  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85256",
      "recalling_firm": "Intuitive Surgical, Inc.",
      "address_1": "1266 Kifer Rd Bldg 101",
      "address_2": "N/A",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide including in the states of AL, AK, AZ, AR, CA, CO, CT, DE, D.C., FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI and the countries of Australia, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands, Norway, South Korea, Spain, Switzerland, United Kingdom.",
      "recall_number": "Z-1891-2020",
      "product_description": "daVinci XI Sureform 60 Reload, 60 mm black 6-Row 4.6 mm, Ref:  48360T-08 , UDI: 00886874113851, Rx only, SterileEO - Product Usage: are intended to be used with da Vinci Surgical Systems for resection, transection, and, or creation of anastomoses in General, Thoracic, Gynecologic, Urologic, and Pediatric surgery.",
      "product_quantity": "N/A",
      "reason_for_recall": "The firm became aware of a manufacturing variation in which black and green reloads in select Stapler Reloads may not deploy individual adjacent staples, which could result in an incomplete staple line.  Failure of a incomplete staple like could potentially lead to an air leak or an anastomotic leak which may require an additional procedural intervention.",
      "recall_initiation_date": "20190701",
      "center_classification_date": "20200505",
      "report_date": "20200513",
      "code_info": "Lot numbers: T10180615; T10180924; T10181003; T11181003; T10181010; T10181016; T10181029; T10181107; T10181109; T10181113",
      "more_code_info": ""
    }
  ]
}