZ2081;  ANEZ3398;  ANFN3946;  ANFP0857;  ANDR1267;  ANDT1055;  ANEN2153;  ANEP2424;  ANDU4039;  ANDT1055;  ANDT0506;  ANEZ3398;  ANCW3262;  ANDT0506;  ANDU2242;  ANEZ3398;  ANDZ2307;  ANDR1267;  ANDW1998;  ANEN2153;  ANDU2242;  ANDZ2307;  ANEX6010;  ANEZ0908;  ANDU4039;  ANFN3946;  ANDZ2307;  ANDT0506;  ANDT1055;  ANFP0857;  ANDU4039;  ANDT1055;  ANDZ2307;  ANEZ2081;  ANEX6010;  ANDT1055;  ANEW4218;  ANDT1055;  ANDW1998;  ANDT0506;  ANDU2242;  ANDU2242;  ANDZ2307;  ANEN2153;  ANCS1158;  ANFP0857;  ANDT1055;  ANEZ3398;  ANDW1998;  ANEP2424;  ANDW1998;  ANDY3289;  ANDY3289;  ANDY4964;  ANDZ2307;  ANET0542;  ANEW3366;  ANEX1697;  ANEX6010;  ANEX6010;  ANEZ0908;  ANEZ3398;  ANFP0857;  ANDT0506;  ANDX1402;  ANDY3289;  ANDR1267;  ANDT0506;  ANDU2242;  ANDU4039;  ANEP2424;  ANEP2424;  ANEW4218;  ANEX6010;  ANEX6010;  ANEZ2081;  ANEZ3398;  ANFP0857;  ANDU4039;  ANDT0506;  ANDT1055;  ANDW1998;  ANEW4218;  ANDU4039;  ANDZ2307;  ANEN2153;  ANDZ2307;  ANES2189;  ANDU2242;  ANDU2242;  ANES2189;  ANDP1357;  ANDR1267;  ANEN2153;  ANEN2153;  ANDR1267;  ANDZ2307;  ANFN3946;  ANDY3289;  ANDZ2307;  ANDX1402;  ANDZ2307;  ANDR1267;  ANDT1055;  ANDX1402;  ANEN2153;  ANEN2153;  ANDU4039;  ANDU4039;  ANEP2424;  ANET0542;  ANFP0857;  ANCW3262;  ANDT0506;  ANDU2242;  ANDU4039;  ANDU4039;  ANDZ2307;  ANDU2242;  ANDT0506;  ANDU4039;  ANDR1267;  ANEZ2081;  ANDR1267;  ANDX1402;  ANEP2424;  ANES2189;  ANDP2690;  ANDU2242;  ANCR1285;  ANDY4964;  ANEX6010;  ANDU4039;  ANEP2424;  ANEX1697;  ANEP2421;  ANDQ3868;  ANDS0752;  ANDT1059;  ANDX1401;  ANDX2425;  ANDX5362;  ANDY1230;  ANDZ1322;  ANDZ2286;  ANEN4897;  ANEN4897;  ANEP0867;  ANEP2421;  ANEV1711;  ANEW4219;  ANFN1432;  ANDN0810;  ANDS0752;  ANDS0752;  ANDV4534;  ANDX5362;  ANDZ1322;  ANDZ2286;  ANEQ4039;  ANEN4897;  ANEQ2718;  ANCW1424;  ANDT0507;  ANDX1401;  ANDQ3022;  ANDV4534;  ANFP0858;  ANFN3947;  ANFN3947;  ANEW4219;  ANEP3930;  ANES4062  ANDX2425;  ANDQ1713;  ANEP2421;  ANEP3930;  ANEZ2082;  ANDQ3022;  ANFP2358;  ANDV4534;  ANEQ4039;  ANER1514;  ANDT0507;  ANDV4534;  ANDY1230;  ANDV4534;  ANDT1059;  ANEP3930;  ANEW4219;  ANCW1424;  ANER1514;  ANCX1867;  ANER1514;  ANDX2425;  ANFN3947;  ANDY1230;  ANDY1230;  ANDZ1322;  ANDY1230;  ANEP0867;  ANEQ4039  ANES2190;  ANDZ1322;  ANEP0867;  ANEW3367;  ANEW4219;  ANES4062;  ANES4062;  ANEV1711;  ANEW3367;  ANEW4219;  ANEX0746;  ANEX1708;  ANEY1228;  ANFN1432;  ANFN3947;  ANFP0858;  ANDQ1713;  ANDQ3022;  ANDQ3022;  ANDQ3868;  ANDX3057;  ANDX5362;  ANCV0494;  ANDN0810;  ANDT0507;  ANDT0507;  ANDX1401;  ANDX1401;  ANDX2425;  ANDX5362;  ANDY1230;  ANDZ2286;  ANEN4897;  ANEP0867;  ANEP3930;  ANEQ2718;  ANEQ4039;  ANEQ4039;  ANEQ4039;  ANER1514;  ANER1514;  ANER1514;  ANES2190;  ANES4062;  ANEX1708;  ANEX3091;  ANDT1059;  ANDQ3868;  ANEP3930;  ANCU0848;  ANDX1401;  ANER1514;  ANDQ3868;  ANDX2425;  ANDY1230;  ANEQ4039;  ANER1514;  ANES4062;  ANEY1228;  ANDT0507;  ANDV4534;  ANDY1230;  ANDN0810;  ANDT0507;  ANDT1059;  ANDX2425;  ANDZ1322;  ANER1514;  ANFP0858;  ANDT0507;  ANEW4219;  ANEZ2082;  ANER1514;  ANEX3091;  ANEW4219;  ANCU0848;  ANDX3057;  ANDZ1322;  ANER1514;  ANES2190;  ANDS0752;  ANER1514;  ANDT0507;  ANDQ1713;  ANDS0752;  ANDX2425;  ANDX3057;  ANER1514;  ANEX3091;  ANFN3947;  ANDX2425;  ANDX2425;  ANER1514;  ANEX3091;  ANDT",
      "center_classification_date": "20210617",
      "distribution_pattern": "Worldwide distribution. US Nationwide, ALGERIA,   ANDORRA,  ARGENTINA,  AUSTRALIA,  AUSTRIA,  AZERBAIJAN,  BAHRAIN,  BELGIUM,  BRAZIL,  BRUNEI,  BULGARIA,  CHILE,  COSTA RICA,  CYPRUS,  CZECH REPUBLIC,  DENMARK,  DEUTSCHLAND,  EGYPT,  FINLAND,  FRANCE,  GEORGIA,  GREECE,  HONG KONG,  HUNGARY,  INDIA,  IRAN,  IRAQ,  IRELAND,  ISRAEL,  ITALY,  JORDAN,  KAZAKHSTAN,  MALAYSIA,  MEXICO,  NETHERLANDS,  NORWAY,  OMAN,  PANAMA,  POLAND,  PORTUGAL,  ROMANIA,  RUSSIA,  SAUDIA ARABIA,  SINGAPORE,  SLOVAKIA (SLOVAK REPUBLIC),  SLOVENIA,  SOUTH AFRICA,  SOUTH KOREA,  SPAIN,  SWEDEN,  SWITZERLAND,  TAIWAN,  TURKEY,  UNITED ARAB EMIRATES (UAE),  UNITED KINGDOM,  VIETNAM",
      "state": "AZ",
      "product_description": "VENOVO Venous Stent System 8F    Product Description(REF)/dimension:  VENEL10040\t/Venovo 8F 10/40/1200mm OUS;  VENEL10060\t/Venovo 8F 10/60/1200mm OUS;  VENEL10080\t/Venovo 8F 10/80/1200mm OUS;  VENEL10100\t/Venovo 8F 10/100/1200mm OUS;  VENEL10120\t/Venovo 8F 10/120/1200mm OUS;  VENEL10140\t/Venovo 8F 10/140/1200mm OUS;  VENEL10160\t/Venovo 8F10/160/1200mm OUS;  VENEL12040\t/Venovo 8F 12/40/1200mm OUS;  VENEL12060\t/Venovo 8F 12/60/1200mm OUS;  VENEL12080\t/Venovo 8F 12/80/1200mm OUS;  VENEL12100\t/Venovo 8F 12/100/1200mm OUS;  VENEL12120\t/Venovo 8F 12/120/1200mm OUS;  VENEL12140\t/Venovo 8F 12/140/1200mm OUS;  VENEL12160\t/Venovo 8F 12/160/1200mm OUS;  VENEM10040\t/Venovo 8F 10/40/800mm OUS;  VENEM10060\t/Venovo 8F 10/60/800mm OUS;  VENEM10080\t/Venovo 8F 10/80/800mm OUS;  VENEM10100\t/Venovo 8F 10/100/800mm OUS;  VENEM10120\t/Venovo 8F 10/120/800mm OUS;  VENEM10140\t/Venovo 8F 10/140/800mm OUS;  VENEM10160\t/Venovo 8F 10/160/800mm OUS;  VENEM12040\t/Venovo 8F 12/40/800mm OUS;  VENEM12060\t/Venovo 8F 12/60/800mm OUS;  VENEM12080\t/Venovo 8F 12/80/800mm OUS;  VENEM12100\t/Venovo 8F 12/100/800mm OUS;  VENEM12120\t/Venovo 8F 12/120/800mm OUS;  VENEM12140\t/Venovo 8F 12/140/800mm OUS;  VENEM12160\t/Venovo 8F 12/160/800mm OUS;  VENUL10040\t/Venovo 8F 10/40/1200mm US;  VENUL10060\t/Venovo 8F 10/60/1200mm US;  VENUL10080\t/Venovo 8F 10/80/1200mm US;  VENUL10100\t/Venovo 8F 10/100/1200mm US;  VENUL10120\t/Venovo 8F 10/120/1200mm US;  VENUL10140\t/Venovo 8F 10/140/1200mm US;  VENUL10160\t/Venovo 8F 10/160/1200mm US;  VENUL12040\t/Venovo 8F 12/40/1200mm US;  VENUL12060\t/Venovo 8F 12/60/1200mm US;  VENUL12080\t/Venovo 8F 12/80/1200mm US;  VENUL12100\t/Venovo 8F 12/100/1200mm US;  VENUL12120\t/Venovo 8F 12/120/1200mm US;  VENUL12140\t/Venovo 8F 12/140/1200mm US;  VENUL12160\t/Venovo 8F 12/160/1200mm US;  VENUM10040\t/Venovo 8F 10/40/800mm US;  VENUM10060\t/Venovo 8F 10/60/800mm US;  VENUM10080\t/Venovo 8F 10/80/800mm US;  VENUM10100\t/Venovo 8F 10/100/800mm US;  VENUM10120\t/Venovo 8F 10/120/800mm US;  VENUM10140\t/Venovo 8F 10/140/800mm US;  VENUM10160\t/Venovo 8F 10/160/800mm US;  VENUM12040\t/Venovo 8F 12/40/800mm US;  VENUM12060\t/Venovo 8F 12/60/800mm US;  VENUM12080\t/Venovo 8F 12/80/800mm US;  VENUM12100\t/Venovo 8F 12/100/800mm US;  VENUM12120\t/Venovo 8F 12/120/800mm US;  VENUM12140\t/Venovo 8F 12/140/800mm US;  VENUM12160\t/Venovo 8F 12/160/800mm US;",
      "report_date": "20210623",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "recall_number": "Z-1898-2021",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "87208",
      "more_code_info": "0507;  ANDX1401;  ANDX2425;  ANDV4534  ANEP3930;  ANES4062;  ANFP0858;  ANDS0752;  ANDV4534;  ANDX1401;  ANEQ4039;  ANEQ4039;  ANDX3057;  ANEZ2082;  ANES4062;  ANEQ4039;  ANDT0507;  ANDX2425;  ANES4062;  ANEW4219;  ANEP3930;  ANEQ2718;  ANCV0494;  ANDQ1713;  ANDQ1713;  ANFN3947;  ANDV4534;  ANDX2425;  ANEQ4039;  ANFN3947;  ANDQ1713;  ANDT1059;  ANEQ4039;  ANER1514;  ANDS0752;  ANDS0752;  ANDX3057;  ANES2190;  ANDZ1322;  ANEP3930;  ANES4062;  ANFN3947;  ANEQ4039;  ANEX3091;  ANER1514;  ANES2190;  ANDX2425;  ANEQ4039;  ANEP3930;  ANES4062;  ANDX3057;  ANDX3057;  ANDT0507;  ANEX1708;  ANEP5317;  ANDR2271;  ANDR2271;  ANDV3567;  ANDX1405;  ANDX2034;  ANDY4960;  ANDZ1332;  ANEP3939;  ANES2191;  ANET0543;  ANEW4260;  ANFN2783;  ANFN3948;  ANCW4216;  ANDN0821;  ANDN0821;  ANDV4528;  ANDZ2304;  ANEN1070;  ANEN3402;  ANEQ4040;  ANBY2393;  ANCQ2599;  ANDV4528;  ANEQ4040;  ANFN3948;  ANFN3948;  ANCW3263;  ANEQ4040;  ANEQ4040;  ANDR2271;  ANFN3948;  ANFN3948;  ANDP3957;  ANEX3093;  ANCW3263;  ANEP3939;  ANFN3948;  ANDV4528;  ANDV4528;  ANDV4528;  ANEQ4040;  ANES2191;  ANDV4528;  ANDR2271;  ANDV4528;  ANCW4216;  ANEP5317;  ANCT1573;  ANDV4528;  ANDZ1332;  ANEP5317;  ANDZ2304;  ANDZ2304;  ANDZ2304;  ANEP5317;  ANEP5317;  ANFN3948;  ANDQ3039;  ANDV4528;  ANDZ1332;  ANDZ1332;  ANEP5317;  ANET0543;  ANET3786;  ANET3786;  ANEW4260;  ANEW4260;  ANEX3093;  ANEX3093;  ANEY1221;  ANEY3210;  ANEY3210;  ANEY3210;  ANFN2783;  ANFN3948;  ANDP3957;  ANDQ1717;  ANDR2271;  ANCT4314;  ANCV0496;  ANDP3957;  ANDQ3039;  ANDV4528;  ANDX1374;  ANDY1238;  ANDY4960;  ANDZ1332;  ANEP3939;  ANEQ2719;  ANEQ4040;  ANEQ4040;  ANEQ4040;  ANEQ4040;  ANER5118;  ANER5118;  ANES2191;  ANEX3093;  ANEX6157;  ANEY1221;  ANFN3948;  ANDZ2304;  ANDY1238;  ANEQ2719;  ANDX1374;  ANDV4528;  ANEQ4040;  ANDP3957;  ANDV4528;  ANDX1374;  ANDY1238;  ANDP3987;  ANDV3567;  ANEP3939;  ANER5118;  ANEY3210;  ANDZ2304;  ANDV4528;  ANEY3210;  ANEZ3431;  ANDZ2304;  ANDV4528;  ANDP3957;  ANES2191;  ANEX6157;  ANEZ3431;  ANER5118;  ANDQ1717;  ANET3786;  ANEY1221;  ANEX6157;  ANFN3948;  ANDV4528;  ANCW4216;  ANDQ1717;  ANDX1405;  ANDQ3039;  ANDX1374;  ANDX2034;  ANDY1238;  ANDZ1332;  ANEN1070;  ANER5118;  ANEQ4040;  ANDR2271;  ANDQ1717;  ANDR2271;  ANDV4528;  ANDV4528;  ANEQ4040;  ANEX6157;  ANEX6157;  ANET3786;  ANEY1221;  ANEP5317;  ANEQ2719;  ANEQ4040;  ANDP3957;  ANEZ3431;  ANDQ3039;  ANDX1405;  ANEP5317;  ANDX1374;  ANDX1374;  ANEQ4040;  ANEQ4040;  ANFN3948;  ANEQ4040;  ANCW3263;  ANEQ2719;  ANEQ4040;  ANER5118;  ANDQ1717;  ANEP5317;  ANEQ4040;  ANES2191;  ANDR2271;  ANDV3567;  ANEP5317;  ANEQ4040;  ANDV3567;  ANDX2034;  ANEN3402;  ANDR2271;  ANET3786;  ANEX6157;  ANET3786;  ANDY1238;  ANDZ2304;  ANET3786;  ANFN3948;  ANEP2435;  ANDQ1723;  ANDQ3040;  ANDU2243;  ANDW4355;  ANDW4355;  ANDW4355;  ANDX1406;  ANDX2035;  ANDX5316;  ANDX5390;  ANDZ2283;  ANEN1071;  ANEN4895;  ANER1515;  ANES4063;  ANEX0672;  ANEX6011;  ANEZ3399;  ANFN1433;  ANCX0889;  ANDP3959;  ANDP3959;  ANDP3959;  ANDP3959;  ANDX2035;  ANDZ2283;  ANEN1071;  ANEQ2720;  ANEQ4041;  ANDW4355;  ANEX0672;  ANFN1433;  ANDQ3040;  ANDV1244;  ANDX2035;  ANDX2035;  ANET3787;  ANDQ3040;  ANEQ2720;  ANER1515;  ANEY3211;  ANEQ2720;  ANDX5316;  ANFN1433;  ANCX0889;  ANCX3891;  ANDQ3879;  ANDU2243;  ANDU2243;  ANDV1244;  ANDX2035;  ANDU2243;  ANDX2035;  ANDV1244;  ANES4063;  ANDQ1723;  ANDV1244;  ANEP2435;  ANEQ2720;  ANEX6011;  ANDV1244;  ANDQ1723;  ANDQ3040;  ANDZ2283;  ANEQ2720;  ANDQ5218;  ANEN4895;  ANEX3081;  ANER1515;  ANES4063;  ANET3787;  ANEW3412;  ANEX3081;  ANEX6011;  ANEY3211;  ANEZ3399;  ANFN1433;  ANFN3949;  ANDQ0326;  ANDQ1723;  ANDQ3040;  ANDQ3879;  ANDP3959;  ANDQ0326;  ANDX5316;  ANDX5316;  ANDX5316;  ANDX5390;  ANDY3308;  ANDZ2283;  ANEQ2720;  ANEQ4947;  ANEQ4947;  ANER1515;  ANES4063;  ANEX3081;  ANEX6011;  ANDU2243;  ANDU2243;  ANEQ4041;  ANCX3891;  ANDW4355;  ANEN4895;  ANDQ3879;  ANDQ5218;  ANDU2243;  ANDV1244;  ANDX5390;  ANDZ2283;  ANEN4895;  ANEP2435;  ANEQ4947;  ANER1515;  ANER1515;  ANER1515;  ANES4063;  ANFN3949;  ANDQ1723;  ANDQ5218;  ANDV1244;  ANCQ1429;  ANDZ2283;  ANEQ4947;  ANER1515;  ANEY3211;  ANDX1406;  ANDX2035;  ANDQ3040;  ANCQ1429;  ANDX5390;  ANDX5390;  ANEX6011;  ANEY3211;  ANDU2243;  ANDX2035;  ANER1515;  ANDQ1723;  ANDW4355;  ANDX1406;  ANER1515;  ANEQ2720;  ANCQ2600;  ANDX1406;  ANEQ4947;  ANEQ4947;  ANCN3578;  ANDX2035;  ANDP3959;  ANDQ1723;  ANDU2243;  ANDX2035;  ANDX2035;  ANEQ2720;  ANEQ4041;  ANEQ4041;  ANEQ4041;  ANER1515;  ANEX6011;  ANES4063;  ANEX6011;  ANDZ2283;  ANEQ2720;  ANDX2035;  ANEP2435;  ANDX5316;  ANEQ4041;  ANEX3081;  ANDW4355;  ANER1515;  ANEX6011;  ANFN1433;  ANDX5390;  ANDQ3879;  ANDX5390;  ANEN1071;  ANES4063;  ANEX0672;  ANEY3211;  ANEQ4947;  ANEX6011;  ANDX1406;  ANES4063;  ANFN1433;  ANEX6011;  ANEZ3399;  ANFN3949;  ANEN2150;  ANDS0751;  ANDS0751;  ANDS4788;  ANDU2244;  ANDV4533;  ANDX3056;  ANDX5361;  ANDY3283;  ANDY4965;  ANDZ1327;  ANDZ2287;  ANEW4220;  ANEW4220;  ANEX1707;  ANFN1434;  ANFN2784;  ANDP3950;  ANDP3950;  ANDX1373;  ANDY4965;  ANEN4880;  ANDS4788;  ANDS4823;  ANEQ4042;  ANCQ1430;  ANDN0809;  ANDS4788;  ANDW4372;  ANEQ4042;  ANDY3283;  ANEQ4042;  ANER1516;  ANEN2150;  ANEN4880;  ANCS1563;  ANCW4217;  ANDS4823;  ANEN2150;  ANFN2784;  ANDS4788;  ANDY4965;  ANEP0877;  ANER1516;  ANDS4823;  ANDS4788;  ANDY3283;  ANFN2784;  ANDW4372;  ANDZ2287;  ANDZ2287;  ANEX1707;  ANDP3950;  ANDX1373;  ANDX3056;  ANDY1229;  ANDY2456;  ANCW4217;  ANDS0751;  ANDS4788;  ANDS4823;  ANDU2244;  ANDV4533;  ANDW4372;  ANDW4372;  ANDW4372;  ANDY2456;  ANDY2456;  ANEP0877;  ANEQ4042;  ANER1516;  ANEX1707;  ANDY3283;  ANDS4788;  ANEN2150;  ANDX1373;  ANEW4220;  ANDS4788;  ANEX1707;  ANEQ4042;  ANDV4533;  ANDS4823;  ANER1516;  ANER1516;  ANEX1707;  ANDS4788;  ANDU2244;  ANEW4220;  ANEW4220;  ANCS3345;  ANCS3345;  ANDS4788;  ANDX1373;  ANDZ2287;  ANDU2244;  ANDU2244;  ANEW4220;  ANER1516;  ANER1516;  ANFN2784;  ANER1516;  ANFN2784;  ANER1516;  ANDW4372;  ANDS4823;  ANEP0877;  ANDV4533;  ANEN4880;  ANEX1707;  ANEN2150;  ANER1516;  ANCS3345;  ANDN0809;  ANEN4880;  ANDY3283;  ANEN4880;  ANEN4880;  ANEW4220;  ANEW4220;  ANEN2150;  ANDS4823;  ANEP0877;  ANEQ4042;  ANEX1707;  ANDS0751;  ANDU2244;  ANER1516;  ANDS4823;  ANDU2244;  ANDS4788;  ANDZ2287;  ANEP0877;  ANDY4965;  ANDZ2287;  ANER1516;  ANEW4220;  ANEX1707;  ANEX1707;  ANEX1707;  ANCQ1430;  ANER1516;  ANFN1434;  ANEN3387;  ANEN1072;  ANEP2430;  ANBZ1288;  ANDX5376;  ANEP2430;  ANCS3433;  ANEN1072;  ANER1517;  ANDS1282;  ANDS1282;  ANDU2245;  ANEP2430;  ANDW2002;  ANDW4389;  ANDX5376;  ANDY3288;  ANCS3433;  ANDU2245;  ANDU2245;  ANDU2245;  ANDW2002;  ANDY3288;  ANDY3288;  ANEN1072;  ANEP2430;  ANEQ4043;  ANER1517;  ANDS1282;  ANEQ4043;  ANEQ4043;  ANER1517;  ANDU2245;  ANDU2245;  ANDW2002;  ANDW2002;  ANEN1072;  ANEN3387;  ANBZ1288;  ANDS1282;  ANDW4389;  ANDS1282;  ANDS1282;  ANEP2430;  ANDX5376;  ANER1517;  ANDS1282;  ANEP2430;  ANEN3387;  ANEQ4043;  ANDX5376;  ANEN1072;  ANEN3387;  ANDU2245;  ANEN3387;  ANDX5376;  ANEN1072;  ANEN3387;  ANFN3952;  ANEN1072;  ANDY3287;  ANDX5375;  ANDY3287;  ANEP2412;  ANDV4536;  ANDY3287;  ANEQ4948;  ANES4065;  ANDS4776;  ANDS4776;  ANDV4536;  ANDX5375;  ANDY3287;  ANEP2412;  ANEP2412;  ANEP2412;  ANEQ4948;  ANDY3287;  ANEP2412;  ANDV4536;  ANDY3287;  ANDX1403;  ANDX5375;  ANCY3335;  ANDV4536;  ANDX1403;  ANDX1403;  ANDX5375;  ANDS4776;  ANDY3287;  ANEQ4948;  ANES4065;  ANEQ4948;  ANES4065;  ANDS4776;  ANDT1686;  ANEZ3427;  ANEZ3427;  ANCY3335;  ANDS4776;  ANDT1686;  ANEQ4948;  ANDX1403;  ANCU1841;  ANEQ4948;  ANEQ4948;  ANEP2412;  ANDY3287;  ANES4065;  ANES4065;  ANEZ3427;  ANEZ3427;",
      "recall_initiation_date": "20201214",
      "postal_code": "85281-2438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}