  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85495",
      "recalling_firm": "NuVasive Inc",
      "address_1": "7475 Lusk Blvd",
      "address_2": "N/A",
      "postal_code": "92121-5707",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The Countries of Argentina, Austria, Belgium, Bulgaria, Canada, Chile, Denmark, Finland, France, Germany, Hong Kong, Hungary, India, Israel, Ireland, Italy, Jordan, Kuwait, Luxembourg, Malaysia, Netherlands, Norway, Philippines, Poland, Portugal, Saudi Arabia, Slovenia/Serbia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-1898-2020",
      "product_description": "NUVASIVE MAGEC 2 Rod X.X mm XX mm, REF MC2-XXXXS   LOT 0000000 AAA - Product Usage: comprised of a sterile single-use spinal rod that is surgically implanted using appropriate NuVasive Reline, Reline Small Stature, and Armada fixation components (i.e. Pedicle screws, hooks and/or connectors, provided non-sterile, to be sterilized by the end user).",
      "product_quantity": "3502 rods",
      "reason_for_recall": "Post-implantation separation of an actuator end cap component that may expose internal components of the actuator, which could lead to hastened degeneration of the internal components and egress of Titanium alloy wear debris and resultant localized tissue discoloration.",
      "recall_initiation_date": "20200213",
      "center_classification_date": "20200505",
      "termination_date": "20220525",
      "report_date": "20200513",
      "code_info": "Part Number:  MC2-4570R  Product:  MAGEC 2 ROD, 4.5mm 70mmOFFSET  Lot Numbers:  8061403AAA  8061411AAA  8061419AAA  8061419AAB  8061508AAA  8061808AAA  8061809AAA  8062301AAA  8062602AAA  8062705AAA  8062810AAA  8062810AAB  8063003AAA  8063011AAA  8072409AAA  8072501AAA  8072505AAA  8072627AAA  8072628AAA  8080805AAA  8081102AAA  8092609AAA  8100810AAA  8101905AAA  8102003AAA  8103165AAA  8110306AAA  8111202AAA  8111504AAA  8111508AAA  8122101AAA  8122610AAA  9011102AAA  9011428AAA  9011429AAA  9021601AAA  9021603AAA  9022703AAA  9031511AAA  9031601AAA  9032002AAA  9032901AAA  9032914AAA  9032915AAA  9042508AAA  9042601AAA  9070104AAA  9070112AAA  9071114AAA  9071521AAA  9071610AAA  9071617AAA  9072215AAA  9072701AAA  9072702AAA  9072705AAA  9073001AAA  9073022AAA  9081501AAA  9081502AAA  9082113AAA  9091905AAA  9091906AAA  9112302AAA  9121924AAA  9121938AAA  9121941AAA  9122622AAA    Part Number:  MC2-4570S  Product:  MAGEC 2 ROD, 4.5mm 70mmSTANDARD  Lot Numbers:  8060727AAA  8061101AAA  8061106AAA  8061203AAA  8061204AAA  8061213AAA  8061214AAA  8061218AAA  8061318AAA  8062307AAA  8062622AAA  8062713AAA  8062902AAA  8063005AAA  8070206AAA  8071001AAA  8071118AAA  8071208AAA  8072508AAA  8072510AAA  8080703AAA  8080902AAA  8081001AAA  8081010AAA  8081011AAA  8081012AAA  8091324AAA  8091401AAA  8091401AAB  8091404AAA  8091701AAA  8092738AAA  8100803AAA  8101015AAA  8101901AAA  8102902AAA  8102902AAB  8102910AAA  8110505AAA  8111322AAA  8111325AAA  8111414AAA  8112108AAA  8120727AAA  8120902AAA  8121328AAA  8121904AAA  8121913AAA  9011112AAA  9011502AAA  9011601AAA  9011611AAA  9011627AAA  9011712AAA  9030102AAA  9030104AAA  9030111AAA  9021401AAA  9021402AAA  9021404AAA  9031501AAA  9032701AAA  9032704AAA  9032708AAA  9032711AAA  9032714AAA  9032801AAA  9032807AAA  9040110AAA  9051701AAA  9051708AAA  9051801AAA  9051802AAA  9051803AAA  9052008AAA  9052010AAA  9061124AAA  9061125AAA  9061126AAA  9061220AAA  9061221AAA  9071726AAA  9071727AAA  9071732AAA  9071826AAA  9071905AAA  9071906AAA  9082101AAA  9082102AAA  9090438AAA  9091601AAA  9091619AAA  9091620AAA  9091701AAA  9091702AAA  9091712AAA  9092414AAA  9092415AAA  9092417AAA  9093011AAA  9100701AAA  9112014AAA  9112015AAA  9112016AAA  9112017AAA  9122059AAA  9122063AAA  9122101AAA    Part Number:  MC2-4590R   Product:  MAGEC 2 ROD, 4.5mm 90mmOFFSET  Lot Numbers:  0010301AAA  0010333AAA  0010334AAA  0010818AAA  8062009AAA  8062101AAA  8062107AAA  8062118AAA  8062123AAA  8062125AAA  8070211AAA  8070601AAA  8070701AAA  8070702AAA  8071113AAA  8071116AAA  8072518AAA  8072520AAA  8073004AAA  8080706AAA  8081313AAA  8082802AAA  8082903AAA  8090705AAA  8091005AAA  8091118AAA  8091122AAA  8092414AAA  8092508AAA  8100505AAA  8111411AAA  8121813AAA  9010909AAA  9012422AAA  9022811AAA  9032503AAA  9051306AAA  9051407AAA  9051411AAA  9051419AAA  9061712AAA  9061722AAA  9062801AAA  9062802AAA  9062803AAA  9062819AAA  9071201AAA  9071717AAA  9072003AAA  9072004AAA  9072706AAA  9073023AAA  9080814AAA  9080815AAA  9081901AAA  9081902AAA  9081911AAA  9092011AAA  9121310AAA  9121311AAA  9121311AAB  9121312AAA    Part Number:  MC2-4590S   Product:  MAGEC 2 ROD, 4.5mm 90mmSTANDARD  Lot Numbers:  8060409AAA  8060601AAA  8061910AAA  8061921AAA  8061931AAA  8061932AAA  8062001AAA  8062014AAA  8062318AAA  8062625AAA  8070223AAA  8070615AAA  8070901AAA  8070902AAA  8071204AAA  8071213AAA  8071706AAA  8071907AAA  8072105AAA  8082309AAA  8082311AAA  8082313AAA  8082401AAA  8082410AAA  8082419AAA  8082701AAA  8082713AAA  8082718AAA  8082814AAA  8090402AAA  8090709AAA  8091013AAA  8092007AAA  8092102AAA  8092402AAA  8092404AAA  8100305AAA  8100410AAA  8111306AAA  8111415AAA  8111421AAA  8121109AAA  8121111AAA  8121205AAA  8121211AAA  8121214AAA  8121312AAA  8121406AAA  9010806AAA  9011803AAA  9011812AAA  9020405AAA  9021216AAA  9021306AAA  9021302AAA  9022204AAA  9031107AAA  9031814AAA  9032005AAA  9032006AAA  9032007AAA  9032014AAA  9032101AAA  9032102AAA  9051011AAA  9051016AAA  9061812AAA  9061819AAA  9061901AAA  9061902AAA  9061911AAA  9061912AAA  9061929AAA  9062018AAA  9062411AAA  9062412AAA  9062413AAA  9062508AAA  9062517AAA  9062519AAA  9062520AAA  9062702AAA  9072313AAA  9072314AAA  9072516AAA  9072601AAA  9080301AAA  9081307AAA  9081312AAA  9090930AAA  9090931AAA  9090956AAA  9091001AAA  9091002AAA  9093010AAA  9111222AAA  9111223AAA  9112013AAA  9121635AAA  9121711AAA  9121712AAA  9121713AAA  9121719AAA    Part Number:  MC2-5070R  Product:  MAGEC 2 ROD, 5.0mm 70mmOFFSET  Lot Numbers:  0011023AAA  0012101AAA  0012102AAA  0012128AAA  0012129AAA  0012201AAA  0012202AAA  0012910AAA  0012922AAA  0013001AAA  0013002AAA  8111511AAA  9022011AAA  9030412AAA  9030415AAA  9030504AAA  7080101AAA  7083002AAA  7090517AAA  7090608AAA  7090903AAA  7091514AAA  7091604AAA  7092101AAA  7100501AAA  7101310AAA  7112214AAA  7121218AAA  8030603AAA  8042406AAA  8042704AAA  9030205AAA  8043013AAA  8050804AAA  8052201AAA  8102504AAA  8102511AAA  8103001AAA  8103102AAA  8110621AAA  8110703AAA  9041817AAA  9041906AAA  9041918AAA  9050905AAA  9050907AAA  9062729AAA  9090517AAA  9090518AAA  9090609AAA  9093018AAA  9100115AAA  9100116AAA  9101501AAA  9101502AAA  9101514AAA  9101516AAA  9101516AAB  9101612AAA  9101613AAA  9101701AAA  9111310AAA  9112214AAA  9112216AAA  9122337AAA  9122338AAA  9122401AAA  9122709AAA    Part Number:  MC2-5070S  Product:  MAGEC 2 ROD, 5.0mm 70mmSTANDARD  Lot Numbers:  0010742AAA  0010743AAA  0010770AAA  0010771AAA  0010773AAA  0011643AAA  0011644AAA  0011714AAA  0011715AAA  0011803AAA  0011804AAA  0012003AAA  0012020AAA  0012024AAA  7080701AAA  7083001AAA  7090112AAA  7090516AAA  7090701AAA  7090902AAA  7091506AAA  7091603AAA  7092303AAA  7092605AAA  7092802AAA  7100409AAA  7101309AAA  7112003AAA  7112803AAA  7121410AAA  7121602AAA  8041114AAA  8042408AAA  8042705AAA  8050905AAA  8051004AAA  8051101AAA  8051107AAA  8051112AAA  8062821AAA  8071403AAA  8071601AAA  8071701AAA  8071902AAA  8072623AAA  8072626AAA  8080403AAA  8080613AAA  8091906AAA  8091909AAA  8100919AAA  8102313AAA  8102501AAA  8103117AAA  8103150AAA  8110501AAA  8110514AAA  8110915AAA  9030206AAA  9030207AAA  9042001AAA  9042002AAA  9042004AAA  9042209AAA  9042214AAA  9042216AAA  9050818AAA  9050909AAA  9051008AAA  9070807AAA  9070808AAA  9082301AAA  9082302AAA  9082613AAA  9082614AAA  9082701AAA  9082702AAA  9082704AAA  9090511AAA  9090512AAA  9090608AAA  9091301AAA  9091302AAA  9100117AAA  9100126AAA  9100301AAA  9100302AAA  9100705AAA  9101803AAA    Part Number:  MC2-5090R  Product:  MAGEC 2 ROD, 5.0mm 90mmOFFSET  Lot Numbers:  0010403AAA  0010404AAA  0011317AAA  0011318AAA  0011319AAA  0011401AAA  0011402AAA  0011506AAA  0011524AAA  0011525AAA  0011601AAA  0011602AAA  7081001AAA  7082502AAA  7083108AAA  7090110AAA  7090805AAA  7091602AAA  7092302AAA  7100401AAA  7102501AAA  7112002AAA  7112213AAA  7121411AAA  8020716AAA  8030601AAA  8042507AAA  8042804AAA  8062818AAA  8073105AAA  8091403AAA  8092702AAA  8092702AAB  8100818AAA  8101159AAA  8102314AAA  8103027AAA  8103050AAA  8112008AAA  8122112AAA  8122606AAA  9011103AAA  9022001AAA  9022114AAA  9031406AAA  9031414AAA  9042301AAA  9042305AAA  9042308AAA  9042308AAB  9042309AAA  9050902AAA  9060107AAA  9060604AAA  9060611AAA  9060614AAA  9061412AAA  9082216AAA  9082305AAA  9082306AAA  9082706AAA  9091201AAA  9091202AAA  9091317AAA  9091318AAA  9100333AAA  9100334AAA  9100407AAA  9100408AAA  9121402AAA    Part Number:  MC2-5090S  Product:  MAGEC 2 ROD, 5.0mm 90mmSTANDARD  Lot Numbers:  7072701AAA  7082501AAA  7083105AAA  7090111AAA  7090702AAA  7090806AAA  7091505AAA  7091601AAA  7092214AAA  7092301AAA  7092506AAA  7100219AAA  7101604AAA  7102419AAA  7110102AAA  7121217AAA  7121601AAA  8020207AAA  8040902AAA  8040917AAA  8042603AAA  8043003AAA  8050908AAA  8051003AAA  8051201AAA  8051402AAA  8051501AAA  8051511AAA  8051511AAB  8071032AAA  8071101AAA  8071105AAA  8071709AAA  8071801AAA  8072110AAA  8072421AAA  8072421AAB  8081610AAA  8081701AAA  8081714AAA  8081715AAA  8082002AAA  8082111AAA  8082805AAA  8082902AAA  8083006AAA  8090609AAA  8091801AAA  8091806AAA  8100403AAA  8102209AAA  8110104AAA  8110515AAA  8110601AAA  8111254AAA  8111901AAA  8112026AAA  8120903AAA  8120904AAA  8120905AAA  8121407AAA  8121915AAA  8121923AAA  8122007AAA  8122011AAA  9010801AAA  9011114AAA  9011816AAA  9011901AAA  9012301AAA  9021802AAA  9021807AAA  9021903AAA  9021906AAA  9021908AAA  9030521AAA  9030610AAA  9031409AAA  9040102AAA  9041005AAA  9041008AAA  9042401AAA  9042412AAA  9042414AAA  9042501AAA  9050819AAA  9052407AAA  9052408AAA  9052409AAA  9052804AAA  9052805AAA  9052816AAA  9052817AAA  9053025AAA  9053112AAA  9060105AAA  9061106AAA  9061120AAA  9061120AAB  9061310AAA  9082201AAA  9082202AAA  9082801AAA  9082802AAA  9101401AAA  9101403AAA  9101415AAA  9101416AAA  9101601AAA  9101610AAA  9110107AAA  9110108AAA  9110110AAA  9110401AAA  9110407AAA  9110408AAA  9110901AAA  9110902AAA  9110903AAA  9110904AAA  9111109AAA  9111110AAA  9112021AAA  9112022AAA  9112023AAA  9112101AAA  9112102AAA  9112611AAA    Part Number:  MC2-5570R  Product:  MAGEC 2 ROD, 5.5mm 70mmOFFSET  Lot Numbers:  8051804AAA  8052101AAA  8052105AAA  8052308AAA  8052309AAA  8060705AAA  8060723AAA  8060724AAA  8070303AAA  8071304AAA  8071316AAA  8072701AAA  8080705AAA  8080705AAB  8081408AAA  8081416AAA  8081511AAA  8081602AAA  8081713AAA  8092602AAA  8100811AAA  8102201AAA  8102212AAA  8102503AAA  8111908AAA  8112015AAA  8112112AAA  8122110AAA  8122607AAA  9021606AAA  9021607AAA  9022102AAA  9022814AAA  9030707AAA  9030810AAA  9040114AAA  9040201AAA  9052304AAA  9052316AAA  9052321AAA  9052406AAA  9060306AAA  9060309AAA  9060317AAA  9060703AAA  9060707AAA  9060714AAA  9060804AAA  9061006AAA  9061018AAA  9061032AAA  9061105AAA  9090444AAA  9091804AAA  9091805AAA  9121316AAA  9121317AAA    Part Number:  MC2-5570S  Product:  MAGEC 2 ROD, 5.5mm 70mmSTANDARD  Lot Numbers:  0012222AAA  0012301AAA  0012302AAA  0012304AAA  0012317AAA  0012318AAA  0012401AAA  0012415AAA  0012419AAA  0012713AAA  0012714AAA  0013115AAA  0013116AAA  8060103AAA  8060503AAA  8060611AAA  8060613AAA  8060702AAA  8061323AAA  8062017AAA  8062121AAA  8062124AAA  8062322AAA  8062628AAA  8062718AAA  8062905AAA  8063010AAA  8070502AAA  8082017AAA  8082022AAA  8082105AAA  8082201AAA  8082204AAA  8082213AAA  8083008AAA  8090401AAA  8092608AAA  8100809AAA  8101904AAA  8102306AAA  8102320AAA  8102323AAA  8102411AAA  8111603AAA  8121051AAA  8121056AAA  8121060AAA  8121330AAA  9011903AAA  9021502AAA  9021520AAA  9022605AAA  9021405AAA  9031410AAA  9031901AAA  9032810AAA  9032812AAA  9032814AAA  9052120AAA  9052121AAA  9052122AAA  9052209AAA  9052210AAA  9060413AAA  9060516AAA  9061224AAA  9061225AAA  9061314AAA  9062003AAA  9062004AAA  9062005AAA  9082001AAA  9082002AAA  9082933AAA  9083001AAA  9091019AAA  9091021AAA  9091116AAA  9121401AAA  9121437AAA  9122102AAA  9122603AAA    Part Number:  MC2-5590R  Product:  MAGEC 2 ROD, 5.5mm 90mmOFFSET  Lot Numbers:  0010335AAA  0010336AAA  0012801AAA  0012802AAA  0012829AAA  0012901AAA  8060107AAA  8060517AAA  8060605AAA  8060608AAA  8062019AAA  8062112AAA  8070505AAA  8071327AAA  8071402AAA  8071608AAA  8071711AAA  8071814AAA  8071910AAA  8072305AAA  8073006AAA  8080217AAA  8080809AAA  8081103AAA  8081104AAA  8082215AAA  8083003AAA  8090504AAA  8090724AAA  8091101AAA  8091211AAA  8091221AAA  8091311AAA  8091702AAA  8092415AAA  8092502AAA  8092505AAA  8100509AAA  9010919AAA  9012002AAA  9012125AAA  9012204AAA  9012425AAA  9021317AAA  9022110AAA  9022801AAA  9032511AAA  9032601AAA  9032603AAA  9032628AAA  9032633AAA  9051431AAA  9051501AAA  9051504AAA  9051506AAA  9051601AAA  9051602AAA  9051603AAA  9060408AAA  9060411AAA  9062106AAA  9062107AAA  9062403AAA  9062622AAA  9062623AAA  9073007AAA  9081401AAA  9081402AAA  9081412AAA  9081417AAA  9081601AAA  9112019AAA  9112020AAA  9122001AAA    Part Number:  MC2-5590S  Product:  MAGEC 2 ROD, 5.5mm 90mmSTANDARD  Lot Numbers:  0010620AAA  0010621AAA  8050118AAA  8050207AAA  8050312AAA  8050703AAA  8060109AAA  8061116AAA  8061126AAA  8061206AAA  8061210AAA  8062023AAA  8062113AAA  8062511AAA  8062702AAA  8062808AAA  8062920AAA  8070512AAA  8072309AAA  8072311AAA  8073005AAA  8080316AAA  8080401AAA  8080612AAA  8080812AAA  8080905AAA  8081316AAA  8081317AAA  8081404AAA  8081503AAA  8081703AAA  8082905AAA  8083101AAA  8090615AAA  8091103AAA  8092107AAA  8092405AAA  8092411AAA  8100101AAA  8100311AAA  8100414AAA  8101813AAA  8110105AAA  8110203AAA  8110223AAA  8111257AAA  8112604AAA  8112610AAA  8120619AAA  8120625AAA  8120722AAA  8121704AAA  8121705AAA  8121717AAA  8121801AAA  8121802AAA  9012003AAA  9012309AAA  9012332AAA  9012402AAA  9012402AAB  9012404AAA  9012410AAA  9022002AAA  9030612AAA  9030613AAA  9030703AAA  9030823AAA  9030826AAA  9030901AAA  9030902AAA  9030904AAA  9031102AAA  9032119AAA  9032120AAA  9032205AAA  9032212AAA  9032213AAA  9032217AAA  9032222AAA  9041701AAA  9041706AAA  9041707AAA  9050704AAA  9050706AAA  9050712AAA  9050806AAA  9051004AAA  9070915AAA  9070916AAA  9071111AAA  9071208AAA  9071209AAA  9071301AAA  9071302AAA  9071517AAA  9072315AAA  9072318AAA  9072319AAA  9072415AAA  9072420AAA  9072421AAA  9080812AAA  9080813AAA  9081301AAA  9081917AAA  9082601AAA  9082602AAA  9090410AAA  9091714AAA  9091801AAA  9111301AAA  9111302AAA  9111306AAA  9112119AAA  9112201AAA  9112209AAA  9112210AAA  9112211AAA  9112523AAA  9120306AAA  9120307AAA  9120401AAA  9121809AAA  9121810AAA  9121815AAA    Part Number:  MC2-6070R  Product:  MAGEC 2 ROD, 6.0mm 70mmOFFSET  Lot Numbers:  8041901AAA  8042002AAA  8042101AAA  8042304AAA  8042614AAA  8043008AAA  8051614AAA  8073109AAA  8082909AAA  8090604AAA  9081508AAA  9082707AAA  9112303AAA  9120511AAA  9120518AAA  9120531AAA  9120632AAA  9120633AAA  9121201AAA  9121202AAA  9122620AAA  9122621AAA    Part Number:  MC2-6070S  Product:  MAGEC 2 ROD, 6.0mm 70mmSTANDARD  Lot Numbers:  8041808AAA  8050203AAA  8050303AAA  8050401AAA  8051711AAA  8051715AAA  8062828AAA  8080101AAA  8080402AAA  8083007AAA  8092001AAA  8092012AAA  8092014AAA  9042611AAA  9061503AAA  9073115AAA  9091803AAA  9102401AAA  9120510AAA  9120517AAA  9120521AAA  9120526AAA  9120528AAA  9121203AAA  9122334AAA  9122602AAA    Part Number:  MC2-6090R  Product:  MAGEC 2 ROD, 6.0mm 90mmOFFEST  Lot Numbers:  0010401AAA  0010402AAA  8041646AAA  8043010AAA  8051105AAA  8051203AAA  8051205AAA  8051409AAA  8051503AAA  8051601AAA  8061303AAA  8071214AAA  8071219AAA  8080103AAA  8080112AAA  8080202AAA  8081508AAA  8090508AAA  8091011AAA  8092703AAA  8100533AAA  8121902AAA  8122106AAA  9011002AAA  9012209AAA  9012424AAA  9031821AAA  9051624AAA  9061510AAA  9072222AAA  9082004AAA  9092001AAA  9120414AAA  9120519AAA  9120901AAA  9121015AAA  9121309AAA    Part Number:  MC2-6090S  Product:  MAGEC 2 ROD, 6.0mm 90mmSTANDARD  Lot Numbers:  8041635AAA  8042102AAA  8042402AAA  8042404AAA  8042703AAA  8050808AAA  8071003AAA  8071005AAA  8071020AAA  8072007AAA  8072011AAA  8072103AAA  8072109AAA  8080105AAA  8080113AAA  8080605AAA  8081605AAA  8081611AAA  8082813AAA  8091704AAA  8091705AAA  8091706AAA  8091915AAA  8100314AAA  8120705AAA  8121805AAA  8121810AAA  9010807AAA  9021209AAA  9022526AAA  9022536AAA  9031113AAA  9031208AAA  9051101AAA  9051103AAA  9051104AAA  9071718AAA  9072424AAA  9100203AAA  9100204AAA  9100207AAA  9112301AAA  9120402AAA  9120416AAA  9120417AAA  9120501AAA  9120502AAA  9121009AAA  9121016AAA  9121439AAA  9121441AAA    Reference \"Attachment 4 NuVasive MAGEC X PN-Lot Numbers.xlsx\" uploaded into Associated Documents.",
      "more_code_info": ""
    }
  ]
}