{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83482",
      "recalling_firm": "Intuitive Surgical Inc",
      "address_1": "950 Kifer Rd",
      "address_2": "N/A",
      "postal_code": "94086-5206",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, District of Columbia, FL, GA, IL, IN, IA, KS. MA, MI, MS, NB, NY, OH, PA, TN, TX, VA, WA and WI and the countries of Australia, Belgium, Brazil, France, Germany, India, Italy, Japan, South Korea, Spain, Sweden, and the United Kingdom.",
      "recall_number": "Z-1896-2020",
      "product_description": "da Vinci Xi Surgical System, IS4000 - Product Usage: The da Vinci Xi surgical system (IS4000) is made of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The Patient Cart is the operative component of the IS4000 system and its primary function is to support the instrument arms and camera arm. The system has four Universal Surgical Manipulators (USM) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. There are 4 USM per da Vinci Xi surgical system.",
      "product_quantity": "270 USM units on a combination of 100 unique systems",
      "reason_for_recall": "Failure of Universal Surgical Manipulators due to a high friction points, resulting in an increase of arm movement to include: feelings of resistance, hesitation, and jerking.",
      "recall_initiation_date": "20190411",
      "center_classification_date": "20200505",
      "termination_date": "20240408",
      "report_date": "20200513",
      "code_info": "UDI: 00886874114216  Device Listing: D215864  Serial Numbers:  SK0736, SK0856, SK1318, SK1943, SK2228, SK2236, SK2239, SK2240, SK2242, SK2244, SK2248, SK2251, SK2255, SK2256, SK2259, SK2260, SK2261, SK2266, SK2269, SK2270, SK2274, SK2275, SK2281, SK2283, SK2285, SK2294, SK2295, SK2297, SK2300, SK2303, SK2304, SK2305, SK2310, SK2311, SK2312, SK2313, SK2314, SK2315, SK2317, SK2318, SK2320, SK2327, SK2329, SK2330, SK2331, SK2333, SK2341, SK2342, SK2343, SK2346, SK2348, SK2349, SK2350, SK2353, SK2355, SK2356, SK2358, SK2359, SK2360, SK2362, SK2363, SK2364, SK2365, SK2367, SK2368, SK2369, SK2370, SK2373, SK2374, SK2375, SK2383, SK2385, SK2389, SK2390, SK2393, SK2397, SK2402, SK2404, SK2405, SK2406, SK2412, SK2418, SK2420, SK2437, SK2449, SK2451, SK2452, SK2453, SK2455",
      "more_code_info": ""
    }
  ]
}