{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83082",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1896-2019",
      "product_description": "Siemens Artis zee biplane-Interventional Fluoroscopic X-Ray System  Model Number: 10094141",
      "product_quantity": "120",
      "reason_for_recall": "Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off,  can result in severe injuries to patients and operating personnel",
      "recall_initiation_date": "20190606",
      "center_classification_date": "20190628",
      "termination_date": "20210209",
      "report_date": "20190710",
      "code_info": "Serial Numbers:  153883  154041  153640  153772  154007  153870  153762  153700  153797  153827  153898  153779  153869  153952  153736  153829  153763  153998  153819  153758  153928  153760  153865  153903  153908  154592  153863  154785  154079  153988  153882  153831  153939  153875  153847  153930  153913  153890  153922  153992  153916  153975  154012  153947  154076  154086  154058  154028  154073  154048  154034  153987  154093  154090  154074  154052  154075  154406  154408  154459  154425  154423  154502  154084  154579  154546  154085  154471  154453  154097  154098  154544  154663  154576  154572  154560  154584  154571  154096  154514  154556  154509  154731  154536  154564  154739  154596  155147  154783  154653  154732  154687  154685  154779  154708  155150  154709  154715  154789  131011  131224  154800  131013  155119  131101  154814  131105  155122  131108  131110  155126  155131  155148  155189  131223  155195  131219  155203  155235  131231"
    }
  ]
}